The Magicore II System is intended to replace missing teeth to restore chewing function. The Magicore II System can be placed in support of single or multiple-unit restorations including; cement retained, or overdenture restorations, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading.
Device Story
Endosseous dental implant system; replaces missing teeth; restores chewing function. Components include fixtures, abutments, cylinders, caps, and screws. Fixtures feature SLA surface treatment (sand-blasted, large-grit, acid-etched) and internal hex/Morse taper connection. Used by dental clinicians in clinical settings for one or two-stage surgical procedures. Fixtures provided sterile; abutments/screws provided non-sterile for end-user sterilization. System supports single/multiple-unit restorations, cement/screw-retained prosthetics, and overdentures. Clinical benefit: provides stable foundation for prosthetic teeth, restoring oral function.
Clinical Evidence
Bench testing only. Fatigue testing performed per ISO 14801:2016. Biocompatibility testing conducted per ISO 10993 series. Sterilization validation performed per ANSI/AAMI ST79 and relevant ISO standards. Surface treatment analysis provided to support substantial equivalence.
Indicated for patients requiring replacement of missing teeth to restore chewing function via single or multiple-unit restorations (cement-retained, screw-retained, or overdenture) and fixed bridgework. Suitable for one or two-stage surgical procedures with delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 28, 2020
InnoBioSurg Co., Ltd. % April Lee Consultant Withus Group Inc 106 Superior Irvine, California 92620
Re: K192197
Trade/Device Name: Magicore II System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: April 23, 2020 Received: April 27, 2020
Dear April Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192197
Device Name Magicore II System
#### Indications for Use (Describe)
The Magicore II System is intended to replace missing teeth to restore chewing function. The Magicore II System can be placed in support of single or multiple-unit restorations including; cement retained, or overdenture restorations, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;"> <svg height="10" width="10"> <rect fill="lightgray" height="10" stroke="black" width="10" x="0" y="0"></rect> </svg> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <svg height="10" width="10"> <rect fill="white" height="10" stroke="black" width="10" x="0" y="0"></rect> </svg> Over The Counter Use (21 CFR 801)</span> |
|| | Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary K192197
#### Submitter
InnoBioSurg Co., Ltd. Eun-Ji Yoo 44-19 , Techno 10-ro, Yuseong-gu Daejeon, 34027 Republic of Korea Email: ejyoo@ibsimplant.com Tel. +82-42-933-2879 Fax. +82-42-933-2881
### Device Information
### Official Correspondent
Withus Group Inc April Lee 106 Superior Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122
- Trade Name: Magicore II System
- Common Name: Endosseous Dental Implant
- Classification Name: Implant, Endosseous, Root-Form
- Product Code: DZE
- . Secondary Product Code: NHA
- Panel: Dental
- . Regulation Number: 21 CFR 872.3640
- Device Class: Class II
- . Date prepared: April 26, 2020
# Predicate Devices:
The subject device is substantially equivalent to the following predicate devices:
### Primary Predicate
K152520, Magicore System manufactured by InnoBioSurg Co., Ltd.
### Reference Device
K072570, Nobelactive Multi Unit Abutment by Nobel Biocare AB K080594, MS System (Narrow Ridge) by Osstem Implant Co., Ltd.
K150344, Dentis Dental Implant System by Dentis Co., Ltd.
K153350. IBS Implant System manufactured by InnoBioSurg Co., Ltd.
K160670, ET US SS Prosthetic system by Osstem Implant Co., Ltd.
K161689, OSSTEM Implant System - Abutment by Osstem Implant Co., Ltd.
K162099, IBS Implant System II manufactured by InnoBioSurg Co., Ltd.
K171027, Dentis Dental Implant System by Dentis Co., Ltd.
K173120, CCM Abutment System manufactured by InnoBioSurg Co., Ltd.
K173575. OsteoReady Dental Implant System by OsteoReady LLC.
{5}------------------------------------------------
### K192197
### Indication for Use:
The Magicore II System is intended to replace missing teeth to restore chewing function. The Magicore II System can be placed in support of single or multiple-unit restorations including; cement retained, or overdenture restorations, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading.
### Device Description:
An endosseous dental implant is a device made of a material such as Ti 6AL 4V Eli (Conforming to ASTM Standard F-136). The Magicore II System consists of dental implants, cylinders, caps and screws for use in one or two-stage dental implant placement and restorations.
The implant-abutment connection is tight and precise fitting with internal hex and Morse taper bevel. The surface of the Magicore II implant is treated with SLA (sand-blasted, large-grit, acid-etched). The Fixture diameters are 4.0, 4.5, 5.0, 5.5, 6.0, 6.5mm and threaded lengths are 7, 8, 9, 10, 11, 12, 13mm in this system.
The contained various abutments and screws in the system are Short Abutment, Magic Multi Abutment, Magic Multi Cylinder, Magic Multi Abutment ST, Abutment Screw, Healing Cap, Magicore Angled Abutment.
| Abutments | Diameters (Ø) | Lengths (mm) |
|--------------------------------------------------------------------------------------|----------------------------------------------------------|------------------------------|
| Short Abutment (Hex, Non-Hex) | 3.5, 3.86, 4.3, 4.6 | 4.55, 5.55, 6.55, 7.55, 8.55 |
| Magic Abutment (Hex, Non-Hex) | 4.57, 4.7, 5.07, 5.57, 5.7, 5.87,<br>6.2, 6.37, 6.5, 7.0 | 4.5, 5.5, 6.5, 7.5, 8.5 |
| | 4.27, 5.21, 5.5, 6.0 | |
| Magic Multi Abutment<br>(Screw type – Hex, Non-Hex, Cemented type –<br>Hex, Non-Hex) | 4.8 | 4.6, 5.6, 6.6, 7.6, 8.6 |
| | | 4.6, 5.6, 6.6, 7.6, 8.6 |
| | | 5, 6, 7, 8 |
| | | 5.2, 6.2, 7.2, 8.2 |
| | 5.8 | 4.7, 5.7, 6.7, 7.7, 8.7 |
| | | 5.1, 6.1, 7.1, 8.1 |
| | | 5.3, 6.3, 7.3, 8.3 |
| Magic Multi Cylinder (Hex, Non-Hex, Post) | 5.0, 6.0 | 10 |
| Magic Multi Abutment ST (Hex, Non-Hex, Post) | 5.0, 6.0 | 10 |
| Healing Cap | 5.2, 6.2 | 4.5 |
| Magicore Angled Abutment | 4.3 | 11.2, 12.2, 13.2, 14.2 |
| Abutment Screw | 2 | 5.2, 7.1 |
The dimension ranges of the abutments are below:
Tolerance of dimension for Fixtures and Abutments shall be within ± 1% range.
Fixtures and abutments are packaged separately. The Fixtures are supplied sterile and the abutments and screw are provided non-sterile. The abutments and screw should be sterilized before use.
### Materials:
- Fixtures are fabricated from Ti-6AI-4V Eli (Conforming to ASTM Standard F-136).
- Short Abutment, Magic Abutment, Magic Multi Abutment, Magic Multi Cylinder, Magic Multi Abutment ST, Abutment Screw, Healing Cap, Magicore Angled Abutment are fabricated from Ti-6Al-4V Eli (Conforming to ASTM Standard F-136).
{6}------------------------------------------------
# Summaries of Technology Characteristics:
# 1) Fixture
| | Subject Device | Primary Predicate | Reference Device | Reference Device | | Subject Device | Reference Device | Reference Device | | Subject Device | Reference Device | |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Product Name | Magicore II System | Magicore System | IBS Implant System II | MS System (Narrow Ridge) | Company | InnoBioSurg Co., Ltd. | Dentis Co., Ltd. | OsteoReady LLC | Company | InnoBioSurg Co., Ltd. | Osstem Co., Ltd. | |
| 510(k)<br>Product code<br>Class | N/A<br>DZE<br>II | K152520<br>DZE<br>II | K162099<br>DZE<br>II | K080594<br>DZE<br>II | Device Name | Magicore II System | Dentis Dental Implant System | OsteoReady Dental Implant System | Device Name | Magicore II System | SS System | |
| Manufacturer | InnoBioSurg Co., Ltd | InnoBioSurg Co., Ltd | InnoBioSurg Co., Ltd. | OSSTEM IMPLANT CO.,LTD | 510(k) | N/A | K171027 | K173575 | 510(k) Number | N/A | K160670 | |
| Indications for<br>use | The Magicore II System is<br>intended to replace missing teeth<br>to restore chewing function. The<br>Magicore II System can be placed<br>in support of single or multiple-<br>unit restorations including;<br>cement retained, screw retained,<br>or overdenture restorations, and<br>terminal or immediate abutment<br>support for fixed bridgework.<br>This system is for one or two<br>stage surgical procedures. This<br>system is intended for delayed<br>loading. | The Magicore System is intended<br>to replace missing teeth to restore<br>chewing function. The Magicore<br>can be placed in support of single<br>or multiple-unit restorations<br>including; cement retained, screw<br>retained, or overdenture<br>restorations, and terminal or<br>immediate abutment support for<br>fixed bridgework. This system is<br>for one or two stage surgical<br>procedures. This system is<br>intended for delayed loading. | The IBS Implant System II is<br>intended to replace missing teeth to<br>restore chewing function.<br>The IBS Implant System II can be<br>placed in support of single or<br>multiple-unit restorations including;<br>cement retained, screw retained, or<br>overdenture restorations, and<br>terminal or immediate abutment<br>support for fixed bridgework. This<br>system is for one or two stage<br>surgical procedures. This system is<br>intended for delayed loading. | The MS System (Narrow Ridge) is<br>intended to use in the treatment of<br>missing mandibular central and<br>lateral incisors to support prosthetic<br>device, such as artificial teeth, in<br>order to restore chewing function in<br>partially edentulous patients. MS<br>System (Narrow Ridge) are<br>intended for single use only and not<br>for immediate loading. | Indications<br>for Use | The Magicore II System is intended to<br>replace missing teeth to restore chewing<br>function. The Magicore II System can be<br>placed in support of single or multiple-<br>unit restorations including; cement<br>retained, screw retained, or overdenture<br>restorations, and terminal or immediate<br>abutment support for fixed bridgework.<br>This system is for one or two stage<br>surgical procedures. This system is<br>intended for delayed loading. | The Dentis Dental Implant System is an endosseous<br>dental implant that is indicated for surgical<br>placement in the upper and lower jaw arches, to<br>provide a root form means for single or multiple-<br>units prosthetic appliance attachment to restore a<br>patient's chewing function. Implants can be placed<br>with a conventional two stage surgical process with<br>an option for transmucosal healing or they can be<br>placed in a single stage surgical process for<br>immediate loading when good primary stability has<br>been achieved and with appropriate occlusal loading. | OsteoReady® Dental Implant System is<br>indicated for use in surgical and<br>restorative applications for placement in<br>the bone of the upper or lower jaw to<br>provide support for prosthetic devices,<br>such as artificial teeth, in order to<br>restore the patient's chewing function.<br>RidgeReady® 3.0 implants are intended<br>to replace a lateral incisor in the maxilla<br>and/or a central or lateral incisor in the<br>mandible. | Indications for Use | The Magicore II System is intended to replace missing teeth to<br>restore chewing function. The Magicore II System can be<br>placed in support of single or multiple-unit restorations<br>including; cement retained, screw retained, or overdenture<br>restorations, and terminal or immediate abutment support for<br>fixed bridgework. This system is for one or two stage surgical<br>procedures. This system is intended for delayed loading. | ET System<br>The HIOSSEN Prosthetic system is intended for use with<br>a dental implant to provide support for prosthetic<br>restorations such as crowns, bridges, or over-dentures.<br>US/SS System<br>The OSSTEM Prosthetic system is intended for use with<br>a dental implant to provide support for prosthetic<br>restorations such as crowns, bridges, or over-dentures. | |
| Design | Image: Magicore II System Design | Image: Magicore System Design | Image: IBS Implant System II Design<br>NR Fix<br>Image: IBS Implant System II Design<br>Magic FC | Image: MS System (Narrow Ridge) Design | Material | TI-6AL-4V ELI | Pure Titanium( Grade 4 ) | Ti-6Al-4V-ELI | Material | TI-6AL-4V ELI | TI-6AL-4V ELI | |
| Composition of<br>Material | Titanium Alloy<br>Ti-6Al-4V Eli<br>ASTM F136 | Titanium Alloy<br>Ti-6Al-4V Eli<br>ASTM F136 | Titanium Alloy<br>Ti-6Al-4V Eli<br>ASTM F136 | Titanium Alloy<br>Ti-6Al-4V Eli<br>ASTM F136 | Design | Image: Magicore II System design | Image: Dentis Dental Implant System design | Image: OsteoReady Dental Implant System design | Design | Image: Magicore II System design | Image: Osstem System design | |
| Connection | Internal Hex<br>Non - Submerged | Internal Hex<br>Non - Submerged | Internal Hex | - | Diameters<br>(Ø) | 3.5, 3.86, 4.3, 4.6 | 3.5 | 4.5 | Diameters (Ø) | 4.27, 4.57, 4.7, 5.07, 5.2, 5.21, 5.5, 5.57, 5.7, 5.87, 6.0, 6.2,<br>6.37, 6.5, 7.0 | 4.8, 6.0 | |
| Endosseous<br>Implant | Tapered, macro threads | Tapered, macro threads | Tapered, macro threads | Micro threaded,<br>One Body Implant | Lengths(mm) | 4.55, 5.55, 6.55, 7.55, 8.55 | 8.3 | 5,7,9,12 | Lengths(mm) | 4.5, 5.5, 6.5, 7.5, 8.5<br>4.6, 5.6, 6.6, 7.6, 8.6 | 6.5, 7, 7.5, 8, 8.5, 9, 9.5, 10 | |
| Range of<br>Diameters<br>(mm) | 4.0, 4.5, 5.0, 5.5, 6.0, 6.5mm | 4.0, 4.5, 5, 5.5, 6, 6.5mm | NR Fix : 3.5 and 3.8<br>Magic FC : 4.0, 4.5, 5.0, 5.5, 6.0,<br>6.5mm | 3.0mm | Surface<br>Treatment | Machine- | Machine- | Machine- | Surface<br>Treatment | Machine- | Machine- | |
| Range of<br>Lengths (mm) | 7, 8, 9, 10, 11, 12, 13mm | 7, 9, 11, 13mm | NR Fix 3.5 : 9, 11 13<br>NR Fix 3.8: 7, 10, 11, 12, 13, 14,<br>15mm<br>Magic FC : 7, 9, 11, 13, 15mm | 10, 11.5, 13, 15mm | Sterilization | End User Sterilization | Steam sterilization by user | End User Sterilization | Sterilization | End User Sterilization | End User Sterilization | |
| Modified<br>Surface | S.L.A | R.B.M | S.L.A | R.B.M | Principle of<br>Operation | Has one component to engage anti<br>rotational hex of implant body and the<br>other component to fixate the abutment<br>and implant body together. | Conventional procedure | Conventional procedure | Principle of Operation | Has one component to engage anti rotational hex of implant<br>body and the other component to fixate the abutment and<br>implant body together. | As general cement retained restoration, it is connected with<br>fixture and cemented crown on the abutment | |
| Surgical<br>Technique | 1 stage and 2 stage, self tapping | 1 stage and 2 stage, self tapping | 1 stage and 2 stage, self tapping | 1 stage and 2 stage, self tapping | SE<br>Discussion | There is no similar product as the subject short abutment in the primary predicate, K152520. So, K171027 was added as reference device.<br>Both subject device and K171027 have similar indication for use and same functions, surface treatment, and general shape (design). The<br>difference between subject and K171027 is material and dimensions. To support the discrepancy, K173575 was added and the difference<br>doesn't affect device's fundamental functions and safety. | | | SE Discussion | There is no similar product as the subject magic abutment in the primary predicate, K152520. So, K160670 was added as<br>reference device. Both subject device and K160670 have similar indication for use and same functions, material, surface<br>treatment, and general shape (design). The difference between subject and K160670 is dimensions. However, the difference<br>doesn't affect device's fundamental functions and safety. | | |
| Gamma<br>Sterilization | Yes | Yes | Yes | Yes | | | | | | | | |
| Principle<br>Operation | Dental implant that can be<br>inserted on a jawbone as a<br>material for dental surgery to<br>support…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.