OsteoReady® Dental Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. RidgeReady® 3.0 implants are intended to replace a lateral incisor in the maxilla and/or a central or lateral incisor in the mandible.
Device Story
System consists of endosseous dental implants, cover screws, healing caps, and abutments; used in surgical and restorative dental procedures to support prosthetic devices. Implants are threaded, tapered, self-tapping, root-form designs; manufactured from Grade 5 Titanium alloy (Ti-6Al-4V ELI) or PEEK. Operated by dental professionals in clinical settings. Device provides mechanical anchorage for artificial teeth; restores patient chewing function. Performance verified via mechanical fatigue testing (ISO 14801) and biocompatibility assessments. System includes various diameters and lengths to accommodate anatomical requirements.
Clinical Evidence
Bench testing only. Evidence includes sterilization validation (ISO 11137, ISO 17665), biocompatibility (ISO 10993-5, ISO 10993-10), fatigue testing (ISO 14801), shelf-life testing (ASTM F-1980), and packaging validation (ASTM D4169). LAL/endotoxin testing performed per USP 161/85. Risk assessment conducted per ISO 14971.
Technological Characteristics
Materials: Grade 5 Titanium alloy (Ti-6Al-4V ELI) per ISO 5832-3 and PEEK. Design: Threaded, tapered, self-tapping, root-form endosseous implants with internal hex connection. Surface: Sand-blasted and acid-etched. Sterilization: Gamma radiation for implants; steam sterilization for abutments. Dimensions: Various diameters (3.0mm to 6.0mm) and lengths (7mm to 15mm).
Indications for Use
Indicated for surgical and restorative applications in the upper or lower jaw to support prosthetic devices (artificial teeth) to restore chewing function. RidgeReady 3.0 implants are indicated for replacing a lateral incisor in the maxilla and/or a central or lateral incisor in the mandible.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
August 17, 2018
OsteoReady LLC % Daniela Levy Regulatory Consultant Sterling Medical Registration 22817 Ventura Blvd. Woodland Hills, California 91364
Re: K173575
Trade/Device Name: OsteoReady® Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: July 12, 2018 Received: July 25, 2018
#### Dear Daniela Levy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Andrew I. Steen -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image contains the logo for OsteoReady, a company that provides practical implant solutions. The logo consists of a stylized blue symbol on the left, resembling a stylized letter "U" made of curved lines. To the right of the symbol is the company name "OsteoReady" in a bold, blue font. Below the company name is the tagline "Practical Implant Solutions" in a smaller, lighter font.
Page 1 of 1
#### Indication for Use Statement
510(k) Number (if known): K173575
Device Name:
SECTION 4 -
OsteoReady® Dental Implant System
Indications for Use (Describe)
Indications for Use:
OsteoReady® Dental Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. RidgeReady® 3.0 implants are intended to replace a lateral incisor in the maxilla and/or a central or lateral incisor in the mandible.
Type of Use (Select one or both, as applicable)
区Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use(21 CFR 801 Subpart
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for OsteoReady. The logo consists of a blue abstract symbol on the left, resembling a stylized eye or target. To the right of the symbol, the word "OsteoReady" is written in a bold, blue font. Below "OsteoReady", the phrase "Practical Implant Solutions" is written in a smaller, lighter blue font.
# 510(k) Summary (21 CFR 807.92)
#### 510(k) Number K173575
| 1 | Submission Owner | OsteoReady LLC | | |
|---|--------------------------------------------------|--------------------------------------|---|----------------------------------|
| | | Dr. Robert Gottlander | | |
| | | 150 Baker Ave. Ext. Ste.110 | | |
| | | Concord, MA 01742 | | |
| | | USA | | |
| | | Phone : 206-812-7727 | | |
| | | Fax : 800-721-4907 | | |
| 2 | Official Correspondent | Sterling Medical Registration | | |
| | Contact Person | Daniela Levy - Regulatory Consultant | | |
| | | 22815 Ventura blvd. | | |
| | | Woodland Hills, CA 91364 | | |
| | | Phone: 1-213-787-3027 | | |
| | | Email: sterlingmedical2017@gmail.com | | |
| 3 | Prepared Date | August 16, 2018 | | |
| 4 | Device Trade Name | OsteoReady® Dental Implant System | | |
| 5 | Regulation Description | Endosseous Dental Implants Abutment | | |
| 6 | Classification | Primary Product Code: | | DZE |
| | | Device Name | | : Implant, endosseous, root-form |
| | | Regulation No | : | 872.3640 |
| | | Class | : | II |
| | | Panel | : | Dental |
| | | Secondary Product Code: | | NHA |
| 7 | Reason for the Premarket Notification Submission | : | | New Device |
- Identification of Legally Marketed Predicate Devices : 8 OsteoReady® Dental Implant System is substantially equivalent to the following: Primary Predicate: MIS Implants System K040807;
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for OsteoReady, a company that provides practical implant solutions. The logo features a stylized blue symbol on the left, resembling a tooth or bone structure. To the right of the symbol is the company name "OsteoReady" in a bold, blue font, with the tagline "Practical Implant Solutions" in a smaller font underneath.
Reference Devices: NobelActive K142260; DENTIN Implants System K120530; MIS K163349; A.B.Dental Implants System K112440; OsteoReady ZircoSeal Abutments K151169;
in terms of intended use, indication for use, raw material, technological characteristics and performance. The predicate devices are a Class II medical device.
- റി Device Description:
OsteoReady® Dental Implant System is consist of endosseous form dental implants, internal hex implants, tapered design; cover screws, healing caps and abutment systems;
Dental Implants:
Performance implants are available as follow: Diameters 3.85, 4.2, 5.0 and 6.0mm with lengths, 7 ( 7mm only to 5&6 dmm), 8, 10, 11.5, 13, & 15 (15mm not for 6.0 dm);
Performance Hybrid Implants are available as follow: Diameter 3.85mm with lengths of 10, 11.5, 13 and 15mm;
RidgeReady Implants are available as follow: Diameter 3.0mm with lengths of 8.0, 10, 12, and 14mm;
Dental Abutments:
Healing Caps, Temporary Abutment, Anatomic Angulated Abutment 15, Anti-Rotation Abutment Slim / Standard, Anatomic Abutment, Angulated Abutment 15, Anti-Rotation Abutment with Collar, Ball Attachments, PEEK Temporary Abutment, O-Ring Abutments, Direct Clip Abutments, attachments for ball.
- 10 Indication for Use:
OsteoReady® Dental Implant System is indicated for use in surqical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. RidgeReady® 3.0 implants are intended to replace a lateral incisor in the maxilla and/or a central or lateral incisor in the mandible.
- 11 Performance Standards or Special Controls
- . ISO 7405 Second edition 2008-12-15 Dentistry - Evaluation of biocompatibility of medical devices used in dentistry.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for OsteoReady, a company that provides practical implant solutions. The logo features a stylized blue icon on the left, resembling a tooth or bone structure. To the right of the icon is the company name, "OsteoReady," in a bold, blue font. Below the name is the tagline "Practical Implant Solutions" in a smaller, lighter blue font.
- ISO 5832-3:1996 Implants for surgery -- Metallic materials -- Part 3: Wrought titanium 6-aluminium 4-vanadium alloy.
- ISO 14801 Second edition 2007-11-15 Dentistry-Implants-Dynamic fatigue test for . endosseous dental implants.
- FDA guidance document: Class II Special Controls Guidance Document: Root-form . Endosseous Dental Implants and Endosseous Dental Abutments - Guidance for Industry and FDA Staff.
- 12 Substantial Equivalence
### Indication for Use Table
| Candidate<br>OsteoReady<br>Dental Implant<br>System | PRIMARY<br>PREDICATE<br>Lance Implant | REFERENCE<br>DEVICE<br>NobelActive 3.0<br>Implants | REFERENCE<br>DEVICE<br>DENTIN Implant<br>System | REFERENCE<br>DEVICE<br>A.B.Dental<br>Devices |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K173575 | K040807 | K142260 | K120530 | K112440 |
| OsteoReady<br>LLC | MIS Implant<br>Technologies Ltd | Nobel Biocare USA<br>LLC | DENTIN Implants<br>Technologies Itd | A.B. Dental<br>Implants System |
| OsteoReady®<br>Dental Implant<br>System is<br>indicated for<br>use in surgical<br>and restorative<br>applications for<br>placement in<br>the bone of the<br>upper or lower<br>jaw to provide<br>support for<br>prosthetic<br>devices, such<br>as artificial<br>teeth, in order to<br>restore the<br>patient's<br>chewing<br>function.<br>RidgeReady®<br>3.0 implants are<br>intended to<br>replace a lateral<br>incisor in the<br>maxilla and/or a<br>central or lateral<br>incisor in the<br>mandible. | The MIS Dental<br>Implant System is<br>indicated for use in<br>surgical and<br>restorative<br>applications for<br>placement in the<br>bone of the upper<br>or lower jaw to<br>provide support for<br>prosthetic devices,<br>such as artificial<br>teeth, in order to<br>restore the patient's<br>chewing function. | NobelActive®<br>implants are<br>endosseous<br>implants intended<br>to be surgically<br>placed in the upper<br>or lower jaw bone<br>for anchoring or<br>supporting tooth<br>replacements to<br>restore patient<br>esthetics and<br>chewing function.<br>NobelActive®<br>implants are<br>indicated for single<br>or multiple unit<br>restorations in<br>splinted or non-<br>splinted<br>applications. This<br>can be achieved by<br>a 2-stage or 1-<br>stage surgical<br>technique in<br>combination with<br>immediate, early or<br>delayed loading<br>protocols, | DENTIN® Dental<br>Implants System is<br>indicated for use in<br>surgical and<br>restorative<br>applications for<br>placement in the<br>bone of the upper<br>or lower jaw to<br>provide support for<br>prosthetic devices,<br>such as artificial<br>teeth, in order to<br>restore the patient's<br>chewing function.<br>DENTIN® Dental<br>Implants System is<br>indicated also for<br>immediate loading<br>when good primary<br>stability is achieved<br>and with<br>appropriate<br>occlusal loading.<br>Two Stage<br>Implants:<br>CLASSIC, RAPID,<br>PRESTIGE.<br>One Stage | The AB Dental<br>Devices implants<br>are intended for<br>surgical placement<br>in the maxillary<br>and/or mandibular<br>arch to support<br>crowns, bridges, or<br>overdentures in<br>edentulous or<br>partially edentulous<br>patients. I7 Integral<br>implant, I5 Conical<br>implant,<br>P15<br>Temporary<br>abutment, P12-T,L<br>Temporary flat<br>connection<br>abutment, and P16<br>Straight adaptor<br>are appropriate for<br>immediate loading<br>when good primary<br>stability is achieved<br>and with<br>appropriate<br>occlusal loading. |
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for OsteoReady. The logo consists of a blue abstract symbol on the left, resembling a stylized "U" shape with concentric curves. To the right of the symbol is the company name, "OsteoReady," in a bold, blue font. Below the name, in a smaller font, is the tagline "Practical Implant Solutions."
| | recognizing<br>sufficient primary<br>stability and<br>appropriate<br>occlusal loading for<br>the selected<br>technique.<br>NobelActive® 3.0<br>implants are<br>intended to replace<br>a lateral incisor in<br>the maxilla and/or a<br>central or lateral<br>incisor in the<br>mandible.<br>NobelActive® 3.0<br>implants are<br>indicated for single<br>unit restorations<br>only. | Implants: PIECE<br>DENTIN®<br>ONEPIECE<br>Implants 3.0 mmd<br>are intended for<br>placement at the<br>mandibular central<br>and lateral incisors<br>and maxillary and<br>lateral incisors.<br>Indicated also for<br>denture<br>stabilization using<br>multiple implants. |
|--|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
#### Discussion
OsteoReady shares the same indication for use as its primary predicate. The only difference is related to statement of RidgeReady® 3.0 implants which are intended to replace a lateral incisor in the maxilla and/or a central or lateral incisor in the mandible. This statement is similar to the reference devices NobelActive K142260 and DENTIN K120530 since these reference devices also introduce the 3.0ø implants, whereas the primary predicate doesn't introduce the 3.0ø implants . Therefore, OsteoReady doesn't introduce any safety or efficacy issues.
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for OsteoReady, a company that provides practical implant solutions. The logo consists of a blue circular icon on the left and the company name "OsteoReady" in blue text on the right. Below the company name is the tagline "Practical Implant Solutions" in a smaller, lighter blue font. The icon is a stylized representation of a tooth or bone structure, with concentric circles forming the outer shape.
## Dental Implants System
| CHARACTERISTIC | Candidate No.1<br>Performance Implants | PRIMARY PREDICATE<br>Lance Implant |
|---------------------------|------------------------------------------------------------------|-----------------------------------------|
| 510k | K173575 | K040807 |
| Company Name | OsteoReady LLC | MIS Implant Technologies Ltd |
| Device Design | Threaded, root form endosseous implants | Threaded, root form endosseous implants |
| Classification | Class 2 872.3640 P.Code DZE | Class 2 872.3640 P.Code DZE |
| Material | GR-5 Titanium Ti-6Al-4V ELI | GR-5 Titanium Ti-6Al-4V ELI |
| Diameters mmd | 3.85, 4.2, , 5, 6 | 3.75, 4.2, 5 |
| Lengths mml | 7 [only to 5 & 6 mmd]<br>8, 10, 11.5, 13<br>15 [Not for 6.0 mmd] | 10, 11.5,13, 16 |
| Connection Type | Internal Hex | Internal Hex |
| Implant Body<br>Contour | Tapered, Self Tapping | Tapered, Self Tapping |
| Coating / Surface | Sand Blast & Acid Etched | Sand Blast & Acid Etched |
| Anatomical Site | Oral Cavity | Oral Cavity |
| Principle of<br>operation | Conventional procedure | Conventional procedure |
| Self tapping | ✓ | ✓ |
| Sterilization | Gamma Ray | Gamma Ray |
| Packaging | Double packaging | Double packaging |
| CHARACTERISTIC | Candidate No.2<br>Performance Hybrid Implants | PRIMARY PREDICATE<br>SEVEN Implants |
|-------------------------|-----------------------------------------------|-----------------------------------------|
| 510k | K173575 | K040807 |
| Company Name | OsteoReady LLC | MIS Implant Technologies Ltd |
| Device Design | Threaded, root form endosseous implants | Threaded, root form endosseous implants |
| Classification | Class 2 872.3640 P.Code DZE | Class 2 872.3640 P.Code DZE |
| Material | GR-5 Titanium Ti-6Al-4V ELI | GR-5 Titanium Ti-6Al-4V ELI |
| Diameters mmd | 3.85 | 3.75, 4.2, 5, 6 |
| Lengths mml | 10, 11.5, 13, 15 | 8, 10, 11.5, 13, 16 (N/A for 6mmd) |
| Connection Type | Internal Hex | Internal Hex |
| Implant Body<br>Contour | Tapered, Self Tapping | Tapered, Self Tapping |
| Coating / Surface | Sand Blast & Acid Etched | Sand Blast & Acid Etched |
| Anatomical Site | Oral Cavity | Oral Cavity |
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for OsteoReady. The logo consists of a blue circular icon on the left and the text "OsteoReady" in blue on the right. Below the text "OsteoReady" is the text "Practical Implant Solutions" in a smaller font size. The logo is simple and professional, and it conveys the company's focus on providing practical solutions for dental implants.
| Principle of<br>operation | Conventional procedure | |
|---------------------------|------------------------|------------------|
| Self tapping | ✓ | ✓ |
| Sterilization | Gamma Ray | Gamma Ray |
| Packaging | Double packaging | Double packaging |
| CHARACTERISTIC | Candidate No.3<br>RidgeReady Implants | REFERENCE DEVICE<br>NobelActive 3.0 Implants |
|------------------------|-----------------------------------------|----------------------------------------------|
| 510k | K173575 | K142260 |
| Company Name | OsteoReady LLC | Nobel Biocare USA LLC |
| Device Design | Threaded, root form endosseous implants | Threaded, root form endosseous implants |
| Classification | Class 2 872.3640 P.Code DZE | Class 2 872.3640 P.Code DZE |
| Material | GR-5 Titanium Ti-6Al-4V EL | GR-5 Titanium Ti-6Al-4V ELI |
| Diameters dmm | 3.0 | 3.0, 3.5, 4.3, 5.0, 5.5 |
| Lengths Imm | 8, 10, 12, 14 | 7, 8.5, 10, 11.5, 13, 15, 18 |
| Connection Type | Internal Hex | Internal Hex |
| Implant Body Contour | Tapered, Self Tapping | Tapered, Self Tapping |
| Surface | Sand Blast & Acid Etched | TiUnite |
| Anatomical Site | Oral Cavity | Oral Cavity |
| Principle of operation | Conventional procedure | Conventional procedure |
| Self tapping | ✓ | ✓ |
| Sterilization | Gamma Ray | Gamma Ray |
| Packaging | Double packaging | Double packaging |
## Dental Abutments System
| | Candidate | Primary Predicate | Reference Device |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Company / 510k | OsteoReady LLC K173575 | | DENTIN 510K 120530 |
| Product Name | Healing Caps | | Healing Caps |
| Dimensions mm | Slim - Length 1,2,3,4,5,6,7<br>Standard - Length 1,2,3,4,5,6,7<br>Wide - Length 2,3,4,5,6,7<br>Narrow Diameter 3.0 - Height 1,3,5 | | Slim - Length 1,2,3,4,5,6,7<br>Standard - Length 1,2,3,4,5,6,7<br>Wide - Length 2,3,4,5,6,7<br>Narrow Diameter 3.0 - Height 1,3,5 |
| Material | Titanium alloy Ti-6Al-4V-ELI. | | Titanium alloy Ti-6Al-4V-ELI. |
| Connection | Internal Hex | | Internal Hex |
| Company / 510k | OsteoReady LLC K173575 | MIS K040807 | DENTIN 510K 120530 |
| Product Name | Anatomic Angulated Titanium<br>Abutment | Esthetic Angulated Abutment | Anatomic Angulated<br>Titanium Abutment |
| Dimensions mm | Height 1,2,3,4 - Angle 15 | Height 1,2,3 - Angle 15 | Height 1,2,3,4 - Angle 15 |
| Material | Titanium alloy Ti-6Al-4V-ELI. | Titanium alloy Ti-6Al-4V-ELI. | Titanium alloy Ti-6Al-4V-ELI. |
| Connection | Internal Hex | Internal Hex | Internal Hex |
| Company / 510k | OsteoReady LLC K173575 | MIS K040807 | DENTIN 510K 120530 |
| Product Name | Anti-Rotation Abutment Slim | Cemented Abutment | Slim Titanium Abutment |
| Dimensions mm | Diameter 3.8<br>Length 5, 7, 9, 12 mm | Diameter 3.75<br>Length 8.5, 11 | Diameter 3.8<br>Length 5, 7, 9, 12, 15 |
| Material | Titanium alloy Ti-6Al-4V-ELI. | Titanium alloy Ti-6Al-4V-ELI. | Titanium alloy Ti-6Al-4V-ELI. |
| Connection | Internal Hex | Internal Hex | Internal Hex |
| | | | |
| Company / 510k | OsteoReady LLC K173575 | MIS K040807 | DENTIN 510K 120530 |
| Product Name | Anti-Rotation Abutment | Cemented Abutment | Titanium Abutment |
| Dimensions mm | Diameter 4.5 | Diameter 4.7 | Diameter 4.5 |
| | Length 5, 7, 9, 12 | Length 11 | Length 7, 9, 12 |
| Material | Titanium alloy Ti-6Al-4V-ELI. | Titanium alloy Ti-6Al-4V-ELI. | Titanium alloy Ti-6Al-4V-ELI. |
| Connection | Internal Hex | Internal Hex | Internal Hex |
| | | | |
| Company / 510k | OsteoReady LLC K173575 | MIS K040807 | DENTIN 510K 120530 |
| Product Name | Anatomic Titanium Abutment | Anatomic transgingival<br>Abutment | Anatomic Titanium<br>Abutment |
| Dimensions mm | Height 1,2,3,4 | Height 1,2,3,4 | Height 1,2,3,4 |
| Material | Titanium alloy Ti-6Al-4V-ELI. | Titanium alloy Ti-6Al-4V-ELI. | Titanium alloy Ti-6Al-4V-ELI. |
| Connection | Internal Hex | Internal Hex | Internal Hex |
| | | | |
| Company / 510k | OsteoReady LLC | MIS K040807 | DENTIN 510K 120530 |
| Product Name | Angulated Titanium Abutment | Angulated Cement Abutment | Angulated Titanium<br>Abutment |
| Dimensions mm | Length 9 - Angle 15 | Length 9,11 - Angle 15 | Length 9,12 - Angle 15 |
| Material | Titanium alloy Ti-6Al-4V-ELI. | Titanium alloy Ti-6Al-4V-ELI. | Titanium alloy Ti-6Al-4V-ELI. |
| Connection | Internal Hex | Internal Hex | Internal Hex |
| | | | |
| | Candidate | Primary Predicate | Reference Device |
| Company / 510k | OsteoReady LLC K173575 | MIS K040807 | DENTIN 510K 120530 |
| Product Name | Anti-Rotation Abutment with<br>Collar | Straight Cement Abutment | Straight Titanium<br>Abutment |
| Dimensions mm | Diameter 4.5 Length 1,2,3,4 | Diameter NP, RP, WP Length<br>1,2,3,4 | Diameter 4.5 Length 1,2,3,4 |
| Material | Titanium alloy Ti-6Al-4V-ELI. | Titanium alloy Ti-6Al-4V-ELI. | Titanium alloy Ti-6Al-4V-ELI. |
| Connection | Internal Hex | Internal Hex | Internal Hex |
| Product Name | Ball Attachment | Ball Attachment | Ball Attachment |
| Dimensions mm | Height 0.5,1,2,3,4,5,6 | Height 1,2,3,4,5 | Height 0.5,1,2,3,4,5,6 |
| Material | Titanium alloy Ti-6Al-4V-ELI. | Titanium alloy Ti-6Al-4V-ELI. | Titanium alloy Ti-6Al-4V-ELI. |
| Connection | Internal Hex | Internal Hex | Internal Hex |
| Related Components | Metal Caps, Silicon Caps | Metal Caps, Silicon Caps | Metal Caps, Silicon Caps |
| Company / 510k | OsteoReady LLC K173575 | MIS K163349 | DENTIN 510K 120530 |
| Product Name | PEEK Temporary Abutment | Temporary plastic abutment | Straight curve minor - PEEK |
| Dimensions mm | Height 1,2,3, | NP,RP Height 1,2,3 | Height 1,2,3, |
| Material | PEEK | PEEK | PEEK |
| Connection | Internal Hex | Internal Hex, conical | Internal Hex |
| Company / 510k | OsteoReady LLC K173575 | MIS K040807 | A.B.Dental Devices K112440 |
| Product Name | O-Ring Abutments | Ball Attachment | Straight Adaptor P16 |
| Dimensions mm | Diameter 3.0 mm | Diameter NP<br>Height 1,2,3 mm | Diameter 3.0 mm<br>Height 1,2,3,4 mm |
| Material | Titanium alloy Ti-6Al-4V-ELI. | Titanium alloy Ti-6Al-4V-ELI. | Titanium alloy Ti-6Al-4V-ELI. |
| Connection | Internal Hex | Internal Hex | Internal Hex |
| Company / 510k | OsteoReady LLC K173575 | | A.B.Dental Devices K112440 |
| Product Name | Direct Clip Abutments | | Straight Adaptor P16 |
| Dimensions mm | Diameter 3.0 mm,<br>Height 1,3,5 | | Diameter 3.0 mm<br>Height 1,2,3,4 mm |
| Material | Titanium alloy Ti-6Al-4V-ELI. | | Titanium alloy Ti-6Al-4V-ELI. |
| Connection | Internal Hex | | Internal Hex |
| | Candidate | Primary Predicate | Reference Device |
| Company / 510k | OsteoReady LLC K173575 | MIS K040807 | |
| Product Name | Straight Abutments | Transgingival Abutment, NP | |
| Dimensions mm | Diameter 3.0 mm,<br>Height 0.5,2.5,4.5 | Diameter NP,<br>Height 1,2,3,4 | |
| Material | Titanium alloy Ti-6Al-4V-ELI. | Titanium alloy Ti-6Al-4V-ELI. | |
| Connection | Internal Hex | Internal Hex | |
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for OsteoReady. The logo consists of a blue circular design on the left, resembling a stylized eye or target. To the right of the circular design is the text "OsteoReady" in a bold, blue font. Below "OsteoReady" is the tagline "Practical Implant Solutions" in a smaller, lighter blue font.
{11}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the logo for OsteoReady, a company that provides practical implant solutions. The logo consists of a blue symbol on the left and the company name "OsteoReady" in blue text on the right. Below the company name, there is a tagline "Practical Implant Solutions" in a smaller font size.
{12}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for OsteoReady, a company that provides practical implant solutions. The logo features a stylized blue symbol on the left, resembling a target or a stylized letter 'U'. To the right of the symbol is the company name, "OsteoReady," in a bold, blue font. Below the company name is the tagline "Practical Implant Solutions" in a smaller, lighter blue font.
13 Summary of Equivalence:
OsteoReady® Dental Implant System: Performance Implants & Performance Hybrid Implants share similarity to MIS Implants System K040807; RidgeReady Implants share similarity to NobelActive K142260; in terms of intended use, indication for use, technological characteristics, design, raw material and performance. The minor difference is regarding to OsteoReady diameter size which is wider, however it doesn't introduce any new safety or efficacy issues. OsteoReady Dental Implant system is made of Titanium alloy Ti-6AI-4V-ELI as in similar to its primary predicate.
OsteoReady® Dental Abutment System shares similarity to MIS Implants System K040807, K163349; DENTIN Implants System K120530; A.B.Dental Implants System K112440; in terms of intended use, indication for use, technological characteristics, design, raw material and performance.
The minor difference is regarding to measurements related few abutments types, however it doesn't introduce any new safety or efficacy issues since similar measurements are already exist in the market.
As demonstrated by the substantial equivalent table, the minor differences of OsteoReady LLc vs. its predicate and reference devices raise no new safety or/and effectiveness issues and therefore OsteoReady® Dental Implant System is considered to be substantial equivalent to its predicate and reference devices.
14 Non Clinical Performance:
Sterilization Validation Test was carried out with accordance to ISO 11137 in order to demonstrate substantial equivalence of OsteoReady® Dental Implants. Test results have demonstrated that the SAL of 10-® was achieved and all testing requirements were met.
LAL/endotoxin testing was carried out with accordance to USP 161 and USP 85. Test results have met the criteria.
OsteoReady will not market its Dental Implant System as "non-pyrogenic".
Shelf Life Test was carried out with accordance to ASTM F-1980 in order to demonstrate substantial equivalence of OsteoReady® Dental Implants of 5 years shelf life.
Steam Sterilization Test was carried out with accordance to ISO 17665 in order to demonstrate substantial equivalence of OsteoReady® Dental Abutments - Test results have demonstrated that the SAL of 10-6 was achieved and all testing requirements were met.
{13}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for OsteoReady. The logo features a blue abstract design on the left, resembling a stylized letter "U" or a series of concentric arcs. To the right of the design is the company name "OsteoReady" in a bold, blue font. Below the name, in a smaller font, is the tagline "Practical Implant Solutions."
Packaging validation test was carried out with accordance to ASTM D4169 in order to validate the safety of the device and packaging during transportation test results were meet the criteria.
Biocompatibility - OsteoReady uses the same raw material as was previously cleared under 510k K151169 titanium alloy Ti6Al4V Eli and uses the same conventional manufacturing process. Cytotoxicity testing was conducted with accordance to ISO 10993-5 to the used raw materials: Titanium alloy and for PEEK Abutments, in order to demonstrate that the manufacturing process did not change the biocompatibility profile. Irritation testing and Sensitization testing were conducted with accordance to ISO 10993-10, all test results met the criteria.
Fatigue test was carried out with accordance to ISO 14801 in order to verify the mechanical connection strength of implant/abutment, results have demonstrated the high performance with the use of OsteoReady® Dental Implants / Abutments.
Surface Test was carried out to ensure the cleanness of the implant surface, surface results were met OsteoReady requirements.
Risk Assessment was carried out with accordance to ISO 14971 and has demonstrated no new safety and/or effectiveness issues.
#### 15 Conclusion:
As verified by substantial equivalence, risk assessment and bench testing OsteoReady® Dental Implant System shares similarity to its predicated devices in terms of intended use, indication for use, raw material, technological characteristics and performance. Therefore, OsteoReady® Dental Implant system is considered to be substantially equivalent to its predicate and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.