ST Internal Implant System

K192347 · Megagen Implant Co., Ltd. · DZE · Nov 14, 2019 · Dental

Device Facts

Record IDK192347
Device NameST Internal Implant System
ApplicantMegagen Implant Co., Ltd.
Product CodeDZE · Dental
Decision DateNov 14, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Intended Use

The ST Internal Implant System is intended to be surgically placed in the maxillary or mandibular arches for the purpose providing prosthetic support for dental restorations (Crown, bridges, and overdentures) in partially or fully edentulous individuals. It is used to restore a patient's chewing function. Smaller implants (less than 6.0 mm) are dedicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Larger implants are dedicated for the molar region and are indicated for delayed loading.

Device Story

The ST Internal Implant System consists of endosseous dental implants (fixtures) and various abutments (EZ Post, Solid, Angled, Milling, Healing, Temporary) and screws. Fixtures are surgically placed into the maxillary or mandibular jawbone to provide prosthetic support for dental restorations. The system replaces missing teeth to restore chewing function. The device is used by dental professionals in a clinical setting. The fixture integrates with the bone (osseointegration), and the abutment is secured to the fixture via an abutment screw to support the final prosthesis (crown, bridge, or overdenture). The system allows for immediate loading of smaller implants when primary stability is achieved, while larger implants are used for delayed loading. The device benefits patients by restoring oral function and aesthetics.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including static compression-strength and fatigue testing conducted in accordance with ISO 14801 and FDA Class II Special Controls guidance.

Technological Characteristics

Materials: CP Ti Grade 4 (fixtures, ASTM F67), Ti-6Al-4V ELI (abutments/screws, ASTM F136). Surface treatment: SLA (fixtures), Anodizing (abutments). Connection: Internal Hex/Conical. Sterilization: Gamma (fixtures), moist heat/autoclave (abutments). Dimensions: Various diameters (3.7-6.8mm) and lengths. Connectivity: None (mechanical).

Indications for Use

Indicated for partially or fully edentulous patients requiring prosthetic support for dental restorations (crowns, bridges, overdentures) in maxillary or mandibular arches. Smaller implants (<6.0 mm) indicated for immediate loading given good primary stability; larger implants indicated for molar region with delayed loading.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. November 14, 2019 Megagen Implant Co. Ltd YouJung Kim Chief Researcher 45, Secheon-ro 7-gil, Dasa-eup, Dalseong-gun Daegu. 42921 REPUBLIC OF KOREA Re: K192347 Trade/Device Name: ST Internal Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: August 14, 2019 Received: August 29, 2019 # Dear YouJung Kim: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Srinivas Nandkumar, Ph.D. Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K192347 Device Name ST Internal Implant System # Indications for Use (Describe) The ST Internal Implant System is intended to be surgically placed in the maxillary or mandibular arches for the purpose providing prosthetic support for dental restorations (Crown, bridges, and overdentures) in partially or fully edentulous individuals. It is used to restore a patient's chewing function. Smaller implants (less than 6.0 mm) are dedicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Larger implants are dedicated for the molar region and are indicated for delayed loading. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <div><span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div>□ Over-The-Counter Use (21 CFR 801 Subpart C)</div> | |-------------------------------------------------------------------------------------------------|----------------------------------------------------------| |-------------------------------------------------------------------------------------------------|----------------------------------------------------------| CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ {3}------------------------------------------------ # 510(k) Summary (K192347) This summary of 510(K) is being submitted in accordance with requirements of 21 CFR Part 807.92. ### Date: November 13, 2019 # 1. MegaGen Implant Co., Ltd. 45, Secheon-ro, 7-gil, Dasa-eup, Dalesong-gun, Daegu, Republic of Korea Tel: + 82-53-222-2828 ### 2. Submission Correspondent YouJung Kim MegaGen Implant Co., Ltd. 45, Secheon-ro, 7-gil, Dasa-eup, Dalesong-gun, Daegu, Republic of Korea Tel: +82-53-222-2985 Fax: +82-53-289-3420 Email: rnd ra4@imegagen.com # 3. Device | ■ Trade Name: | ST Internal Implant System | | |--------------------------------|--------------------------------|--| | ■ Common Name: | Endosseous Dental Implant | | | ■ Classification Name: | Implant, Endosseous, Root-Form | | | ■ Classification Product Code: | DZE | | | ■ Subsequent Product Code: | NHA | | | ■ Classification regulation: | Class II, 21 CFR 872.3640 | | ### 4. Predicate Device: - " Primary Predicate Device: K182448 – AnyRidge Octa 1 Implant System - . Reference Devices: K123988 – AnyOne Internal Implant System {4}------------------------------------------------ # 5. Description: The ST Internal Fixture is a substructure of a dental implant system made of CP Ti Grade 4 with the surface treated by SLA method. The abutment is a superstructure of a dental implant system which is attached to the implants made of Ti-6A1-4V ELI with the surface treated by anodizing method(except healing and temporary abutment). The ST Internal Implant System is intended to be placed in the upper or lower jaw to support prosthetic devices, such as artificial teeth, and to restore a patient's chewing function. The system is consisted of the following components. | Component | Content | |------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ST Internal<br>Fixture | Description<br>ST Internal Fixtures are made of titanium. The osseointegrated implant fixtures placed in the anterior or posterior site of maxillary or mandibular jawbone considering bone quality and bone quantity. The dental implants which used in conjunction with other prosthetic restore lost chewing ability, improve appearance.<br>Material Composition<br>CP Ti Grade 4<br>Dimension<br>(Diameter &Length)<br>Ø 3.70 x 8.50, 10.00, 11.50, 13.00, 15.00 mm<br>Ø 4.20 x 7.00, 8.50, 10.00, 11.50, 13.00, 15.00 mm<br>Ø 4.60 x 7.00, 8.50, 10.00, 11.50, 13.00, 15.00 mm<br>Ø 5.10 x 7.00, 8.50, 10.00, 11.50, 13.00, 15.00 mm<br>Ø 6.00 x 7.00, 8.50, 10.00, 11.50, 13.00 mm<br>Ø 6.80 x 7.00, 8.50, 10.00, 11.50, 13.00 mm | | EZ Post<br>Abutment | Description<br>EZ Post Abutment is used in conjunction with fixture to provide support for cement and screw retained type final prosthesis.<br>Material Composition<br>Ti-6A1-4V ELI<br>Dimension<br>(Diameter &Length)<br>Ø 4.60 x 5.50, 7.00 mm<br>Ø 5.00 x 4.00, 5.50, 7.00 mm<br>Ø 6.00 x 4.00, 5.50, 7.00 mm<br>Ø 7.00 x 5.50 mm<br>Angulation<br>0° | | Solid<br>Abutment | Description<br>Solid Abutment is used in conjunction with fixture to provide support for final prosthesis, and used in cement retained type prosthesis only.<br>Material Composition<br>Ti-6A1-4V ELI<br>Dimension<br>(Diameter &Length)<br>Ø 4.00 x 10.00, 11.00, 11.50, 12.00, 12.50, 13.00, 13.50, 14.00,<br>14.50, 15.00, 15.50, 16.00, 17.00 mm<br>Ø 4.00 x 10.40, 11.40, 11.90, 12.40, 12.90, 13.40, 13.90, 14.40,<br>14.90, 15.40, 15.90, 16.40, 17.40 mm<br>Ø 4.60 x 10.00, 11.00, 11.50, 12.00, 12.50, 13.00, 13.50, 14.00,<br>14.50, 15.00, 15.50, 16.00, 17.00 mm<br>Ø 4.60 x 10.40, 11.40, 11.90, 12.40, 12.90, 13.40, 13.90, 14.40,<br>14.90, 15.40, 15.90, 16.40, 17.40 mm<br>Ø 5.00 x 10.40, 11.40, 11.9কিন্তo, 12.40, 12.90, 13.40, 13.90, 14.40,<br>14.90, 15.40, 15.90, 16.40, 17.40 mm<br>Ø 6.00 x 10.40, 11.40, 11.90, 12.40, 12.90, 13.40, 13.90, 14.40,<br>14.90, 15.40, 15.90, 16.40, 17.40 mm<br>Ø 7.00 x 11.90, 12.90, 13.90, 14.90, 15.90 mm<br>Angulation<br>0° | | Angled<br>Abutment | Description<br>Angled Abutment is used for correcting the prosthetic angulation of implant.<br>Material Composition<br>Ti-6A1-4V ELI<br>Dimension<br>(Diameter &Length)<br>Ø 4.30 x 12.60, 14.60 mm<br>Ø 4.50 x 12.50, 14.50 mm<br>Ø 5.00 x 12.50, 14.50 mm<br>Ø 6.00 x 12.50, 14.50 mm<br>Angulation<br>17° | {5}------------------------------------------------ | Milling<br>Abutment | Description | Milling Abutment is used in conjunction with fixture to provide<br>support for cement and screw retained type final prosthesis and used<br>for establishing an adequate safety margin from occlusal line by hand<br>milling of the post part. | |-----------------------|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Material Composition | Ti-6A1-4V ELI | | | Dimension<br>(Diameter &Length) | Ø 4.00 x 14.50, 14.60 mm<br>Ø 5.00 x 12.00, 14.50 mm<br>Ø 6.00 x 14.50 mm<br>Ø 7.00 x 14.50 mm | | | Angulation | 0° | | Cover<br>Screw | Description | Cover Screw is used for protecting the inner structure of a fixture and<br>the exposed fixture platform after fixture placement. | | | Material Composition | Ti-6A1-4V ELI | | | Dimension<br>(Diameter &Length) | Ø 3.10 x 5.30mm<br>Ø 3.60 x 5.90mm | | | Angulation | 0° | | Healing<br>Abutment | Description | Healing Abutment helps to form suitable emergence profile during<br>period of gingival healing. | | | Material Composition | Ti-6A1-4V ELI | | | Dimension<br>(Diameter &Length) | Ø 4.30 x 7.50, 8.50, 8.60, 9.50, 9.60, 10.60, 11.50, 12.60, 13.50,<br>14.60 mm<br>Ø 4.80 x 7.50, 8.50, 8.60, 9.50, 9.60, 10.60, 11.50, 12.60, 13.50,<br>14.60 mm<br>Ø 5.30 x 8.60, 9.60, 10.60, 12.60, 14.60 mm<br>Ø 6.30 x 8.60, 9.60, 10.60, 12.60, 14.60 mm<br>Ø 7.30 x 8.60, 9.60, 10.60, 12.60, 14.60 mm | | | Angulation | 0° | | | | | | Temporary<br>Abutment | Description | Temporary Abutment is used in conjunction with fixture to provide<br>support for provisional restoration. Temporary Abutment has knurled<br>surface on the top part, which allows for better retention of resin or<br>wax. | | | Material Composition | Ti-6A1-4V ELI | | | Dimension<br>(Diameter &Length) | Ø 4.00 x 13.60, 15.60 mm<br>Ø 4.50 x 13.50, 15.50 mm | | | Angulation | 0° | | Abutment<br>Screw | Description | Abutment Screw is used for securing the abutment to the endosseous<br>implant. | | | Material Composition | Ti-6A1-4V ELI | | | Dimension<br>(Diameter &Length) | Ø 2.15 x 10.20 mm<br>Ø 2.35 x 8.40 mm | ### 6. Indication for use: The ST Internal Implant System is intended to be surgically placed in the maxillary or mandibular arches for the purpose providing prosthetic support for dental restorations (Crown, bridges, and overdentures) in partially or fully edentulous individuals. It is used to restore a patient's chewing function. Smaller implants (less than 6.0 mm) are dedicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Larger implants are dedicated for the molar region and are indicated for delayed loading. {6}------------------------------------------------ # 7. Basis for Substantial Equivalence The ST Internal Implant System is substantially equivalent to the predicate device in terms of indication for use, technical characteristics and function. They are material and have similar design. The size range and design of the subject device slightly differ from the predicate device however it is very minor not affecting substantial equivalence. In order to demonstrate the difference in design does not raise any new issues in safety and effectiveness, each performance test on the subject and predicate device have been performed to figure out the physical property(e.g. fatigue limit) with combination of the fixture and the angled abutment, in consideration of the worst case according to "ISO 14801:2007" and "Section 8 of Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutment". The test result shows that the subject device is better able to withstand the fatigue and loading. Based on the detailed comparison charts below and test results provided in this submission, we conclude that the subject device is substantially equivalent to the predicate devices. | | Subject Device | Predicate Device | Reference Device | |---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k)<br>Number | Not yet | K182448 | K123988 | | Device Name | ST Internal Fixture | AnyRidge Octa 1 Implant System | AnyOne Internal Implant System | | Manufacturer | MegaGen Implant Co., Ltd. | MegaGen Implant Co., Ltd. | MegaGen Implant Co., Ltd. | | Indications<br>for Use<br>Statement | The ST Internal Implant System is<br>intended to be surgically placed in<br>the maxillary or mandibular arches<br>for the purpose providing prosthetic<br>support for dental restorations<br>(Crown, bridges, and overdentures)<br>in partially or fully edentulous<br>individuals.<br>It is used to restore a patient's<br>chewing function. Smaller implants<br>(less than 6.0 mm) are dedicated for<br>immediate loading when good<br>primary stability is achieved and<br>with appropriate occlusal loading.<br>Larger implants are dedicated for the<br>molar region and are indicated for<br>delayed loading. | The AnyRidge Octa 1 Implant<br>System is intended to be surgically<br>placed in the maxillary or<br>mandibular arches for the purpose of<br>providing prosthetic support for<br>dental restorations (Crown, bridges,<br>and overdentures) in partially or<br>fully edentulous individuals. It is<br>used to restore a patient's chewing<br>function in the following situations<br>and with the clinical protocols:<br>-Delayed loading.<br>-Immediate loading when good<br>primary stability is achieved and<br>with appropriate occlusal loading.<br>Larger implants are dedicated for the<br>molar region. | The AnyOne Internal Implant<br>System is intended to be surgically<br>placed in the maxillary or<br>mandibular molar areas for the<br>purpose providing prosthetic support<br>for dental restorations (Crown,<br>bridges, and<br>overdentures) in partially or fully<br>edentulous individuals.<br>It is used to restore a patient's<br>chewing function. Smaller implants<br>(less than 6.0 mm) are dedicated for<br>immediate loading when good<br>primary stability is achieved and<br>with appropriate occlusal loading.<br>Larger implants are dedicated for the<br>molar region and are indicated<br>for delayed loading. | | Appearance | | | | | Material | CP Ti Grade 4 | CP Ti Grade 4 | CP Ti Grade 4 | | Sterilization | Gamma sterilization | Gamma sterilization | Gamma sterilization | | Diameter(Ø) | 3.7, 4.2, 4.6, 5.1, 6.0, 6.8mm | 3.6, 4.0, 4.4, 4.7, 4.8, 5.0,<br>5.5mm | 3.9, 4.3, 4.8, 5.3, 5.8, 6.3, 6.8<br>7.3, 7.8, 8.3 mm | | Length (mm) | 7.0, 8.5, 10, 11.5, 13, 15mm | 7.0, 7.7, 9.2, 10.7, 12.2, 14.2,<br>17.2mm | 7.0, 8.0, 9.5, 11.0, 12.5,<br>14.5mm | | Surface | Sand-blasted, Large grit, Acid- | Sand-blasted, Large grit, Acid- | Sand-blasted, Large grit, Acid- | | treatment | etched (S.L.A) | etched (S.L.A) | etched (S.L.A) | | Implant-to-<br>abutment<br>connection | Hex | Octa | Hex | | Feature | - Submerged implant<br>- Tapered body<br>- cutting edge with selftapping<br>- 0.7, 0.8mm thread pitch | - Submerged implant<br>- Tapered body<br>- cutting edge with self-tapping<br>- 0.8mm thread pitch | - Submerged implant<br>- Tapered body<br>- cutting edge with self-tapping<br>- 0.8mm thread pitch | {7}------------------------------------------------ | Principle of<br>operation | This product is a tapered body<br>fixture which is inserted in the<br>alveolar bone.<br>It replaces the functions of the<br>missing teeth as a dental implant<br>fixture. | This product is a tapered body<br>fixture which is inserted in the<br>alveolar bone.<br>It replaces the functions of the<br>missing teeth as a dental implant<br>fixture. | This product is a tapered body<br>fixture which is inserted in the<br>alveolar bone.<br>It replaces the functions of the<br>missing teeth as a dental implant<br>fixture. | |------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Shelf Life | 5 Years | 5 Years | 5 Years | | Substantial Equivalence Discussion | | | | ### 1. Similarities - The subject device has the same characteristic for the following compared to the predicate/reference devices. Manufacturer, Indication for use, Material, Sterilization Method, Surface treatment, Implant-to-abutment - connection, Principle of operation and Shelf life #### Differences 2. - The subject device has the different characteristic for the following compared to the predicate/reference devices. Diameter, Length and Feature - ### 3. Discussion for difference - Diameter: The diameters of the subject device are slightly different with predicate device however the subject device lies within the range of reference device excepting for diameter 3.7 can be covered with diameter 3.6 included in predicate device. There is a minimal difference in the diameter but the substantial equivalence was demonstrated by the fatigue test stated below. - Feature: - The subject device has a same feature as a predicate/reference device excluding a thread pitch. A thread pitch(0.7mm) is added in subject device, however there is a just 0.1mm difference in screw pitch between 0.7mm and 0.8mm. It does not cause a matter in substantial equivalence since the size difference is very minor. : In order to demonstrate the differences do not raise an issue in substantial equivalence, each fatigue test on the subject and predicate device have been performed to figure out the physical property(e.g. fatigue limit), with combination of the fixture and the angled abutment, in consideration of the worst case according to "ISO 14801" and "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutment". The test result shows that the subject device is better able to withstand the fatigue and loading in spite of the size and feature differences. In conclusion, the subject device is substantially equivalent to the predicate device since the differences are minor and theses have been identified via the fatigue test the differences do not impact substantial equivalence. {8}------------------------------------------------ ### EZ Post Abutment | | Subject Device | Predicate Device | Reference Device | |------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k)<br>Number | Not yet | K182448 | K123988 | | Device Name | EZ Post Abutment<br>for ST Internal Implant System | EZ Post Abutment<br>AnyRidge Octa 1 Implant System | EZ Post Abutment<br>for AnyOne Internal Implant System | | Manufacturer | MegaGen Implant Co., Ltd. | MegaGen Implant Co., Ltd. | MegaGen Implant Co., Ltd. | | Indications for<br>Use Statement | The ST Internal Implant System is<br>intended to be surgically placed in<br>the maxillary or mandibular arches<br>for the purpose providing prosthetic<br>support for dental restorations<br>(Crown, bridges, and overdentures)<br>in partially or fully edentulous<br>individuals.<br>It is used to restore a patient's<br>chewing function. Smaller implants<br>(less than 6.0 mm) are dedicated for<br>immediate loading when good<br>primary stability is achieved and<br>with appropriate occlusal loading.<br>Larger implants are dedicated for the<br>molar region and are indicated for<br>delayed loading. | The AnyRidge Octa 1 Implant<br>System is intended to be surgically<br>placed in the maxillary or<br>mandibular arches for the purpose of<br>providing prosthetic support for<br>dental restorations (Crown, bridges,<br>and overdentures) in partially or<br>fully edentulous individuals. It is<br>used to restore a patient's chewing<br>function in the following situations<br>and with the clinical protocols:<br>-Delayed loading.<br>-Immediate loading when good<br>primary stability is achieved and<br>with appropriate occlusal loading.<br>Larger implants are dedicated for the<br>molar region. | The AnyOne™ Internal Implant<br>System is intended to be surgically<br>placed in the maxillary or<br>mandibular molar areas for the<br>purpose providing prosthetic support<br>for dental restorations (Crown,<br>bridges, and overdentures) in<br>partially or fully edentulous<br>individuals. It is used to restore a<br>patient's chewing function. Smaller<br>implants (less than Ø6.0 mm) are<br>dedicated for immediate loading<br>when good primary stability is<br>achieved and with appropriate<br>occlusal loading. Larger implants are<br>dedicated for the molar region and<br>are indicated for delayed<br>loading. | | Appearance | | | | | Diameter | 4.6, 5.0, 6.0, 7.0mm | 4.0, 5.0, 6.0, 7.0mm | 4.5, 5.5, 6.5 mm | | Post Height | 4.0, 5.5, 7.0 mm | 4.0, 5.5, 7.0 mm | 4.0, 5.5mm | | Gingival<br>Height | 1.0, 2.0, 3.0, 4.0, 5.0mm | 0.8, 1.8, 2.8, 3.8, 4.8mm | 1.0, 1.5, 2.5, 3.5, 4.5, 5.5mm | | Connection<br>Interface | Hex, Non-Hex | Octa, non-octa | Hex, Non-Hex | | Surface<br>treatment | Anodizing | Anodizing | Anodizing | | Sterilization | Non-sterile; intended for terminal<br>sterilization via moist<br>heat(autoclave) | Non-sterile; intended for terminal<br>sterilization via moist<br>heat(autoclave)…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...