Magicore System

K152520 · Innobiosurg Co., Ltd. · DZE · Jan 25, 2016 · Dental

Device Facts

Record IDK152520
Device NameMagicore System
ApplicantInnobiosurg Co., Ltd.
Product CodeDZE · Dental
Decision DateJan 25, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Magicore System is intended to replace missing teeth to restore chewing function. The Magicore can be placed in support of single or multiple-unit restorations including; cement retained, or overdenture restorations, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading.

Device Story

Magicore System is an endosseous dental implant system comprising fixtures, abutments, and screws. Fixtures are surgically placed into the jawbone to support dental restorations; used in one or two-stage procedures. System features internal hex and Morse taper bevel connections; surface treated with Resorbable Blasted Media (RBM). Fixtures are provided sterile; abutments and accessories are provided non-sterile and require sterilization by the clinician. Used by dental professionals in clinical settings to restore chewing function in patients with missing teeth. Output is a stable foundation for prosthetic restorations. Benefits include restoration of oral function and aesthetics.

Clinical Evidence

Bench testing only. Biocompatibility evaluation confirmed materials conform to FDA consensus standards. Sterilization validation performed per ISO 11137-1:2006 and ISO 11137-2:2006. Shelf-life validated via real-time aging per ASTM 1980.

Technological Characteristics

Materials: Ti-6Al-4V ELI (ASTM F-136). Surface: RBM (Resorbable Blasted Media). Connection: Internal hex and Morse taper bevel. Fixture diameters: 4.0-6.5mm; lengths: 7-13mm. Sterilization: Gamma radiation. Components: Fixtures, abutments, healing caps, screws. Abutment cap material: PolyDiacetate.

Indications for Use

Indicated for patients with missing teeth requiring replacement to restore chewing function. Supports single or multiple-unit restorations, including cement-retained or overdenture restorations, and terminal or immediate abutment support for fixed bridgework. For one or two-stage surgical procedures; intended for delayed loading.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a representation of the human form, composed of three interconnected profiles. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 25, 2016 InnoBioSurg Co., Ltd. c/o Ms. April Lee Consultant WithUS Group Inc. 2531 Pepperdale Drive Rowland Heights, California 91748 Re: K152520 Trade/Device Name: Magicore System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: December 16, 2015 Received: December 22, 2015 #### Dear Ms. Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Page 2 - Ms. April Lee Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Tina Kiang -S for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K152520 Device Name Magicore System Indications for Use (Describe) The Magicore System is intended to replace missing teeth to restore chewing function. The Magicore can be placed in support of single or multiple-unit restorations including; cement retained, or overdenture restorations, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for IBS Implant Innobiosurg Co., Ltd. The logo features a stylized letter "G" in blue with a yellow accent at the top left. Next to the "G" is the text "IBS" in bold, blue letters, followed by "implant" in a smaller font. Below "implant" is the text "Innobiosurg Co.,Ltd" in a smaller, lighter font. # InnoBioSt 44-19. Techno 10-ro. Yuseong-gu. Daeieon. 305-510. Korea Tel. 82- 42-933-2879 / Fax. 82-42-933-2881 ## 510(k) Summary #### Submitter InnoBioSurg Co., Ltd. Bo-reumYoo 44-19, Techno 10-ro, Yuseong-gu, Daejeon, 305-510 Republic of Korea Email: bryoo@ibsimplant.com Tel.+82-42-933-2879 Fax.+82-42-933-2881 # Official Correspondent WithUS Group Inc April Lee 2531 Pepperdale Drive Rowland Heights, CA 91748 USA Email:withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122 ## Device Information - Trade Name: Magicore System - Common Name: Dental Implant System - Classification Name: Endosseous dental implant - Primary Product Code: DZE - Secondary Product Code: NHA - . Panel: Dental - Regulation Number: 872.3640 - Device Class: Class II - Date prepared: 1/22/2016 ## General Description An endosseous dental implant is a device made of a material such as Ti 6AL 4V Eli(Conforming to ASTM Standard F-136). The Magicore System is consists of dental implants, abutments, and screws for use in one or two-stage dental implant placement and restorations. The implant-abutment connection is tight and precise fitting with internal hex and Morse taper bevel. The surface of the system has been treated with RBM (Resorbable Blasted media). The Fixture diameters are 4, 4.5, 5, 5, 6, 6.5mm and lengths are 7, 9, 11, 13mm in this system. The contained various abutments and accessories in the system are abutment screw, magicore abutment &cap, healing cap (color anodizing: yellow, blue, green, and purple). The abutments are provided straight only and that they are no intended to be modified to provide an angle correction. Fixtures and abutments are packaged separately. The Fixtures are supplied sterile and the abutments and accessories are provided non-sterile. The abutments and accessories should be sterilized before use. ## Indication for Use The Magicore System is intended to replace missing teeth to restore chewing function. The Magicore can be placed in support of single or multiple-unit restorations including; cement retained, or overdenture restorations, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures and not for immediate loading. This system is intended for delayed loading. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for IBS implant. The logo consists of the letters "IBS" in a bold, dark blue font, with a yellow dot above the "I". To the right of "IBS" is the word "implant" in a smaller, dark blue font. Below "implant" is the text "Innobiosurg Co.,Ltd" in a smaller, light yellow font. ## Materials: The dental implants are fabricated from Ti 6AL 4V Eli (Conforming to ASTM Standard F-136). # Non-Clinical Data: Non-clinical tests followed the recommendations in the "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implant and Endosseous Dental Implant abutments". Sterilization Validation testing performed in accordance with ISO 11137-1:2006, ISO 11137-2:2006 and surface treatment analysis were used to support the decision of substantial equivalence. A biocompatibility evaluation was conducted and since all the materials were used in the company's own predicate or conform to an FDA consensus standard, no additional biocompatibility testing was deemed necessary. Shelf-life validation was conducted on the packaging of the sterile device according to ASTM 1980 to validate the 5 year shelf life via real-time aging. ## Predicate Device: ## 1) Comparison to Predicate Devices: (implant) - . K073247, US/SS/GS SYSTEM manufactured by OSSETEM Implant Co., Ltd. - (Reference Device) K140806, IBS Implant System manufactured by Innobiosurg Co., Ltd. | | Subject device | Predicate device | Reference device | |---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Name | Magicore System | US/SS/GS SYSTEM | IBS Implant System | | 510(k)<br>Product code<br>Class | K152520<br>DZE,NHA<br>II | K073247<br>DZE<br>II | K140806<br>DZE/NHA<br>II | | Manufacturer | InnoBioSurg Co., Ltd. | OSSETEM Co., Ltd. | InnoBioSurg Co., Ltd. | | Indication for<br>use | The Magicore System<br>is intended to replace<br>missing teeth to restore<br>chewing function. The<br>Magicore can be placed<br>in support of single or<br>multiple-unit<br>restorations including; | Intended for use in partially<br>or fully edentulous<br>mandible sand maxillae, in<br>support of single or<br>multiple-unit restorations<br>including; cement-<br>retained, screw- retained,<br>oroverdenture restorations | Intended to replace missing<br>teeth to restore chewing<br>function. The IBS Implant<br>System can be placed in<br>support of single or multiple-<br>unit restorations including;<br>cement retained, screw- | | | cement retained, screw<br>retained, or overdenture<br>restorations, and terminal<br>or immediate abutment<br>support for fixed<br>bridgework. This system is<br>for one or two stage<br>surgical procedures. This<br>system is intended for<br>delayed loading. | And terminal or<br>intermediate<br>Abutment support for<br>fixed bridgework. This<br>system is dedicated<br>For one and two stage<br>surgical procedures and not<br>dedicated<br>For immediate loading.<br>This system is intended for<br>Delayed<br>loading. | retained, or<br>overdenture restorations, and<br>terminal or immediate<br>abutment support for fixed<br>bridgework. This system is<br>for one or two stage surgical<br>procedures and not for<br>immediate loading. This<br>system is intended for delayed<br>loading. | | Design | Image: Implant 1 | Image: Implant 2 | Image: Implant 3 | | Composition<br>of Material | Titanium Alloy<br>Ti-6Al-4V Eli<br>ASTM F136 | Titanium Alloy<br>Ti-6A1-4V Eli<br>ASTM F136 | Titanium Alloy<br>Ti-6A1-4V Eli<br>ASTM F136 | | Neck Length | 1,2,3mm | 1.8mm, 2.8mm | | | Fixture Type | NON Submerged | NON Submerged | Submerged | | Endosseous<br>Implant | Tapered,<br>macro threads | Tapered,<br>Macro threads | Tapered, macro and<br>micro threads | | Components | Various abutments<br>and accessories | Various abutments<br>and accessories | Various abutments<br>and accessories | | Range of<br>Diameters<br>(mm) | 4, 4.5, 5, 5.5, 6,<br>6.5mm | 4.0 - 6.0mm | 3.8, 4.3, 4.8, 5.3, 5.8,6.3 mm | | Range of<br>Lengths(mm) | 7 mm - 13 mm | 7 mm - 13 mm | 7 mm – 15 mm | | Modified<br>Surface | R.B.M. | R.B.M. | R.B.M. | | Surgical<br>Technique | 1 stage and 2 stage,<br>self tapping | 1 stage and 2 stage,<br>self tapping | 1 stage and 2 stage, self<br>tapping | | Gamma<br>Sterilization | Yes | Yes | Yes | | Body Contact | Direct | Direct | Direct | | Contact<br>Duration | Class C (>30days) | Class C (>30days) | Class C (>30days) | | | Subject device | Predicate device | Reference device | | Product Name | Magicore System | US/SS/GS SYSTEM | IBS Implant System | | 510(k)<br>Product code<br>Class | K152520<br>DZE/NHA<br>II | K073247<br>DZE<br>II | K140806<br>DZE/NHA<br>II | | Manufacturer | InnoBioSurg Co., Ltd. | OSSETEM Co., Ltd. | InnoBioSurg Co., Ltd. | | Indication for<br>use | The Magicore System is<br>intended to replace<br>missing teeth to restore<br>chewing function. The<br>Magicore can be placed<br>in support of single or<br>multiple-unit<br>restorations including;<br>cement retained, screw<br>retained, or overdenture<br>restorations, and<br>terminal or immediate<br>abutment support for<br>fixed bridgework. This<br>system is for one or two<br>stage surgical<br>procedures. This system<br>is intended for delayed<br>loading. | Intended for use in partially<br>or fully edentulous<br>mandibles and maxillae, in<br>support of single or<br>multiple-unit restorations<br>including ;cement-<br>retained, screw- retained,<br>or overdenture restorations,<br>and terminal or<br>intermediate abutment<br>support for fixed<br>bridgework. This system is<br>dedicated<br>For one and two stage<br>surgical procedures and not<br>dedicated for immediate<br>loading. This system is<br>intended for delayed loading. | Intended to replace missing<br>teeth to restore chewing<br>function. The IBS Implant<br>System can be placed in<br>support of single or multiple-<br>unit restorations including:<br>cement retained, screw<br>retained, or overdenture<br>restorations, and terminal or<br>immediate abutment support<br>for fixed bridgework. This<br>system is for one or two stage<br>surgical procedures and not<br>for immediate loading. This<br>system is intended for delayed<br>loading. | | Abutment<br>design | Image: Abutment design | Image: Abutment design | Image: Abutment design | | Diameter | 3.5, 3.86, 4.3, 4.6mm | | 4.0, 4.5, 5.0, 5.5, 6.0,<br>6.5mm | | length | 9.6, 10.6, 11.6mm | | 10.6, 11.6, 12.6, 13.6, 14.6,<br>15.6mm | | Healing cap<br>Design | Image: Healing cap design 1 | Image: Healing cap design 2 | Image: Healing cap design 3 | | Diameter | 5.5mm | | 4.0, 4.5, 5.0, 5.5, 6.0mm | | length | 7.3mm | | 9.1, 10.1, 11.1, 12.1, 13.1,<br>14.1mm | | Abutment cap<br>design | Image: Abutment cap design 1 | Image: Abutment cap design 2 | Image: Abutment cap design 3 | | diameter | 5.5, 6.0, 6.5, 7.0mm | | 4.5, 5.0, 5.5, 6.0, 6.5,<br>7.0mm | | length | 5.5, 6.5, 7.5, 8.5,<br>9.5mm | | 5.1, 6.6, 8.1mm | | Closing screw<br>design | Image: Closing screw design 1 | Image: Closing screw design 2 | Image: Closing screw design 3 | | diameter | 3.5mm | | 3.4mm | | length | 5.65mm | | 5.65mm | | Composition<br>of Material | Titanium Alloy<br>Ti-6Al-4V Eli<br>ASTM F136 | Titanium Alloy<br>Ti-6Al-4V Eli<br>ASTM F136 | Titanium Alloy<br>Ti-6Al-4V Eli<br>ASTM F136 | | Materials of<br>Magicore<br>Abutment<br>Cap | PolyDiacetate | Polycarbonate polymer | PolyDiacetate | | Contact<br>Duration | Class C (>30days) | Class C (>30days) | Class C (>30days) | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for IBS implant. The logo consists of the letters "IBS" in a bold, sans-serif font, followed by the word "implant" in a smaller, italicized font. Below the word "implant" is the text "Innobiosurg Co.,Ltd" in a smaller font. The logo is primarily blue, with a small yellow accent on the "I". InnoBioSurg Co., Ltd. 44-19, Techno 10-ro, Yuseong-gu, Daejeon, 305-510, Korea 11 Tel. 82-42-933-2879 / Fax. 82-42-933-2881 {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for IBS implant. The logo consists of the letters "IBS" in a bold, sans-serif font, with a small yellow circle above the "I" and a blue square around the "I". To the right of "IBS" is the word "implant" in a smaller, sans-serif font. Below "implant" is the text "Innobiosurg Co.,Ltd" in a smaller, lighter font. InnoBioSurg Co., Ltd. 44-19, Techno 10-ro, Yuseong-gu, Daejeon, 305-510, Korea 14-19, Tel. 82- 42-933-2879 / Fax. 82-42-933-2881 # 2) Comparison chart for abutments {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for IBS implant. The logo consists of a stylized letter "i" and "G" in blue, followed by the text "IBS implant" in a bold, dark blue font. Below the word "implant" is the text "Innobiosurg Co.,Ltd" in a smaller, lighter font. InnoBioSurg Co., Ltd. 44-19, Techno 10-ro, Yuseong-gu, Daejeon, 305-510, Korea 11 Tel. 82-42-933-2879 / Fax. 82-42-933-2881 {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for IBS implant. The logo consists of the letters "IBS" in a bold, sans-serif font, with the "I" having a yellow square in the upper left corner. To the right of "IBS" is the word "implant" in a smaller, sans-serif font. Below "implant" is the text "Innobiosurg Co.,Ltd" in an even smaller, sans-serif font. #### Substantial Equivalence Discussion The indications for use, material, general shape design, surface, connection to abutment method, application method and sterilization method between the Magicore System and the predicate devices are similar. The differences between them are the external shape of the fixture and fixture diameters. The subject device and primary predicate have slightly different Indications for Use language. However, the difference in language does not change the intended use or substantial equivalence. Any differences between the Mazicore System and predicate device do not raise new types of safety or effectiveness issues. Accordingly we can claim the substantially equivalence of the Magicore System to predicate device. #### Conclusion The Magicore System, subject device of this submission, constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental scientific technology as its predicate devices. Therefore, Magicore System and its predicates are substantially equivalent.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...