OsteoReady ® ZircoSeal® Dental Abutments are intended to be used with an endosseous implant to aid prosthetic rehabilitation in order to restore the patient's chewing function. OsteoReady® ZircoSeal® Abutments are compatible with DENTIN® Dental Implant System - Prestige Implant System diameters 3.75, 4.2, 5 & 6;
Device Story
OsteoReady ZircoSeal Abutments are straight, anti-rotation dental abutments fabricated from titanium alloy (Ti-6Al-4V ELI). They feature a zirconium nitride surface treatment (trade name ZircoSeal). The device is intended for use with endosseous dental implants to support prosthetic rehabilitation and restore chewing function. The abutments are provided non-sterile and must be sterilized by the clinician via autoclave before use. They are single-use devices. The clinician selects the appropriate abutment based on the implant system and dimensions, then secures it to the implant to serve as a base for the final dental prosthesis. The device does not provide angle correction and is not intended for chairside or CAD/CAM modification. By providing a stable connection between the implant and the prosthesis, the device enables the restoration of oral function for the patient.
Clinical Evidence
No clinical data included. Bench testing included steam sterilization validation (achieving SAL 10^-6 per ISO 17665-1/2 and ANSI/AAMI ST79:2010) and biocompatibility testing per ISO 10993-5. Surface treatment evaluation was conducted per FDA guidance for Endosseous Dental Implants and Abutments.
Technological Characteristics
Material: GR-5 Titanium (Ti-6Al-4V ELI) per ISO 5832-3. Surface: Zirconium nitride (ZircoSeal). Design: Straight, anti-rotation, internal hex connection. Dimensions: Diameters 3.75mm and 4.5mm; lengths 5-13mm or cuff heights 1-4mm. Sterilization: Autoclave (provided non-sterile).
Indications for Use
Indicated for patients requiring prosthetic rehabilitation via endosseous dental implants to restore chewing function. Compatible with DENTIN Prestige Implant System (diameters 3.75, 4.2, 5, and 6).
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 8, 2015
OsteoReady LLC c/o Ms. Daniela Levy Regulatory Consultant Sterling Medical Registration 22817 Ventura Blvd Woodland Hills, California 91364
Re: K151169
Trade/Device Name: OsteoReady® ZircoSeal® Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: August 26, 2015 Received: September 9, 2015
Dear Ms. Levy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image contains the logo for OsteoReady, a company that provides practical implant solutions. The logo consists of a stylized blue icon on the left, resembling concentric waves forming a shape. To the right of the icon is the company name, "OsteoReady," in a bold, dark blue font. Below the company name is the tagline "Practical Implant Solutions" in a smaller, lighter blue font.
Page 1 of
| SECTION 4 - | Indication for Use Statement |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| 510(k) Number (if known): | |
| K151169 | |
| Device Name: | |
| OsteoReady® ZircoSeal® Abutments | |
| Indications for Use ( <i>Describe</i> ) | |
| Indications for Use: | |
| OsteoReady ® ZircoSeal® Dental Abutments are intended to be used with an endosseous<br>implant to aid prosthetic rehabilitation in order to restore the patient's chewing function. | |
| OsteoReady® ZircoSeal® Abutments are compatible with DENTIN® Dental Implant System<br>- Prestige Implant System diameters 3.75, 4.2, 5 & 6; | |
Type of Use (Select one or both, as applicable)
区Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use(21 CFR 801 Subpart
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
-
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image displays the logo for OsteoReady, a company specializing in practical implant solutions. Below the logo, the text "SECTION 5 - 510(k) Summary (21 CFR 807.92)" is present. This suggests the image is part of a document related to regulatory compliance, specifically a 510(k) summary as required by the Code of Federal Regulations for medical devices.
# 510(k) Number K151169
| 1 | Submission | OsteoReady LLC | | | |
|---|---------------------------------------------------------------------------------|--------------------------------------------------|------------|----------|--|
| | Owner | Michael D. Silverman, DMD, DICOI | | | |
| | | 106 Lenora Street | | | |
| | | Seattle, WA 98121 | | | |
| | | USA | | | |
| | | Phone : 206-812-7727 | | | |
| | | Fax : 800-721-4907 | | | |
| 2 | Official Correspondent | Sterling Medical Registration | | | |
| | Contact Person | Daniela Levy - Regulatory Consultant | | | |
| | | 22817 Ventura blvd. | | | |
| | | Woodland Hills, CA 91364 | | | |
| | | Phone: 1-213-787-3027 | | | |
| | | Email: daniela@sterlingmedicalregistration.com | | | |
| 3 | Date Prepared | October 05, 2015 | | | |
| 4 | Device Trade Name | OsteoReady® ZircoSeal® Abutments | | | |
| 5 | Regulation Description | Endosseous Dental Implant Abutment | | | |
| 6 | Classification | Device Name : Abutment,Implant,Dental,Endosseous | | | |
| | | Product Code | : | NHA | |
| | | Regulation No | : | 872.3630 | |
| | | Class | : | II | |
| | | Panel | : | Dental | |
| 7 | Reason for the Premarket Notification Submission | : | New Device | | |
| 8 | Identification of Legally Marketed Predicate Devices : | | | | |
| | OsteoReady® ZircoSeal® Abutments are substantially equivalent to the following: | | | | |
#### Primary Predicate
- PerioType X-Pert (K123386); .
Reference Predicate
- . AB Dental Devices Implants and Accessories (K112440)
- A.B. Dental Devices (K051719) .
-
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for OsteoReady, a company that provides practical implant solutions. The logo consists of a stylized blue icon on the left, resembling a series of concentric arcs. To the right of the icon is the company name, "OsteoReady," in a bold, blue font. Below the company name is the tagline "Practical Implant Solutions" in a smaller, lighter blue font.
- IQ Implant System(K130829)
in terms of intended use, indication for use, raw material, technological characteristics and performance.
The predicate devices are a Class II medical device.
- တ Device Description:
OsteoReady® ZircoSeal® Abutments are Straight Anti-Rotation Abutments made of titanium alloy. OsteoReady® ZircoSeal® Abutments are available in the following measurements:
Slim Anti-Rotation Titanium Abutment with ZircoSeal®: Length 5,7,9,11,13 mm with Diameter 3.75 mm.
Anti-Rotation Titanium Abutment with ZircoSeal®: Length 5,7,9,11,13 mm Diameter 4.5 mm.
Anti-Rotation Straight Titanium Abutment with ZircoSeal®: Shoulder Length 1,2,3,4 mm Diameter 4.5 mm.
OsteoReady® ZiroSeal® Abutments are straight only and are not intended to be modified chairside or by CAD/CAM to provide an angle correction.
The ZircoSeal surface treatment is the trade name of zirconium nitride.
The ZircoSeal® simply indicates the signature zirconium nitride surface treatment of the abutment. The trade name is unrelated to claims of "sealing" of implant/abutment connection platform.
- 10 Indication for Use:
OsteoReady ® ZircoSeal® Dental Abutments are intended to be used with an endosseous implant to aid prosthetic rehabilitation in order to restore the patient's chewing function. OsteoReady® ZircoSeal® Abutments are compatible with DENTIN® Dental Implant System - Prestige Implant System diameters 3.75, 4.2, 5 & 6;
- 11 Performance Testing and Special Controls
Steam validation test was conducted in order to ensure safety assurance level of ( SAL) 10-6 related to OsteoReady® ZircoSeal® Abutments system.
Biocompatibility test was conducted according to ISO 10993-5, test results have met all applicable criteria. Additional biocompatibility justification was provided attesting that the raw material and surface treatments of the subject abutment is the same as the predicates.
Clinical Testing - No clinical data is included in this submission.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for OsteoReady, a company that provides practical implant solutions. The logo consists of a stylized blue icon on the left, resembling a tooth or bone structure. To the right of the icon is the company name "OsteoReady" in a bold, blue font. Below the company name, in a smaller font, is the tagline "Practical Implant Solutions."
- ISO 7405 Second edition 2008-12-15 Dentistry Evaluation of biocompatibility . of medical devices used in dentistry.
- ISO 5832-3:1996 Implants for surgery -- Metallic materials -- Part 3: . Wrought titanium 6-aluminium 4-vanadium alloy.
- Sterilization validation according to to ISO 17665-1/2 and ANSI/AAMI . ST79:2010. Test results have demonstrated that the SAL of 10-6 was achieved and all testing requirements were met.
- . Surface treatment evaluation per the FDA guidance document for Endosseous Dental Implants and Abutments including chemical analysis and abrasion characteristics of the surface.
FDA guidance document: Class II Special Controls Guidance Document: Rootform Endosseous Dental Implants and Endosseous Dental Abutments - Guidance for Industry and FDA Staff.
| Substantial<br>Equivalent Table<br>CHARACTERISTIC | Candidate<br>Slim Anti Rotation<br>Abutment with ZircoSeal ® | REFERENCE PREDICATE<br>DEVICE<br>P3 Anti rotation abutment | REFERENCE<br>PREDICATE DEVICE<br>Slim Straight Abutment |
|---------------------------------------------------|------------------------------------------------------------------|------------------------------------------------------------|---------------------------------------------------------|
| 510k | | K112440, K051719 | K130829 |
| Company Name | OsteoReady LLC | A.B.Dental Devices Itd. | IQ Implants Itd |
| Device Design | Endosseous dental implant<br>abutment | Endosseous dental implant<br>abutment | Endosseous dental implant<br>abutment |
| Classification | Class II 872.3630<br>P.Code NHA | Class II 872.3630<br>P.Code NHA | Class II 872.3630<br>P.Code NHA |
| Material | GR-5 Titanium Ti-6Al-4V ELI | GR-5 Titanium Ti-6Al-4V ELI | GR-5 Titanium Ti-6Al-4V ELI |
| Diameters mm | 3.75 | 3.75 | 3.75 |
| Lengths<br>mm | 5, 7, 9, 11, 13 | 5, 7, 9, 12, 15 | 5, 7, 9, 11 |
| Implant / Abutment<br>connection | Internal Hex | Internal Hex | Internal Hex |
| Surface | ZircoSeal<br>[Surface is similar to PerioType<br>X-pert K123386] | CNC Machined | CNC Machined |
| Anatomical Site | Oral Cavity | Oral Cavity | Oral Cavity |
## 12 Substantial Equivalence
-
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for OsteoReady, a company that provides practical implant solutions. The logo consists of a stylized blue symbol on the left, resembling a tooth or bone structure. To the right of the symbol is the company name, "OsteoReady," in a bold, blue font. Below the name is the tagline "Practical Implant Solutions" in a smaller, lighter font.
| Indication for Use | OsteoReady ® ZircoSeal® Dental Abutments are intended to be used with an endosseous implant to aid prosthetic rehabilitation in order to restore the patient's chewing function. OsteoReady® ZircoSeal® Abutments are compatible with DENTIN® Dental Implant System - Prestige Implant System diameters 3.75, 4.2, 5 & 6. | The AB Dental Devices implants are intended for surgical placement in the maxillary mandibular and/or arch to support crowns, bridges, or overdentures in edlentulous or partially edentulous patients. | IQ Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. IQ Implant System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading. |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization | Provided non sterile prior to their use sterilize in an autoclave. | Provided non sterile prior to their use sterilize in an autoclave. | Provided non sterile prior to their use sterilize in an autoclave. |
| Use | Single use | Single use | Single use |
| Substantial Equivalent<br>Table<br>CHARACTERISTIC | Candidate<br>Anti Rotation Abutment with<br>ZircoSeal ® | REFERENCE PREDICATE DEVICE<br>P3* Abutment Anti rotation |
|---------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k | | K112440, K051719 |
| Company Name | OsteoReady LLC | A.B.Dental Devices Itd. |
| Device Design | Endosseous dental implant abutment | Endosseous dental implant abutment |
| Classification | Class II 872.3630<br>P.Code NHA | Class II 872.3630<br>P.Code NHA |
| Material | GR-5 Titanium Ti-6Al-4V ELI | GR-5 Titanium Ti-6Al-4V ELI |
| Diameters mm | 4.5 | 4.5 |
| Lengths<br>mm | 5, 7, 9, 11, 13 | 5, 7, 9, 12, 15 |
| Implant / Abutment connection | Internal Hex | Internal Hex |
| Surface | ZircoSeal [Surface is similar to PerioType X-<br>pert K123386] | CNC Machined Surface |
| Anatomical Site | Oral Cavity | Oral Cavity |
| Indication for Use | OsteoReady ® ZircoSeal® Dental<br>Abutments are intended to be used with<br>an endosseous implant to aid prosthetic<br>rehabilitation in order to restore the<br>patient's<br>chewing<br>function.<br>OsteoReady® ZircoSeal® Abutments are<br>compatible with DENTIN® Dental Implant<br>System - Prestige Implant System<br>diameters 3.75, 4.2, 5 & 6. | The AB Dental Devices implants<br>are<br>intended for surgical placement in the<br>maxillary mandibular and/or arch to<br>support crowns, bridges, or overdlentures<br>in edlentulous or partially edentulous<br>patients. |
| Sterilization | Provided non sterile prior to their use<br>sterilize in an autoclave. | Provided non sterile prior to their use<br>sterilize in an autoclave. |
| Use | Single use | Single use |
・
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for OsteoReady. The logo features a stylized blue icon on the left, resembling a series of concentric arcs. To the right of the icon is the company name, "OsteoReady," in a bold, blue font. Below the name is the tagline "Practical Implant Solutions" in a smaller, lighter font.
| Substantial Equivalent Table<br>CHARACTERISTIC | Candidate<br>Anti Rotation Straight Abutment with<br>ZircoSeal ® | PRIMARY PREDICATE DEVICE<br>PerioType X-pert, ZircoSeal® Abutments<br>NAS |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k | | K123386 |
| Company Name | OsteoReady LLC | Clinical House Europe Gmbh |
| Device Design | Endosseous dental implant abutment | Endosseous dental implant abutment |
| Classification | Class II 872.3630<br>P.Code NHA | Class II 872.3630<br>P.Code NHA |
| Material | GR-5 Titanium Ti-6Al-4V ELI | Titanium Grade 4 |
| Diameters mm | 4.5 | 4.2, 5.5 |
| Lengths<br>mm | Cuff height 1, 2, 3, 4 | Cuff height 1, 2, 3, 4 |
| Implant / Abutment connection | Internal Hex | Internal Hex |
| Surface | ZircoSeal | ZircoSeal |
| Anatomical Site | Oral Cavity | Oral Cavity |
| Indication for Use | OsteoReady ® ZircoSeal® Dental<br>Abutments are intended to be used with<br>an endosseous implant to aid prosthetic<br>rehabilitation in order to restore the<br>patient's chewing function.<br>OsteoReady® ZircoSeal® Abutments<br>are compatible with DENTIN® Dental<br>Implant System - Prestige Implant<br>System diameters 3.75, 4.2, 5 & 6. | The PerioType X-Pert Abutments are<br>straight and angled dental implant<br>abutments to be used in conjunction with<br>the PerioType X-Pert dental implant fixture<br>to aid in prosthetic rehabilitation. |
| Sterilization | Provided non sterile prior to their use<br>sterilize in an autoclave. | Provided non sterile prior to their use<br>sterilize in an autoclave. |
| Use | Single use | Single use |
#### Summary of Equivalence:
OsteoReady® ZircoSeal® Abutment System shares similarity to A.B.Dental Devices Itd. K112440, K051719; IQ Implants K130829; PerioType X-Pert K123386 in terms of intended use, indication for use, technological characteristics and performance.
The differences are that A.B.Dental Devices Itd. K112440, K051719; IQ Implants K130829; use CNC machined surface to their abutments while OsteoReady uses the ZircoSeal® surface treatment which is identical surface to its other predicate device the PerioType X-Pert K123386. In addition, PerioType X-Pert K123386 uses titanium grade 4 while OsteoReady uses titanium alloy as in similar to its other predicate devices A.B.Dental Devices Itd. K112440, K051719; IQ Implants K130829; The verbiage of the Indications for Use of the subject device is slightly different than that of the declared predicates; however, these slight differences in wording does not
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for OsteoReady, a company that provides practical implant solutions. The logo consists of a blue circular graphic on the left, resembling a stylized wave or ripple effect. To the right of the graphic is the company name "OsteoReady" in a bold, blue font. Below the name, in a smaller font, is the tagline "Practical Implant Solutions."
change the intended use of the subject device has compared to the declared predicate. As demonstrated by the substantial equivalent table, the differences raise no new safety or/and effectiveness issues, thus OsteoReady® ZircoSeal® Abutments System considered to be substantial equivalent to its predicate devices.
Risk Assessment was conducted with accordance to ISO 14971:2012 Medical devices -Application of risk management to medical devices Class II Special Controls Guidance Document.
Therefore, OsteoReady® ZircoSeal® Abutments system is considered to be substantially equivalent to its predicate devices.
Conclusion:
As verified by substantial equivalence, risk assessment and bench testing OsteoReady® ZircoSeal® Abutment System shares similarity with its predicate devices in terms of intended use, indication for use, raw material, technological characteristics and performance.
Thus OsteoReady® ZircoSeal® Abutment System is considered to be substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.