IR Type Implant System

K171185 · Biotem Co., Ltd. · DZE · Dec 1, 2017 · Dental

Device Facts

Record IDK171185
Device NameIR Type Implant System
ApplicantBiotem Co., Ltd.
Product CodeDZE · Dental
Decision DateDec 1, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Intended Use

IR Type Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. IR Type Implant System is for one stage surgical procedures. It is intended for delayed loading. IR Implant System is intended only for straight placement with no correction of angulation.

Device Story

Endosseous dental implant system; surgically placed in maxilla or mandible bone; replaces missing teeth. System comprises titanium fixtures and abutments; utilizes RBM (Resorbable Blast Media) surface treatment. Used in clinical dental settings by trained professionals. Device provides structural support for dental restorations; delayed loading protocol. Benefits include restoration of oral function and aesthetics. No electronic or software components.

Clinical Evidence

Bench testing only. Includes sterilization validation (ISO 11137, ISO 17665), surface characterization (SEM/chemical analysis), and biocompatibility testing (ISO 10993-5 cytotoxicity, ISO 10993-10 sensitization, USP <85> endotoxin).

Technological Characteristics

Material: CP Ti Gr 4 (ASTM F67). Design: Internal octa tapered. Surface: RBM treated. Sterilization: Gamma (fixtures) and steam (abutments). Dimensions: Diameters 3.5-5.5mm, Lengths 7.5-15.0mm. No software or electronic components.

Indications for Use

Indicated for partially or fully edentulous adult patients requiring single or multiple-unit dental restorations, including cemented, screw-retained, or overdenture restorations, and fixed bridgework support. Contraindicated for immediate loading; intended for straight placement only.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue and "ADMINISTRATION" in blue below. December 1, 2017 Biotem Co., Ltd. % Joyce Bang Consultant Provision Consulting Group, Inc. 14071 Peyton Dr. #967 Chino Hills, California 91709 Re: K171185 Trade/Device Name: IR Type Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: September 28, 2017 Received: October 31, 2017 Dear Joyce Bang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Andrew I. Steen -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K171185 Device Name IR Type Implant System #### Indications for Use (Describe) IR Type Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. IR Type Implant System is for one stage surgical procedures. It is intended for delayed loading. IR Implant System is intended only for straight placement with no correction of angulation. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------------------------|--| | <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | | | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## K171185510(k) Summary #### 15. 510(K) SUMMARY #### Submitter: Hong Koo Yeo Biotem Co., Ltd . 57, Sasang-ro , 401 beon-gil , Sasang-gu , Busan , Korea Tel : +82-70-8220-4934 Fax : +82-51-908-8257 #### Official Correspondent/ US Agent: Joyce Bang Provision Consulting Group, Inc. 14071 Peyton dr. #967 Chino Hills, CA 91709 #### Prepared Date: 11-30-2017 #### Device Information: Device Name: IR Type Implant System Classification Name: Endosseous Dental Implant Common Name: Endosseous Dental Implant Classification: Class II Product Code: DZE Subsequent Product Code: NHA Regulation number: 21 CFR 872.3640 #### Predicate Device: SS System (K062051) manufactured by Osstem Implant Co., Ltd. #### Indications for use IR Type Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. IR Type Implant System is for one stage surgical procedures. It is intended for delayed loading. R Implant System is intended only for straight placement with no correction of angulation. ### Device Description IR Type Implant System is a dental implant system made of of CP Ti Gr. 4 intended to be surgically placed in the bone of the upper or lower jaw arches for loading after a conventional healing period. The implant may be used to replace one or more missing teeth. The system is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics. The surface of the system has been treated with R.B.M (Resorbable Blast Media). {4}------------------------------------------------ | Endosseous Diameter | Endosseous Length(mm) | |---------------------|----------------------------------| | Ø3.5 | 8.5, 10.0, 11.5, 13.0, 15.0 | | Ø 4.1 | 7.5, 8.5, 10.0, 11.5, 13.0, 15.0 | | Ø 4.8 | 7.5, 8.5, 10.0, 11.5, 13.0, 15.0 | | Ø 5.5 | 7.5, 8.5, 10.0, 11.5, 13.0, 15.0 | {5}------------------------------------------------ ## Substantial Equivalence Comparison Chart | | | Subject Device | Predicate Device | |-----------------------|-----------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Number | | Not available yet | K062051 | | Device Name | | IR Type Implant System | SS System | | Manufacturer | | Biotem Co., Ltd. | Osstem Implant Co., Ltd. | | Indications for Use | | IR Type Implant System is indicated<br>for use in partially or fully<br>edentulous mandibles and maxillae,<br>in support of single or multiple-unit<br>restorations including; cemented<br>retained, screw retained, or<br>overdenture restorations, and<br>terminal or intermediate abutment<br>support for fixed bridgework. IR<br>Type Implant System is for one stage<br>surgical procedures. It is intended for<br>delayed loading. IR Implant System is<br>intended only for straight placement<br>with no correction of angulation. | SS System is intended for use in<br>partially or fully edentulous mandibles<br>and maxillae, in support of single or<br>multiple-unit restorations including;<br>cemented retained, screw retained, or<br>overdenture restorations, and terminal<br>or intermediate abutment support for<br>fixed bridgework. SS<br>System is for one stage surgical<br>procedures. It is not for immediate<br>load. | | Design | | Internal Octa Tapered | Internal Octa Tapered | | Material | | CP Ti Gr 4 ASTM F67 | CP Ti Gr 4 ASTM F67 | | Sterilization | | Gamma sterilization | Gamma sterilization | | Fixture Diameter (mm) | | Ø 3.5, 4.1, 4.8, 5.5 | Ø 3.5-6.0 | | Fixture Length (mm) | | 7.5, 8.5, 10.0, 11.5, 13.0, 15.0 | 6.0-15.0 | | | Diameters | Ø 4.8, 6.0, 6.5 mm | Ø 4.8, 6.0, 6.5 mm | | | Lengths | 2.0, 3.0, 4.0, 5.5, 6.0, 7.0mm | 2.0, 4.0, 5.5, 7.0mm | | Abutment | Angled | 0° | 20° | | Attachment | | Various abutments and components | Various abutments and components | | Surface treatment | | RBM | RBM | | Product Code | | DZE NHA | DZE NHA | {6}------------------------------------------------ #### Basis for Substantial Equivalence IR Type Implant System has a substantially equivalent intended use as the identified predicate (K062051). Both are used for mandible and maxilla endosseous dental implant and accessories. The IR Type Implant System is similar in fundamental scientific technology to the predicate device in that they all have been designed, manufactured and tested in compliance with FDA's Class II special controls guidance document root-form endosseous dental implants and endosseous dental implant abutments. The subject and predicate devices are similar in size, surface treatment and materials. When compared with predicate devices, no new questions of substantially equivalence have been raised for the IR Type Implant System. #### Non- Clinical Testing The subject device was tested to evaluate its performance as below. - Sterilization Validation testing for sterile devices (fixtures) has been performed in accordance with ISO11137, ISO 11737-1& ISO 11737-2 for gamma sterilization - Steam Sterilization validation for non-sterile devices (abutments) has been performed in accordance with ISO 17665-1 and IS 17665-2. • Surface Characteristics Test Report - Chemical and SEM image analyses have been performed to verify that there is no residual after RBM treatment on the fixtures. - · Cytotoxicity Test performed according to ISO 10993-5:2009 - · Sensitization test performed according to ISO 10993-10:2010 - LAL Endotoxin lot release testing according to USP <85> - · Shelf Life Testing performed according to ISO 11607 Those tests have been performed to evaluate the substantial equivalence in the surface characteristics compared to the predicate device. The result of the above tests have met the criteria of the standard, and proved the substantial equivalence with the predicate device. Non- clinical testing consisted of performance of testing in accordance with the FDA guidance "Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments." The results of the non-clinical testing demonstrate that the subject device is substantially equivalent to the predicate device #### Conclusion The subject devices and the predicate devices have the similar intended use and have the similar technological characteristics. The subject and predicate implants are all made of commercially pure titanium and have the similar surface treatments. Overall, the IR Implant System has the following similarities to the predicate devices: - has the similar me intended use, - · uses the similar operating principle, - · incorporates the similar basic design, - incorporates the similar material and the surface treatment. Based on the similarities, we conclude that the IR Implant System is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...