MBA Biotech Implant System
K243002 · MBA Biotech Co., Ltd. · DZE · Jun 6, 2025 · Dental
Device Facts
| Record ID | K243002 |
| Device Name | MBA Biotech Implant System |
| Applicant | MBA Biotech Co., Ltd. |
| Product Code | DZE · Dental |
| Decision Date | Jun 6, 2025 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.3640 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
MBA Biotech Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. Wide Fixture System is intended to be used in the molar region.
Device Story
MBA Biotech Implant System is a two-stage endosseous screw-type dental implant system used to replace missing teeth. The system comprises a fixture (implant body), cover screw, and various abutments (cemented, angled, and abutment screws). The fixture is surgically placed into the mandible or maxilla by a dental professional. The device functions as an anchor for dental prosthetics. The fixture features an internal hexagonal connection and an SLA (sand-blasted, large-grit, acid-etched) surface to promote osseointegration. Abutments are connected to the fixture to support final or temporary restorations. The system is intended for delayed loading, meaning the prosthetic is attached after a healing period. The device benefits patients by restoring oral function and aesthetics in edentulous sites.
Clinical Evidence
Bench testing only. Evidence includes fatigue testing per ISO 14801:2016, sterilization validation per ANSI/AAMI ST79 and ISO 17665, and MRI safety assessment based on scientific rationale and literature. Biocompatibility, shelf-life, and surface characterization data were leveraged from the predicate device (K211090).
Technological Characteristics
Endosseous root-form dental implant. Fixtures: Pure Titanium (ASTM F67), SLA surface treatment. Abutments/Screws: Ti-6Al-4V ELI (ASTM F136). Connection: Internal hex. Sterilization: Gamma irradiation (fixtures), end-user moist heat (abutments). Dimensions: Fixture diameters 3.75-6.75mm, lengths 7-15mm.
Indications for Use
Indicated for partially or fully edentulous adult patients requiring single or multiple unit dental restorations (cemented, screw-retained, or overdenture) or fixed bridgework support. Wide fixtures are indicated for the molar region. Intended for delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
- IZEN Implant system (K211090)
Related Devices
- K210039 — CORE1 Implant System · Cybermed, Inc. · Jan 7, 2022
- K202479 — IBS Implant System · Innobiosurg Co., Ltd. · Sep 23, 2020
- K181477 — MS Lide Implant Systems · Hangzhou Minsheng Lide Medical Science & Technology Co., Ltd. · Oct 20, 2020
- K233896 — JDentalCare Dental Implant System: JDEvolution; JDEvolution plus; JDIcon; JDIcon plus · Jdentalcare Srl · Aug 16, 2024
- K222707 — IH Implant System · Sewonmedix, Inc. · Jul 7, 2023
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 6, 2025
MBA Biotech Co., Ltd.
% April Lee
Regulatory Consultant
Withus Group Inc
106 Superior
Irvine, California 92620
Re: K243002
Trade/Device Name: MBA Biotech Implant System
Regulation Number: 21 CFR 872.3640
Regulation Name: Endosseous Dental Implant
Regulatory Class: Class II
Product Code: DZE, NHA
Dated: September 27, 2024
Received: May 5, 2025
Dear April Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243002 - April Lee
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K243002 - April Lee
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Andrew I. Steen -S
Andrew I. Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243002
Device Name
MBA Biotech Implant System
Indications for Use (Describe)
MBA Biotech Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. Wide Fixture System is intended to be used in the molar region.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
K243002
Page 1 of 7
# 510(K) Summary
## Submitter
MBA Biotech Co., Ltd.
Kim Sung Youn
913F, 124, Sagimakgol-ro, Jungwon-gu, Seongnam-si,
Gyeonggi-do, Republic of Korea
Email: mbabiotech@naver.com
Tel. +82-31-776-3621
## Official Correspondent
Withus Group Inc.
April Lee
106 Superior, Irvine, CA 92620
USA
Email: withus6664@gmail.com
Phone: 1-909-274-9971
## Device Information
- Trade Name: MBA Biotech Implant System
- Common Name: Dental Implant System
- Classification Name: Implant, endosseous, root-form
- Primary Product Code: DZE
- Secondary Product Code: NHA
- Panel: Dental
- Regulation Number: 872.3640
- Device Class: Class II
- Date Prepared: 06/06/2025
## Predicate Devices:
- K211090, IZEN Implant system
## Indications for Use:
MBA Biotech Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. Wide Fixture System is intended to be used in the molar region.
## Device Description:
The MBA Biotech Implant System is used to replace missing teeth in various situations ranging from a single tooth loss to the complete loss of teeth. It is two stage endosseous screw type implant with internal hexagonal connection.
This system consists of the fixture, cover screw, and various abutments. Only the subject abutments can be used with the subject fixtures.
The Fixture is made of Pure Titanium of ASTM F67 and the surface of the fixture is treated with the SLA(Sand-blasted, Large grit, Acid-etched surface). Fixture is provided sterile.
{5}
K243002
Page 2 of 7
The dimensions of fixture are as following:
| No. | Device Name | Dimension Ranges |
| --- | --- | --- |
| 1 | A3 Implant Bone Level | Ø3.75, 4.20, 4.25, 4.6, 5.05, 5.9, 6.75 (D) X 7, 8.5, 10, 11.5, 13, and 15mm |
| 2 | A3 Implant Active Bone Level | Ø3.75, 4.20, 4.25, 4.6, 5.05 (D) X 7, 8.5, 10, 11.5, 13, and 15mm |
Tolerance of dimension shall be within $\pm 1\%$ range.
The dimensions of abutments are as following:
| No. | Device Name | Dimension Ranges | Angulation |
| --- | --- | --- | --- |
| 1 | Cover Screw | Ø3.0(D) X 5.0 mm(L)
Ø3.4(D) X 6.0, 6.6 mm(L)
Ø3.6(D) X 5.7 mm(L)
Ø3.9(D) X 6.7, 7.3 mm(L) | 0° |
| 2 | Cemented Abutment(Hex) | Ø4.5(D) X 9.08, 9.20, 10.08, 10.20, 10.58, 10.70, 11.08, 11.20, 11.58, 11.70, 12.08, 12.20, 12.58, 12.70, 13.08, 13.20, 13.58, 13.70, 14.08, 14.20, 14.58, 14.70, 15.08, 15.20, 15.58, 15.70, 16.58, 16.70 mm(L) | 0° |
| | | Ø5.0(D) X 9.20, 10.20, 10.70, 11.20, 11.70, 12.20, 12.70, 13.20, 13.70, 14.20, 14.70, 15.20, 15.70, 16.70 mm(L) | |
| | | Ø6.0(D) X 9.20, 10.20, 10.70, 11.20, 11.70, 12.20, 12.70, 13.20, 13.70, 14.20, 14.70, 15.20, 15.70, 16.70 mm(L) | |
| 3 | Cemented Abutment (Non-Hex) | Ø4.5(D) X 8.48, 9.20, 9.48, 9.98, 10.20, 10.48, 10.70, 10.98, 11.20, 11.48, 11.70, 11.98, 12.20, 12.48, 12.98, 12.70, 13.20, 13.48, 13.70, 13.98, 14.20, 14.48, 14.98, 15.70, 15.20, 15.98, 16.70 | 0° |
| | | Ø5.0(D) X 9.20, 10.20, 10.70, 11.20, 11.70, 12.20, 12.70, 13.20, 13.70, 14.20, 14.70, 15.20, 15.70, 16.70 mm(L) | |
| | | Ø6.0(D) X 9.20, 10.20, 10.70, 11.20, 11.70, 12.20, 12.70, 13.20, 13.70, 14.20, 14.70, 15.20, 15.70, 16.70 mm(L) | |
| 4 | Angled Abutment(Hex) | Ø4.5(D) X 11.7, 13.7 mm(L)
Ø5.0(D) X 11.7, 13.7 mm(L) | 15° |
| | | Ø4.5(D) X 11.58, 13.58 mm(L) | 17° |
| | | Ø4.5(D) X 11.7, 13.7 mm(L)
Ø5.0(D) X 11.7, 13.7 mm(L) | 25° |
| 5 | Angled Abutment(Non-Hex) | Ø4.5(D) X 11.7, 13.7 mm(L)
Ø5.0(D) X 11.7, 13.7 mm(L) | 15° |
| | | Ø4.5(D) X 10.98, 12.98 mm(L) | 17° |
| | | Ø4.5(D) X 11.7, 13.7 mm(L)
Ø5.0(D) X 11.7, 13.7 mm(L) | 25° |
| 6 | Abutment Screw | Ø2.3(D) X 8.35 mm(L) | 0° |
Tolerance of dimension shall be within $\pm 1\%$ range.
{6}
K243002
Page 3 of 7
The Abutments have below featured:
| Name | Uses | Surface | Connection |
| --- | --- | --- | --- |
| Cover Screw | It is used for protecting inner hole and connecting part with exposed upper part of structure during the hea ling period after inserting dental implant fixture. | N/A | Screw Retained |
| Cemented Abutment | It is an abutment for making cement and combination maintenance-type prosthetics by taking a fixture level impression. It is an abutment for making cement and combination-retaining prosthesis by taking a fixture level impression. | N/A | Screw Retained, Internal Hex, Non-Hex |
| Angled Abutment | The Abutment is connected with fixture and it supports prosthesis which restores tooth function. | N/A | Screw Retained, Internal Hex, Non-Hex |
| Abutment Screw | Connection body to connect abutment to fixture | N/A | Screw Retained |
Tolerance of dimension for Abutments shall be within ± 1% range.
All abutments are provided non-sterilized.
## Materials:
- Fixtures are fabricated from Pure titanium of ASTM F67
- Cemented Abutment, Angled Abutment and Abutment Screw are fabricated from Ti-6Al-4V of ASTM F136
## Summaries of Technological Characteristics & Substantial Equivalence Discussion
Fixture
| | Subject Device | Primary Predicate |
| --- | --- | --- |
| K number | K243002 | K211090 |
| Manufacturer | MBA Biotech Co., Ltd. | Izenimplant Co., Ltd. |
| Trade Name | MBA Biotech Implant System | ZENEX Implant System |
| Model | A3 Implant Bone Level
A3 Implant Active Bone Level | ZENEX MULTI Fixture
ZENEX PLUS Fixture |
| Design | | |
| Indications for Use | The MBA Biotech Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. Wide Fixture System is intended to be used in the molar region. | The ZENEX Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including: cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. Wide Fixture System is intended to be used in the molar region. |
| Length(mm) | Mini Ø3.75 x L8.5, 10, 11.5, 13, 15
Ø4.20 x L8.5, 10, 11.5, 13, 15
Regular Ø4.25 x L7, 8.5, 10, 11.5, 13, 15
Ø4.6 x L7, 8.5, 10, 11.5, 13, 15 | Mini Ø3.75 x L8.5, 10, 11.5, 13, 15
Regular Ø4.25 x L7, 8.5, 10, 11.5, 13, 15
Ø4.6 x L7, 8.5, 10, 11.5, 13, 15
Wide Ø5.05 x 7, 8.5, 10, 11.5, 13, 15 |
{7}
K243002
Page 4 of 7
| | Wide Ø5.05 x 7, 8.5, 10, 11.5, 13, 15
Ø5.9 x 7, 8.5, 10, 11.5, 13
Ø6.75 x 7, 8.5, 10, 11.5, 13 | Ø5.4 x 7, 8.5, 10, 11.5, 13
Ø5.9 x 7, 8.5, 10, 11.5, 13
Ø6.75 x 7, 8.5, 10, 11.5, 13 |
| --- | --- | --- |
| Surface Treatment | Sandblasted and acid-etched | Sandblasted and acid-etched |
| Material | Pure Titanium Grade 4
(ASTM F67) | Pure Titanium Grade 4
(ASTM F67) |
| Shelf life | 5 years | 5 years |
| Sterilization | Gamma Irradiation | Gamma Irradiation |
| Comparison | The subject devices and basic conditions have the same characteristics, including indications for Use, general design, materials, abutment connection, surface treatment, and sterilization method. Since both devices are manufactured by the same manufacturer, it is substantially equivalent. | |
## Cover Screw
| | Subject Device | Primary Predicate |
| --- | --- | --- |
| K number | K243002 | K211090 |
| Manufacturer | MBA Biotech Co., Ltd. | Izenimplant Co., Ltd. |
| Trade Name | MBA Implant System | ZENEX Implant System |
| Model | Cover Screw | Cover Screw |
| Design | | |
| Indication for Use | The MBA Biotech Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. Wide Fixture System is intended to be used in the molar region. | The ZENEX Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including: cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. Wide Fixture System is intended to be used in the molar region. |
| Diameter(Ø) | 3.0~3.9 | 3.0~3.9 |
| Length(mm) | 5.0~7.3 | 5.0~7.3 |
| Surface Treatment | No Treatment | No Treatment |
| Material | Ti 6Al 4V ELI(ASTM F136) | Ti 6Al 4V ELI(ASTM F136) |
| Sterilization | Gamma Sterilization | Gamma Sterilization |
| Comparison | The subject device and predicate device have the same characteristics: purpose, general design, materials, connection structures, and diameters, etc. The difference lies in the presence or absence of coating and length differences. These differences do not impact product's fundamental technologies; therefore, subject device and predicate device are substantially equivalent. | |
{8}
K243002
Page 5 of 7
## Cemented Abutment
| | Subject Device | Primary Predicate |
| --- | --- | --- |
| K number | K243002 | K211090 |
| Manufacturer | MBA Biotech Co., Ltd. | Izenimplant Co., Ltd. |
| Trade Name | MBA Implant System | ZENEX Implant System |
| Model | Cemented Abutment | Cemented Abutment |
| Design | | |
| Hex | Hex, Non-Hex | Hex, Non-Hex |
| Diameter(Ø) | 4.5~6.5 | 4.5~6.5 |
| G/H(mm) | 1.0~6.8 | 1.0~7.0 |
| P/H(mm) | 5.5~7.0 | 4.0~7.0 |
| Angle | 0° | 0° |
| Surface Treatment | Non coated | Partial TiN coated in upper |
| Material | Ti 6Al 4V ELI(ASTM F136) | Ti 6Al 4V ELI(ASTM F136) |
| Sterilization | End User Sterilization | End User Sterilization |
| Comparison | The subject device and predicate device have the same characteristics: purpose, general design, materials, connection structures, and diameters, etc. The difference lies in the presence or absence of coating and length differences. These differences do not impact product's fundamental technologies; therefore, subject device and predicate device are substantially equivalent. | |
## Angled Abutment
| | Subject Device | Primary Predicate |
| --- | --- | --- |
| K number | K243002 | K211090 |
| Manufacturer | MBA Biotech Co., Ltd. | Izenimplant Co., Ltd. |
| Trade Name | MBA Biotech Implant System | ZENEX Implant System |
| Model | Angled Abutment | Angled Abutment |
| Design | | |
| Hex | Hex, Non-Hex | Hex, Non-Hex |
| Diameter(Ø) | 4.5~5.0 | 4.5~5.7 |
| P/H(mm) | 7.0 | 7.0 |
| Angle | 15°~25° | 15°~25° |
| Surface Treatment | Non coated | Partial TiN coated in upper |
| Material | Ti 6Al 4V ELI(ASTM F136) | Ti 6Al 4V ELI(ASTM F136) |
| Sterilization | End User Sterilization | End User Sterilization |
| Comparison | The subject device and predicate device have the same characteristics: purpose, general design, materials, connection structures, and diameters, etc. The difference lies in the presence or absence of coating and length differences. These differences do not impact product's fundamental technologies; therefore, subject device and predicate device are substantially equivalent. | |
{9}
K243002
Page 6 of 7
## Abutment Screw
| | Subject Device | Primary Predicate |
| --- | --- | --- |
| K number | K243002 | K211090 |
| Manufacturer | MBA Biotech Co., Ltd. | Izenimplant Co., Ltd. |
| Trade Name | MBA Biotech Implant System | ZENEX Implant System |
| Model | Abutment Screw | Abutment Screw |
| Design | | |
| Diameter(Ø) | 2.3 | 2.3 |
| Length(mm) | 8.35 | 8.35 |
| Angle | 0° | 0° |
| Material | Ti 6Al 4V ELI(ASTM F136) | Ti 6Al 4V ELI(ASTM F136) |
| Surface Treatment | Non-coating | Non-coating |
| Sterilization | End User Sterilization | End User Sterilization |
| Comparison | The Subject device and primary predicate have same characteristics such as indications for Use, design, material, dimensions, and sterilization method.
The difference between the two devices is design. These differences do not impact product's fundamental technologies; therefore, subject device and predicate device are substantially equivalent. | |
## Non-Clinical Test Data
Below tests were performed on subject device:
- Fatigue Testing under the worst-case scenario according to ISO 14801:2016
- End User Sterilization Validation Test Report on Abutments according to ANSI/AAMI ST79, ISO 17665-1, ISO 17665-2, ISO 11737-1, ISO 11737-2, and ISO 11138-1
Below tests were performed for predicate devices and leveraged for the subject device:
- Sterilization Validation Test on Fixtures according to ISO 11137-1,2,3 referenced in K211090
- Shelf-Life Test on Fixtures according to ASTM F1980 referenced in K211090
- Biocompatibility testing on fixtures according to ISO 10993-1:2009, ISO 10993-3:2014, ISO 10993-5:2009, ISO 10993-6:2007, ISO 10993-10:2010 and ISO 10993-11:2006 referenced in K211090
- Biocompatibility testing on Abutments made with Ti-6Al-4V ELI according to ISO 10993-1:2009, ISO 10993-3:2014, ISO 10993-5:2009, ISO 10993-6:2007, ISO 10993-10:2010 and ISO 10993-11:2006 referenced in K211090
- Bacterial Endotoxin Test Report on Fixtures according to ANSI/AAMI ST72:2011, USP <161>, and USP <85> referenced in K211090
- End User Sterilization Testing on Abutments made with Ti-6Al-4V ELI with ISO 17665-1,2 referenced in K211090
The results of the above tests have met the criteria of the standards and demonstrated the substantial equivalence with the predicate device.
The surface modification information with SLA (Sandblasted with Large-grit and Acid-etching) for fixtures was provided. To compare surface modification between the subject and predicate devices,
{10}
K243002
Page 7 of 7
K211090, surface roughness, surface composition analysis, and SEM imaging were provided, and it demonstrated the substantial equivalence. The surface treatment is identical to K211090 and is being leveraged from K211090.
The Sterilization validation test and shelf-life test for fixtures were performed for predicate device, K211090 and leveraged for the subject device because the material, sterilization method, packaging methods, and manufacturing process of both products are exactly same.
The Biocompatibility Test was conducted on the predicate device, K211090 and leveraged for the subject device because both products are manufactured by the same manufacturer with the same materials and manufacturing process. It demonstrates that the subject device is biocompatible and substantial equivalence with the predicate.
End User Moist Heat Sterilization Validation Test Report on Abutments according to ANSI/AAMI ST79, ISO 17665-1, -2, ISO 11737-1, -2 and ISO 11138-1 was leveraged from K211090.
## MR Environment Condition
Non-clinical worst-case MRI review was performed to evaluate the metallic MBA Biotech Implant System in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan, "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices", Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system including all variations (all compatible implant bodies, dental abutments, and fixation screws) and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.
The non-clinical testing results demonstrate that the subject device is substantially equivalent to the predicate device.
## Conclusion
MBA Biotech Implant System constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental scientific technology as its predicate devices. Therefore, MBA Biotech Implant System and its predicates are substantially equivalent.