The Magic UCLA Abutment System is intended to replace missing teeth to restore chewing function. The Magic UCLA Abutment System can be placed in support of single or multiple-unit restorations including: cement retained, screw retained, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading.
Device Story
The Magic UCLA Abutment System consists of dental abutments (screw-retained and cement-retained types, hex and non-hex) used to connect dental implants to prosthetic restorations like crowns, bridges, and overdentures. The device is used in clinical settings by dental professionals. It functions as a mechanical interface; the abutment is attached to the implant via a screw, providing a base for the final prosthesis. The system is provided non-sterile and requires steam sterilization by the end-user before placement. It is not intended for angulation correction. The device benefits patients by restoring chewing function in edentulous or partially edentulous cases.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing leveraged from predicate devices, including biocompatibility (ISO 10993 series), galvanic reaction testing (CoCr and Ti Grade 5), and steam sterilization validation (ISO 17665, ANSI/AAMI ST79).
Indicated for partially or fully edentulous patients requiring replacement of missing teeth to restore chewing function via single or multiple-unit restorations (cement/screw retained, fixed bridgework). Suitable for one or two-stage surgical procedures with delayed loading.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 3, 2020
InnoBioSurg Co., Ltd. % April Lee Consultant Withus Group Inc. 106 Superior Irvine, California 92620
Re: K202418
Trade/Device Name: Magic UCLA Abutment System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: November 3, 2020 Received: November 3, 2020
Dear April Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew Steen Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K202418
Device Name Magic UCLA Abutment System
#### Indications for Use (Describe)
The Magic UCLA Abutment System is intended to replace missing teeth to restore chewing function. The Magic UCLA Abutment System can be placed in support of single or multiple-unit restorations including: cement retained, screw retained, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;"><b>Prescription Use</b></span> (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
#### Submitter
InnoBioSurg Co., Ltd. Sun-Mi Park 44-19, Techno 10-ro, Yuseong-gu Daejeon, 34027 Republic of Korea Email: sumpark@ibsimplant.com Tel. +82-42-933-2879 Fax. +82-42-933-2881
### Official Correspondent
Withus Group Inc. April Lee 106 Superior Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122
### Device Information
- Trade Name: Magic UCLA Abutment System ●
- Common Name: Dental Abutment System
- Classification Name: Endosseous dental implant abutment
- Product Code: NHA
- Panel: Dental
- Regulation Number: 872.3630
- Device Class: Class II
- Date prepared: 12/02/2020
#### Predicate Devices:
The subject device is substantially equivalent to the following predicate devices:
#### Primary Predicate
K173120, CCM Abutment System manufactured by InnoBioSurg Co., Ltd.
#### Reference Device
K153350, IBS Implant System manufactured by InnoBioSurg Co., Ltd.
{4}------------------------------------------------
### General Description
The purpose of this submission is to add abutments to the previously cleared device. CCM Abutment System (K173120). The added abutments are below:
Abutments
- . Magic UCLA Screw Retained Type (Hex, Non-Hex)
- Magic UCLA Cement Retained Type (Hex, Non-Hex) ●
The Magic UCLA Abutment System is used with a dental implant to provide support to prosthetic restorations such as crowns, bridges, and overdentures in partially or fully edentulous patients.
The subject device is compatible with the following implants:
| K Number | Compatible Implant |
|----------|---------------------------------|
| K152520 | Magicore System A L A P S M A P |
The dimension ranges of the abutments are below:
| Abutments | Diameter<br>(mm) | Length (mm) |
|---------------------------------|------------------|-------------|
| Magic UCLA Screw Retained Type | 4.5 | 10.2 |
| | 5.2 | 10.4 |
| | 6.2 | 10.5 |
| Magic UCLA Cement Retained Type | 4.5 | 10.5 |
| | 5.2 | 11.1 |
| | 6.2 | 11.1 |
Tolerance of dimensions for Abutments shall be within ±1%.
The Magic UCLA Abutment System is provided non-sterile and packaged separately. The abutments should be sterilized before use. The Magic UCLA Screw Retained Type and the Magic UCLA Cement Retained Type are not intended to be cast at angulation or placed to provide angular correction.
#### Indications for Use
The Magic UCLA Abutment System is intended to replace missing teeth to restore chewing function. The Magic UCLA Abutment System can be placed in support of single or multiple-unit restorations including: cement retained, screw retained, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading.
#### Materials
Abutments are fabricated from Co-Cr-Mo Alloy with Poly Diacetate.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for IBS Implant. The logo is primarily blue and features a stylized "IBS" with a yellow star-like shape above the "I". Below the blue "IBS", the word "IMPLANT" is written in a smaller, sans-serif font.
# InnoBioSurg Co., Ltd. 44-19, Techno 10-ro, Yuseong-gu, Daejeon, 34027, Korea
Tel. 82-42-933-2879 / Fax. 82-42-933-2881
### Summaries of Technological Characteristics:
| Magic UCLA Screw Retained Type (Hex, Non-Hex) | | | |
|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Subject Device | Primary Predicate Device | Reference Predicate Device |
| | Magic UCLA Abutment System | CCM Abutment System | IBS Implant system |
| 510k | K202418 | K173120 | K153350 |
| Part Name | Magic UCLA Screw Retained Type | Burn out core cylinder | UCLA Abutment |
| Materials | Co-Cr-Mo Alloy<br>Poly Diacetate | Co-Cr-Mo Alloy<br>Poly Diacetate | Titanium Alloy<br>Poly Diacetate |
| Manufacturer | InnoBioSurg Co., Ltd. | InnoBioSurg Co., Ltd. | InnoBioSurg Co., Ltd. |
| Indications for Use | The Magic UCLA Abutment System is<br>intended to replace missing teeth to<br>restore chewing function. The Magic<br>UCLA Abutment System can be placed<br>in support of single or multiple-unit<br>restorations including: cement retained,<br>screw retained, and terminal or<br>immediate abutment support for fixed<br>bridgework. This system is for one or<br>two stage surgical procedures. This<br>system is intended for delayed loading. | The CCM Abutment System is intended<br>to replace missing teeth to restore<br>chewing function. The CCM Abutment<br>System can be placed in support of<br>single or multiple-unit restorations<br>including: cement retained, screw<br>retained, and terminal or immediate<br>abutment support for fixed bridgework.<br>This system is for one or two stage<br>surgical procedures. This system is<br>intended for delayed loading. | The IBS Implant System is intended to<br>replace missing teeth to restore chewing<br>function. The IBS Implant can be placed in<br>support of single or multiple-unit<br>restorations including: cement retained,<br>screw retained, or overdenture restorations,<br>and terminal or immediate abutment support<br>for fixed bridgework. This system is for one<br>or two stage surgical procedures. This<br>system is intended for delayed loading. |
| Principle of Operation | A screw retained restoration type of<br>abutment using a screw to fix a<br>prosthesis. | A screw retained restoration type of<br>abutment using a screw to fix a<br>prosthesis. | A screw retained restoration type of<br>abutment using a screw to fix a prosthesis. |
| Hex | | | |
| Dimensions | 4.5, 5.2, 6.2mm (Ø) X 10.2, 10.4,<br>10.5mm (L) | 5, 6mm (Ø) X 8.5, 8.65mm (L) | 3.5, 4, 4.5, 5, 5.5mm (Ø) X 14, 15,<br>16, 17mm (L) |
| Restoration angulations | No Angle | No Angle | No Angle |
| Design | | | |
| Non-Hex | | | |
| Dimensions | 4.5, 5.2, 6.2mm (Ø) X 10.2, 10.4,<br>10.5mm (L) | 5, 6mm (Ø) X 8.5, 8.65mm (L) | 3.5, 4, 4.5, 5, 5.5mm (Ø) X 14, 15,<br>16, 17mm (L) |
| Restoration angulations | No Angle | No Angle | No Angle |
| Design | | | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for IBS Implant. The logo consists of the letters "IBS" in a stylized, blocky font, with a small yellow star above the "I". Below the letters, the word "IMPLANT" is written in a smaller, sans-serif font. The letters "IBS" are in blue.
# InnoBioSurg Co., Lt
44-19, Techno 10-ro, Yuseong-gu, Daejeon, 34027, Korea Tel. 82- 42-933-2879 / Fax. 82-42-933-2881
#### Similarities
The Magic UCLA Screw Retained Type is substantially equivalent in indications for use, fundamental scientific technology, principle of operation, functions, and materials as the identified predicates.
#### Differences
Compared to the primary predicate device, new abutments with various dimensions are added to the subject system. To support the dimension differences between the primary predicate device, K153350 was added as the reference device. The dimensions of the subject device are in range of the ceference device and the slight difference does not affect the application and fundamental scientific technology of the device. Therefore, it is concluded that the Magic UCLA Screw Retained Type is substantially equivalent with the predicate devices.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo for a company called "IBS Implant". The logo consists of the letters "IBS" stacked on top of each other in a stylized font, with the word "IMPLANT" written below. The letters "IBS" are dark blue, and there is a yellow star-like shape above the "I". The word "IMPLANT" is also dark blue.
InnoBioSurg Co., Ltd.
44-19, Techno 10-ro, Yuseong-gu, Daejeon, 34027, Korea
Tel. 82- 42-933-2879 / Fax. 82-42-933-2881
| Magic UCLA Cement Retained Type (Hex, Non-Hex) | | | |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device | Primary Predicate Device | Reference Predicate Device |
| Device Name | Magic UCLA Abutment System | CCM Abutment System | IBS Implant system |
| 510k | K202418 | K173120 | K153350 |
| Part Name | Magic UCLA Cement Retained Type | Burn out core cap | UCLA Abutment |
| Materials | Co-Cr-Mo Alloy<br>Poly Diacetate | Co-Cr-Mo Alloy<br>Poly Diacetate | Titanium Alloy<br>Poly Diacetate |
| Manufacturer | InnoBioSurg Co., Ltd. | InnoBioSurg Co., Ltd. | InnoBioSurg Co., Ltd. |
| Indications for Use | The Magic UCLA Abutment System is<br>intended to replace missing teeth to<br>restore chewing function. The Magic<br>UCLA Abutment System can be placed<br>in support of single or multiple-unit<br>restorations including: cement retained,<br>screw retained, and terminal or<br>immediate abutment support for fixed<br>bridgework. This system is for one or<br>two stage surgical procedures. This<br>system is intended for delayed loading. | The CCM Abutment System is intended<br>to replace missing teeth to restore<br>chewing function. The CCM Abutment<br>System can be placed in support of<br>single or multiple-unit restorations<br>including: cement retained, screw<br>retained, and terminal or immediate<br>abutment support for fixed bridgework.<br>This system is for one or two stage<br>surgical procedures. This system is<br>intended for delayed loading. | The IBS Implant System is intended to<br>replace missing teeth to restore chewing<br>function. The IBS Implant can be placed in<br>support of single or multiple-unit<br>restorations including: cement retained,<br>screw retained, or overdenture restorations,<br>and terminal or immediate abutment suppor<br>for fixed bridgework. This system is for one<br>or two stage surgical procedures. This<br>system is intended for delayed loading. |
| Principle of Operation | A screw retained restoration type<br>of abutment using a screw to fix a<br>prosthesis. | A screw retained restoration type<br>of abutment using a screw to fix a<br>prosthesis. | A screw retained restoration type of<br>abutment using a screw to fix a prosthesis. |
| Hex | | | |
| Dimensions | 4.5, 5.2, 6.2mm (Ø) X 10.5,<br>11.1mm (L) | 5, 6mm (Ø) X 9.15mm (L) | 3.5, 4,4.5,5, 5.5mm (Ø) X 14, 15,<br>16, 17mm (L) |
| Restoration angulations | No Angle | No Angle | No Angle |
| Design | | | |
| Image: dental implant | Image: dental implant | Image: dental implants | |
| Non-Hex | | | |
| Dimensions | 4.5, 5.2, 6.2mm (Ø) X 10.5,<br>11.1mm (L) | 5, 6mm (Ø) X 9.15 mm (L) | 3.5,4,4.5, 5, 5.5mm (Ø) X 14, 15,<br>16, 17mm (L) |
| Restoration angulations | No Angle | No Angle | No Angle |
| Design | | | |
| Image: dental implant | Image: dental implant | Image: dental implants | |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a logo for IBS Implant. The logo is composed of the letters "IBS" stacked on top of each other in a stylized, sans-serif font. The letters are dark blue, and there is a yellow star-like shape in the upper-left corner of the "I". Below the letters, the word "IMPLANT" is written in a smaller, sans-serif font, also in dark blue.
# InnoBioSurg Co., Ltd
44-19, Techno 10-ro, Yuseong-gu, Daejeon, 34027, Korea Tel. 82- 42-933-2879 / Fax. 82-42-933-2881
#### Similarities
The Magic UCLA Cement Retained Type is substantially equivalent in indications for use, fundamental scientific technology, principle of operation, functions, and materials as the identified predicates.
#### Differences
Compared to the primary predicate device, new abutments with various dimensions are added to the subject system. To support the dimension differences between the primary predicate device. K153350 was added as the reference device. The dimensions of the subject device are in range of the reference device and the slight difference does not affect the application and fundamental scientific technology of the device, it is concluded that the Magic UCLA Cement Retained Type is substantially equivalent with the predicate devices.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for IBS Implant. The logo is primarily blue and features the letters "IBS" stacked on top of each other. The "I" has a yellow star-like shape in the upper left corner. Below the letters, the word "IMPLANT" is written in a smaller, sans-serif font.
#### Non-Clinical Data
Non-clinical testing was not performed for the subject device because the subject device and predicate devices are substantially equivalent in indications, fundamental scientific technology, materials, and design. By performing the product's design controls activity summary, it is concluded that any differences do not impact the product's safety and effectiveness and demonstrate substantial equivalence.
The tests below were performed for the predicate devices and leveraged for the subject device:
- Biocompatibility tests according to ISO 10993-1:2009, ISO 10993-3:2014, ISO 10993-5:2009, . ISO 10993-10:2010, and ISO 10993-11:2006 referenced in K173120
- Galvanic Reaction Test between CoCr alloy and non-precious metals referenced in K173120 ●
- End User Steam Sterilization Test according to ISO 17665-1:2006, -2:2009, and ● ANSI/AAMI ST79 referenced in K192197
The results of the above tests have met the criteria of the standards, and demonstrated substantial equivalence with the predicate devices.
The Biocompatibility Test was conducted on the primary predicate and leveraged for the subject device because both products are manufactured with the same materials and undergo the same manufacturing process. It demonstrates that the subject device is biocompatible and substantially equivalent with the predicate.
The galvanic reaction testing between the CoCr alloy and non-precious dental alloys, Ti Grade 5, was performed on the primary predicate and compatible implants. Per FDA guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments", corrosion potential of each CoCr and Ti Grade 5 alloy and couple potential for assembled CoCr and Ti Grade 5 alloy were assessed.
The end user sterilization test was performed for the predicate device. K192197, and leveraged for the subject device because the product category, materials, manufacturing process, facility, and packaging is the exactly the same as the predicate, K192197.
Mechanical testing such as fatigue testing was not performed for the subject device because per FDA Guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments", the Magic UCLA Abutment System is not intended to be cast at angulation or placed to provide angular correction.
The non-clinical testing results demonstrate that the subject device is substantially equivalent to the predicate devices.
#### Conclusions
The Magic UCLA Abutment System, subject device of this submission, constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental scientific technology as the predicate devices. Therefore, the Magic UCLA Abutment System is substantially equivalent.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.