CPS AIM UNIVERSAL SLITTABLE INNER CATHETER

K130252 · St Jude Medical · DQY · Mar 20, 2013 · Cardiovascular

Device Facts

Record IDK130252
Device NameCPS AIM UNIVERSAL SLITTABLE INNER CATHETER
ApplicantSt Jude Medical
Product CodeDQY · Cardiovascular
Decision DateMar 20, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2

Intended Use

The St. Jude Medical CPS Aim™ Universal slittable inner catheter (subselector/cannulator) is designed for intracardiac access of the coronary sinus and subselection of the venous system of the heart, and to serve as a conduit during implantation for delivery of contrast medium and St. Jude Medical devices (such as guidewires and implantable left heart leads). In addition the CPS Aim Universal slittable inner catheter (subselector/cannulator) can work with outer guide catheters as a system

Device Story

The CPS Aim™ Universal slittable inner catheter is a percutaneous introducer used during left heart lead delivery procedures. It functions as a conduit for contrast medium, guidewires, and implantable leads, and can be used in conjunction with outer guide catheters. The device features a braid-reinforced PEBAX shaft with a PTFE-lined inner diameter, an atraumatic distal soft tip with an embedded radiopaque marker for fluoroscopic visualization, and a hydrophilic coating for lubricity. The proximal hub includes a retention cap, integrated valve, and sideport assembly with a 3-way stopcock. It is operated by a physician in a clinical setting (e.g., cardiac catheterization lab) to facilitate venous system cannulation. The device provides a pathway for lead placement, directly assisting the physician in cardiac device implantation, thereby benefiting the patient by enabling successful lead delivery.

Clinical Evidence

Bench testing only. Verification and validation activities confirmed that the device meets all predetermined design and performance specifications.

Technological Characteristics

Braid-reinforced PEBAX shaft; PTFE-lined inner diameter; hydrophilic coating; distal soft tip with embedded marker band; proximal hub with retention cap, integrated valve, and 3-way stopcock. Available in 59 and 65 cm lengths. Non-powered, mechanical device.

Indications for Use

Indicated for patients requiring intracardiac access to the coronary sinus and venous system of the heart for the delivery of contrast media, guidewires, and implantable left heart leads.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # MAR 20 2013 ### 510(k) Summary The purpose of this Traditional 510(K) is to gain clearance for the CPS Aim™ Universal slittable inner catheter. The key design features of the CPS Aim™ Universal slittable inner catheter have only undergone a minimal change to increase the inner diameter of the catheter as compared to the predicate, CPS Aim® SL slittable inner catheter (K090613). | Submitter: | St. Jude Medical,CRMD | |---------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------| | Address: | 15900 Valley View Ct<br>Sylmar, CA 91342 | | Phone: | 818 493 2960 | | Fax: | 818 493 3615 | | Contact Person: | Colleen Canan | | Trade Name/Proprietary<br>Name: | CPS Aim™ Universal slittable inner catheter | | Common Name: | Percutaneous Catheter<br>Model Numbers: DS2N026-59, DS2N027-59, DS2N028-59, DS2N026-65, DS2N027-65, DS2N028-65, DS2N029-65, DS2N030-65 | | Classification: | Class II, 21 CFR 870.1250 | | Legally marketed device<br>to which your firm is<br>claiming equivalence: | CPS Aim® SL slittable inner catheter (K090613) | ### Device Description: The device description of the CPS Aim™ Universal slittable inner catheter is as follows. The CPS Aim™ Universal slittable inner catheter (subselector/cannulator) is used to facilitate left heart lead delivery procedures. It is an introducer that is used to cannulate the coronary venous system and act as a conduit for contrast medium. implantable coronary leads, or other devices. The CPS Aim™ Universal slittable inner catheters will be available in the same working lengths, 59 and 65 cm as the predicate and will be available with the same number of curves as the predicate. The key design features of the CPS Aim™ Universal slittable inner catheters have only undergone a minimal change to increase the inner diameter of the catheter and the accessories have not changed. The key design features are listed below: > CPS Aim™ Universal Catheters Inner Catheter Traditional 510(k) St. Jude Medical {1}------------------------------------------------ - . Braid reinforced, varying durometer PEBAX shaft with molded proximal hub. - Inner diameter of the catheter is PTFE lined . - Atraumatic distal soft tip. - . Embedded marker band on soft tip for fluoroscopic visibility. - Outside surface of the catheter shaft is coated with a hydrophilic coating to . provide lubricity during use. - . Hub of the catheter is fitted with a retention cap and an integrated valve that includes a sideport assembly with 3-way stopcock. - The catheters are available in cannulator and sub-selector models. . - Accessories such as VBT used to assist the insertion of SJM Devices . (leads, guidewires, inner catheters, etc) #### The indication for use is as follows: The St. Jude Medical CPS Aim™ Universal slittable inner catheter (subselector/cannulator) is designed for intracardiac access of the coronary sinus and subselection of the venous system of the heart, and to serve as a conduit during implantation for delivery of contrast medium and St. Jude Medical devices (such as guidewires and implantable left heart leads). In addition the CPS Aim Universal slittable inner catheter (subselector/cannulator) can work with outer guide catheters as a system. #### Technological Characteristics of the Device Compared to the Predicate Device: The device has the same technological characteristics as the currently marketed CPS Aim® SL slittable inner catheter, with only minimal changes to the inner diameter, outer diameter, valve bypass tool, and design of shaft and hub to accommodate change in diameter of the catheter . Where differences exist between the subject device and the predicate devices performance testing demonstrated that these differences do not adversely affect the safety and effectiveness of the subject devices. #### Non-clinical Test Summary: Completion of all verification and validation activities demonstrated that the candidate devices meet their predetermined design and performance specifications and that the products are substantially equivalent to the predicate devices (Appendix 1). > CPS Aim™ Universal Catheters Inner Catheter Traditional 510(k) St. Jude Medical {2}------------------------------------------------ ## Conclusion (Statement of Equivalence): The results of the verification and validation tests and the risk analysis have demonstrated the CPS Aim™ Universal slittable inner catheter functions in accordance with product specifications. St. Jude Medical considers the CPS Aim™ Universal slittable inner catheter to be substantially equivalent to the legally marketed predicate device, the CPS Aim® SL slittable inner catheter cleared in 510(k) K090613. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an eagle or bird with outstretched wings. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 20, 2013 St. Jude Medical, CRMD C/O Colleen Canan 15900 Valley View Ct. Sylmar, CA 91342 Re: K130252 Trade/Device Name: CPS AIMTM Universal Slittable Inner Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Catheter, Percutaneous Regulatory Class: Class II Product Code: DQY Dated: January 29, 2013 Received: February 1, 2013 Dear Ms. Canan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, OwenPFaris-S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use Statement | 510(k)<br>Number<br>(if known) | K130252 | |--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | CPS Aim <sup>™</sup> Universal slittable inner catheter | | Indications<br>for Use | The St. Jude Medical CPS Aim <sup>™</sup> Universal slittable inner catheter<br>(subselector/cannulator) is designed for intracardiac access of the<br>coronary sinus and subselection of the venous system of the heart, and to<br>serve as a conduit during implantation for delivery of contrast medium<br>and St. Jude Medical devices (such as guidewires and implantable left<br>heart leads). In addition the CPS Aim Universal slittable inner catheter<br>(subselector/cannulator) can work with outer guide catheters as a system | PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED # Concurrence of CDRH, Office of Device Evaluation (ODE) OR Prescription Use X (Per 21.CFR 801. 109) Over-The-Counter Use Cwer Zi Owen P. Faris -S 2013.03.20 12:14:29 -04'00' CPS Aim™ Universal Catheters Inner Catheter Traditional 510(k) St. Jude Medical
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...