GEN5 and GEN5+ Dental Implant System

K251938 · Paragon Implant Mfg., LLC · DZE · Oct 30, 2025 · Dental

Device Facts

Record IDK251938
Device NameGEN5 and GEN5+ Dental Implant System
ApplicantParagon Implant Mfg., LLC
Product CodeDZE · Dental
Decision DateOct 30, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Intended Use

GEN5 and GEN5 Implants: GEN5 and GEN5+ implants are two-piece implants for one-stage or two-stage surgical procedures. These implants are intended for use in partially or fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework. Implants can be indicated for immediate function when initial implant stability has been achieved and with appropriate occlusal loading. Short (<9mmL) Implants: Indicated for single tooth (mandibular and maxillary central and lateral incisors), multiple tooth replacements or denture stabilization. GEN5 and GEN5+ Abutment System: The GEN5 and GEN5+ Abutment System is intended for use with dental implants in the maxillary and/or mandibular arches to provide support for crowns, bridges, or overdentures for edentulous or partially edentulous patients, using traditional crown & bridge techniques.

Device Story

The GEN5 and GEN5+ Dental Implant System comprises root-form endosseous dental implants, abutments, and prosthetic components. Implants are two-piece titanium alloy (Ti-6Al-4V ELI) assemblies with an internal hex connection, available in various diameters and lengths. The GEN5+ variant includes a pre-attached extender healing abutment for single-stage surgical procedures. Implants feature a microtextured blasted surface and an anodized coronal collar. Components are used by dental surgeons to support crowns, bridges, or overdentures in edentulous or partially edentulous patients. The system is supplied sterile (gamma irradiated) or non-sterile for end-user steam sterilization. Clinicians use the system to restore patient chewing function and aesthetics. The device provides a stable foundation for prosthetic attachments, with the extender collar offering trans-mucosal support or vertical flexibility for abutment connection.

Clinical Evidence

No clinical data was included in this submission. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993/7405), sterilization validation (ISO 11137/17665), fatigue testing (ISO 14801), torsional/pullout load testing (ASTM F543), MRI compatibility (ASTM F2052/F2119/F2182/F2213), and surface characterization (SEM/EDS). Short implant performance was evaluated via in silico analysis and pullout testing.

Technological Characteristics

Materials: Ti-6Al-4V ELI (ASTM F136-13), PEEK. Design: Threaded root-form endosseous implant with internal hex connection. Surface: Soluble blast media (SBM) microtextured body; anodized coronal collar with horizontal grooves. Energy: Mechanical. Connectivity: N/A. Sterilization: Gamma irradiation (sterile components) or steam sterilization (non-sterile components).

Indications for Use

Indicated for partially or fully edentulous patients requiring single or multiple-unit restorations, fixed bridgework, or overdentures. Short implants (<9mm) indicated for single tooth (mandibular/maxillary incisors), multiple tooth replacement, or denture stabilization. Contraindicated for patients without adequate bone quality/quantity for primary stability.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION Paragon Implant Mfg., LLC Renee Bennett Regulatory Affairs Manager 27030 Malibu Hills Road Calabasas, California 91301 October 30, 2025 Re: K251938 Trade/Device Name: GEN5 and GEN5+ Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: September 26, 2025 Received: September 29, 2025 Dear Renee Bennett: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251938 - Renee Bennett Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K251938 - Renee Bennett Page 3 Sincerely, Sherrill Lathrop Blitzer for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (7/17) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K251938 | | | Device Name GEN5 and GEN5+ Dental Implant System | | | Indications for Use (Describe) GEN5 and GEN5 Implants: GEN5 and GEN5+ implants are two-piece implants for one-stage or two-stage surgical procedures. These implants are intended for use in partially or fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework. Implants can be indicated for immediate function when initial implant stability has been achieved and with appropriate occlusal loading. Short (<9mmL) Implants: Indicated for single tooth (mandibular and maxillary central and lateral incisors), multiple tooth replacements or denture stabilization. GEN5 and GEN5+ Abutment System: The GEN5 and GEN5+ Abutment System is intended for use with dental implants in the maxillary and/or mandibular arches to provide support for crowns, bridges, or overdentures for edentulous or partially edentulous patients, using traditional crown & bridge techniques. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} 510(k) Summary K251938 GEN5 and GEN5+ Dental Implant System Page 1 of 23 510(k) Summary Paragon Implant Mfg., LLC GEN5 and GEN5+ Dental Implant System October 29, 2025 Manufacturer Name Paragon Implant Mfg., LLC 27030 Malibu Hills Road Calabasas, CA 91301, U.S.A. Telephone (818) 475-4675 Official Contact Renee Bennett, Manager of Regulatory Affairs Paragon Implant Mfg., LLC 27030 Malibu Hills Road Calabasas, CA 91301 Telephone: +1-818-475-4675 Email: rbennett@paragon-implant.com Device Name and Classification Trade/Proprietary Name GEN5™ and GEN5+™ Dental Implant System Common Name Implant, Endosseous, Root-Form Regulation Number 21 CFR § 872.3640 Regulation Name Endosseous dental implant Regulatory Class Class II Product Code(s) DZE; NHA Legally Marketed Predicate Device(s) Primary Predicate Device (DZE - Implants) K192221 (Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, Legacy4 fixture-mounts), Implant Direct Sybron Manufacturing, LLC Reference Device (DZE - Implants) K202344 (TiUltra Implants and Xeal Abutments), Nobel Biocare AB K241972 (BLUEDIAMOND IMPLANT), MegaGen Implant Co., Ltd. Primary Predicate Device (NHA - Abutments) K192221 (Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, Legacy4 fixture-mounts), Implant Direct Sybron Manufacturing, LLC Reference Devices (NHA - Abutments) K161655 (On1 Concept), Nobel Biocare AB {5} 510(k) Summary K251938 GEN5 and GEN5+ Dental Implant System Page 2 of 23 K052600 [Zimmer Dental Hex-Lock Prepared Abutment (Straight &amp; Angled)], Zimmer Dental Inc. K090234 (Spectra System Dental Implants 2008), Implant Direct LLC K061319 (Spectra System), Implant Direct LLC K060063 (Legacy Abutment System), Implant Direct LLC K143011 (2014 InterActive/ SwishActive System), Implant Direct Sybron Manufacturing LLC ## INDICATIONS FOR USE STATEMENT GEN5 and GEN5 Implants: GEN5 and GEN5+ implants are two-piece implants for one-stage or two-stage surgical procedures. These implants are intended for use in partially or fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework. Implants can be indicated for immediate function when initial implant stability has been achieved and with appropriate occlusal loading. Short (&lt;9mmL) Implants: Indicated for single tooth (mandibular and maxillary central and lateral incisors), multiple tooth replacements or denture stabilization. GEN5 and GEN5+ Abutment System: The GEN5 and GEN5+ Abutment System is intended for use with dental implants in the maxillary and/or mandibular arches to provide support for crowns, bridges, or overdentures for edentulous or partially edentulous patients, using traditional crown &amp; bridge techniques. ## SUBJECT DEVICE DESCRIPTION The GEN5™ and GEN5+ Dental Implant System consists of a root form Endosseous dental implant, Endosseous dental implant abutment, and various other prosthetic/restorative components (medical devices) to plan, implant, and support restorations for edentulous patients. GEN5 and GEN5+ Dental Implants are supplied as a two-piece implant assembly for use in a two-stage or single-stage surgical procedures intended for surgical implantation in edentulous or partially edentulous mandibles or maxilla for attachment of complete denture prostheses, or for fixed or removable bridgework or as a free-standing single tooth replacement. The standard GEN5 Implants are for use in a two-stage surgical procedure. The first surgery is for implant placement, and the second surgery is a few weeks prior to start of prosthetic rehabilitation. By the addition of an abutment of any type at time of implant placement, the implant functions as a one-stage, two-piece system. The GEN5 Implant and GEN5+ Implant are the exact same dental implant (internal hex implant). The difference between the two is that the GEN5+ Implant includes a pre-attached hex engaging Extender {6} 510(k) Summary K251938 GEN5 and GEN5+ Dental Implant System Page 3 of 23 Healing Abutment (accessory restorative component) screw-retained on the GEN5 Implant prior to packaging requiring only a single surgical procedure prior to prosthetic rehabilitation. The top (coronal portion) of the GEN5 Implant body is with a 2.0mm straight neck. The depth gauge lines at 1mm, 2mm and 2.5mm (or at 1mm and 2mm on the ≤9mm length) from the top of the implant facilitate the placement level with the crest of the ridge or 1mm above the crest of the ridge for all implants except the 3.7mmD x 7mmL. The lower apical aspect of the Implant is tapered with double-lead progressively deeper buttress threads. Three cutting flutes extend over the tapered portion of the implant's body. The GEN5+ Implants offer the additional flexibility of a 2mm extender collar that can serve as the trans-mucosal collar of an abutment or can be removed for abutment connection directly to the top of the implant for vertical flexibility. The GEN5 and GEN5+ dental implant body is available in five diameter sizes (3.7mm, 4.2mm, 4.7mm, 5.2mm, and 5.7mm) with two platform diameters (3.5mm or 4.5mm), and five lengths (7mm, 9mm, 11mm, 12.5mm, and 14mm). GEN5 and GEN5+ Implants have 2.0mmL of their coronal surface anodized with either a gold (3.5mmD) or a rose gold (4.5mmD) color for aesthetic purposes to help identify the two platform diameters while the remainder of the Implant has gone through a surface treatment (blasting) to create a microtextured surface roughness. These Implants (and the devices included in the packaging) are sold sterile. The GEN5 and GEN5+ Implants are manufactured from titanium alloy (Ti-6Al-4V ELI) complying with standard ASTM F136-13 - Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant. The GEN5 and GEN5+ have an internal hex for abutment connection, which is designed to receive multiple abutment variations expanding its restorative options and allowing for the implant to be used for support of attachment-retained overdentures and cement-retained or screw-retained prostheses. The GEN5 Implants are provided with an Internal Hex Carrier while the GEN5+ Implants (a GEN5 Implant pre-fitted with an Extender Healing Abutment) are packaged with an Extender Carrier and an Extender Carrier Fixation Screw. The devices are sold in gamma irradiated sterile packaging which consists of an outer vial and an inner vial which is sealed with a threaded cap. The Extender Carriers and Extender Carrier Fixation Screws are all manufactured from titanium alloy (Ti-6Al-4V ELI) complying with standard ASTM F136-13 - Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant). The various devices included in the GEN5 and GEN5+ Dental Implant System (such as Endosseous dental implant abutments and various other prosthetic/restorative components) used to plan, implant, and support restorations for fully and partially edentulous patients are offered in compatible platform dimensions to enable a complete dental restoration. The prosthetic and restorative components (Straight Contoured, Angled-Contoured, ASC, Multi-Unit Abutments, Titanium Multi-Unit Abutment Copings, Titanium Temporary, Healing, Extenders, Cover Screws, Transfers, and Analogs) are all manufactured from titanium alloy (Ti-6Al-4V ELI) and anodized. {7} 510(k) Summary K251938 GEN5 and GEN5+ Dental Implant System Page 4 of 23 The prosthetic components (PEEK Temporary Abutments) are all manufactured from polyetheretherketone (PEEK), ## PERFORMANCE DATA Non-clinical data submitted to demonstrate substantial equivalence included: - Chemical characterization and biocompatibility validations per ISO 10993 and ISO 7405 - Sterilization validation for all devices sold sterile per ANSI/AAMI/ISO 11137-2/. - Cleaning and sterilization validations for all devices sold non-sterile per ISO 17664-1, AAMI TIR12, and ISO17665. - Fatigue (Dynamic Load) testing following ISO 14801 and Class II FDA Special Controls Guidance for worst-case dental implant system configurations containing angled abutments. - Torsional yield, driving (insertion) torque, pullout load testing per ASTM F543 A1, A2, A3. - Simulation and Physical MRI validations per ASTM F2052, ASTM F2119, ASTM F2182, and ASTM F2213. - Accelerated aging per ASTM F1980-21 Because the packaging is a vial with a cap, determination of shelf-life by means of conventional package integrity testing after aging is not applicable. Therefore, shelf-life of the sterile barrier was determined by membrane filtration sterility testing of the interior of the vial according to ISO 11737-2 after accelerated aging according to ASTM F1980. - Simulated transportation conditioning was conducted per ASTM D4169 and ASTM F7386. - Surface characterization of the anodized surface including surface area, surface roughness, surface thickness confirms that the anodization does not affect the surface roughness of the implant collar compared to the machined surface, and the surface characteristics are similar to those of the Nobel TiUltra implant body (K202344). - Surface treatment analysis of the blasted surface by performing SEM (Scanning Electron Microscopy) and EDS (Energy Dispersive X-ray Spectroscopy) testing conducted according to “Root-form Endosseous Dental Implants and Endosseous Dental Abutments - Class II Special Controls Guidance Document for Industry and FDA Staff” and FDA Guidance Document "Recommended Content and Format of Non-Clinical Bench Performance Testing Information in Premarket Submissions" to validate cleaning protocols after blasting. - Endotoxin Testing was conducted for endotoxins using the kinetic turbidimetric and chromogenic techniques in accordance with ANSI/AAMI ST72:2019, USP &lt;161&gt;, and USP &lt;85&gt;. - Testing of short implants was conducted in silico [using SolidWorks 2023 (Dassault Systèmes)] to evaluate available implant surface area (SA) and initial bone-to-implant contact (BIC) for bone level and 1mm supracrestal placement, before and after 3mm of resorption. - Pullout load testing was conducted on short implants to evaluate indirect BIC using bone level placement. - No clinical data was included in this submission. ## EQUIVALENCE TO MARKETED DEVICES The proposed GEN5 and GEN5+ Dental Implant System consists of root form Endosseous dental implants, Endosseous dental implant abutments, and various other prosthetic/restorative components (medical devices) to plan, implant, and support restorations for edentulous patients. {8} 510(k) Summary K251938 GEN5 and GEN5+ Dental Implant System Page 5 of 23 The threaded root form Implants are manufactured from Ti-6Al-4V ELI per complying with standard ASTM F136-13. Both the proposed and predicate device (K192221) are anodized and blasted with a soluble blast media to create a microtextured surface roughness below the bone level region of the Implant. The conical connection of the proposed Implants has the identical angulation as the predicate device (K192221). The implant-abutment conical connection is identical to that of the predicate devices (K192221). The proposed Implant diameters correspond to the various diameters offered by the predicate devices (K192221). The proposed Implants have similar straight/tapered body shape as the predicate device (K192221). The top straight neck portion of the proposed device is different from the predicate device (K192221) in that the proposed device has horizontal grooves at 1mm, 2mm and 2.5mm from the top of the implant while the predicate device (K192221) does not. The difference is the proposed device has a machined anodized collar, with the anodized portion being substantially equivalent to the reference device (K202344), dimensionally, in surface roughness, in surface area and surface thickness, and the machined collar of reference device (K241972) in length. The predicate device (K192221) has quadruple-lead micro-threads on the coronal aspect of the implant. The lower portion of the proposed and predicate device (K192221) both have buttress threads but differ in that the proposed device has double-lead threads while the predicate device (K192221) has triple-lead threads. The lower tapered portion of the proposed device has two cutting flutes while the predicate device (K192221) has three cutting grooves. The body diameters of the proposed Implants are identical to those of the predicate device (K192221). The only exception is that GEN5 and GEN5+ Implants are not offered in the 3.2mm and 7.0mm body diameters. The lengths of the proposed Implants are similar to those of the predicate device (K192221). GEN5 and GEN5+ Implants are within the similar range offered at 7mm, 9mm, 11mm, 12.5mm, and 14mm length diameters. The platforms of the proposed device are similar to the predicate device (K192221). The only exception is that GEN5 and GEN5+ Implants are not offered in the 3.0mm and 5.7mm platform diameters. The additional devices provided inside the packaging differ from the proposed device and predicate device (K192221). Both the proposed device and the predicate device (K192221) are packaged in a vial and cap with labels. The Implants, Healing Abutments, and Cover Screw of the proposed and predicate device(K192221) are single-use and delivered sterile (gamma irradiated) while all other components are single-use and delivered non-sterile and sterilized by the end user via steam sterilization per the Instructions for Use (IFU). Table 1. Substantial Equivalence (IMPLANTS) outlines the Intended Use and Technological Characteristics of each predicate and reference device CONCLUSION {9} 510(k) Summary GEN5 and GEN5+ Dental Implant System Page 6 of 23 The subject device, the primary predicate device, and the reference devices have similar intended use, have similar technological characteristics, and are made of the similar materials. The subject device and the reference devices encompass similar range of physical dimensions, are packaged in the similar materials and are sterilized using similar methods. The data included in this submission demonstrates substantial equivalence to the predicate devices listed above. Table 1. Substantial Equivalence (IMPLANTS) Table 2. Substantial Equivalence (ABUTMENTS) {10} 510(k) Summary K2S193B GEN5 and GEN5+ Dental Implant System | Design Parameter | Subject Device: K243960 GEN5™ and GEN5+™ Dental Implant System | Primary Predicate Device: (K192221) Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, Legacy4 fixture-mounts, Implant Direct Sybron Manufacturing, LLC | Reference Device: (K202344) TIUltra Implants and Xeal Abutments, Nobel Biocare AB | Reference Device: (K241972) BLUEDIAMOND IMPLANT, MegaGen Implant Co., Ltd. | | --- | --- | --- | --- | --- | | Regulation | 21 C.F.R. § 872.3640 (Endosseous dental implants) | 21 C.F.R. § 872.3640 (Endosseous dental implants) | 21 C.F.R. § 872.3640 (Endosseous dental implants) | 21 C.F.R. § 872.3640 (Endosseous dental implants) | | Classification | Class II | Class II | Class II | Class II | | Product code | DZE | DZE | DZE | DZE | | Materials | Titanium 6AL-4V ELI | Titanium 6AL-4V ELI | Commercially pure titanium | CP Ti Grade 4 | | Reason for Predicate/Reference | N/A | • Same General Design • Same Body Diameters • Same Connection Type • Same Sterilization • Same Material • Same/similar lengths, within 6mm–16mm size range of predicate • Same/similar Screw size GEN5 and GEN5+ do not offer M1.6 screw size • Same/similar Platform within size range of predicate • Difference(s): GEN5+ Implant: Supplied with pre-attached Implant Healing Abutment, Carrier and screw • Difference(s) Cutting Flutes Double Lead-Thread vs. Triple Lead-Thread • Difference(s) predicate has micro-thread, subject has machined anodized collar | Collar design, 2mm anodized machined endosseous collar with horizontal grooves | 2.0 mm machined, unthreaded collar | | General Design | Threaded Root form implant | Threaded Root form implant | Threaded Root form implant | Threaded Root form implant | | Implant Diameters | 3.7mm, 4.2mm, 4.7mm, 5.2mm, 5.7mm | 3.2mm, 3.7mm, 4.2mm, 4.7mm, 5.2mm, 5.7mm, 7.0mm | NobelActive TIUltra 3.0, 3.5, 4.3, 5.0, 5.5mm NobelReplace CC TIUltra 3.5, 4.3, 5.0mm NobelParallel CC TIUltra 3.75, 4.3, 5.0, 5.5mm | Ø 4.0 Ø 4.4 Ø 4.7 Ø 4.8 Ø 5.1 | | Implant Lengths | 7mm, 9mm, 11mm, 12.5mm, and 14mm | 6mm (only simplyLegacy), 8mm, 10mm, 11.5mm, 13mm, 16mm | NobelActive TIUltra 7.0, 8.5, 10.0, 11.5, 13.0, 15.0, 18.0mm NobelReplace CC TIUltra Externa : 7.5, 8.5, 10.0, 11.5, 13.0, 15.0, 18.0mm 8.0, 10.0, 11.5, 13.0, 16.0mm NobelParallel CC TIUltra 6.5, 8.0, 9.5, 11.0, 12.5, 14.5, 17.5mm | 7.0mm, 8.0mm, 9.0mm, 10mm, 11mm, 13mm, 15mm | | Implant Platform | For 3.7mm, 4.2mm implants: 3.5mm For 4.7mm, 5.2mm, 5.7mm: 4.5mm | 3.0mm, 3.5mm, 4.5mm, 5.7mm | Internal conical connection with hex interface | N/A | | Cutting Flutes | Double (2) | 3 (Legacy2/4); 2 (Legacy3) | N/A | N/A | | Connection Type | Internal hex | Internal hex | Internal hex | Internal Octa | | Screw Size | 1-72 UNF 2A | M1.6 and 1-72 UNF 2A | N/A | N/A | | Implant Assembly | GEN5 Implant Supplied with friction-fit Carrier; GEN5+ Implant: Supplied with pre-attached Implant Healing Abutment, Carrier and screw. | Implant is assembled with one-piece (Legacy2, 3) or two-piece (Legacy4) fixture-mount. simplyLegacy2, 3: No carrier or fixture mount. | N/A | N/A | | Sterilization | Implants, Healing Abutments, and Cover Screws supplied sterile; Provided sterile All other devices supplied non-sterile. Devices steam sterilized (when applicable) by end user as per Instructions for Use (IFU) | Supplied sterile; fixture-mounts sterilized by end user when used as abutments. | Provided sterile | Provided sterile | | Surface | Resorbable Blast Media, 2.0mm machined, grooved, anodized section only at coronal end | SBM (Soluble Blast Media) implant: HA blasted; HA (Hydroxyapatite) implant: HA blasted/HA coated | Anodic oxidation | Sand-blasted, Large grit, Acid-etched (S.L.A) Machined collar | | Surface Topography | Two level surface: Level 0 (collar): Sa (area roughness) • 0.6μm ± .2 μm Thickness (Thk) • Gold 0.098 μm ± .014 μm • Rose Gold 0.110 μm ± .015 μm Level 1 (body): Sa (area roughness) 1.5 μm -2.3 μm Thk | Level 1 (body): • Sa (area roughness) 1.5 μm -2.3 μm | TIUltra – Three level surface: • Level 0 (collar): o Sa (area roughness) = 0.5 ± 0.3μm o Thickness (Thk) = 0.166±0.008μm • Level 1 (transition): o Sa = 0.8 ± 0.3μm o Thk = 7.5±0.3μm • Level 2 (body): o Sa = 1.5 ± 0.4μm o Thk = 12.0±1.2μm | N/A | | Image | | | | | {11} 510(k) Summary K2S1938 GEN5 and GEN5+ Dental Implant System Page 8 of 23 | Design Parameter | Subject Device: K243960 GEN5™ and GEN5+™ Dental Implant System | Primary Predicate Device: (K192221) Legacy2, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, Legacy4 fixture-mounts, Implant Direct Sybron Manufacturing, LLC | Reference Device: (K202344) TiUltra Implants and Xeal Abutments, Nobel Biocare AB | Reference Device: (K241972) BLUEDIAMOND IMPLANT, MegaGen Implant Co., Ltd. | | --- | --- | --- | --- | --- | | Indications for Use | Indications for Use: GEN5 and GEN5 Implants: GEN5 and GEN5+ implants are two-piece implants for one-stage or two-stage surgical procedures. These implants are intended for use in partially or fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework. Implants can be indicated for immediate function when initial implant stability has been achieved and with appropriate occlusal loading. Short (<9mmL) Implants: Indicated for single tooth (mandibular and maxillary central and lateral incisors), multiple tooth replacements or denture stabilization. GEN5 and GEN5+ Abutment System: The GEN5 and GEN5+ Abutment System is intended for use with dental implants in the maxillary and/or mandibular arches to provide support for crowns, bridges, or overdentures for edentulous or partially edentulous patients, using traditional crown & bridge techniques. | Indications for Use: Legacy2, simplyLegacy2, Legacy3, simplyLegacy3, and Legacy4 dental implants are two-piece implants for one-stage or two-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework. Implants can be indicated for immediate loading when good primary stability has been achieved and with appropriate occlusal loading. *Narrow (3.2mmD) implants: Indicated for single-tooth replacement (mandibular central and lateral incisors; maxillary lateral incisors), multiple-tooth replacements or denture stabilization. *Short (<10mm) 3.7mm implants: Indicated for single-tooth (mandibular and maxillary central and lateral incisors), multiple tooth replacements or denture stabilization. The Legacy 2, Legacy3, and Legacy4 fixture-mounts are intended for use with the corresponding dental implants (Legacy2, Legacy3, and Legacy4, respectively). The fixture-mounts can function as an abutment. As an abutment, fixture-mounts are intended for use with dental implants in the maxillary and/or mandibular arches to provide support for crowns or bridges for edentulous or partially edentulous patients. *Fixture-mounts as an abutment for narrow (3.2mmD) implants: Indicated for single-tooth replacement of mandibular central and lateral incisors and maxillary lateral incisors. *Fixture-mounts as an abutment for short (8mm) 3.7mmD implants: Indicated for tooth replacement of mandibular and maxillary central and lateral incisors. Legacy2, simplyLegacy2, Legacy3, simplyLegacy3, and Legacy4 implants are compatible with the following abutments. Manufacturer: Implant Direct; Abutment Line: Legacy; Platform Diameter: 3.0mm, 3.5mm, 4.5mm, 5.7mm | Indications for Use: NobelActive TiUltra implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting tooth replacements to restore patient esthetics and chewing function. NobelActive TiUltra implants are indicated for single or multiple unit restorations in splinted or non-splinted applications. This can be achieved by a 2-stage or 1-stage surgical technique in combination with immediate, early or delayed loading protocols, recognizing sufficient primary stability and appropriate occlusal loading for the selected technique. NobelActive TiUltra 3.0 implants are intended to replace a lateral incisor in the maxilla and/or a central or lateral incisor in the mandible. NobelActive TiUltra 3.0 implants are indicated for single-unit restorations only. NobelReplace CC TiUltra NobelReplace CC TiUltra implants are endosseous dental implants intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. The NobelReplace CC TiUltra implants are indicated for single or multiple unit restorations. The NobelReplace CC TiUltra implant may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied. NobelParallel CC TiUltra NobelParallel CC TiUltra implants are endosseous implants intended to be surgically placed in the upper or lower jaw bone for anchoring or supporting replacements to restore patient esthetics and chewing function. NobelParallel CC TiUltra implants are indicated for single or multiple restorations in splinted or non-splinted applications. This can be achieved by a 2-stage or 1-stage surgical techniques in combination with immediate, early or delayed loading protocols, recognizing sufficient primary stability and appropriate occlusal loading for the selected technique. Implants with <7mm length are for delayed loading only when appropriate stability has been achieved. | The BLUEDIAMOND IMPLANT is intended to be surgically placed in the maxillary or mandibular molar areas for the purpose providing prosthetic support for dental restorations (Crown, bridges, and overdentures) in partially or fully edentulous individuals. It is used to restore a patient's chewing function in the following situations and with the clinical protocols: - Delayed loading - Immediate loading when good primary stability is achieved and with appropriate occlusal loading. For the BLUEDIAMOND IMPLANTS with a Thread Length of 5mm, it is indicated for fixed or removable reconstruction in situations of moderate to severely atrophic jawbone and with adequate bone quality that allows primary stability after implant insertion, where a longer implant cannot be placed due to limited vertical bone height. The recommended healing time before loading is between 10 to 12 weeks. It is specifically recommended for: - Fixed partial dentures/splinted units (one implant per unit) - Pontic cases in combination with at least one longer implant - Fully edentulous cases with at least one 5 mm Short Implant in combination with 2 longer implants in the anterior region and at least four total implants. | | Sterilization | Gamma irradiation | Gamma irradiation | Gamma Irradiation | Gamma Irradiation | | Usage | Single-patient, single-use | Single-patient, single-use | Single-patient, single-use | Single-patient, single-use | {12} 510(k) Summary K251938 GEN5 and GEN5+ Dental Implant System Page 9 of 23 | Table 2. Substantial Equivalence (ABUTMENTS) | | | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Implant Diameter | Straight Contoured Abutment | 15-degree Angled Abutment | Titanium Temporary Abutment | Plastic (PEEK) Temporary Abutment | Angled Screw Channel (ASC) Abutments (Engaging) | Angled Screw Channel (ASC) Abutments (Non-Engaging) | Straight Multi-Unit Abutments (MUA) | Angled Multi-Unit Screw-Receiving Abutment | Extender Healing Abutment | Healing Abutment | Cover Screw | | Device Image | | | | | | | | | | | | | Connection | Internal Hex | Internal Hex | Internal Hex | Internal Hex | Internal Hex | Internal Hex | Internal Hex | Internal Hex | Internal Hex | Internal Hex | Internal Hex | | Supplied Sterile | Sold non-sterile; reprocessed by end user | Sold non-sterile; reprocessed by end user | Sold non-sterile; reprocessed by end user | Sold non-sterile; reprocessed by end user | Sold non-sterile; reprocessed by end user | Sold non-sterile; reprocessed by end user | Sold non-sterile; reprocessed by end user | Sold non-sterile; reprocessed by end user | Sold non-sterile; reprocessed by end user | Sold non-sterile; reprocessed by end user | Sold non-sterile; reprocessed by end user | Predicates: Legacy™ System Abutment Types: | Implant Diameter | Straight Contoured Abutment (K090234) (K061319) | 15-degree Angled Abutment (K090234) (K061319) | Titanium Temporary Abutment (K090234) (K060063) | Plastic (PEEK) Temporary Abutment (K090234) | Angled Screw Channel (ASC) Abutments (Engaging) (K090234) (K061319) | Angled Screw Channel (ASC) Abutments (Non-Engaging) (K090234) (K060063) | Straight Multi-Unit Abutments (MUA) (K090234) (K061319) | Angled Multi-Unit Screw-Receiving Abutment (K143011) (K061319) | Healing Abutment (K090234) | Cover Screw (K090234) | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Device Image | | | | | | | | | | | | Connection | Internal Hex | Internal Hex | Internal Hex | Internal Hex | Internal Hex | Internal Hex | Internal Hex | Internal Hex | Internal Hex | Internal Hex | | Supplied Sterile | Sold non-sterile; reprocessed by end user | Sold non-sterile; reprocessed by end user | Sold non-sterile; reprocessed by end user | Sold non-sterile; reprocessed by end user | Sold non-sterile; reprocessed by end user | Sold non-sterile; reprocessed by end user | Sold non-sterile; reprocessed by end user | Sold non-sterile; reprocessed by end user | Sold non-sterile; reprocessed by end user | Sold non-sterile; reprocessed by end user | {13} 510(k) Summary K251938 GEN5 and GEN5+ Dental Implant System Page 10 of 23 Predicates for GEN5+ Extender: Nobel On1™ Base (K161655) | Table 2. Substantial Equivalence (Extender Healing Abutment) | | | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Straight Contoured Abutment | Angled Abutment | Titanium Temporary Abutment | Plastic (PEEK) Temporary Abutment | Angled Screw Channel (ASC) Abutments (Engaging) | Angled Screw Channel (ASC) Abutments (Non-Engaging) | Straight Multi-Unit Abutments (MUA) | Angled Multi-Unit Screw-Receiving Abutment | Extender Healing Abutment | Healing Abutment | Cover Screw | | Device Image | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | | N/A | N/A | | Connection | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | Internal Conical Connection | N/A | N/A | | Supplied Sterile | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | Sterile | N/A | N/A | Predicates for GEN5+ Extender: Hex-Lock Prepared Abutment (Straight &amp; Angled) - [Tapered Screw Vent (TSV) Abutments (friction fit)]: (K052600) | Table 2. Substantial Equivalence (Extender Healing Abutment) | | | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Straight Contoured Abutment | Angled Abutment | Titanium Temporary Abutment | Plastic (PEEK) Temporary Abutment | Angled Screw Channel (ASC) Abutments (Engaging) | Angled Screw Channel (ASC) Abutments (Non-Engaging) | Straight Multi-Unit Abutments (MUA) | Angled Multi-Unit Screw-Receiving Abutment | Extender Healing Abutment | Healing Abutment | Cover Screw | | Device Image | | | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | | Connection | internal hex friction-fit connection | internal hex friction-fit connection | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | | Supplied Sterile | Sterile | Sterile | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | {14} 510(k) Summary K251938 GEN5 and GEN5+ Dental Implant System Page 11 of 23 # Technological Characteristics Comparison Table – Abutment Designs | Comparison | Subject Device K243960 (GEN5 and GEN5+ Dental Implant System) | Predicate Device K192221 (Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, Legacy4 fixture-mounts) Implant Direct Sybron Manufacturing, LLC | Reference Device K161655 (On1 Concept) Nobel Biocare AB | Reference Device K052600 (Zimmer Dental Hex-Lock Prepared Abutment (Straight & Angled) Zimmer Dental Inc. | Reference Device K090234 (Spectra System Dental Implants 2008) Implant Direct LLC | Reference Device K061319 (Spectra System) Implant Direct LLC | Reference Device K060063 (Legacy Abutment System) Implant Direct LLC | Reference Device K143011 (2014 InterActive/ SwishActive System) Implant Direct Sybron Manufacturing LLC | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Product Code | NHA | NHA | NHA | NHA | NHA | NHA | NHA | NHA | | Regulation | 21 C.F.R. §872.3640, 21 C.F.R. §872.3630 | 21 C.F.R. §872.3640, 21 C.F.R. §872.3630 | 21 C.F.R. §872.3630 | 21 C.F.R. §872.3630 | 21 C.F.R. §872.3640, 21 C.F.R. §872.3630 | 21 C.F.R. §872.3640 | 21 C.F.R. §872.3630 | 21 C.F.R. §872.3640, 21 C.F.R. §872.3630 | | Classification | Class II | Class II | Class II | Class II | Class II | Class II | Class II | Class II | | Reason for Predicate / Reference | Not Applicable | Legacy Extender Healing Abutments • Same General Design • Same Intended Use • Same Diameters • Same Connection Type • Same Sterilization • Same Usage • Same Material • Same/similar Screw size - GEN5 and GEN5+ do not offer M1.6 screw size • Same Platform size • Same Collar height • Same Post height • Different finish: The predicate is not anodized. | On1 Base (Extender) • Same General Design • Same Intended Use • Same Usage • Same Material • Same/Similar Platform sizes • Similar Collar height • Both are anodized. | Hex-Lock Abutments (Straight and Angled) • Same General Design • Same Intended Use • Same/Similar Diameters • Same connection Type • Same Usage • Same Material • Same screw size • Same Platform size | • Same General Design • Same Intended Use • Same Sterilization • Same Usage • Same Material • Same Connection Type • Same Sterilization • Same Connection Type • Same Angle • Same/Similar Screw size - GEN5 and GEN5+ do not offer M1.6 screw size • Same Platform size • Same Collar height • Different Post height; Subject approx. 9.7mm; Predicate approx. 9.0mm • Same finish | • Same General Design • Same Intended Use • Same Sterilization • Same Usage • Same Material • Same Connection Type • Same Sterilization • Same Connection Type • Same Angle • Same/Similar Screw size - GEN5 and GEN5+ do not offer M1.6 screw size • Same Platform size • Same Collar height • Different Post height; Subject approx. 9.7mm; Predicate approx. 9.0mm • Same finish | Legacy Angled Multi-Unit Screw-Receiving Abutments • Same General Design • Same Intended Use • Same Diameters • Same Connection Type • Same Sterilization • Same Use • Same Material • Same Angle • Same/similar Screw size - GEN5 and GEN5+ do not offer M1.6 screw size • Same Platform size • Same Collar Height • Different Post height; Subject approx. 9.7mm; Predicate approx. 9.0mm • Same finish | | {15} 510(k) Summary K251938 GEN5 and GEN5+ Dental Implant System Page 12 of 23 | Comparison | Subject Device K243960 (GEN5 and GEN5+ Dental Implant System) | Predicate Device K192221 (Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, Legacy4 fixture-mounts) Implant Direct Sybron Manufacturing, LLC | Reference Device K161655 (On1 Concept) Nobel Biocare AB | Reference Device K052600 (Zimmer Dental Hex-Lock Prepared Abutment (Straight & Angled) Zimmer Dental Inc. | Reference Device K090234 (Spectra System Dental Implants 2008) Implant Direct LLC | Reference Device K061319 (Spectra System) Implant Direct LLC | Reference Device K060063 (Legacy Abutment System) Implant Direct LLC | Reference Device K143011 (2014 InterActive/ SwishActive System) Implant Direct Sybron Manufacturing LLC | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | | • Different finish: The predicate is not anodized. **Legacy Healing Abutments** • Same Platform sizes; Predicate offers additional sizes. • Same Collar height; Predicate offers additional size. • Same/Similar Lengths; Predicate offers additional lengths. • Similar Widths: Subject range: 3.7mm – 5.7mm; Predicate 3.8mm – 6.6mm • Same finish **Legacy Straight Contoured Abutments** • Same Diameters; Predicate offers additional diameter. • Same Platform sizes and additional platform size is within the predicate additional size | 2.4mm; predicate approx. 2.7mm • Same Collar height • Same Post height • Same Finish **Legacy Angled Abutments** • Same Diameters; Predicate offers additional diameter. • Same Platform sizes; Predicate offers an additional platform size. • Same Collar height • Same Post height • Same Finish **Legacy Straight Multi-Unit Abutment (MUA)** • Same Diameters • Same Platform sizes; Predicate offers an additional platform size. • Same/Difference Lengths: Both offer 1mm, 2mm, 4mm; subject also offers 3mm and 5mm. • Different Finish: Subject anodizes; predicate none. | | | {16} 510(k) Summary K251938 GEN5 and GEN5+ Dental Implant System Page 13 of 23 | Comparison | Subject Device K243960 (GEN5 and GEN5+ Dental Implant System) | Predicate Device K192221 (Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, Legacy4 fixture-mounts) Implant Direct Sybron Manufacturing, LLC | Reference Device K161655 (On1 Concept) Nobel Biocare AB | Reference Device K052600 (Zimmer Dental Hex-Lock Prepared Abutment (Straight & Angled) Zimmer Dental Inc. | Reference Device K090234 (Spectra System Dental Implants 2008) Implant Direct LLC | Reference Device K061319 (Spectra System) Implant Direct LLC | Reference Device K060063 (Legacy Abutment System) Implant Direct LLC | Reference Device K143011 (2014 InterActive/ SwishActive System) Implant Direct Sybron Manufacturing LLC | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | | range 3.0mm – 5.7mm. • Same Collar height • Same Post height • Same Finish Legacy Angled Abutments • Same Diameters; Predicate offers additional diameter. • Same Platform sizes; Predicate offers an additional platform size. • Same Collar height • Same Post height • Same Finish Legacy Straight Multi-Unit Abutment (MUA) • Same Diameters • Same Platform sizes; Predicate offers an additional platform size. • Same/Difference Lengths: Both offer 1mm, 2mm, 4mm; subject also offers 3mm and 5mm. | Legacy Angled Multi-Unit Screw-Receiving Abutments • Same General Design • Same Diameters • Same Connection Type • Same Sterilization • Same Usage • Same Material • Same Angle • Same Screw size • Same Platform size • Same/Similar Collar Height: Both have 2mm; Subject has 1mm; Predicate has additional sizes offered. • Same Post height • Different finish: The predicate is not anodized. | | | {17} 510(k) Summary K251938 GEN5 and GEN5+ Dental Implant System Page 14 of 23 | Comparison | Subject Device K243960 (GEN5 and GEN5+ Dental Implant System) | Predicate Device K192221 (Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, Legacy4 fixture-mounts) Implant Direct Sybron Manufacturing, LLC | Reference Device K161655 (On1 Concept) Nobel Biocare AB | Reference Device K052600 (Zimmer Dental Hex-Lock Prepared Abutment (Straight & Angled) Zimmer Dental Inc. | Reference Device K090234 (Spectra System Dental Implants 2008) Implant Direct LLC | Reference Device K061319 (Spectra System) Implant Direct LLC | Reference Device K060063 (Legacy Abutment System) Implant Direct LLC | Reference Device K143011 (2014 InterActive/ SwishActive System) Implant Direct Sybron Manufacturing LLC | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | | • Different Finish: Subject anodizes; predicate none. Legacy Temporary Abutments • Same/Similar Diameters; Predicate offers additional sizes. • Same Platform sizes; Subject offers additional 4.7mm • Same Collar height • Different Post height: Subject Approx 9.7mm; Predicate approx. 9.0mm. • Same finish Legacy PEEK Abutments • Same Diameters; Predicate offers additional sizes. • Same Platform size; Predicate offers additional sizes. • Same Collar height • Different Post height; Subject approx. 8mm; Predicate approx. 10.1mm. | | | | {18} 510(k) Summary K251938 GEN5 and GEN5+ Dental Implant System Page 15 of 23 | Comparison | Subject Device K243960 (GEN5 and GEN5+ Dental Implant System) | Predicate Device K192221 (Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, Legacy4 fixture-mounts) Implant Direct Sybron Manufacturing, LLC | Reference Device K161655 (On1 Concept) Nobel Biocare AB | Reference Device K052600 (Zimmer Dental Hex-Lock Prepared Abutment (Straight & Angled) Zimmer Dental Inc. | Reference Device K090234 (Spectra System Dental Implants 2008) Implant Direct LLC | Reference Device K061319 (Spectra System) Implant Direct LLC | Reference Device K060063 (Legacy Abutment System) Implant Direct LLC | Reference Device K143011 (2014 InterActive/ SwishActive System) Implant Direct Sybron Manufacturing LLC | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | | • Same finish: None | | | | | Prosthesis Attachment | Screw – or cement-retained | Screw – or cement-retained | Screw – or cement-retained | Screw – or cement-retained | Screw – or cement-retained | Screw – or cement-retained | Screw – or cement-retained | Screw – or cement-retained | | Restoration | Single or multi-unit | Single or multi-unit | Single or multi-unit | Single or multi-unit | Single or multi-unit | Single or multi-unit | Single or multi-unit | Single or multi-unit | | Sterilization Method – Sterile Devices (Implants, Healing Abutments, and Cover Screws) | Gamma Sterilization | Gamma Sterilization | Gamma Sterilization | Gamma Sterilization | Gamma Sterilization | Gamma Sterilization | N/A | Gamma Sterilization | | Sterilization Method – Non-sterile Devices (All Other Devices) | Steam Sterilization by end user prior to use per IFU | Steam Sterilization by end user prior to use per IFU | Packaged as Sterile | Packaged as Sterile; Steam Sterilization by end user prior to use per IFU | Steam Sterilization by end user prior to use per IFU | Steam Sterilization by end user prior to use per IFU | Steam Sterilization by end user prior to use per IFU | Steam Sterilization by end user prior to use per IFU | | Usage | Single use | Single use | Single Use | Single Use | Single use | Single use | Single use | Single use | | Abutment Material | Ti-6Al-4V ELI alloy; PEEK | Ti-6Al-4V ELI alloy | Ti-6Al-4V ELI alloy | Ti-6Al-4V ELI alloy | Ti-6Al-4V ELI alloy | Ti-6Al-4V ELI alloy | Ti-6Al-4V ELI alloy | Ti-6Al-4V ELI alloy | | Screw Size | 1-72 UNF 2A and M1.4 | 1-72 UNF 2A and M1.6 | M1.6 and M1.4 | 1-72 UNF 2A | 1-72 UNF 2A and M1.6 | 1-72 UNF 2A | 1-72 UNF 2A and M1.6 | 1-72 UNF 2A and M1.6 | | Abutment Screw Material | Ti-6Al-4V ELI alloy | Ti-6Al-4V ELI alloy | Ti-6Al-4V ELI alloy | Ti-6Al-4V ELI alloy | Ti-6Al-4V ELI alloy | Ti-6Al-4V ELI alloy | Ti-6Al-4V ELI alloy | Ti-6Al-4V ELI alloy | | Use with Implant Diameters | 3.7mm; 4.2mm; 4.7mm; 5.2mm; 5.7mm | 3.2mm, 3.7mm, 4.2mm, 4.7mm, 5.2mm, 5.7mm, 7.0mm | 3.5mm, 3.75, 4.3mm, 5.0mm, 5.5mm | 3.7mm, 4.1mm, 4.7mm, 6.0mm | 3.2mm, 3.7mm, 4.2mm, 4.7mm, 5.2mm, 5.7mm, 7.0mm | 3.0mm, 3.2mm, 3.7mm, 4.2mm, 4.3mm, 4.7mm, 5.0mm, 5.2mm, 5.7mm, 7.0 | 3.2mm, 3.7mm, 4.2mm, 4.7mm, 5.2mm, 5.7mm, 7.0mm | 3.2mm, 3.7mm, 5.0mm | | Cover Screws | Platform Diameter: | N/A | N/A | N/A | Platform Diameter: | N/A | N/A | N/A | {19} 510(k) Summary K251938 GEN5 and GEN5+ Dental Implant System Page 16 of 23 | Comparison | Subject Device | Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device | Reference Device | | --- | --- | --- | --- | --- | --- | --- | --- | | K243960 (GEN5 and GEN5+ Dental Implant System) | K192221 (Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, Legacy4 fixture-mounts) Implant Direct Sybron Manufacturing, LLC | K161655 (On1 Concept) Nobel Biocare AB | K052600 (Zimmer Dental Hex-Lock Prepared Abutment (Straight & Angled) Zimmer Dental Inc. | K090234 (Spectra System Dental Implants 2008) Implant Direct LLC | K090234 (Spectra System Dental Implants 2008) Implant Direct LLC | K061319 (Spectra System) Implant Direct LLC | K143011 (2014 InterActive/ SwishActive System) Implant Direct Sybron Manufacturing LLC | | | 3.5mm and 4.5mm Width: 3.4mmm 4.4mm, 5.8mm Length: 1.0mm, 2.0mm, 3.0mm Finish: Anodized | | | | 3.0mm, 3.5mm, 4.5mm, 5.7mm Width: 3.5mm, 4.5mm, 5.7mm Finish: None | | | | Healing Abutments | Platform: 3.5mm, 4.5mm Collar Height: 3mm, 5mm Angle: Straight Width: 3.7mm, 4.7mm, 5.7mm Length: 3.0mm, 5.0mm Angle: Straight Finish: | N/A | N/A | N/A | Platform: 3.0mm, 3.5mm, 4.5mm,5.7mm Collar Height: 3mm, 5mm Angle: Straight Width: 3.8mmD, 4.8mm, 5.8mm, 6.6mm Length: 3.0mm, 5.0mm, 7.0mm Angle: Straight | N/A | N/A | {20} 510(k) Summary K251938 GEN5 and GEN5+ Dental Implant System Page 17 of 23 | Comparison | Subject Device | Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device | Reference Device | | --- | --- | --- | --- | --- | --- | --- | --- | | K243960 (GEN5 and GEN5+ Dental Implant System) | K192221 (Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, Legacy4 fixture-mounts) Implant Direct Sybron Manufacturing, LLC | K161655 (On1 Concept) Nobel Biocare AB | K052600 (Zimmer Dental Hex-Lock Prepared Abutment (Straight & Angled) Zimmer Dental Inc. | K090234 (Spectra System Dental Implants 2008) Implant Direct LLC | K090234 (Spectra System Dental Implants 2008) Implant Direct LLC | K061319 (Spectra System) Implant Direct LLC | K143011 (2014 InterActive/ SwishActive System) Implant Direct Sybron Manufacturing LLC | | | Anodized | | | | Finish: Anodized | | | | Extender Healing Abutments | Platform: 3.5mm, 4.5mm, 5.7mm Collar Height: 1mm, 2mm Post Height: Approx. 7.6mm Interface Length: 2.0mm Finish: Anodized | Platform: 3.5mm, 4.5mm, 5.7mm Collar Height: 1mm, 2mm Post Height: Approx. 7.6mm Finish: None | Platform: 3.5mm (NP), 4.3mm (RP), 5.0mm (WP) Collar Height: 1.75mm, 2.5mm Finish: Anodized | Friction-Fit | N/A | N/A | N/A | | Straight Contoured Abutments | Platform: 3.5mm, 4.5mm, 4.7mm, 5.7mm Diameter: 3.7mm, 4.7mm, 5.7mm Collar Height: 1mm, 2mm Post Height: Approx, 7.6mm Angle: | N/A | N/A | N/A | Platform: 3.0mm, 3.5mm, 4.5mm, 5.7mm Diameter: 3.7mm, 4.7mm, 5.7mm, 6.5mm Collar Height: 1mm, 2mm Post Height: Approx. 7.6mm Angle: | Platform: 3.5mm, 4.5mm, 5.0mm, 6.0mm Diameter: 3.7mm, 4.7mm, 5.7mm, 6.5mm Collar Height: 1mm, 2mm Post Height: Approx. 7.6mm Angle: | N/A | {21} 510(k) Summary K251938 GEN5 and GEN5+ Dental Implant System Page 18 of 23 | Comparison | Subject Device | Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device | Reference Device | | --- | --- | --- | --- | --- | --- | --- | --- | | K243960 (GEN5 and GEN5+ Dental Implant System) | K192221 (Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, Legacy4 fixture-mounts) Implant Direct Sybron Manufacturing, LLC | K161655 (On1 Concept) Nobel Biocare AB | K052600 (Zimmer Dental Hex-Lock Prepared Abutment (Straight & Angled) Zimmer Dental Inc. | K090234 (Spectra System Dental Implants 2008) Implant Direct LLC | K090234 (Spectra System Dental Implants 2008) Implant Direct LLC | K061319 (Spectra System) Implant Direct LLC | K060063 (Legacy Abutment System) Implant Direct LLC | | Straight Hex Width: Approx 2.4mm Interface Length: Approx 1.9mm Finish: Anodized | | | | Straight Hex Width: Approx 2.7mm Interface Length: Approx 1.9mm Finish: Anodized | Straight Hex Width: Approx 2.7mm Interface Length: Approx 1.9mm Finish: Anodized | | | | Angled Abutments | Platform: 3.5mm, 4.5mm, 4.7mm Diameter: 4.7mm, 5.7mm Collar Height: Approx. 0.75mm, 2mm Post Height: Min. 7.75mm Hex Width: Approx 2.4mm Interface Length: Approx 1.9mm Angle: | N/A | N/A | Platform: 3.0mm, 3.5mm, 4.5mm, 5.7mm Diameter: 4.7mm, 5.7mm, 6.5mm Collar Height: 0.75mm, 2mm Post Height: Min. 7.75mm Hex Width: Approx 2.7mm Interface Length: Approx 1.9mm Angle: | Platform: 3.0mm, 3.5mm, 4.5mm, 5.7mm Diameter: 4.7mm, 5.7mm, 6.5mm Collar Height: 0.75mm, 2mm Post Height: Min. 7.75mm Hex Width: Approx 2.7mm Interface Length: Approx 1.9mm Angle: | N/A | N/A | {22} 510(k) Summary K251938 GEN5 and GEN5+ Dental Implant System Page 19 of 23 | Comparison | Subject Device | Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device | Reference Device | Reference Device | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | K243960 (GEN5 and GEN5+ Dental Implant System) | K192221 (Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, Legacy4 fixture-mounts) Implant Direct Sybron Manufacturing, LLC | K161655 (On1 Concept) Nobel Biocare AB | K052600 (Zimmer Dental Hex-Lock Prepared Abutment (Straight & Angled) Zimmer Dental Inc. | K090234 (Spectra System Dental Implants 2008) Implant Direct LLC | K061319 (Spectra System) Implant Direct LLC | K060063 (Legacy Abutment System) Implant Direct LLC | K143011 (2014 InterActive/ SwishActive System) Implant Direct Sybron Manufacturing LLC | | | Angled 15 degree Finish: Anodized | | | | Angled 15 degree Finish: None | Angled 15 degree Finish: None | | | | | Angled Screw Channel (ASC) Engaging Abutments | Platform: 3.5mm, 4.5mm, 4.7mm Diameter: 3.7mm, 4.7mm, 5.2mm, 5.7mm Collar Height: 0.5mm, 2.0mm Post Height: 4.75mm Hex Width: Approx. 2.4mm Interface Length: Approx. 1.9mm Angle: Straight Finish: Anodized | N/A | N/A | N/A | N/A | N/A | N/A | N/A | | Angled Screw Channel (ASC) | Platform: | N/A | N/A | N/A | N/A | N/A | N/A | N/A | {23} 510(k) Summary K251938 GEN5 and GEN5+ Dental Implant System Page 20 of 23 | Comparison | Subject Device | Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device | Reference Device | Reference Device | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | K243960 (GEN5 and GEN5+ Dental Implant System) | K192221 (Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, Legacy4 fixture-mounts) Implant Direct Sybron Manufacturing, LLC | K161655 (On1 Concept) Nobel Biocare AB | K052600 (Zimmer Dental Hex-Lock Prepared Abutment (Straight & Angled) Zimmer Dental Inc. | K090234 (Spectra System Dental Implants 2008) Implant Direct LLC | K061319 (Spectra System) Implant Direct LLC | K060063 (Legacy Abutment System) Implant Direct LLC | K143011 (2014 InterActive/ SwishActive System) Implant Direct Sybron Manufacturing LLC | | | Non-Engaging Abutments | 3.5mm, 4.5mm, 4.7mm Diameter: 3.7mm, 4.7mm, 5.2mm, 5.7mm Collar Height: 0.5mm, 2mm Post Height: 4.75mm Angle: Straight Width: 3.7mm, 4.7mm Finish: Anodized | | | | | | | | | Straight Multi-Unit Abutment (MUA) | Platform: 3.5mm, 4.5mm, 4.8mm Diameter: 4.8mm Collar Height: | N/A | N/A | N/A | Platform: 3.0mm, 3.5mm Diameter: 4.8mm Collar Height: 1mm, 2mm, 4mm Angle: | Platform: 4.5mm, 5.7mm Diameter: 4.8mm Collar Height: 1mm, 2mm, 4mm Angle: | N/A | N/A | {24} 510(k) Summary K251938 GEN5 and GEN5+ Dental Implant System Page 21 of 23 | Comparison | Subject Device | Predicate Device | Preference Device | Reference Device | Reference Device | Reference Device | Reference Device | | --- | --- | --- | --- | --- | --- | --- | --- | | K243960 (GEN5 and GEN5+ Dental Implant System) | K192221 (Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, Legacy4 fixture-mounts) Implant Direct Sybron Manufacturing, LLC | K161655 (On1 Concept) Nobel Biocare AB | K052600 (Zimmer Dental Hex-Lock Prepared Abutment (Straight & Angled) Zimmer Dental Inc. | K090234 (Spectra System Dental Implants 2008) Implant Direct LLC | K090234 (Spectra System Dental Implants 2008) Implant Direct LLC | K061319 (Spectra System) Implant Direct LLC | K143011 (2014 InterActive/ SwishActive System) Implant Direct Sybron Manufacturing LLC | | | 1.0mm, 2.0mm, 3.0mm, 4.0mm, 5.0mm Angle: Straight Finish: Anodized | | | | Straight Finish: None | Straight Finish: None | | | Angled Multi-Unit Screw-Receiving Abutments | Platform: 3.5mm, 4.5mm, 4.8mm Diameter: 4.8mm, 5.5mm, 7.0mm Collar Height: 1mm, 2mm Post Height: Approx 2mm Hex Width: Approx. 2.4mm Interface Length: Approx 1.9mm Angle: | N/A | N/A | N/A | N/A | Platform: 3.5mm, 4.5mm Diameter: Approx 4.8mm Collar Height: 2.0mm, 3.0mm, 4.0mm Post Height: Approx 2mm Hex Width: Approx. 2.7mm Interface Length: Approx 1.9mm Angle: 15 degrees; 30 degrees Finish: | N/A | {25} 510(k) Summary K251938 GEN5 and GEN5+ Dental Implant System Page 22 of 23 | Comparison | Subject Device | Predicate Device | Reference Device | Reference Device | Reference Device | Reference | Finish: | | --- | --- | --- | --- | --- | --- | --- | --- | | K243960 (GEN5 and GEN5+ Dental Implant System) | K192221 (Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, Legacy4 fixture-mounts) Implant Direct Sybron Manufacturing, LLC | K161655 (On1 Concept) Nobel Biocare AB | K052600 (Zimmer Dental Hex-Lock Prepared Abutment (Straight & Angled) Zimmer Dental Inc. | K090234 (Spectra System Dental Implants 2008) Implant Direct LLC | K061319 (Spectra System) Implant Direct LLC | K060063 (Legacy Abutment System) Implant Direct LLC | None | | | 17 degrees; 30 degrees Finish: Anodized | | | | | | Finish: None | | Temporary Abutments | Platform: 3.5mm, 4.5mm, 4.7mm Diameter: 4.7mm, 5.2mm, 5.7mm Collar Height: Approx. 1mm Post Height: Approx. 9.7mm Angle: Straight Finish: Anodized | N/A | N/A | N/A | Platform: 3.0mm, 3.5mm, 4.5mm, 5.7mm Diameter: 3.7mm, 4.2mm, 4.7mm, 5.2mm, 5.7mm; 6.5mm Collar Height: Approx. 1mm Post Height: Approx, 9.0mm Angle: Straight Finish: None | N/A | Platform: 3.5mm, 4.5mm Diameter: 3.5mm, 5.7mm, 6.5mm Collar Height: Approx. 1mm Post Height: Approx, 9.0mm Angle: Straight Finish: Anodized | | Plastic (PEEK) Temporary Abutments | Platform: 3.5mm, 4.5mm Diameter: | N/A | N/A | N/A | Platform: 3.5mm, 4.5mm, 5.7mm Diameter: | N/A | N/A | {26} 510(k) Summary K251938 GEN5 and GEN5+ Dental Implant System Page 23 of 23 | Comparison | Subject Device | Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device | Reference Device | | --- | --- | --- | --- | --- | --- | --- | --- | | K243960 (GEN5 and GEN5+ Dental Implant System) | K192221 (Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, Legacy4 fixture-mounts) Implant Direct Sybron Manufacturing, LLC | K161655 (On1 Concept) Nobel Biocare AB | K052600 (Zimmer Dental Hex-Lock Prepared Abutment (Straight & Angled) Zimmer Dental Inc. | K090234 (Spectra System Dental Implants 2008) Implant Direct LLC | K061319 (Spectra System) Implant Direct LLC | K060063 (Legacy Abutment System) Implant Direct LLC | K143011 (2014 InterActive/ SwishActive System) Implant Direct Sybron Manufacturing LLC | | 4.7mm, 5.2mm, 5.7mm Collar Height: 1.0mm Post Height: Approx 8mm Angle: Straight Hex Width: Approx 2.4mm Interface Length: Approx 1.9mm | | | | | 3.7mm, 4.2mm, 4.7mm, 5.2mm, 5.7mm; 6.5mm Collar Height: 1.0mm Post Height: Approx 10.1mm Angle: Straight: Hex Width: Approx 2.7mm Interface Length: Approx 1.9mm | | |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...