K190958 · Jjgc Industria E Comercio DE Materiais Dentarios S.A. · DZE · Nov 9, 2019 · Dental
Device Facts
Record ID
K190958
Device Name
Neodent Implant System
Applicant
Jjgc Industria E Comercio DE Materiais Dentarios S.A.
Product Code
DZE · Dental
Decision Date
Nov 9, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. The Neodent GM Helix LG implants can be placed bicortically in cases of reduced bone density. The Neodent GM Helix LG implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants.
Device Story
Neodent Implant System consists of endosseous dental implants (GM Helix LG) designed for surgical placement in upper or lower jaw bone. Implants provide support for prosthetic devices (artificial teeth) to restore chewing function. Device features include conical body, trapezoidal threads, and Grand Morse prosthetic interface. Used by dental surgeons in clinical settings. Implants are single-use, provided sterile. Clinical benefit includes restoration of chewing function and esthetics. Output is the physical implant acting as an anchor for prosthetic restorations. Decision-making relies on surgeon assessment of bone density and primary stability; bicortical placement is an option for reduced bone density cases.
Clinical Evidence
Bench testing only. Performance data includes dynamic fatigue testing per ISO 14801, static torsional loading, insertion torque testing in synthetic bone (types II, III, IV), and axial pull-out strength comparisons. Biocompatibility and biological safety assessments conducted per ISO 10993-1, 10993-5, 10993-12, and 10993-18. Sterilization validated per ISO 11137-2 to SAL 1x10^-6.
Indicated for patients requiring dental restoration via endosseous implants in the upper or lower jaw. Suitable for single or multiple unit restorations, including immediate loading when primary stability is achieved. GM Helix LG implants are indicated for bicortical placement in reduced bone density and restricted to splinted, multiple-unit applications using at least two implants.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
K201225 — Neodent Implant System - GM Helix Implants 7.0 · Jjgc Industria E Comercio DE Materiais Dentarios S.A. · Sep 4, 2020
K223638 — Neodent Implant System - Helix Short Implant System · Jjgc Industria E Comercio DE Materiais Dentarios S.A. · Jun 23, 2023
K133592 — NEODENT IMPLANT SYSTEM, TITAMAX TI EX ACQUA AND TI DRIVE ACQUA, TITAMAX SMART EX ACQUA AND DRIVE SMART ACQUA, TITAMAX CM · Neodent USA, Inc. · Aug 22, 2014
K203382 — Neodent Implant System-Easy Pack · Jjgc Industria E Comercio DE Materiais Dentarios S.A. · Mar 15, 2021
K180536 — Neodent Implant System GM Line · Jjgc Industria E Comercio DE Materiais Dentarios S.A. · Aug 30, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 9, 2019
JJGC Industria e Comercio de Materiais Dentarios SA % Jennifer M. Jackson Director Regulatory Affairs Straumann 60 Minuteman Road Andover, Massachusetts 01810
Re: K190958
Trade/Device Name: Neodent Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: October 10, 2019 Received: October 11, 2019
Dear Jennifer M. Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Srinivas Nandkumar, Ph.D. Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190958
Device Name Neodent Implant System
#### Indications for Use (Describe)
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. The Neodent GM Helix LG implants can be placed bicortically in cases of reduced bone density. The Neodent GM Helix LG implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) Summary | |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ADMINISTRATIVE INFORMATION | |
| Sponsor | JJGC Indústria e Comércio de Materiais Dentários SA<br>(dba Neodent)<br>Av. Juscelino Kubitschek de Oliveira, 3291<br>Curitiba, Paraná, Brazil 81270-200<br>Registration No.: 3008261720<br>Owner/Operator No.: 10031702 |
| Contact Person | Jennifer M. Jackson, MS<br>Director of Regulatory Affairs,<br>Straumann USA<br>E-Mail: jennifer.jackson@straumann.com<br>Telephone (978) 747-2509 |
| Date Prepared | 08/Nov/2019 |
| Preparer / Alternate Contact | Mariana Soares Hartmann<br>Regulatory Affairs Analyst<br>mariana.hartmann@neodent.com |
| DEVICE NAME AND CLASSIFICATION | |
| Trade/ Proprietary Name<br>Common Name | Neodent Implant System<br>Endosseous dental implant |
| Classification Name | Implant, Endosseous, Root-Form |
| Classification Regulations<br>Product Code | 21 CFR 872.3640, Class II<br>DZE |
| Classification Panel<br>Reviewing Branch | Dental Products Panel<br>Dental Devices Branch |
| PREDICATE DEVICE INFORMATION | |
| Primary Predicate Device | K160119, NobelSpeedy® Groovy, Nobel Biocare AB |
| Reference Devices | K163194, Neodent Implant System - GM Line, JJGC Indústria<br>e Comércio de Materiais Dentários S.A<br>K180536, Neodent Implant System - GM Line, JJGC Indústria<br>Comércio de Materiais Dentários S.A<br>K190718, Neodent GM Zygomatic Implants – GM Line, JJGC<br>Indústria e Comércio de Materiais Dentários S.A |
{4}------------------------------------------------
### INDICATIONS FOR USE
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. The Neodent GM Helix LG implants can be placed bicortically in cases of reduced bone density. The Neodent GM Helix LG implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants.
# SUBJECT DEVICE DESCRIPTIONS
The subject dental implant devices are single use devices, provided sterile by Gamma Radiation, made of commercially pure Titanium grade 4 (ASTM F67 – ISO 5832-2). The GM Helix LG Implant is a long conical implant with external diameters of 3.75 and 4.0 mm and lengths of 20, 22.5 and 25 mm. It has trapezoidal threads, conical apex with spherical tip and helical chambers, cylindrical body and Grand Morse prosthetic interface. The GM Helix LG implants share the same design characteristics as the existing GM Helix but are longer than the previously cleared range of implants. Indicated to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function.
| | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | GM Helix LG (K190958)<br>Neodent Implant System<br>JJGC Indústria e Comércio de Materiais Dentários S.A. | K160119<br>NobelSpeedy® Groovy Nobel<br>Biocare AB |
| Indications for<br>Use | The Neodent Implant System is intended to be surgically<br>placed in the bone of the upper or lower jaw to provide<br>support for prosthetic devices, such as artificial teeth, to<br>restore chewing function. It may be used with single-<br>stage or two-stage procedures, for single or multiple unit<br>restorations, and may be loaded immediately when good<br>primary stability is achieved and with appropriate occlusal<br>loading. The Neodent GM Helix LG implants can be placed<br>bicortically in cases of reduced bone density. The Neodent<br>GM Helix LG implants are only indicated for multiple unit<br>restorations in splinted applications that utilize at least<br>two implants. | NobelSpeedy® Groovy implants are endosseous implants<br>intended to be surgically placed in the upper or lower jaw bone<br>for anchoring or supporting tooth replacements to restore<br>patient esthetics and chewing function. NobelSpeedy® Groovy<br>implants are indicated for single or multiple unit restorations in<br>splinted or non- splinted applications. This can be achieved by a<br>2- stage or 1-stage surgical technique in combination with<br>immediate, early or delayed loading protocols, recognizing<br>sufficient primary stability and appropriate occlusal loading for<br>the selected technique. Implants allow also for bi-cortical<br>anchorage in cases of reduced bone density NobelSpeedy®<br>Groovy implants 20, 22, 25 mm when placed in the maxilla are<br>only indicated for multiple unit restorations in splinted<br>applications that utilize at least two implants. |
| Design | Threaded root-form implant with internal GM Morse taper<br>connection with internal hex. | Single lead thread with groove, tapered apex with bone<br>cutting flutes and external hex. |
| Device Material | Commercially Pure Titanium (ASTM F67) | Commercially Pure Titanium |
| Reusable | No | No |
| Platform<br>Diameter | 3.75; 4.0 mm | 4.0 mm |
| Length | 20; 22.5; 25 mm | 20; 22; 25 mm |
| Surface | Machined, acid-etched NeoPoros surface. | TiUnite |
# TECHNOLOGICAL CHARACTERISTIC COMPARISON TABLE
{5}------------------------------------------------
| | Sterilization Method | Gamma Irradiation to an SAL of 1x10-6 | Gamma Irradiation |
|-----------|-----------------------------------------------|-----------------------------------------------|-------------------|
| Placement | Single or Bicortically in mandible or maxilla | Single or Bicortically in mandible or maxilla | |
## SUBSTANTIAL EQUIVALENCE DISCUSSION
The subject devices and the primary predicate device have equivalent intended use and Indications for Use statements. Both can be placed single or bicortically in mandible or maxilla. The subject devices and the primary predicate devices also share the same material of composition, the same sterilization method and range of length and present similar designs and platform diameter. The difference between the subject devices and the predicate devices is the Ø3.75 mm device. This change in technology is supported by the torsion test and pull out test, both included as appendices of this submission. The subject devices have the same thread design and the same implant-to-abutment interface design as the reference predicate devices.
## PERFORMANCE DATA
Dynamic fatigue test per ISO 14801 was performed to determine the fatigue strength for implant construct assembled with prosthetic abutment for multi-unit prosthesis, assembled with GM Helix LG Implants, according to FDA Guidance.
Torsion Test was performed to evaluate the GM Helix LG Implant under static torsional loading. Insertion test was performed to evaluate the insertion torque of the GM Helix LG Implant when inserted into sawbones material representing bone type II, III and IV.
Pull-out testing was performed to compare the Axial Pull-out Strength between the subject devices and the predicate devices.
Sterilization of the subject implants via gamma irradiation using a protocol of 25 kGy minimum dose has been performed according to the requirements established by ISO 11137-2. A minimum Sterility Assurance Level (SAL) of 1 x 10 ° has beenvalidated.
Biological Safety Assessment guided by ISO 10993-1, Cytotoxicity guided by ISO 10993-5 and Chemical characterization guided by ISO 10993-18 were performed for GM Helix LG implants. Biocompatibility sample preparation was performed per ISO 10993-12.
# CONCLUSION
The data included in this submission demonstrate that the subject devices are substantially equivalent to the predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.