Excelsior XT-17 Microcatheter, Excelsior XT-17 Flex Microcatheter, Excelsior XT-17 Pre-Shaped Microcatheter, Excelsior XT-17 Flex Pre-Shaped Microcatheter

K142565 · Stryker Neurovascular · DQY · Nov 14, 2014 · Cardiovascular

Device Facts

Record IDK142565
Device NameExcelsior XT-17 Microcatheter, Excelsior XT-17 Flex Microcatheter, Excelsior XT-17 Pre-Shaped Microcatheter, Excelsior XT-17 Flex Pre-Shaped Microcatheter
ApplicantStryker Neurovascular
Product CodeDQY · Cardiovascular
Decision DateNov 14, 2014
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Intended Use

Stryker Neurovascular's Excelsior XT-17 Microcatheters are intended to assist in the delivery of diagnostic agents, such as contrast media, and therapeutic agents, such as occlusion coils, into the peripheral, coronary and neuro vasculature.

Device Story

Single-lumen microcatheter; used to access distal vasculature; facilitates delivery of diagnostic agents (contrast media) and therapeutic agents (occlusion coils). Operated by physicians trained in endovascular procedures in hospital/clinical settings. Device features graded shaft stiffness (flexible tip to semi-rigid proximal section) for tracking over guidewires; radiopaque tip for fluoroscopic visualization; Hydrolene coating to reduce friction. Used with guide catheter and steerable guidewire. Physician navigates device through vasculature; monitors position via fluoroscopy; delivers agents through luer fitting. Benefits include precise navigation to distal sites and controlled delivery of therapeutic/diagnostic materials.

Clinical Evidence

Bench testing only. Verification testing included tip shape retention, flexibility, surface defect analysis, static/dynamic burst pressure, tensile strength, hub integrity, DMSO compatibility, and introducer sheath compatibility. Biocompatibility testing performed per EN ISO 10993-1:2009 + AC:2010, including cytotoxicity, hemolysis, physiochemical tests, FTIR, and latex testing.

Technological Characteristics

Single-lumen catheter; polymers, stainless steel, Hydrolene coating. Graded shaft stiffness. Radiopaque tip. Sterilized via Ethylene Oxide (EO). Patient contact < 24 hours. Includes introducer sheath and steam shaping mandrel. No software or electronic components.

Indications for Use

Indicated for patients requiring delivery of diagnostic or therapeutic agents into peripheral, coronary, or neuro vasculature.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract image of three faces in profile, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 14, 2014 Stryker Neurovascular Ms. Yoko Enrile Principle Regulatory Affairs Specialist 47900 Bayside Parkway Fremont, CA 94538 US Re: K142565 Trade/Device Name: Excelsior XT-17 Microcatheter, Excelsior XT-17 Flex Microcatheter, Excelsior XT-17 Pre-Shaped Microcatheter, Excelsior XT-17 Flex Pre-Shaped Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheters Regulatory Class: Class II Product Code: DQY Dated: September 11, 2014 Received: September 12, 2014 Dear Ms. Enrile: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Carlos L. Pena -S Carlos Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K142565 Device Name Excelsior XT -17 Microcatheter Indications for Use (Describe) Stryker Neurovascular's Excelsior XT -17 Microcatheters are intended to assist in the delivery of diagnostic agents, such as contrast media, and therapeutic agents, such as occlusion coils, into the peripheral, coronary and neuro vasculature. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the Stryker Neurovascular logo. The word "Stryker" is in a bold, sans-serif font on the top line. Below that, the word "Neurovascular" is in a smaller, sans-serif font. The logo is black and white. # 510(k) Summary | Summary Date | November 12, 2014 | | |---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------| | Submitter Name and<br>Address | Stryker Neurovascular<br>47900 Bayside Parkway<br>Fremont, CA. 94538 | | | Contact Person: | Yoko Y Enrile<br>Principal Regulatory Affairs Specialist<br>Phone: 510-413-2619<br>Fax: 510-413-2588<br>Email: yoko.enrile@stryker.com | | | Trade Name: | Excelsior® XT-17TM Microcatheter | | | Common Name: | Percutaneous Catheter; Microcatheter | | | Classification Name: | Percutaneous catheters are currently classified as Class II devices<br>per 21 CFR 870.1250. No performance standards have been<br>established under Section 514 of the Food, Drug and Cosmetic Act<br>for these devices. | | | Product Code: | DQY | | | Legally Marketed<br>Predicate Device: | Reference (Clearance Date) | Device | | | K123452 (3 January 2013) | Tracker-17 Microcatheter | {4}------------------------------------------------ ### Device Description and Characteristics: | Stryker Neurovascular Excelsior® XT-17™ Microcatheters are | | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------| | single lumen devices designed to aid the physician in accessing distal<br>vasculature when used with a guide catheter and steerable guidewire. | | | Graded shaft stiffness ranging from a highly flexible tip to a semi-<br>rigid proximal section aids the physician in tracking over selectively<br>placed guidewires. A luer fitting located on the microcatheter hub is<br>used for the attachment of accessories. A radiopaque tip facilitates<br>fluoroscopic visualization. Stryker Neurovascular hydrophilically<br>coated Excelsior XT-17 Microcatheters are coated on the outer<br>surface with Hydrolene® Coating that reduces friction during<br>manipulation in the vessel. | | | The Excelsior XT-17 Microcatheter is a single use device made of<br>polymers, stainless steel, and Hydrolene® Coating, sterilized with<br>Ethylene Oxide (EO) gas. Patient contact duration is less than 24<br>hours, and patient contact materials are polymers and Hydrolene®<br>Coating. | | | The Excelsior XT-17 Microcatheter is used in health care<br>facility/hospital by physicians trained in performing endovascular<br>procedures. | | | Accessories: | The Excelsior XT-17 Microcatheter is packaged with an introducer<br>sheath and a steam shaping mandrel accessory. | ### Indications for Use / Intended Use: Stryker Neurovascular's Excelsior XT-17 Microcatheters are intended to assist in the delivery of diagnostic agents, such as contrast media, and therapeutic agents, such as occlusion coils, into the peripheral, coronary and neuro vasculature. ### Principle of Operations: The Excelsior XT-17 Microcatheter is used to access and navigate the vasculature as well as to deliver diagnostic and therapeutic agents. The Excelsior XT-17 Microcatheter is inserted with a guide catheter and guided through the vasculature with the aid of a guidewire (over-the-wire microcatheters). {5}------------------------------------------------ ## Technological Characteristics and Product Feature Comparison: Stryker Neurovascular's Excelsior XT-17 Microcatheter is equivalent to the predicate device in terms of: - functionality ● - method of operation - intended use ● - . indication for use - biological safety . A tabular comparison of the specific technological characteristics between the predicate device and the proposed device is provided below. {6}------------------------------------------------ ### 510(k) Summary (cont.) ## Product feature comparison for the Excelsior® XT-17™ Microcatheter: | Characteristics | Results | |-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------| | Outer Shaft Material (Prox.) | Same as predicate | | Outer Shaft Material (Mid) | Same as predicate | | Outer Shaft Material (Distal) | Same as predicate | | Outer Shaft Material<br>(1.5 cm fluorosaver marker) | Material same as predicate but without colorant | | Inner Shaft Material | Same as predicate | | Liner | Same as predicate | | Hub/Luer | Internal hub dimension for XT-17 is slightly smaller | | Shaft Design | • Offered with a fluorosaver marker<br>• Offered with pre-shaped distal tips | | Tip Marker (Proximal) | Same as predicate | | Tip Marker (Distal) | Same material, smaller OD | | Coating | Same as predicate | | Distal Shaft Length | Same as predicate | | Mid shaft length | Same as predicate | | Mid transition length | Same as predicate | | Prox. Shaft Length | Same as predicate | | Proximal ID / OD | Same as predicate | | Distal ID / OD | Same ID, smaller OD | | Effective Length | Same as predicate | | Hydrolene Uncoated Length/<br>Coated length | Same as predicate | | Indications for Use | Same as predicate | | Sterilization Method | Same as predicate | | Packaging Pouch Materials | Offered with Nylon/PE | | Packaging Carton | Same as predicate | | Packaging Configuration | • Straight type (Same as Excelsior XT-27 Microcatheter)<br>• Pre-Shaped (Offered with new packaging design) | | Accessories | Introducer sheath added | | Closure Strip | Same as predicate | | DMSO Compatibility | DMSO compatibility information added | Risk assessment of the modifications in the form of design and use failure modes and effects analysis (design and use FMEAs) has been conducted in accordance with EN ISO 14971:2012. Stryker Neurovascular has determined the modifications to the predicate device raise no new questions of safety or effectiveness. {7}------------------------------------------------ Performance data: Design verification of the Excelsior XT-17 consisted of: Functional testing (T=0): Verification testing to assess: Tip Shape Retention ● ● Tip steam shape shrinkage . Flexibility of the Distal Shaft ● Softness of the Distal Tip . Tip shape retention (only applicable to Preshaped catheters) - Surface defects along effective length - Surface extraneous matter along effective length - . Static Rupture / Leakage under High Static Pressure - . Dynamic Burst - . Tensile Strength - Catheter Hub - Compatibility with DMSO ● - Compatibility with Peel-away introducer sheath ● - Proximal Fluoro-saver marker visibility ● Shelf life testing (Product and Packaging) ### Distribution /shipping Challenge Conditioning and Testing ### Packaging Verification Testing ### Biocompatibility testing: The biocompatibility evaluation for the proposed Excelsior XT-17 Microcatheter was performed according to EN ISO10993-1:2009 + AC: 2010. The battery of testing comprised the following. - Cytotoxicity - Hemolysis - Physiochemical Test - FTIR ● - Latex ● {8}------------------------------------------------ Conclusion: The subject modifications do not alter the intended use or indications for use, or the fundamental scientific technology of the predicate device Tracker-17 Microcatheter. The risk assessment of the modifications conducted in accordance with EN ISO 14971: 2012 and successful verification testing demonstrate that the proposed device Excelsior XT-17 Microcatheter was found to have a safety and effectiveness profile that is similar to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...