ATLAS IMPLANT SYSTEM - WIDE PLUS

K090054 · Cowellmedi Co., Ltd. · DZE · Jul 29, 2009 · Dental

Device Facts

Record IDK090054
Device NameATLAS IMPLANT SYSTEM - WIDE PLUS
ApplicantCowellmedi Co., Ltd.
Product CodeDZE · Dental
Decision DateJul 29, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Atlas Implant System - Wide Plus is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. Also, there are no angled abutments intended to be used with this system. Smaller (e 4.5, e 5.5) implants are dedicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Larger (e 6.0 to e 8.0) implants are dedicated for the molar region and not indicated for immediate loading.

Device Story

Atlas Implant System - Wide Plus is an endosseous dental implant system; titanium alloy construction; surgically placed into maxilla or mandible bone. System supports single/multiple-unit restorations, cemented/overdenture restorations, and fixed bridgework. Used by dental clinicians in clinical settings. Smaller diameter implants (4.5-5.5mm) allow immediate loading given primary stability; larger diameter implants (6.0-8.0mm) intended for molar region without immediate loading. No angled abutments are used. Device provides structural foundation for prosthetic teeth, restoring oral function and aesthetics.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Manufactured from TI6A1-4V ELI alloy per ASTM and ISO standards. Root-form endosseous dental implant. Dimensions range from 4.5mm to 8.0mm diameter. Non-software device.

Indications for Use

Indicated for partially or fully edentulous adult patients requiring single or multiple-unit dental restorations, including cemented, overdenture, or fixed bridgework support. Smaller implants (4.5-5.5mm) suitable for immediate loading with primary stability; larger implants (6.0-8.0mm) for molar region, not for immediate loading.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K09 0054 # JUL 2 9 2009 11: == (1)2 ## 510(K) Summary Submitter ﺎﺳ ਪ੍ਰ Cowellmedi Co., Ltd. Kim Young Tae 155-4, Gamjeon - 2 Dong, Sasang-Gu, 617-801, Busan, South Korea Phone: 82-51-314-2028 Fax: 82-51-314-2026 Official Correspondent Kodent Inc. Jung Bae Bang 13340 E. Firestone Blvd. Suite J Santa Fe Springs, CA 90670 Email: kodentinc@kodent.co.kr Phone: 562-404-8466 Fax: 562-404-2757 #### Device Information Product Name: Atlas Implant System - Wide Plus Common Name: Endosseous Dental Implant Classification Name: Implant, Endosseous, Root-Form Product Code: DZE Regulation Number: 872.3640 Device Class: Class II #### Device Description The Atlas Implant System - Wide Plus is a dental implant made of titanium metal intended to be surgically placed in the bone of the upper and / or lower jaw arches. This system is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics. #### Indication for Use The Atlas Implant System - Wide Plus is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. Also, there are no angled abutments intended to be used with this system. Smaller (e 4.5, e 5.5) implants are dedicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Larger (e 6.0 to e 8.0) implants are dedicated for the molar region and not indicated for immediate loading. {1}------------------------------------------------ #### Materials This device are manufactured from TI6A1-4V ELI alloy following ASTM and ISO standards. ## Predicate Devices The subject device is substantially equivalent to the following predicate devices: - ◆ Atlas Implant System (K071148) manufactured by Cowellmedi Co., Ltd. - Rescue External Implant System (K081302) manufactured by Megagen Implant Co., Ltd. . - US Ultra Wide System (K073465) manufactured by Osstem Co., Ltd. . ## Comparison to Predicate Devices Testing and other comparisons have established that the subject of Atlas Implant System - Wide Plus is substantially equivalent in design, materials, indications and intended use, packaging, labeling, and performance to other predicate devices of the type currently marketed in the U.S. {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling the perimeter, with the acronym "USA" included. In the center is a stylized design, possibly representing a symbol or emblem associated with health and human services. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Cowellmedi Company, Limited C/O Mr. Jung Bae Bang Regulatory Affairs Kodent Incorporated 13340 East Firestone Boulevard, Suite J Santa Fe Springs, California 90670 JUL 29 2009 Re: K090054 Trade/Device Name: Atlas Implant System - Wide Plus Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: June 9, 2009 Received: June 9, 2009 Dear Mr. Bang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2- Mr. Bang Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Susan Turner Susan Runner, B.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ ## Indication for Use ## 510(K) Number (if known): K090054 Device Name: Atlas Implant System - Wide Plus ## Indication for Use: The Atlas Implant System - Wide Plus is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. Also, there are no angled abutments intended to be used with this system. Smaller (e 4.5, e 5.5) implants are dedicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Larger (o 6.0 to o 8.0) implants are dedicated for the molar region and not indicated for immediate loading. Prescription Use AND/OR Over-The-Counter (Part 21 CFR 801 Subpart D) (Per 21 CFR 801 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 Kein Muelly for MSR (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...