The CCM Abutment System is intended to replace missing teeth to restore chewing function. The CCM Abutment System can be placed in support of single or multiple-unit restorations including; cement retained, screw retained, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading.
Device Story
CCM Abutment System consists of Co-Cr-Mo alloy UCLA abutments, burn-out core cylinders/caps, and titanium alloy abutment screws; used with dental implants to support prosthetic restorations (crowns, bridges, overdentures). Device is non-sterile, intended for single use, and requires steam sterilization by the clinician. Abutments are straight (zero angulation) and designed for delayed loading. Clinicians (dentists/specialists) use the system in surgical/restorative procedures; the plastic component is burned out during casting. Output is a custom-fabricated final abutment providing structural support for the prosthesis, restoring chewing function for the patient.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-1, -3, -5, -10, and -11. Galvanic reaction testing conducted between Co-Cr alloy and Ti Gr.5. End-user steam sterilization validation leveraged from predicate K140806 due to identical materials, manufacturing, and packaging.
Technological Characteristics
Materials: Co-Cr-Mo alloy and Poly Diacetate (burn-out components), Ti-6AL-4V Eli (screws). Principle: Screw-retained dental abutment. Dimensions: Various diameters (4.0-6.0mm) and lengths (8.5-17mm). Connectivity: None. Sterilization: Non-sterile, steam sterilization required by user. Software: None.
Indications for Use
Indicated for partially or fully edentulous patients requiring dental implant abutments for single or multiple-unit restorations, including cement-retained, screw-retained, or fixed bridgework. Contraindicated for immediate loading.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.
March 16, 2018
InnoBioSurg Co., Ltd. April Lee Consultant Withus Group Inc. 2531 Pepperdale Drive Rowland Heights, California 91748
Re: K173120
Trade/Device Name: CCM Abutment System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: February 14, 2018 Received: February 14, 2018
## Dear April Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Andrew I. Steen -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K173120
Device Name CCM Abutment System
#### Indications for Use (Describe)
The CCM Abutment System is intended to replace missing teeth to restore chewing function. The CCM Abutment System can be placed in support of single or multiple-unit restorations including, cement retained, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
EF
PSC Publishing Services (301) 443-6740
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{4}------------------------------------------------
# 510(k) Summary
#### Submitter
InnoBioSurg Co., Ltd. Bo-reum Yoo 44-19, Techno 10-ro, Yuseong-gu, Daejeon, 34027 Republic of Korea Email: bryoo@ibsimplant.com Tel. +82-42-933-2879 Fax. +82-42-933-2881
### Official Correspondent
Withus Group Inc April Lee 106 Superior Irvine CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122
## Device Information
- Trade Name: CCM Abutment System
- Common Name: Dental Abutment System
- . Classification Name: Endosseous dental implant abutment
- Product Code: NHA
- . Panel: Dental
- Regulation Number: 872.3630
- Device Class: Class II
- Date prepared: 03/16/2018
## Primary Predicate
K153350, IBS Implant System by Innobiosurg Co., Ltd.
## Reference Predicates
- K140806,IBS Implant System by Innobiosurg Co., Ltd
- K162099, IBS Implant System II by Innobiosurg Co., Ltd.
- K161244, s-Clean OneQ-SL Narrow Implant System by Dentis Co., Ltd.
## Device Description
CCM Abutment System is used with a dental implant to provide support to prosthetic restorations such as crown, bridge and overdentures in partially or fully edentulous patients. CCM Abutment System consists of UCLA Abutment and titanium abutment screws. All subject abutments are intended to be cast at zero angulations and placed on dental implant bodies which do not require angle correction.
## UCLA Abutment
UCLA Abutments are pre-manufactured prosthetic abutments directly connected to the endosseous dental implant intended for permanent restoration, for either single or multiple tooth screw retained restorations. The lower part of the abutment which connects directly to the implant is made of Co-Cr-Mo alloy, and the upper part is made of plastic (Poly Diacetate). The plastic part is dissolved once the casting is done. The abutment screw is connected to the implant by a prosthetic screw, supplied with the abutment, and made of Ti-6AL-4V Eli. UCLA Abutments are intended for no angulation and straight implantation only. The abutments are supplied non-sterile, to be steam sterilized by the user according to the labeling, and intended for single use.
{5}------------------------------------------------
## Multi-unit CCM Cylinder
Multi-unit CCM Cylinders are used in conjunction with screw retained type Abutment to provide support for screw type final prosthesis, and for fabrication of custom abutment for screw retained restorations.
The subject device is compatible with the following Implants:
| K number | Compatible Implants |
|----------|---------------------|
| K140806 | IBS Implant system |
| K152520 | Magicore System |
| K153350 | IBS Implant System |
The CCM Abutment system is provided non-sterile and packaged separately. The abutments should be sterilized before use. Abutment screw (OIAS400) was cleared from K140806.
The UCLA abutments, Burn out core cylinder and cap are not intended to be cast at angulation or placed to provide angular correction. Plastic sleeve of UCLA Abutment can replace the Non-precious metal (Co-Cr Alloy). All finished devices are final abutments.
The diameters and lengths of the CCM Abutment System are below:
## 1) UCLA Abutment
The gingival height could be 1-4mm, angulation straight only, length of abutment post of 4-10mm. Assembled torque: 30N·cm
| No | Diameter (Ø) | Cuff | Length | Remark |
|----|--------------|------|--------|-------------|
| 1 | | 1 mm | 14 mm | |
| 2 | 4 mm | 2 mm | 15 mm | |
| 3 | | 3 mm | 16 mm | |
| 4 | | 4 mm | 17 mm | |
| 5 | | 1 mm | 14 mm | |
| 6 | 4.5 mm | 2 mm | 15 mm | |
| 7 | | 3 mm | 16 mm | |
| 8 | | 4 mm | 17 mm | Hex/Non-Hex |
| 9 | | 1 mm | 14 mm | |
| 10 | 5 mm | 2 mm | 15 mm | |
| 11 | | 3 mm | 16 mm | |
| 12 | | 4 mm | 17 mm | |
| 13 | | 1 mm | 14 mm | |
| 14 | 5.5 mm | 2 mm | 15 mm | |
| 15 | | 3 mm | 16 mm | |
| 16 | | 4 mm | 17 mm | |
{6}------------------------------------------------
## 2) Burn out core cylinder (Screw type)
No Angulation straight only, length of abutment post of 4-9mm.
#### Assembled torque: 30N·cm
| No | Diameter (Ø) | Length | Remark |
|----|--------------|---------|-------------|
| 1 | 5 mm | 8.5 mm | Hex/Non-Hex |
| 2 | 6 mm | 8.65 mm | |
### 3) Burn out core cap (Cemented type)
No Angulation straight only, length of abutment post of 4-9mm.
## Assembled torque: 30N·cm
| No | Diameter (Ø) | Length | Remark |
|----|--------------|---------|-------------|
| 1 | 5 mm | 9.15 mm | Hex/Non-Hex |
| 2 | 6 mm | 9.15 mm | |
## 4) Burn out core T cylinder (Screw type)
No Angulation straight only, length of abutment post of 4-9mm.
Assembled torque: 30N·cm
| No | Diameter (Ø) | Length | Remark |
|----|--------------|----------|-------------|
| 1 | 5 mm | 10.65 mm | Hex/Non-Hex |
| 2 | 6 mm | 10.8 mm | |
## 5) Burn out core U cylinder (Screw type)
No Angulation straight only, length of abutment post of 4-9mm.
Assembled torque: 30N·cm
| No | Diameter (Ø) | Length | Remark |
|----|--------------|----------|-------------|
| 1 | 5 mm | 12.25 mm | Hex/Non-Hex |
| 2 | 6 mm | 12.4 mm | |
## 6) Burn out core U cap (Cemented type)
No Angulation straight only, length of abutment post of 4-9mm.
Assembled torque: 30N·cm
| No | Diameter (Ø) | Length | Remark |
|----|--------------|----------|-------------|
| 1 | 5 mm | 12.95 mm | |
| 2 | 6 mm | 12.95 mm | Hex/Non-Hex |
{7}------------------------------------------------
## 7) Multi-unit CCM Cylinder
No Angulation straight only, length of abutment post of 4-9mm.
Assembled torque: 30N·cm
| No | Diameter (Ø) | Length | Remark |
|----|--------------|--------|-------------|
| 1 | 4.8 mm | 12 mm | Hex/Non-Hex |
## Indication for Use
The CCM Abutment System is intended to replace missing teeth to restore chewing function. The CCM Abutment System can be placed in support of single or multiple-unit restorations including; cement retained, screw retained, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading.
## Non-Clinical Data:
The subject device was tested to evaluate its substantial equivalence according to the following standards.
- Biocompatibility tests according to ISO 10993-1:2009, ISO 10993-3:2014, ISO 10993-5:2009, ● ISO 10993-10:2010 and ISO 10993-11:2006.
- Galvanic Reaction Test between CoCr alloy and non-precious metal ●
Below tests were performed for predicate devices and leveraged for the subject device:
- End User Steam Sterilization Test according to ISO 17665-1:2006, -2:2009 and ANSI/AAMI ● ST79 referenced in K140806
The results of the above tests have met the criteria of the standards, and demonstrated the substantial equivalence with the predicate device.
Biocompatibility Test was conducted for the subject devices and it demonstrates that the subject device is biocompatible and substantial equivalence with the predicate device, K161244.
The galvanic reaction testing between the CoCr alloy and non-precious dental alloys, Ti Gr.5. was performed for subject device and compatible implants. As the FDA guidance of "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implant and Endosseous Dental Implant abutments", corrosion potential of each CoCr alloy and Ti Gr.5 and couple potential for assembled CoCr alloy and Ti Gr.5 were assessed.
The end user sterilization test was performed for predicate device, K140806 and leveraged for the subject device because the product category, material, manufacturing process, facility, and packaging is the exactly the same as the predicate, K140806.
The non-clinical testing results demonstrate that the subject device is substantially equivalent to the predicate device.
{8}------------------------------------------------
# Comparison to Predicate Devices:
| UCLA Abutment | | | |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Subject device | Primary Predicate device | Reference Predicate device |
| | CCM Abutment System | IBS Implant system | s-Clean OneQ-SL Narrow Implant System |
| 510k | N/A | K153350 | K161244 |
| Part Name | UCLA Abutment<br>(Hex, Non-Hex) | UCLA Abutment | S-clean CCM Abutment<br>(Narrow) |
| Material | Co-Cr-Mo Alloy<br>Poly Diacetate | Titanium Alloy<br>Poly Diacetate | Co-Cr-Mo Alloy |
| Manufacturer | Innobiosurg Co.,Ltd | Innobiosurg Co.,Ltd | Dentis Co.,Ltd |
| Indications for use | The CCM Abutment System is<br>intended to replace missing<br>teeth to restore chewing<br>function. The CCM Abutment<br>System can be placed in<br>support of single or multiple-<br>unit restorations including;<br>cement retained, screw<br>retained, and terminal or<br>immediate abutment support<br>for fixed bridgework. This<br>system is for one or two stage<br>surgical procedures. This<br>system is intended for delayed<br>loading. | The IBS Implant System is<br>intended to replace missing<br>teeth to restore chewing<br>function. The IBS Implant<br>can be placed in support of<br>single or multiple-unit<br>restorations including;<br>cement retained, screw<br>retained, or overdenture<br>restorations, and terminal or<br>immediate abutment support<br>for fixed bridgework. This<br>system is for one or two<br>stage surgical procedures.<br>This system is intended for<br>delayed loading. | The s-Clean One-Q-SL Narrow Implant System<br>(3.0, 3.3mm) may be used as an artificial root<br>structure for single tooth replacement of<br>mandibular central and lateral incisors and<br>maxillary lateral incisors. The implants may be<br>restored immediately<br>1) with a temporary prosthesis that is not in<br>functional occlusion,<br>2) when splinted together as an artificial root<br>structure for multiple tooth replacement of<br>mandibular incisors, or<br>3) for denture stabilization using multiple<br>implants in the anterior mandible and maxilla.<br>The implants may be placed in immediate<br>function when good primary stability has been<br>achieved and with appropriate occlusal loading. |
| Dimension | Ø 4mm, Ø 4.5mm, Ø 5mm, Ø<br>5.5mm (D) X 14,15,16,17 mm<br>(L) | Ø3.5mm, Ø4mm, Ø4.5mm,<br>Ø5mm, Ø5.5mm (D) X<br>14,15,16,17 mm (L) | Ø 4.0 mm (D) x 14.5<br>and 15 mm (L) |
{9}------------------------------------------------
| Restoration angulations | No Angle | No Angle | No Angle |
|-------------------------|---------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Design | | | |
| Principle of Operation | a screw retained restoration<br>type of abutment using a screw<br>to fix a prosthesis | a screw retained restoration<br>type of abutment using a<br>screw to fix a prosthesis | a screw retained restoration type of abutment<br>using a screw to fix a prosthesis |
| Burn out core cylinder (Hex and Non-Hex) | | | |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Subject device | Primary Predicate device | Reference Predicate device |
| | CCM Abutment System | IBS Implant system | s-Clean OneQ-SL Narrow Implant System |
| 510k | N/A | K153350 | K161244 |
| Part Name | Burn out core cylinder | UCLA Abutment | S-clean CCM Abutment<br>(Narrow) |
| Material | Co-Cr-Mo Alloy<br>Poly Diacetate | Titanium Alloy<br>Poly Diacetate | Co-Cr-Mo Alloy |
| Manufacturer | Innobiosurg Co.,Ltd | Innobiosurg Co.,Ltd | Dentis Co.,Ltd |
| Indications for use | The CCM Abutment System is<br>intended to replace missing teeth to<br>restore chewing function. The<br>CCM Abutment System can be<br>placed in support of single or<br>multiple-unit restorations<br>including; cement retained, screw<br>retained, and terminal or<br>immediate abutment support for<br>fixed bridgework. This system is<br>for one or two stage surgical | The IBS Implant System is<br>intended to replace missing teeth<br>to restore chewing function. The<br>IBS Implant can be placed in<br>support of single or multiple-unit<br>restorations including; cement<br>retained, screw retained, or<br>overdenture restorations, and<br>terminal or immediate abutment<br>support for fixed bridgework.<br>This system is for one or two | The s-Clean One-Q-SL Narrow Implant<br>System (3.0, 3.3mm) may be used as an<br>artificial root structure for single tooth<br>replacement of mandibular central and lateral<br>incisors and maxillary lateral incisors. The<br>implants may be restored immediately<br>1) with a temporary prosthesis that is not in<br>functional occlusion,<br>2) when splinted together as an artificial root<br>structure for multiple tooth replacement of<br>mandibular incisors, or |
| | procedures. This system is<br>intended for delayed loading. | stage surgical procedures. This<br>system is intended for delayed<br>loading. | 3) for denture stabilization using multiple<br>implants in the anterior mandible and<br>maxilla.<br>The implants may be placed in immediate<br>function when good primary stability has<br>been achieved and with appropriate occlusal<br>loading. |
| Principle of<br>Operation | a screw retained restoration type of<br>abutment using a screw to fix a<br>prosthesis | a screw retained restoration type<br>of abutment using a screw to fix a<br>prosthesis | a screw retained restoration type of abutment<br>using a screw to fix a prosthesis |
| Hex | | | |
| Dimension | Ø 5mm, Ø 6mm (D) X 8.5, 8.65<br>mm (L) | Ø3.5mm, Ø4mm, Ø4.5mm,<br>Ø5mm, Ø5.5mm (D) X<br>14,15,16,17 mm (L) | Ø 4.0 mm (D) x 14.5<br>and 15 mm (L) |
| Restoration<br>angulations | No Angle | No Angle | No Angle |
| Design | Image: abutment | Image: abutment | Image: abutment |
| Non-Hex | | | |
| Dimension | Ø 5mm, Ø 6mm (D) X 8.5, 8.65<br>mm (L) | Ø3.5mm, Ø4mm, Ø4.5mm,<br>Ø5mm, Ø5.5mm (D) X<br>14,15,16,17 mm (L) | Ø 4.0 mm (D) x 14.5<br>and 15 mm (L) |
| Restoration<br>angulations | No Angle | No Angle | No Angle |
| Design | Image: abutment | Image: abutment | Image: abutment |
| Burn out core cap | | | |
| Device Name | Subject device | Primary Predicate device | Reference Predicate device |
| | CCM Abutment System | IBS Implant system | s-Clean OneQ-SL Narrow Implant System |
| 510k | N/A | K153350 | K161244 |
| Part Name | Burn out core cap | UCLA Abutment | S-clean CCM Abutment<br>(Narrow) |
| Material | Co-Cr-Mo Alloy<br>Poly Diacetate | Titanium Alloy<br>Poly Diacetate | Co-Cr-Mo Alloy |
| Manufacturer | Innobiosurg Co.,Ltd | Innobiosurg Co.,Ltd | Dentis Co.,Ltd |
| Indications for use | The CCM Abutment System is<br>intended to replace missing teeth<br>to restore chewing function. The<br>CCM Abutment System can be<br>placed in support of single or<br>multiple-unit restorations<br>including; cement retained,<br>screw retained, and terminal or<br>immediate abutment support for<br>fixed bridgework. This system is<br>for one or two stage surgical<br>procedures. This system is<br>intended for delayed loading. | The IBS Implant System is intended<br>to replace missing teeth to restore<br>chewing function. The IBS Implant<br>can be placed in support of single or<br>multiple-unit restorations including;<br>cement retained, screw retained, or<br>overdenture restorations, and<br>terminal or immediate abutment<br>support for fixed bridgework. This<br>system is for one or two stage<br>surgical procedures. This system is<br>intended for delayed loading. | The s-Clean One-Q-SL Narrow Implant<br>System (3.0, 3.3mm) may be used as an<br>artificial root structure for single tooth<br>replacement of mandibular central and lateral<br>incisors and maxillary lateral incisors. The<br>implants may be restored immediately<br>1) with a temporary prosthesis that is not in<br>functional occlusion,<br>2) when splinted together as an artificial root<br>structure for multiple tooth replacement of<br>mandibular incisors, or<br>3) for denture stabilization using multiple<br>implants in the anterior mandible and maxilla.<br>The implants may be placed in immediate<br>function when good primary stability has been<br>achieved and with appropriate occlusal<br>loading. |
| Principle of<br>Operation | a screw retained restoration type<br>of abutment using a screw to fix<br>a prosthesis | a screw retained restoration type of<br>abutment using a screw to fix a<br>prosthesis | a screw retained restoration type of abutment<br>using a screw to fix a prosthesis |
| Hex | | | |
| Dimension | Ø 5mm, Ø 6mm (D) X 9.15 mm<br>(L) | Ø3.5mm, Ø4mm, Ø4.5mm, Ø5mm,<br>Ø5.5mm (D) X 14,15,16,17 mm (L) | Ø 4.0 mm (D) x 14.5<br>and 15 mm (L)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.