IH Implant System
K222707 · Sewonmedix, Inc. · DZE · Jul 7, 2023 · Dental
Device Facts
| Record ID | K222707 |
| Device Name | IH Implant System |
| Applicant | Sewonmedix, Inc. |
| Product Code | DZE · Dental |
| Decision Date | Jul 7, 2023 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.3640 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
IH Implant System is device made of titanium alloy indicated for in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained restorations and terminal or interminal abutment support for fixed bridgework. IH Implant System is for stage surgical procedures. It is intended for delayed loading.
Device Story
IH Implant System comprises endosseous threaded titanium alloy fixtures and various abutments (healing, cement, multi-unit). Fixtures are surgically implanted into the mandible or maxilla by a dental clinician to replace tooth roots. The system supports single or multiple-unit restorations, including cemented retained restorations and fixed bridgework. The device is intended for delayed loading in one or two-stage surgical procedures. The fixture surface is treated via alumina sandblasting and acid etching to enhance bone contact. Abutments connect to the fixture to support soft tissue healing or prosthetic rehabilitation. The system provides a stable foundation for dental prosthetics, benefiting patients by restoring oral function and aesthetics in edentulous areas.
Clinical Evidence
Bench testing only. Evidence includes fatigue testing per ISO 14801:2016, biocompatibility testing per ISO 10993-1 and ISO 10993-5, endotoxin testing per USP <85>/<161>, and sterilization validation per ISO 11137/ISO 17665. MRI safety was evaluated via scientific rationale and literature review for magnetically induced displacement force and torque.
Technological Characteristics
Endosseous threaded implants and abutments made of Pure Titanium Grade 4 (ASTM F67) and Titanium Alloy (ASTM F136). Features internal conical/hex connections, tapered body, and self-tapping cutting edges. Surface treatment: alumina sandblasting and acid etching. Sterilization: Gamma radiation. Connectivity: Mechanical interface for dental prosthetics.
Indications for Use
Indicated for partially or fully edentulous patients requiring single or multiple-unit restorations in the mandible or maxilla, including cemented retained restorations and fixed bridgework support. Contraindications not specified.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
- IH Implant System (K153521)
Reference Devices
- ET III SA Fixture (K140934)
- ET III SA Ultra Wide Fixture (K140934)
- ET II SA Fixture (K140934)
- OSSTEM Implant System-Abutment (K161689)
- Multi Angled Abutment (K123755)
Related Devices
- K233896 — JDentalCare Dental Implant System: JDEvolution; JDEvolution plus; JDIcon; JDIcon plus · Jdentalcare Srl · Aug 16, 2024
- K202039 — Honorst Implant System · Medimecca Co., Ltd. · Sep 25, 2020
- K222142 — BR SLA Type Implant System · Biotem Co., Ltd. · Jun 24, 2023
- K170608 — UF(II) Implant System · Dio Corporation · Mar 30, 2018
- K210039 — CORE1 Implant System · Cybermed, Inc. · Jan 7, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Sewon Medix Inc. Jinsoo Lee CEO #29, Sa-sang-ro, 375beon-gil, Sa-sang-gu Sa-sang-gu, Busan 46700 Korea. South
7/7/2023
# Re: K222707
Trade/Device Name: IH Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: June 01, 2023 Received: June 08, 2023
Dear Jinsoo Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Sherrill Lathrop Blitzer
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222707
Device Name IH Implant System
#### Indications for Use (Describe)
IH Implant System is device made of titanium alloy indicated for in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained restorations and terminal or interminal abutment support for fixed bridgework. IH Implant System is for stage surgical procedures. It is intended for delayed loading.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------------|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the alphanumeric string "K222707" in a simple, sans-serif font. The characters are uniformly sized and spaced, creating a clear and legible sequence. The text is presented in black against a plain white background, ensuring high contrast and readability.
Image /page/3/Picture/1 description: The image shows the logo for Sewon Medix Dental Implant System. The logo features a stylized geometric shape in shades of orange and blue on the left. To the right of the shape, the words "SEWON MEDIX" are written in a bold, blue sans-serif font. Above the company name, the words "DENTAL IMPLANT SYSTEM" are written in a smaller, gray sans-serif font.
#### 510(k) Summary
#### Date: July 7, 2023 K22707
# 1. Submitter
SEWONMEDIX Inc. JinSoo, Lee #29, Sa-sang-ro, 375beon-gil, Sa-sang-gu, Busan, Republic of Korea TEL: +82-51-303-1713 FAX: +82-51-303-1714 info@sewonmedix.com
## 2. Device Information
- . Trade Name: IH Implant System
- . Common Name: Dental Implant System
- . Classification Name: Endosseous dental implant
- . Primary Product Code: DZE
- . Secondary Product Code: NHA
- . Panel: Dental
- Regulation Number: 872.3640
- . Device Class: Class II
#### 3. Predicate Device
#### Primary Predicate Device
IH Implant System, SEWONMEDIX Inc., K153521
#### Reference Devices
HIOSSEN Implant System, HIOSSEN Inc., K140934 OSSTEM Implant System-Abutment, OSSTEM Implant Co., Ltd., K161689 Multi Angled Abutment, OSSTEM Implant Co., Ltd. K123755
#### 4. General Description
## 4-1) IH2 SLA Fixture
IH2 SLA Fixtures are endosseous threaded implants available in diameters of 3.7, 4.2, 4.6, 5.1 and 5.95mm. IH2 SLA Fixture is dental implant fixture that is used to replace root part of the tooth on full or partial edentulous patients. The surface treatment was carried out by sand blasting with alumina (A12O3) powder followed by acid etching on the surface of machined titanium in order to provide roughness on the surface that would increase the surface area for the bone contact. The implants are compatible with SEWONMEDIX restorative components featuring the internal conical connection.
| Platform | Fixture Diameter | Height(mm) | | | | | |
|----------|------------------|------------|-----|----|------|----|----|
| | | 7 | 8.5 | 10 | 11.5 | 13 | 15 |
| Mini | Ø3.7 | - | ○ | ○ | ○ | ○ | ○ |
| Regular | Ø3.8 | ● | ○ | ○ | ○ | ○ | ○ |
| | Ø4.2 | ● | ○ | ○ | ○ | ○ | ○ |
| | Ø4.5 | ● | ○ | ○ | ○ | ○ | ○ |
| | Ø5.0 | ● | ○ | ○ | ○ | ○ | ○ |
| | Ø5.95 | ● | ● | ● | ● | ● | - |
< Table a: Identification of subiect device and primary predicate device >
(●: Subject Device / ○: primary predicate device)
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Sewon Medix Dental Implant System. The logo features a stylized, geometric design in shades of orange and blue on the left. To the right of the design is the company name, "SEWONMEDIX," in bold, blue capital letters, with the words "DENTAL IMPLANT SYSTEM" in smaller, gray letters above the company name.
### 4-2) IH Healing Abutment
IH Healing Abutments is premanufactured dental implant abutments which can be directly connected to the endosseous dental implant to support healing of the surrounding soft tissue. The following table summarizes the matching of various healing abutments and Abutment Gingival height, including key information of each type of IH Healing Abutment according to the connection type.
| Diameter | Height(mm) | | | | | | | | |
|----------|------------|-----|-----|-----|-----|------|------|------|------|
| | 6.5 | 7.5 | 8.4 | 8.5 | 9.5 | 10.4 | 10.5 | 11.5 | 12.4 |
| Ø4 | ● | ○ | ○ | ● | ○ | ○ | ● | ○ | ○ |
| Ø4.5 | ● | ○ | ○ | ● | ○ | ○ | ● | ○ | ○ |
| Ø5 | ● | ○ | ○ | ● | ○ | ○ | ● | ○ | ○ |
| Ø5.5 | ● | ○ | ○ | ● | ○ | ○ | ● | ○ | ○ |
| Ø6 | ● | ○ | ○ | ● | ○ | ○ | ● | ○ | ○ |
| Ø6.5 | ● | ○ | ○ | ● | ○ | ○ | ● | ○ | ○ |
| Ø9 | ● | ● | ● | ● | ● | ● | ● | ● | ● |
#### < Table b: Identification of subject device and primary predicate device >
: Subject Device / ○: primary predicate device)
#### 4-3) IH Cement Abutment
A Premanufactured dental implant abutment directly connected to the endosseous dental implant intended for use as an aid in prosthetic rehabilitation.
Cement retained restoration: IH Cement Abutment
IH Prosthetic System is device made of titanium alloy intended for use as an aid in prosthetic restoration. Its surfaces are partially TiN-Coated or uncoated.
| | | Height(mm) | | |
|----------|----------|------------|-----|---|
| Platform | Diameter | 4 | 5.5 | 7 |
| Mini | Ø4.5 | - | O | ● |
| Regular | Ø4.5 | ● | O | ● |
| | Ø5 | O | O | O |
| | Ø5.5 | O | O | O |
| | Ø6 | O | O | O |
| | Ø6.5 | O | O | O |
< Table c: Identification of subject device and primary predicate device >
(●: Subject Device / ○: primary predicate device)
#### 4-4) IH Multi-unit Abutment
A pre-manufactured dental implant abutment to be directly connected to an endosseous dental implant intended for use as an aid in prosthetic rehabilitation.
The IH Multi-unit Abutment is made of titanium alloy.
< Table d: Identification of subject device and primary predicate device >
Straight type ।
| Connect Diameter | Gingival Height(mm) | | | |
|------------------|---------------------|-----|-----|-----|
| | 1.5 | 2.5 | 3.5 | 4.5 |
| Ø 2.88 | ● | ● | ● | ● |
| Ø 3.43 | ○ | ○ | ○ | ○ |
Angled type =
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Sewonmedix Dental Implant System. The logo features a stylized flower-like graphic in shades of orange and blue on the left. To the right of the graphic is the company name "SEWONMEDIX" in bold, blue letters, with the words "DENTAL IMPLANT SYSTEM" in smaller, gray letters above the company name.
#29, Sa-sang-ro, 375beon-gil, Sa-sang-gu, Busan, Korea Tel : +82.51.303.1713/ Fax : +82.51.303.1714 / www.sewonmedix.com
| Connect Diameter | Angle(°) | Gingival Height(mm) | | | | | |
|------------------|----------|---------------------|---|-----|---|-----|---|
| | | 2.5 | 3 | 3.5 | 4 | 4.5 | 5 |
| Ø 2.88 | 17 | ● | ● | - | ● | - | - |
| Ø 2.88 | 30 | - | - | ● | ● | - | ● |
| Ø 3.43 | 17 | ○ | ● | ○ | ● | - | - |
| Ø 3.43 | 30 | - | - | ○ | ● | ○ | ● |
(●: Subject Device / ○: primary predicate device)
### 5. Indications for Use
IH Implant System is device made of titanium alloy indicated for in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained restorations and terminal or interminal abutment support for fixed bridgework. IH Implant System is for stage surgical procedures. It is intended for delayed loading.
#### 6. Substantial Equivalence
#### ● Substantial Equivalence (Fixture System)
| Item | Subject device | Primary predicate device | Reference device |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | ET III SA Fixture / ET III SA<br>Ultra Wide Fixture / ET II SA |
| Device name | IH2 SLA Fixture | IH2 SLA Fixture | Fixture |
| | | | |
| 510(k) No. | N/A | K153521 | K140934 |
| Design | Image: design of IH2 SLA Fixture | Image: design of IH2 SLA Fixture | Image: design of ET III SA Fixture / ET III SA Ultra Wide Fixture / ET II SA Fixture |
| | IH Implant System is device | IH Implant System is device | The HIOSSEN Implant System is |
| Indication for Use | made of titanium and titanium<br>alloy indicated for in partially or<br>fully edentulous mandibles and<br>maxillae, in support of single or<br>multiple-unit restorations<br>including; cemented retained or<br>screw retained restorations and<br>terminal or interminal abutment<br>support for fixed bridgework. IH<br>Implant System is for single and<br>two stage surgical procedures. It<br>is intended for delayed loading. | made of titanium and titanium<br>alloy indicated for in partially or<br>fully edentulous mandibles and<br>maxillae, in support of single or<br>multiple-unit restorations<br>including; cemented retained or<br>screw retained restorations and<br>terminal or interminal abutment<br>support for fixed bridgework. IH<br>Implant System is for single and<br>two stage surgical procedures. It<br>is intended for delayed loading. | indicated for use in partially or<br>fully edentulous mandibles and<br>maxillae, in support of single or<br>multiple-unit restorations<br>including; cemented retained,<br>screw retained, or overdenture<br>restorations, and final or<br>temporary abutment support for<br>fixed bridgework. It is intended<br>for delayed loading. ETIII SA<br>Ultra-Wide Fixture is intended to<br>be used in the molar region. |
| Surgery Type | One or two stage Surgery | One or two stage Surgery | One or two stage Surgery |
| Structure | - Internal Hex connected<br>- Submerged fixture<br>- Tapered body shape<br>- Cutting edge with self-tapping<br>- 0.8mm thread pitch | - Internal Hex connected<br>- Submerged fixture<br>- Tapered body shape<br>- Cutting edge with self-tapping<br>- 0.8mm thread pitch | - Internal Hex connected<br>- Submerged fixture<br>- Tapered body shape & straight<br>body shape |
| Diameter (D) | 3.7, 4.2, 4.6, 5.1, 5.95 | 3.7, 4.2, 4.6, 5.1 | 3.5, 3.75, 3.77, 4.2, 4.25, 4.45,<br>4.6, 4.25, 4.63, 4.65, 4.9, 5.0,<br>5.05, 5.08, 5.1, 5.92, 5.95, 6, 6.8 |
| Length (mm) | 7.0, 8.5, 10, 11.5, 13, 15 | 8.5, 10, 11.5, 13, 15 | 6.2, 7.2, 8.7, 10.2, 11.7, 13.2,<br>15.2, 18.2 |
| Material of<br>Fixture | Pure Titanium Grade 4<br>(ASTMF67-06) | Pure Titanium Grade 4<br>(ASTMF67-06) | Pure Titanium Grade 4<br>(ASTMF67-06) |
{6}------------------------------------------------
K222707
Image /page/6/Picture/1 description: The image shows the logo for Sewon Medix Dental Implant System. The logo features a stylized geometric shape in orange and blue on the left, followed by the company name "SEWONMEDIX" in bold, blue letters. Above the company name, the words "DENTAL IMPLANT SYSTEM" are written in smaller, gray font.
## SEWONMEDIX Inc.
#29, Sa-sang-ro, 375beon-gil, Sa-sang-gu, Busan, Korea Tel : +82.51.303.1713/ Fax : +82.51.303.1714 / www.sewonmedix.com
| Surface | SLA | SLA | SLA |
|---------------|-------------------|-------------------|-------------------|
| Sterilization | Radiation Sterile | Radiation Sterile | Radiation Sterile |
| Shelf life | 5 years | 5 years | 8 years |
As previously cleared in K153521 except with added new models of diameter 5.95 or length 7.0, new packaging types (IH2 Fixture with Pre-Mount type). The subject device has the same indications for use as the primary predicate device (K153521) and there is no difference that would impact substantial equivalence.
| Item | Subject device | Primary predicate device | Reference device |
|-----------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| Device name | IH Cover Screw | IH Cover Screw | Cover Screw |
| 510(k) No. | N/A | K153521 | K140934 |
| Design | Image: Subject device | Image: Primary predicate device | Image: Reference device |
| Device<br>Description | Used to protect the exposed<br>platform of the implant during<br>healing period. | Used to protect the exposed<br>platform of the implant during<br>healing period. | Used to protect the exposed<br>platform of the implant during<br>healing period. |
| Diameter (D) | 3.13, 3.6 | 3.13/3.6 | 3.03, 3.58, 3.25, 3.4, 3.75, 3.9 |
| Length (mm) | 5/5.2/5.8 | 5/5.8 | 5.25, 5.9, 6.25, 6.85, 6.9, 7.5 |
| Material | Pure Titanium Gr. 4<br>(ASTMF67) | Pure Titanium Gr. 4<br>(ASTMF67) | Pure Titanium (ASTM F 67) |
| Surface | Anodizing | Anodizing | Anodizing |
As previously cleared in K153521 except with added new models of length 5.2mm.
#### ● Substantial Equivalence (IH Prosthetic System)
| Item | Subject device | Primary predicate device |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | IH Healing Abutment | IH Healing Abutment |
| 510(k) No. | N/A | K153521 |
| Design | Image: Subject device design | Image: Primary predicate device design |
| Description | IH Healing Abutment is used to make a soft<br>tissue shape before setting up prosthetics and<br>removing cover screw after osseointergration. | IH Healing Abutment is used to make a soft<br>tissue shape before setting up prosthetics and<br>removing cover screw after osseointergration. |
| Connection Type | Screw Connected | Screw Connected |
| Diameter (D) | 4/4.5/5/5.5/6/6.5/9 | 4/4.5/5/5.5/6/6.5 |
| Height(mm) | 6.5/7.5/8.4/8.5/9.5/10.4/10.5/11.5/12.4 | 7.5/8.4/9.5/10.4/11.5/12.4 |
| Material | Titanium Alloy (ASTM F 136) | Titanium Alloy (ASTM F 136) |
| Surface | None | None |
As previously cleared in K153521 except with added new models of diameter 9 and height 6.5, 8.5, 10.5, 11.5mm.
| Item | Subject device | Primary predicate device |
|------|----------------|--------------------------|
|------|----------------|--------------------------|
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Sewon Medix Dental Implant System. The logo features a stylized, segmented circular design in shades of orange and blue on the left. To the right of the design is the company name "SEWON MEDIX" in bold, blue sans-serif font. Above the company name, in smaller gray font, are the words "DENTAL IMPLANT SYSTEM."
#29, Sa-sang-ro, 375beon-gil, Sa-sang-gu, Busan, Korea Tel:+82.51.303.1713/ Fax:+82.51.303.1714 / www.sewonmedix.com
| Device name | IH Cement Abutment | IH Cement Abutment |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| 510(k) No. | NA | K153521 |
| Design | Image: IH Cement Abutment | Image: IH Cement Abutment |
| Description | A Premanufactured dental implant abutment<br>directly connected to the endosseous dental<br>implant intended for use as an aid in prosthetic<br>rehabilitation.<br>Cement retained restoration: IH Cement<br>Abutment<br>IH Prosthetic System is device made of<br>titanium, titanium alloy intended for use as an<br>aid in prosthetic restoration. Its surfaces are<br>partially TiN-Coated or uncoated. | IH Cement Abutment to fabricate a prosthesis of<br>internal single & bridge cement retained type. |
| Connection Type | External Hex-Connected | External Hex-Connected |
| Diameter (D) | 4.5/5/5.5/6/6.5 | 4.5/5/5.5/6/6.5 |
| G/H Length (mm) | 1/1.5/2/2.5/3/3.5/4/4.5/5 | 1/1.5/2/2.5/3/3.5/4/4.5/5 |
| Post Length (mm) | 4/5.5/7 | 4/5.5/7 |
| Material | Titanium Alloy (ASTM F 136) | Titanium Alloy (ASTM F 136) |
| Surface | Partial TiN coated | Partial TiN coated |
As previously cleared in K153521 except with added new models of diameter 4.5 x G/H x 4 and 7 mm.
| Item | Subject device | Primary predicate device | Reference device |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Device name | IH Multi-unit Abutment<br>(Straight type) | IH Multi-unit Abutment<br>(Straight type) | Multi Abutment |
| 510(k) No. | N/A | K153521 | K161689 |
| Design | Image: Subject device | Image: Primary predicate device | Image: Reference device |
| Description | A pre-manufactured dental<br>implant abutment to be directly<br>connected to an endosseous<br>dental implant intended for use as<br>an aid in prosthetic rehabilitation.<br>The IH Multi-unit Abutment is<br>made of titanium alloy. | IH Multi-Unit Abutment is used<br>for screw retained multiple case. | Multi Abutment is used for<br>edentulous mandible or maxilla<br>to make<br>full denture. |
| Connection<br>Type | Screw & External Hex-<br>Connected | Screw & External Hex-<br>Connected | Screw & External Hex-<br>Connected |
| Diameter (D) | 4.8 | 4.8 | 4.8 |
| Connection<br>Diameter (Ø) | 2.88, 3.43 | 3.43 | 2.88, 3.43 |
| G/H Length<br>(mm) | 1.5/2.5/3.5/4.5 | 1.5/2.5/3.5/4.5 | 1/2/3/4/5 |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for SEWONMEDIX Dental Implant System. The logo features a geometric design on the left, composed of orange and blue triangles arranged in a circular pattern. To the right of the geometric design, the text "SEWONMEDIX" is displayed in a bold, blue sans-serif font. Above the company name, the words "DENTAL IMPLANT SYSTEM" are written in a smaller, gray font.
#29, Sa-sang-ro, 375beon-gil, Sa-sang-gu, Busan, Korea Tel : +82.51.303.1713/ Fax : +82.51.303.1714 / www.sewonmedix.com
| Material | Titanium Alloy (ASTM F 136) | Titanium Alloy (ASTM F 136) | Titanium Alloy (ASTM F 136) |
|----------|-----------------------------|-----------------------------|-----------------------------|
| Surface | None | None | Partial TiN coated |
IH Multi-unit Abutment (Straight, Angled) is as previously cleared in K153521 except added New models of angled type of gingival height 3, 4, 5 and Mini type.
| Item | Subject device | Primary predicate device | Reference device |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Device name | IH Multi-unit Abutment<br>(Angled type) | IH Multi-unit Abutment<br>(Angled type) | Multi Angled Abutment |
| 510(k) No. | N/A | K153521 | K123755 |
| Design | Image: IH Multi-unit Abutment (Angled type) | Image: IH Multi-unit Abutment (Angled type) | Image: Multi Angled Abutment |
| Description | A pre-manufactured dental<br>implant abutment to be directly<br>connected to an endosseous<br>dental implant intended for use as<br>an aid in prosthetic rehabilitation.<br>The IH Multi-unit Abutment is<br>made of titanium alloy. | IH Multi-Unit Abutment is used<br>for screw retained multiple case. | The Multi Angled Abutment is<br>device made of titanium alloy<br>intended for use as an aid in<br>prostrhtic restoration. |
| Connection<br>Type | Screw & External Hex-<br>Connected | Screw & External Hex-<br>Connected | Screw & External Hex-<br>Connected |
| Diameter (Ø) | 4.8 | 4.8 | 4.8 |
| Connection<br>Diameter (Ø) | 2.88, 3.43 | 3.43 | 2.88, 3.43 |
| G/H Length<br>(mm) | 2.5/3/3.5/4/4.5/5 | 2.5/3.5/4.5 | 2.5/3/3.5/4/4.5/5 |
| Angle(°) | 17, 30 | 17, 30 | 17, 30 |
| Material | Titanium Alloy (ASTM F 136) | Titanium Alloy (ASTM F 136) | Titanium Alloy (ASTM F 136) |
| Surface | None | Partial TiN coated | None |
IH Multi-unit Abutment (Straight, Angled) is as previously cleared in K153521 except added New models of angled type of gingival height 3, 4, 5
#### . Substantial Equivalence Review:
Subject devices in this submission are substantially equivalent to the identified prior clearances , therefore, manufacturing process including raw material, machining, cleaning and similar design and technological characteristics as the primary predicate device (K153521).
| Type | Product name | Identity |
|-------------------|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Fixture<br>System | IH2 SLA Fixture | - It was cleared under K153521<br>- New models of diameter 5.95 or length 7.0 are added.<br>- New packaging types (IH2 Fixture with Pre-Mount type) is added. |
| | IH Cover Screw | It was cleared under K153521 except with added new models of length<br>5.2mm. |
Refer to the difference of additional products for detail as below table.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for SEWONMEDIX Dental Implant System. The logo features a stylized, geometric shape in blue and orange on the left. To the right of the shape, the words "SEWONMEDIX" are written in large, bold, blue letters. Above the company name, the words "DENTAL IMPLANT SYSTEM" are written in smaller, gray letters.
#29, Sa-sang-ro, 375beon-qil, Sa-sang-gu, Busan, Korea Tel : +82.51.303.1713/ Fax : +82.51.303.1714 / www.sewonmedix.com
| IH<br>Prosthetic<br>System | IH Healing Abutment | - It was cleared under K153521<br>- New models of diameter 9 and height 6.5, 8.5, 10.5, 11.5mm is added. |
|----------------------------|----------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| | IH Cement Abutment | - It was cleared under K153521<br>- New model with a diameter of 4.5 x G/H x 4 &7mm is added. |
| | IH Multi-unit Abutment<br>(Straight, Angled) | - It was cleared under K153521<br>- New models of angled type of gingival height 3, 4, 5 and Mini type is added. |
| | IH Multi-unit Abutment<br>Screw | Add length 7.5, 8, 9.1 or diameter 2, 2.05 of existing device |
IH Implant System has been subjected to performance and product validations prior to release. The differences between the subject device and predicate devices are detailed shape and detailed dimension of diameter and length.
#### 8 Summary of non-clinical testing
The subject device was tested to evaluate its substantial equivalence according to the following standards.
- Endotoxin testing on the Subject device or suitable test specimens was performed following USP<85> and ● USP<161> according to the endotoxin sampling plan
- Biological Evaluation of the Subject device was performed according to ISO 10993-1. Cytotoxicity Testing per ISO . 10993-5. Biocompatibility information is also leveraged from primary predicate device(K153521).
- · Fatigue Testing according to ISO 14801:2016 under the worst-case scenario
- · Gamma sterilization validation Test Report according to ISO 11137-1 and ISO 11137-2
- · End User Sterilization Validation Test Report according to ANSVAAMI ST79, ISO 17665-2, ISO 11737-1, ISO 11737-2, ISO 11138-1 and ISO 11138-3
- · Shelf Life Test according to ASTM F1140, ASTM F1929, and ASTM F2096, ASTM F1980 and ISO 11607
The non-clinical testing results demonstrate that the subject device is substantially equivalent to the predicate devices. Non-clinical worst-case MRI review was performed to evaluate the metallic devices in the MRI environment using scientific rationale and published literature (i.e., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system to include all variations (all compatible implant bodies, dental abutments, and fixation screws) and material compositions. The rationale addressed parameters per the FDA Guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment", including magnetically induced displacement force and torque.
#### 9 Conclusion
The comparison of similarities and differences between the proposed devices and the respective predicate devices demonstrate that the subject device is substantially equivalent to the identified predicate devices.