VISTA BRITE TIP CATHETERS
Device Facts
| Record ID | K962830 |
|---|---|
| Device Name | VISTA BRITE TIP CATHETERS |
| Applicant | Cordis Corp. |
| Product Code | DQY · Cardiovascular |
| Decision Date | Oct 17, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.1250 |
| Device Class | Class 2 |
Intended Use
The guiding catheter is intended for use for intravascular introduction of interventional/diagnostic devices into the coronary and peripheral vascular systems.
Device Story
Single-lumen guiding catheter; nylon body reinforced with stainless steel braid wire; transition segments utilize varying durometer nylons for stiffness gradient; distal 'Brite Tip' segment composed of pellethane with radiopaque filler for atraumatic navigation. Used by physicians in clinical settings for coronary and peripheral vascular access; facilitates delivery of diagnostic/interventional devices. Device provides trackability and support during procedures.
Clinical Evidence
Bench testing only; biocompatibility testing successfully completed.
Technological Characteristics
Single-lumen catheter; nylon body; stainless steel braid wire reinforcement; pellethane tip with radiopaque filler; variable durometer segments for stiffness transition. Non-powered, manual device.
Indications for Use
Indicated for patients requiring intravascular introduction of interventional or diagnostic devices into coronary and peripheral vascular systems.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
- Cordis Vista Brite Tip (K962830)
Related Devices
- K971572 — VISTA BRITE TIP CATHETERS · Cordis Corp. · Jul 14, 1997
- K962362 — CORDIS ENVOY & VISTA BRITE TIP GUIDING CATHETERS · Cordis Corp. · Aug 8, 1996
- K963962 — VISTA BRITE TIP CATHETERS · Cordis Corp. · May 7, 1997
- K965211 — VISTA BRITE TIP CATHETERS · Cordis Corp. · Mar 11, 1997
- K992673 — VISTA BRITE TIP CATHETERS · Cordis Corp. · Oct 12, 1999