JR IMPLANT SYSTEM

K130694 · LK Consulting Group USA, Inc. · DZE · Jun 19, 2014 · Dental

Device Facts

Record IDK130694
Device NameJR IMPLANT SYSTEM
ApplicantLK Consulting Group USA, Inc.
Product CodeDZE · Dental
Decision DateJun 19, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Intended Use

The JR Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The JR Implant System is for single and two stage surgical procedures. These systems are intended for delayed loading.

Device Story

JR Implant System consists of endosseous dental implants and abutments used to replace missing teeth. Implants are surgically placed into the upper or lower jaw bone by a dental professional. The system utilizes a bone-level, tapered design with an internal tri-lobe connection and a straight-walled neck with micro-threads for a crestal seal. The fixture body features a Resorbable Blast Media (RBM) surface treatment to facilitate integration. Following a conventional healing period, the implant is loaded with prosthetic restorations. The device provides a foundation for single or multiple-unit restorations, including cemented or screw-retained crowns and overdentures, or serves as an abutment for fixed bridgework. Benefits include restoration of oral function and aesthetics for edentulous patients.

Clinical Evidence

Bench testing only. Biocompatibility testing (ISO Cytotoxicity, Sensitization, and Intracutaneous reactivity) confirmed materials are non-toxic and non-sensitizing. Sterilization validation and surface treatment analysis were performed.

Technological Characteristics

Materials: Ti 6Al 4V ELI (ASTM F136). Design: Bone-level, tapered body, internal tri-lobe connection, straight-walled neck with micro-threads. Surface: RBM (Resorbable Blast Media) on fixture; anodized abutments. Sterilization: Gamma radiation. Platform sizes: 3.5mm, 4.3mm, 5.0mm, 6.0mm.

Indications for Use

Indicated for partially or fully edentulous patients requiring single or multiple-unit restorations (cement-retained, screw-retained, or overdenture) or fixed bridgework support in the mandible or maxilla. Suitable for single or two-stage surgical procedures with delayed loading.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(K) SUMMARY (K130694) 1. Submitter / Applicant: June / 1 8/2014 KJ Meditech Co., Ltd. 959-21 Daechon-dong, Buk-gu, Gwang-ju, 500-470, South Korea Phone : +82-62-972-5476 Fax : +82-62-973-2809 #### 2. Submission Correspondent: Priscilla Chung LK Consulting Group USA, Inc. 2651 E Chapman Ave Ste 110, Fullerton CA 92831 Tel: 714-202-5789 Email: juhee.c@lkconsultinggroup.com #### 3. Device Name : | Trade Name | : | JR Implant System | |---------------------|---|--------------------------------------| | Common Names | : | Abutment, Dental, Endosseous implant | | Classification Name | : | Implant, Endosseous, Root-Form | | Regulation | : | 21 CFR 872.3640, DZE | #### 4. Predicate Device : NobelReplace Tapered Conical Connection (K062566) Manufactured by Nobel Biocare USA LLC #### 5. Device Description : The JR Implant system is made of Titanium 6AL 4V ELI alloy intended to be surgically placed in the bone of the upper or lower jaw arches for loading after a conventional healing period. The implants may be used to replace one or more missing teeth. The systems are similar to #### 510(k) Summary {1}------------------------------------------------ KJ Meditech Co., Ltd other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics. The surface of the implant has been treated with R.B.M (Resorbable Blast Media). The fixtures are offered in the following length and diameter. - Narrow Platform Fixture: 3.5mm Dia. x 8.6mm(L) / 12.1mm(L) / 13.6mm(L) / 16.6mm(L) - Regular Platform Fixture : 4.3 mm Dia. x 8.6mm(L) / 12.1mm(L) / 13.6mm(L) / 13.6mm(L) - Wide Platform Fixture : 5.0mm Dia x 8.6mm(L) / 12.1mm(L) / 13.6mm(L) / 16.6mm(L) - 6.0 Platform Fixture : 6.0mm Dia. x 8.6mm(L) / 12.1mm(L) / 13.6mm(L) / 13.6mm(L) There are five different types of abutments offered and they are provided straight only. - Esthetic Abutment - EZ Abutment - Healing Abutment - Cover Screw - Temporary Abutment #### 5 Intended For Use : The JR Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The JR Implant System are for single and two stage surgical procedures. These systems are intended for delayed loading. | Item | Subject Device | Predicate Device | |-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(K) Number | N/A | K062566 | | Device Name | JR Implant System | NobelReplace Tapered Conical Connection | | Manufacturer | KJ Meditech Co., Ltd. | Nobel Biocare USA LLC | | Indications for Use | Intended for use in partially or fully<br>edentulous mandibles and maxillae, in<br>support of single or multiple-unit<br>restorations including; cement-retained,<br>screw-retained, or overdenture restorations.<br>and terminal or intermediate abutment<br>support for fixed bridgework. | Intended for use in partially or fully<br>edentulous mandibles and maxillae, in<br>support of single or multiple-unit<br>restorations including; cement-retained,<br>screw-retained, or overdenture restorations,<br>and terminal or intermediate abutment<br>support for fixed bridgework. | | | | | | Design | • Implant Type: Bone Level Implant<br>• Connection Type: Internal tri-lobe<br>connection<br>• Neck Design: Straight walled neck with<br>micro-thread provides crestal seal.<br>• Body Design: Tapered design enables<br>placement near impringing anatomical<br>structures while maximizing prosthetic<br>table diameter for natural emergence<br>profile. | • Implant Type: Bone Level Implant<br>• Connection Type: Internal tri-channel<br>connection<br>• Neck Design: Straight walled neck with<br>circumferential thread provides crestal<br>seal.<br>• Body Design: Tapered design enables<br>placement near impringing anatomical<br>structures while maximizing prosthetic<br>table diameter for natural emergence<br> | | Fixture<br>Appearance | Image: Implant with RBM Treatment on the fixture body | Image: Implant with TiUnite surface treatment | | Fixture Surface<br>Treatment | RBM Treatment on the fixture body | TiUnite® surface treatment. | | Fixture Material | Ti 6Al 4V ELI, ASTM F136 | Pure titanium, ASTM F67 | | Fixture Sterile | Yes | Yes | | Sterilization Method | Gamma | Gamma | | Fixture Diameters | 3.5mm 4.3mm 5.0mm 6.0mm | 3.5mm 4.3mm 5.0mm 6.0mm | | Fixture Lengths | 8mm - 16.0 mm | 8mm - 16.0 mm | | Abutment Surface<br>Treatment | Anodizing | Anodizing | | Abutment material | Ti 6Al 4V ELI, ASTM F136 | Pure titanium, ASTM F67 | | Cover screw Surface<br>Treatment | Anodizing | Anodizing | | Cover screw material | Ti 6Al 4V ELI, ASTM F136 | Pure titanium, ASTM F67 | | Healing Abut Surface<br>Treatment | Anodizing | Anodizing | | Healing Abut material | Ti 6Al 4V ELI, ASTM F136 | Pure titanium, ASTM F67 | | Temporary Abut Surface<br>Treatment | Anodizing | Anodizing | | Temporary Abut material | Ti 6Al 4V ELI, ASTM F136 | Pure titanium, ASTM F67 | | Abutment<br>Type | For Narrow Platform Fixture<br>For Regular Platform Fixture<br>For Wide Platform Fixture<br>For 6.0 Platform Fixture | For Narrow Platform Fixture<br>For Regular Platform Fixture<br>For Wide Platform Fixture<br>For 6.0 Platform Fixture | | Cover Screw Type | For Narrow Platform Fixture<br>For Regular Platform Fixture<br>For Wide Platform Fixture<br>For 6.0 Platform Fixture | For Narrow Platform Fixture<br>For Regular Platform Fixture<br>For Wide Platform Fixture<br>For 6.0 Platform Fixture | | Healing Abut Type | For Narrow Platform Fixture<br>For Regular Platform Fixture<br>For Wide Platform Fixture | For Narrow Platform Fixture<br>For Regular Platform Fixture<br>For Wide Platform Fixture | | | For 6.0 Platform Fixture | For 6.0 Platform Fixture | | Temporary Abutment | For Narrow Platform Fixture<br>For Regular Platform Fixture<br>For Wide Platform Fixture<br>For 6.0 Platform Fixture | For Narrow Platform Fixture<br>For Regular Platform Fixture<br>For Wide Platform Fixture<br>For 6.0 Platform Fixture | | Product Code | DZE | DZE | #### 6. Substantial Equivalence : {2}------------------------------------------------ KJ Meditech Co., Ltd 510(k) Summary An - Relation of a 1 Page 3 of 4 {3}------------------------------------------------ #### KJ Meditech Co., Ltd The JR Implant system is substantially equivalent in intended use, design and performance to the predicate devices. The differences between the subject device and the predicate device are the titanium grade and surface treatment of the fixture; however, testing data such as biocompatibility testing provided in the submission proves that this difference would not raise issues in safety and performance. Therefore, we claim that the proposed device is substantially equivalent to the predicate device. #### 7. Performance Testing · Biocompatibility testing on the proposed JR Implant System has been completed. The biocompatibility test results show that the materials used in the design and manufacture of the components of the proposed device are non-toxic and non-sensitizing to biological tissues consistent with its intended use. The following biocompatibility tests were completed. -ISO Cytotoxicity -ISO Sensitization -ISO Intracutaneous reactivity · Sterilization validation and surface treatment analysis were also conducted on JR Implant System. #### 8. Conclusion Based on the information provided in this premarket notification KJ Meditech Co., Ltd concludes that the JR Implant System is substantially equivalent to the predicate device as described herein. {4}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains an emblem of an eagle with outstretched wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged around the top and left side of the circle. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 19, 2014 KJ Meditech Co., Ltd. C/O Ms. Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 2651 East Chapman Avenue, Suite 110 Fullerton, CA 92833 Re: K130694 Trade/Device Name: JR Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: May 14, 2014 Received: May 20, 2014 #### Dear Ms. Chung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 - Ms. Chung Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mary S. Runner -S Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indication for use Statement K130694 510(k) Number ( If known ) : Device Name : JR Implant System Indication for use : The JR Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The JR Implant System is for single and two stage surgical procedures. These systems are intended for delayed loading. Prescription Use ---------X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter (Part 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) # Sheena A. Green -S 2014.06.19 10:58:57 -04'00' Indication for use statement Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...