Dentis SQ-SL AXEL Fixture
K253493 · Dentis Co., Ltd. · DZE · Mar 4, 2026 · Dental
Device Facts
| Record ID | K253493 |
| Device Name | Dentis SQ-SL AXEL Fixture |
| Applicant | Dentis Co., Ltd. |
| Product Code | DZE · Dental |
| Decision Date | Mar 4, 2026 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.3640 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
Dentis SQ -SL AXEL Fixture is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. This system is intended for delayed loading.
Device Story
Threaded endosseous dental implant; fabricated from pure titanium (ASTM F67); surface treated with SLA (sandblasted, large-grit, acid-etched). Placed in alveolar bone by dental surgeons to replace missing teeth; provides foundation for prosthetic restorations (crowns, bridges, overdentures). Device features internal hex connection for abutment interface. System includes Mini and Regular variants with varying diameters and lengths. Provided sterile via gamma irradiation. Clinical benefit: restoration of masticatory function and aesthetics in edentulous patients. Healthcare providers use the device to support prosthetic components; output is a stable, osseointegrated anchor for dental restorations.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including fatigue testing (ISO 14801) and SEM/EDS surface characterization. Biocompatibility, sterilization, and shelf-life data were leveraged from predicate devices (K192688, K153639) based on identical materials and manufacturing processes.
Technological Characteristics
Endosseous root-form dental implant; material: CP Titanium Grade 4 (ASTM F67); surface: SLA (sandblasted, large-grit, acid-etched); connection: internal hex; sterilization: gamma irradiation; shelf life: 8 years. Dimensions: Ø3.8-7.0mm, lengths 7.0-17.5mm.
Indications for Use
Indicated for partially or fully edentulous mandibles and maxillae in patients requiring single or multiple unit restorations (cemented, screw-retained, or overdenture) or fixed bridgework support. Dedicated for one or two-stage surgery; intended for delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
- s-Clean SQ-SL Implant System Regular (K192688)
Reference Devices
- HIOSSEN Implant System (K140934)
- OneQ-SL s-Clean Implant System (K153639)
- S-Clean SQ-SL Implant System Mini (K202773)
- Dentis s-Clean s-Line Mini (K210080)
- Dentis s-Clean s-Line (K210134)
Related Devices
- K233163 — ZENEX Implant System_Short · Izenimplant Co., Ltd. · Dec 19, 2023
- K210132 — s-Clean SQ-SL Implant System Regular · Dentis Co., Ltd. · Feb 17, 2021
- K202046 — LOTA SLA Dental Implant System and LOTA HA Dental Implant System · Kj Meditech Co., Ltd. · Mar 1, 2022
- K061500 — SICACE DENTAL IMPLANT SYSTEM AND SIC ANGLED ABUTMENTS · Sic Invent AG · Aug 8, 2006
- K153639 — OneQ-SL s-Clean Implant System · Dentis Co., Ltd. · Apr 8, 2016
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
Dentis Co., Ltd.
% April Lee
Regulatory Consultant
Withus Group, Inc.
106 Superior
Irvine, California 92620
March 4, 2026
Re: K253493
Trade/Device Name: Dentis SQ-SL AXEL Fixture
Regulation Number: 21 CFR 872.3640
Regulation Name: Endosseous Dental Implant
Regulatory Class: Class II
Product Code: DZE
Dated: October 24, 2025
Received: February 2, 2026
Dear April Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253493 - April Lee
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K253493 - April Lee
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Andrew I. Steen -S
Andrew I. Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K253493
Device Name
Dentis SQ-SL AXEL Fixture
Indications for Use (Describe)
Dentis SQ -SL AXEL Fixture is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. This system is intended for delayed loading.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
K253493
Page 1 of 6
# 510(K) Summary
## Submitter
Dentis Co., Ltd.
Ju Mi Hong
99, Seongseoseo-ro, Dalseo-gu
Daegu, 42718 Korea
Email: hjm1002@dentis.co.kr
Tel. +82-53-589-3644
Fax. +82-53-289-7922
## Official Correspondent
Withus Group Inc.
April Lee
106 Superior,
Irvine, CA 92620 USA
Email: withus6664@gmail.com
Phone: 1-909-274-9971
Fax: 1-909-460-8122
## Device Information
- Trade Name: Dentis SQ-SL AXEL Fixture
- Common Name: Endosseous dental implant
- Classification Name: implant, endosseous, root-form
- Product Code: DZE
- Panel: Dental
- Regulation Number: 872.3640
- Device Class: Class II
- Date Prepared: 03/03/2026
## Predicate Devices:
### Primary Predicate
- K192688, s-Clean SQ-SL Implant System Regular by Dentis Co., Ltd.
### Reference Device
- K140934, HIOSSEN Implant System by HIOSSEN, Incorporated
- K153639, OneQ-SL s-Clean Implant System by Dentis Co., Ltd.
- K202773, S-Clean SQ-SL Implant System Mini by Dentis Co., Ltd.
- K210080, Dentis s-Clean s-Line Mini by Dentis Co., Ltd.
- K210134, Dentis s-Clean s-Line by Dentis Co., Ltd.
## Indication for Use:
Dentis SQ -SL AXEL Fixture is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. This system is intended for delayed loading.
{5}
K253493
Page 2 of 6
# Device Description:
Dentis SQ-SL AXEL Fixture is two types as Mini and Regular according to the connection. Dentis SQ-SL AXEL Fixture is a thread type implant made of Pure titanium according to ASTM F67 which will be placed in the alveolar bone to replace the function of the missing tooth. This device has connection between the upper prosthesis and the internal Hex.
The surface of fixture is treated with SLA (Sandblasted with Large-grit and Acid-etching).
The dimensions of fixtures are as following:
| No. | Device Name | Dimension Ranges |
| --- | --- | --- |
| 1 | SQ-SL AXEL Fixture | Ø4.2, 4.6, 5.0, 5.4, 6.0, 7.0 (D) X 7.0, 7.5, 8.5, 9.5, 11.5, 13.5, 15.5, 17.5mm |
| 2 | SQ-SL AXEL Fixture Mini | Ø3.8, 4.2 (D) X 7.5, 8.5, 9.5, 11.5, 13.5, 15.5, 17.5mm |
Tolerance of dimension shall be within $\pm 1\%$ range.
Dentis SQ-SL Fixture is compatible with K192688 and K210134 as below:
| 510(K) | Abutment Name | Diameter(Ø) | Angulation | Length(mm) |
| --- | --- | --- | --- | --- |
| K192688 | s-Clean Cover Screw | Ø3.6mm | 0 | 5.9mm |
| | s-Clean (TiN Half Coating) Sole Abutment S-Line | Ø4.5, 5.5, 6.5 and 7.5 | 0 | 11.6, 12.6 and 13.6 |
| K210134 | s-Clean (TiN Half Coating) Angled Abutment | Ø4.5 | 15 | 12.6 |
Dentis SQ-SL Fixture Mini is compatible with K210080 as below:
| 510(K) | Abutment Name | Diameter (Ø) | Angulation | Length(mm) |
| --- | --- | --- | --- | --- |
| K210080 | s-Clean Cover Screw Mini | Ø3.2mm | 0 | 5.0mm |
| | s-Clean Healing Abutment S-Line Mini | Ø4.3, 4.8 and 5.8 | 0 | 7.61, 8.61, 9.61, 10.61, 11.61, 12.61 and 14.61 |
| | s-Clean (TiN Half Coating) Sole Abutment Mini | Ø4.5 and 5.5 | 0 | 11.01, 12.01, 12.51, 13.01, 13.521, 14.01, 14.51, 15.01, 15.51, 16.01 and 17.01 |
| | s-Clean (TiN Half Coating) Couple Abutment Mini | Ø4.0, 4.5 and 5.5 | 0 | 8.35, 8.6, 9.35, 9.6, 9.85, 10.1, 10.35, 10.6, 11.1, 11.35, 11.6, 12.1, 12.85, 13.1, 13.35, 13.6, 14.35 and 14.6 |
| | s-Clean (TiN Half Coating) Angled Abutment Mini | Ø4.0 and 4.5 | 15 | 12.09, 12.34, 12.51 and 12.76 |
Dentis SQ-SL AXEL Fixture is provided sterilized.
Dentis SQ-SL AXEL Fixture is packaged with Cover screw that was cleared in FDA as K192688 and K210080.
# Materials:
- SQ-SL AXEL Fixture and SQ-SL AXEL Fixture Mini are fabricated from Pure titanium of ASTM F67
{6}
K253493
Page 3 of 6
Summaries of Technological Characteristics & Substantial Equivalence Discussion
SQ-SL AXEL Fixture
| | Subject Device | Primary Predicate | Reference Device |
| --- | --- | --- | --- |
| K number | NA | K192688 | K140934 |
| Manufacturer | Dentis Co., Ltd | Dentis Co., Ltd | HIOSSEN Inc. |
| Trade Name | Dentis SQ-SL AXEL Fixture | s-Clean SQ-SL Implant System Regular | Osstem Implant System |
| Product Name | SQ-SL AXEL Fixture | s-Clean SQ-SL Fixture | ETII SA Fixture |
| Design | | | |
| Indications for Use | Dentis SQ-SL AXEL Fixture is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. This system is intended for delayed loading. | s-Clean SQ-SL Implant System Regular is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures and not dedicated for immediate loading. This system is intended for delayed loading. | The HIOSSEN Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multipleunit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. ETIII SA Ultra-Wide Fixture is intended to be used in the molar region. |
| Diameter and Length | Regular: Ø4.2 x 7, 7.5, 8.5, 9.5, 11.5, 13.5, 15.5, 17.5mm
Ø4.6 x 7, 7.5, 8.5, 9.5, 11.5, 13.5, 15.5, 17.5mm
Ø5.0 x 7, 7.5, 8.5, 9.5, 11.5, 13.5, 15.5, 17.5mm
Ø5.4 x 7, 7.5, 8.5, 9.5, 11.5, 13.5mm
Wide: Ø6.0 x 7, 7.5, 8.5, 9.5, 11.5, mm
Ø7.0 x 7, 7.5, 8.5, 9.5, 11.5, mm | Ø4.1 x 7, 7.5, 9.5, 11.5, 13.5mm
Ø4.35 x 7, 7.5, 9.5, 11.5, 13.5mm
Ø4.8 x 7, 7.5, 9.5, 11.5, 13.5mm
Ø5.8 x 7, 7.5, 9.5, 11.5mm
Ø6.8 x 7, 7.5, 9.5, 11.5mm
Ø7.8 x 7, 7.5, 9.5, 11.5mm | Ø3.5 x 8.7, 10.2,11.7, 13.2, 15.2, 18.2mm
Ø4.2 x 7.2, 8.7, 10.2, 11.7, 13.2, 15.2, 18.2mm
Ø4.45 x 7.2, 8.7,10.2, 11.7, 13.2, 15.2mm
Ø5.0x 6.2mm
Ø4.9 x 7.2, 8.7, 10.2, 11.7, 13.2, 15.2mm |
| Surface Treatment | SLA | SLA | SA |
| Material | CP Titanium Gr4 (ASTM F67) | CP Titanium Gr4 (ASTM F67) | CP Titanium Gr4 (ASTM F67) |
| Sterilization | Gamma Irradiation | Gamma Irradiation | Gamma Irradiation |
| Shelf Life | 8years | 8years | 8years |
| Comparison | The Subject Device and primary device, K192688 have same characteristics such as indications for Use, design, surface treatment, material, sterilization, and shelf life. The difference between subject device and primary predicate is dimension. The subject device has longer lengths fixtures, such as 15.5 and 17.5mm. To support this difference, K140934 was added because the dimension combination covers all range of the subject device's dimensions. Therefore, it is substantially equivalent. | | |
{7}
K253493
Page 4 of 6
# SQ-SL AXEL Fixture Mini
| | Subject Device | Reference Device | Reference Device |
| --- | --- | --- | --- |
| K number | NA | K202773 | K140934 |
| Manufacturer | Dentis Co., Ltd | Dentis Co., Ltd | HIOSSEN Inc. |
| Trade Name | Dentis SQ-SL AXEL Mini Fixture | s-Clean SQ-SL Implant System Mini | Osstem Implant System |
| Product Name | SQ-SL AXEL Mini Fixture | s-Clean SQ-SL Mini Fixture | ETII SA Fixture |
| Design | | | |
| Indications for Use | Dentis SQ-SL Mini Fixture is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. This system is intended for delayed loading. | s-Clean SQ-SL Implant System Mini is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. This system is intended for delayed loading. | The HIOSSEN Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multipleunit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. ETIII SA Ultra-Wide Fixture is intended to be used in the molar region. |
| Diameter and Length | Ø3.8 x 7.5, 8.5, 9.5, 11.5, 13.5, 15.5, 17.5mm
Ø4.2 x 7.5, 8.5, 9.5, 11.5, 13.5, 15.5, 17.5mm | Ø3.7 x 7.5, 9.5, 11.5, 13.5mm
Ø4.1 x 7.5, 9.5, 11.5, 13.5mm | Ø3.5 x 8.7, 10.2, 11.7, 13.2, 15.2, 18.2mm
Ø4.2 x 7.2, 8.7, 10.2, 11.7, 13.2, 15.2, 18.2
Ø4.45 x 7.2, 8.7, 10.2, 11.7, 13.2, 15.2mm
Ø5.0x 6.2mm
Ø4.9 x 7.2, 8.7, 10.2, 11.7, 13.2, 15.2mm |
| Surface Treatment | SLA | SLA | SA |
| Material | CP Titanium Gr4 (ASTM F67) | CP Titanium Gr4 (ASTM F67) | CP Titanium Gr4 (ASTM F67) |
| Sterilization | Gamma Irradiation | Gamma Irradiation | Gamma Irradiation |
| Shelf Life | 8years | 8years | 8years |
| Comparison | The Subject Device and reference device, K202773 have same characteristics such as indications for Use, design (dual screw joint), surface treatment, material, sterilization, and shelf life. The difference between subject device and primary predicate is dimension. The subject device has longer lengths fixtures, such as 15.5 and 17.5mm. To support this difference, K140934 was added because the dimension combination covers all range of the subject device's dimensions. Therefore, it is substantially equivalent. | | |
{8}
K253493
Page 5 of 6
# Non-Clinical Test Data
Below tests were performed on subject device:
- Fatigue Testing according to ISO 14801:2016 under the worst-case scenario
- SEM/EDS analysis
Below tests were performed for predicate devices and leveraged for the subject device:
- Sterilization Validation Test on Fixtures according to ISO 11137-1,2,3 referenced in K192688
- Shelf-Life Test on Fixtures according to ASTM F1980 referenced in K153639
- Biocompatibility testing on fixtures according to ISO 10993-1:2009, ISO 10993-3:2014, ISO 10993-5:2009, ISO 10993-6:2007, ISO 10993-10:2010 and ISO 10993-11:2006 referenced in K153639
- Bacterial Endotoxin Test Report on Fixtures according to ANSI/AAMI ST72:2011, USP <161>, and USP <85> referenced in K192688
The results of the above tests have met the criteria of the standards and demonstrated the substantial equivalence with the predicate device.
Non-clinical test data was conducted in accordance with FDA Guidance “Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments”. Clinical testing was not necessary to establish substantial equivalency of the device.
The surface modification information with SLA (Sandblasted with Large-grit and Acid-etching) for fixtures was provided. The implant body surface was characterized using Scanning Electron Microscopy (SEM) and Energy Dispersive X-ray Spectroscopy (EDS).
The Sterilization validation test and shelf-life test for fixtures were performed for cleared device, K192688 and K153639 and leveraged for the subject device because the material, sterilization method, packaging methods, and manufacturing process of both products are exactly same.
The Biocompatibility Test was conducted on the predicate device, K153639 and leveraged for the subject device because both products are manufactured with the same materials and manufacturing process. It demonstrates that the subject device is biocompatible and substantial equivalence with the predicate.
Dynamic fatigue tests were newly performed on the subject device with the compatible abutments to demonstrate the mechanical value according to the FDA guidance document “Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments” and ISO 14801. The Results of fatigue testing showed that subject devices are substantially equivalent under the worst-case scenario.
The non-clinical testing results demonstrate that the subject device is substantially equivalent to the predicate device.
{9}
K253493
Page 6 of 6
## MR Environment Condition
Non-clinical worst-case MRI review was performed to evaluate the metallic. Dentis SQ-SL AXEL Fixture in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry 0., Jana G. Delfino, and Sunder Rajan. 'Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices.' Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system including all variations (all compatible implant bodied, dental abutments and, fixation screws) and material composition. Rationale addressed parameters per the FDA Guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.
## Conclusion
Dentis SQ-SL AXEL Fixture constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental scientific technology as its predicate devices. Therefore, Dentis SQ-SL AXEL Fixture and its predicates are substantially equivalent.