HERO I DENTAL IMPLANT SYSTEM

K121098 · Kj Meditech Co., Ltd. · DZE · Jan 17, 2013 · Dental

Device Facts

Record IDK121098
Device NameHERO I DENTAL IMPLANT SYSTEM
ApplicantKj Meditech Co., Ltd.
Product CodeDZE · Dental
Decision DateJan 17, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Hero I Dental Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The Hero I Dental Implant System is for single and two stage surgical procedures. The system is intended for delayed loading.

Device Story

Hero I Dental Implant System consists of threaded, root-form endosseous implants; fabricated from Titanium 6AL 4V ELI alloy; surface treated with Resorbable Blast Media (RBM). Designed for surgical placement into maxilla or mandible bone to support dental prosthetics. Features internal hexagon abutment connection with beveled interface for prosthetic stability; straight-walled neck; tapered body. Used by dental professionals in clinical settings for single or two-stage surgical procedures; intended for delayed loading. Provides structural foundation for single/multiple-unit restorations or fixed bridgework. Benefits patients by replacing missing teeth and restoring oral function.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing and comparison of technological characteristics, materials, and design to predicate devices.

Technological Characteristics

Material: Titanium 6AL 4V ELI alloy. Surface: Resorbable Blast Media (RBM) treated. Design: Threaded, root-form, bone-level implant; internal hexagon abutment connection with beveled interface; straight-walled neck; tapered body. Sterilization: Not specified.

Indications for Use

Indicated for partially or fully edentulous adult patients requiring single or multiple-unit restorations (cemented, screw-retained, or overdenture) or fixed bridgework support in the mandible or maxilla. Suitable for single or two-stage surgical procedures with delayed loading.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K121098 ## 510(k) Summary ## JAN 1 7 2013 This summary of 510(K) - safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92. Date: 11/09/2012 - 1. Submitter | | Submitter | |---------|----------------------------------------------------------------| | Name | KJ Meditech Co., Ltd. | | Address | 959-21 Daechon-dong, Buk-gu, Gwang-ju, 500-470,<br>South Korea | | Phone | +82-62-972-5476 | | Fax | +82-62-973-2809 | | Contact | Huykki Moon, CEO | 2. U.S Agent/Contact Person LK Consulting Group 951 Starbuck St. Unit J, Fullerton, CA 92833 Priscilla Juhee Chung Phone: 714.869.3080 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com 3. Device Trade Name: Hero I Dental Implant System Common Name: Dental Implant Classification Name: Endosseous Dental Implant System Product Code: DZE, NHA Classification regulation: 21CFR872.3640 2. Predicate Device: RENOVA™ Internal Hex Implant System by Lifecore Biomedical, Inc. (K032774) OsseoFuse Dental Implant System by Dynamic Innovations Inc. (K110577) Hero II and IS Dental Implant System by KJ Meditech Co., Ltd. (K121047) {1}------------------------------------------------ #### 3. Description: The Hero I Dental Implant System is made of Titanium 6AL 4V ELI alloy intended to be surgically placed in the bone of the upper or lower jaw arches for loading after a conventional healing period. The implant may be used to replace one or more missing teeth. The system is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition an performance characteristics. The surface of the system has been treated with R.B.M (Resorbable Blast Media). #### 4. Indication for use: The Hero I is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The Hero I Dental Implant System is for single and two stage surgical procedures. The system is intended for delayed loading. - 5. Basis for Substantial Equivalence The Hero I Dental Implant system has the same intended use as the identified predicate devices. The Hero I is similar in fundamental scientific technology in that they are all threaded, root form implants constructed of titanium with RBM roughened surfaces. Thev all share same internal hexagon abutment connection system with internal beveled interface for maximum prosthetic stability. The subject and the predicate devices are bone-level implants that share similar body shape design such as straight walled neck and tapered body design. The subject and the predicate devices are similar in size, materials, surface treatment, and are sterilization method. When compared with the predicate devices, no new question of safety or effectiveness has been raised for the Hero I Dental Implant System. - 6. Conclusion The subject device and the predicate devices have the same intended use and have the same technological characteristics. The subject and the predicate implant are all made of commercially pure titanium and have the same surface treatment. Overall, the Hero I Dental Implant system has the following similarities to the predicate devices: - * has the same intended use. - * uses the same operating principle, - * incorporates the same basic design, - * incorporates the same material and the surface treatment. Based on the similarities, we conclude that the Hero I Dental Implant system is substantially equivalent to the predicate devices. 510(k) Submission. {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's symbol, which is a stylized depiction of an abstract human figure embracing a sphere. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the symbol. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 #### January 17, 2013 KJ Meditech Company C/O Ms. Priscilla Chung Regulatory Affairs Consultant LK Consulting Group 951 Starbuck Street Unit J FULLERTON CA 92833 Re: K121098 Trade/Device Name: Hero I Dental Implant System Regulation Number: 12 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: December 21, 2012 Received: December 26, 2012 #### Dear Ms. Chung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Ms. Priscilla Chung Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Cintron lo mio Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K121098 Hero I Dental Implant System Device Name: ### Indications For Use: The Hero I Dental Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The Hero I Dental Implant System is for single and two stage surgical procedures. The system is intended for delayed loading. ## Prescription Use (Per 21 CFR 801 Subpart D) AND ### Over-The Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ## Concurrence of CDRH, Office of Device Evaluation (ODE) | Susan Runner DDS, MA | |----------------------| | 2013.01.14 | | 11:51:34 -05'00' | (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices K121098 510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...