Magicore System

K212517 · Innobiosurg Co., Ltd. · DZE · Oct 15, 2021 · Dental

Device Facts

Record IDK212517
Device NameMagicore System
ApplicantInnobiosurg Co., Ltd.
Product CodeDZE · Dental
Decision DateOct 15, 2021
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Magicore System is intended to replace missing teeth to restore chewing function. The Magicore System can be placed in support of single or multiple-unit restorations including; cement retained, screw retained, or overdenture restorations, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading.

Device Story

Endosseous dental implant system; includes fixtures, abutments, cylinders, caps, and screws. Inputs: surgical site preparation; device provides structural support for dental restorations. Operation: fixtures placed into bone; abutments connected via internal hex/Morse taper bevel. Used in dental clinics by licensed dentists. Output: stable foundation for prosthetic teeth. Benefits: restores chewing function for edentulous patients. Fixtures feature RBM (Resorbable Blasted Media) surface treatment. System supports one or two-stage surgical protocols.

Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing, including sterilization validation (ISO 11137, ISO 17665), biological assessment (ISO 10993-1), shelf-life testing (ASTM F1980), and fatigue testing (ISO 14801) leveraged from predicate device submissions.

Technological Characteristics

Materials: Ti-6Al-4V ELI (ASTM F136) and Co-Cr-Mo alloy/Poly Diacetate. Design: Endosseous root-form implant with internal hex/Morse taper connection. Surface: RBM (Resorbable Blasted Media). Dimensions: Various diameters and lengths. Sterilization: Gamma (fixtures) or end-user (abutments).

Indications for Use

Indicated for patients requiring replacement of missing teeth in partially or fully edentulous mandibles and maxillae to restore chewing function. Supports single or multiple-unit restorations (cement/screw-retained, overdentures) and fixed bridgework. For one or two-stage surgical procedures; delayed loading.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION". October 15, 2021 InnoBioSurg Co., Ltd. % April Lee Consultant Withus Group Inc 106 Superior Irvine, California 92620 Re: K212517 Trade/Device Name: Magicore System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: September 10, 2021 Received: September 17, 2021 Dear April Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K212517 Device Name Magicore System Indications for Use (Describe) The Magicore System is intended to replace missing teeth to restore chewing function. The Magicore System can be placed in support of single or multiple-unit restorations including; cement retained, or overdenture restorations, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading. Type of Use (Select one or both, as applicable) | <div> <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |-------------------------------------------------------------------------------------------------------------------------| | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) Summary #### Submitter InnoBioSurg Co., Ltd. JongHyuk Seo 44-19, Techno 10-ro, Yuseong-gu Daejeon, 34027 Republic of Korea Email: jhseo@ibsimplant.com Tel. +82-42-933-2879 Fax. +82-42-933-2881 ## Official Correspondent Withus Group Inc April Lee 106 Superior Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122 ### Device Information - Trade Name: Magicore System - Common Name: Endosseous Dental Implant - . Classification Name: Implant, Endosseous, Root-Form - . Product Code: DZE - . Secondary Product Code: NHA - . Panel: Dental - . Regulation Number: 872.3640 - Date prepared: 10/14/2021 . ### Predicate Devices: The subject device is substantially equivalent to the following Reference Devices: Primary Predicate K201981, Magicore System manufactured by InnoBioSurg Co., Ltd. Reference Device K153639. OneO-SL s-Clean Implant System manufactured by Dentis Co., Ltd. K192197, Magicore II System manufactured by InnoBioSurg Co., Ltd. K201621, Magicore II System manufactured by InnoBioSurg Co., Ltd. K202418, Magic UCLA Abutment System manufactured by InnoBioSurg Co., Ltd. ### Indications for Use The Magicore System is intended to replace missing teeth to restore chewing function. The Magicore System can be placed in support of single or multiple-unit restorations including; cement retained, screw retained, or overdenture restorations, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading. {4}------------------------------------------------ # Device Description This submission is to add new fixtures and abutments to the previously cleared device, Magicore System (K201981), Magicore II System (K201621), Magicore II System (K192197), and Magic UCLA Abutment System (K202418). The fixtures and abutments in this system are below: 1) Fixture - . Magicore - Magicore (Cutting Edge) ● ### 2) Abutment - . Magic Multiunit Abutment (Screw type - Hex, Non-Hex & Cemented type - Hex, Non-Hex) - Magic Multiunit UCLA Cylinder ● - Magic Multiunit Cap ● - Magic Abutment (Screw type Hex, Non-Hex & Cemented type Hex, Non-Hex) ● - Magic UCLA Cement Retained Type (Hex, Non-Hex) - Magic Cylinder (Hex, Non-Hex, Post) - Magic Multiunit Cylinder (Hex, Non-Hex, Post) - Magicore Healing Cap ● - Magicore Healing Cap Screw ● - Cylinder Screw . An endosseous dental implant is a device made of a material such as Ti 6AL 4V Eli (Conforming to ASTM Standard F-136). The Magicore System consists of dental implants, Abutments, cylinders, caps and screws for use in one or two-stage dental implant placement and restorations. The implant-Abutment connection is tight and precise fitting with internal hex and Morse taper bevel. The surface of the Magicore implant is treated with RBM (Resorbable Blasted media). Below is the fixture dimension range: | Fixture | Platform<br>Diameters (Ø) | Fixture<br>Diameters (Ø) | Cuff<br>Lengths (mm) | Implantable<br>Lengths (mm) | |-----------------------------------------------------------|---------------------------|-------------------------------------|----------------------|-----------------------------| | Magicore<br>(Cleared in<br>K201981) | 5.2 | 4.0, 4.5 | 1, 2, 3, 4 | 7, 8, 9, 10, 11, 12, 13 | | | 5.2 | 5.0, 5.5, 6.0, 6.5<br>(Newly Added) | | | | | 5.7 | 4.5, 5.0, 5.5, 6.0, 6.5 | | | | | 5.7 | 7.0, 7.5, 7.8<br>(Newly Added) | | | | Magicore,<br>Cutting Edge type<br>(Cleared in<br>K201981) | 5.2 | 4.0, 4.5 | 1, 2, 3, 4 | | | | 5.2 | 5.0, 5.5, 6.0, 6.5<br>(Newly Added) | | | | | 5.7 | 4.5, 5.0, 5.5, 6.0, 6.5 | | | | | 5.7 | 7.0, 7.5, 7.8<br>(Newly Added) | | | {5}------------------------------------------------ | Fixture<br>(Subject Device) | Platform<br>Diameter (Ø) | Fixture<br>Diameters (Ø) | Cuff<br>Lengths (mm) | Implantable<br>Lengths (mm) | |---------------------------------|--------------------------|--------------------------|----------------------|-----------------------------| | Magicore | 5.2 | 5.0, 5.5, 6.0, 6.5 | | 7, 8, 9, 10, 11, 12, 13 | | | 5.7 | 7.0 | | 7, 8, 9, 10 | | | | 7.5 | | 7,8,9 | | | | 7.8 | | 7,8 | | Magicore<br>(Cutting Edge type) | 5.2 | 5.0, 5.5, 6.0, 6.5 | 1, 2, 3, 4 | 7, 8, 9, 10, 11, 12, 13 | | | 5.7 | 7.0 | | 7, 8, 9, 10 | | | | 7.5 | | 7,8,9 | | | | 7.8 | | 7,8 | The subject fixtures are provided sterile. The subject fixtures are compatible with the abutment in this submission and the abutments of K192197, K201621, K201981 and K202418. Below is the abutment dimension range: | Abutments | Diameters (Ø) | Length (mm) | Angulation (°) | |------------------------------------------------------------------------------------------------------------------|-------------------------------------|-----------------------------------------------------------------------|--------------------| | Magic Multiunit Abutment<br>(Screw type - Hex, Non-Hex,<br>Cemented type – Hex, Non-Hex)<br>(Cleared in K201621) | 4.8, 5.8 | 1.5, 2.5, 3.5, 4.5 | 5, 10, 15, 20, 25 | | | | 2.3 (Length Change) | 0<br>(Newly Added) | | Magic Multiunit UCLA Cylinder<br>(Cleared in K202418) | 4.5, 5.0, 5.2, 6.0, 6.2, 6.3 | 8.5, 8.75, 8.9, 10.2, 10.4,<br>10.5 | - | | | 5.4, 6.4 (Newly Added) | 11.2 (Newly Added) | | | Magic Cylinder<br>(Cleared in K201621) | 5.0, 6.0 | 10 | - | | | | 10.6 (Length Change) | | | Magic Multiunit Cylinder<br>(Cleared in K201621) | 5.0, 6.0 | 10 | - | | | 5.4, 6.4 (Newly Added) | | | | Magic Multiunit Cap<br>(Cleared in K192197) | 5.2, 6.2 | 4.5 | - | | | 5.4, 6.4 (Newly Added) | 4.3 (Newly Added) | | | Magic Abutment<br>(Screw type - Hex, Non-Hex,<br>Cement type - Hex, Non-Hex)<br>(Cleared in K192197) | 4.27, 4.7, 5.21, 5.5, 6.0 | 4.5, 5.5, 6.5, 7.5, 8.5<br>4.6, 5.6, 6.6, 7.6, 8.6 | - | | | 5.2, 5.7, 6.2, 6.7<br>(Newly Added) | 4.51, 5.51, 6.5, 6.51, 7.5,<br>7.51, 8.5, 8.51, 10.5<br>(Newly Added) | | | Magic UCLA Cement<br>Retained Type<br>(Cleared in K202418) | 4.5, 5.2, 6.2 | 8.5, 8.75, 8.9, 10.2, 10.4,<br>10.5 | - | | | | 12.05, 12.85<br>(Length Change) | | | Magicore Healing Cap<br>(Cleared in K192197) | 5.2, 6.2 | 4.5 | - | | | 5.3, 5.8, 6.3, 6.8<br>(Newly Added) | 2.8, 3.8, 4.8, 5.8, 6.8<br>(Newly Added) | | | Magicore Healing Cap Screw | 2.0 | 5.2, 7.1 | | {6}------------------------------------------------ | (Cleared in K173120) | | 5.4, 6.4, 7.4, 8.4, 9.4<br>(Newly Added) | | |------------------------------------------------------------------------------------------|--------------------|------------------------------------------------------|----------------------| | Cylinder Screw<br>(Cleared in K192197) | 2.0 | 5.2, 7.1 | - | | | | 4.9 (Length Change), 7.1 | | | Abutments (Subject Device) | Diameters (Ø) | Length (mm) | Angulation (°) | | Magic Multiunit Abutment<br>(Screw type - Hex, Non-Hex,<br>Cemented type – Hex, Non-Hex) | 4.8, 5.8 | 2.3 | 0, 5, 10, 15, 20, 25 | | Magic Multiunit UCLA Cylinder | 5.4, 6.4 | 11.2 | - | | Magic Cylinder | 5.0, 6.0 | 10.6 | - | | Magic Multiunit Cylinder | 5.0, 6.0, 5.4, 6.4 | 10 | - | | Magic Multiunit Cap | 5.4, 6.4 | 4.3 | - | | Magic Abutment<br>(Screw type - Hex, Non-Hex,<br>Cement type - Hex, Non-Hex) | 5.2, 5.7, 6.2, 6.7 | 4.51, 5.51, 6.5, 6.51, 7.5,<br>7.51, 8.5, 8.51, 10.5 | - | | Magic UCLA Cement<br>Retained Type | 4.5, 5.2, 6.2 | 12.05, 12.85 | - | | Magicore Healing Cap | 5.3, 5.8, 6.3, 6.8 | 2.8, 3.8, 4.8, 5.8, 6.8 | - | | Magicore Healing Cap Screw | 2.0 | 5.4, 6.4, 7.4, 8.4, 9.4 | - | | Cylinder Screw | 2.0 | 4.9, 7.1 | - | The subject abutments are compatible with the fixtures in this submission and the fixtures of K192197 and K201981. The abutments are provided non-sterile and packaged separately. The abutments should be sterilized before use. {7}------------------------------------------------ | | Subject Device | Primary predicate | Reference Device | |-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer | InnoBioSurg Co., Ltd | InnoBioSurg Co., Ltd | Dentis | | Device Name | Magicore System | Magicore System | OneQ-SL s-Clean Implant System | | 510(k) No. | N/A | K201981 | K153639 | | Indications for use | The Magicore System is intended to replace missing teeth to restore chewing function. The Magicore System can be placed in support of single or multiple-unit restorations including; cement retained, screw retained, or overdenture restorations, and terminal or immediate abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading. | The Magicore System is intended to replace missing teeth to restore chewing function. The Magicore System can be placed in support of single or multiple-unit restorations including; cement retained, screw retained, or overdenture restorations, and terminal or immediate Abutment support for fixed bridgework. This system is for one or two stage surgical procedures. This system is intended for delayed loading. | The OneQ-SL s-Clean Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate Abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures and not dedicated for immediate loading. This system is intended for delayed loading. | | Design | Image: Cutting Edge, Non-Cutting Edge | Image: Cutting Edge, Non-Cutting Edge | Image: Cutting Edge | | Composition of Material | Titanium Alloy<br>Ti-6Al-4V Eli<br>ASTM F136 | Titanium Alloy<br>Ti-6Al-4V Eli<br>ASTM F136 | CP Titanium Grade 4<br>(ASTM F67) | | Connection | Internal Hex<br>Non - Submerged | Internal Hex<br>Non - Submerged | Internal Hex<br>Submerged | | Endosseous Implant | Tapered, macro threads | Tapered, macro threads | Tapered & Straight Body | | Platform Diameters (Ø) | 5.2 5.7 | 4.7 5.2 5.7 | Regular: 3.7, 3.9, 4.2, 4.7, 5.2<br>Wide: 6.0, 7.0, 8.0 | | Fixture Diameters (Ø) | 5.0, 5.5, 6.0, 6.5 7.0, 7.5, 7.8 | 4.0, 4.5, 5.0 4.0, 4.5 4.5, 5.0, 5.5, 6.0, 6.5 | | | Cuff Lengths<br>(mm) | 1, 2, 3, 4 | 1, 2, 3, 4 | - | | Neck Lengths<br>(mm) | 2 | 2 | - | | Implantable<br>Lengths (mm) | 7, 8, 9, 10,<br>11, 12, 13 | 7, 8, 9, 10, 11, 12, 13 | Regular: 7.8, 10, 12, 14<br>Wide: 7,8, 10, 12 | | Thread pitch | 1.15P | 1.15P | - | | Modified<br>Surface | R.B.M | R.B.M | S.L.A | | Surgical<br>Technique | 1 and 2 stage, self-tapping | 1 and 2 stage, self-tapping | 3 sided cutting edge with<br>self-tapping | | Gamma<br>Sterilization | Yes | Yes | Yes | | SE Discussion | The Magicore System has same device characteristics with the Primary predicate (K201981)<br>such as intended use, material, functions, general shape (Design), surface treatment, structure<br>and applied production method. Platform/D 5.2 (Fixture/D 5.0,5.5,6.0.5,6.5) and P/D 5.7<br>(F/D 7.0,7.5,7.8) was added in the Primary predicate device.<br>The subject fixture has bigger diameters such as 7.0, 7.5 and 7.8mm than the primary<br>predicate. To support this discrepancy, the reference device, K153639 was added and the<br>difference doesn't affect device's fundamental functions and safety since the diameter of the<br>subject device is within the range of the reference device's diameters. Therefore, the subject<br>device is substantial equivalent. | | | #### Summaries of Technology Characteristics: 1) Fixture {8}------------------------------------------------ {9}------------------------------------------------ # 2) Abutment # < Magic Multiunit Abutment (Screw type – Hex, Non-Hex & Cemented type – Hex, Non-Hex)> | | Subject Device | | Reference Device | | |---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|--------------------------|----------------| | Manufacturer | InnoBioSurg Co., Ltd. | | InnoBioSurg Co., Ltd. | | | Device Name | Magicore System | | Magicore II System | | | Abutment Name | Magic Multiunit Abutment | | Magic Multiunit Abutment | | | 510(k) No. | NA | | K201621 | | | Material | TI-6AL-4V ELI | | TI-6AL-4V ELI | | | Design | Image: Hex | Image: Non-Hex | Image: Hex | Image: Non-Hex | | | Hex | Non-Hex | Hex | Non-Hex | | Diameters (Ø) | 4.8, 5.8 | | 4.8, 5.8 | | | Gingiva Height (mm) | 2.3 | | 1.5, 2.5, 3.5, 4.5 | | | Angulation (°) | 0 | | 5, 10, 15, 20, 25 | | | Surface Treatment | Machine- | | Machine- | | | Sterilization | End User Sterilization | | End User Sterilization | | | SE Discussion | The subject device is same in indications for Use, fundamental scientific<br>technology, principle of operation, design, technology, functions, dimensions, and<br>materials with the identified reference device.<br>The Gingiva Height of previously cleared device (K201621) has changed. Also,<br>the abutments with 0 ° angulation are added to the previously cleared device. Since<br>the primary predicate's abutment is worst case (largest angulation), this difference<br>does not impact product's safety and effectiveness. | | | | #### <Magic Multiunit UCLA Cylinder> | | Subject Device | Reference Device | |-------------------|------------------------------------------------------|-----------------------------------------| | Manufacturer | InnoBioSurg Co., Ltd. | InnoBioSurg Co., Ltd. | | Device Name | Magicore System | Magic UCLA Abutment System | | Abutment Name | Magic Multiunit UCLA Cylinder | Magic UCLA Screw Retained Type | | 510(k) No. | NA | K202418 | | Material | Co-Cr-Mo Alloy<br>Poly Diacetate | Co-Cr-Mo Alloy<br>Poly Diacetate | | Design | Image: Two abutment designs, one white and one black | Image: Two abutment designs, both white | | Diameters (Ø) | 5.4, 6.4 | 4.5, 5.0, 5.2, 5.5, 6.0, 6.2, 6.3 | | Length (mm) | 11.2 | 8.5, 8.75, 8.9, 10.2, 10.4, 10.5 | | Surface Treatment | Machine- | Machine- | | Sterilization | End User Sterilization | End User Sterilization | {10}------------------------------------------------ | SE Discussion | The subject device is same in indications for use, material, fundamental scientific<br>technology, principle of operation, design, technology, functions with the identified<br>reference device.<br>The differences between the subject and reference device are the design and<br>dimensions. However, it does not affect device's fundamental functions and safety;<br>therefore, it is substantial equivalent. | |---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| <Magic Multiunit Cap> | | Subject Device | Reference Device | |-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------| | Manufacturer | InnoBioSurg Co., Ltd. | InnoBioSurg Co., Ltd. | | Device Name | Magicore System | Magicore II System | | Abutment Name | Magic Multiunit Cap | Magic Multiunit Abutment Cap | | 510(k) No. | NA | K192197 | | Material | TI-6AL-4V ELI | TI-6AL-4V ELI | | Design | Image: Subject Device Design | Image: Reference Device Design | | Diameters ( $\varnothing$ ) | 5.4, 6.4 | 5.2, 6.2 | | Length (mm) | 4.3 | 4.5 | | Surface Treatment | Machine- | Machine- | | Sterilization | End User Sterilization | End User Sterilization | | SE Discussion | The subject device is same in indications for use, fundamental scientific<br>technology, principle of operation, general design, technology, functions, and<br>materials with the identified reference device.<br>The differences between the subject and reference device are the design and<br>dimensions. However, it does not affect device's fundamental functions and<br>safety; therefore, it is substantial equivalent. | | < Magic Abutment (Screw type- Hex, Non-Hex & Cemented type- Hex, Non-Hex)> | | Subject Device | Reference Device | |---------------|-----------------------------------------------------------|-----------------------------------------------------| | Company | InnoBioSurg Co., Ltd. | InnoBioSurg Co., Ltd. | | Device Name | Magicore System | Magicore II System | | Abutment Name | Magic Abutment | Magic Abutment | | 510(k) | N/A | K192197 | | Material | TI-6AL-4V ELI | TI-6AL-4V ELI | | Design | Image: Hex<br>Image: Non-Hex | Image: Hex<br>Image: Non-Hex | | Diameters (Ø) | 5.2, 5.7, 6.2, 6.7 | 4.27, 4.7, 5.21, 5.5, 6.0 | | Lengths(mm) | 4.51, 5.51, 6.5, 6.51, 7.5, 7.51, 8.5,<br>8.51, 9.5, 10.5 | 4.5, 5.5, 6.5, 7.5, 8.5,<br>4.6, 5.6, 6.6, 7.6, 8.6 | {11}------------------------------------------------ | Surface<br>Treatment | Machine- | Machine- | |----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Sterilization | End User Sterilization | End User Sterilization | | SE Discussion | The subject device is same in indications for Use, fundamental scientific<br>technology, principle of operation, general design, technology, functions, and<br>materials with the identified reference device.<br>The difference between the subject and reference device is the design and lengths<br>of the device. However, it does not affect device's fundamental functions and<br>safety; therefore, it is substantial equivalent. | The subject device is same in indications for Use, fundamental scientific<br>technology, principle of operation, general design, technology, functions, and<br>materials with the identified reference device.<br>The difference between the subject and reference device is the design and lengths<br>of the device. However, it does not affect device's fundamental functions and<br>safety; therefore, it is substantial equivalent. | # <Magic UCLA Cement Retained Type (Hex, Non-Hex)> | | Subject Device | Primary Predicate | |-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer | InnoBioSurg Co., Ltd.…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...