The CPS Direct™ Universal slittable outer guide catheter is designed for intracardiac access of the venous system of the heart and to serve as a conduit during implantation for the delivery of contrast medium and other devices (including implantable left heart leads and delivery tools), and support of fluids where minimizing blood loss is essential. In addition, the CPS Direct Universal slittable outer quide catheters can work with inner catheters as a system.
Device Story
CPS Direct™ Universal slittable outer guide catheter facilitates intracardiac access to the coronary venous system during cardiac resynchronization therapy (CRT). Device acts as a conduit for contrast medium, implantable left heart leads, and delivery tools; supports fluid delivery while minimizing blood loss. Used in clinical settings by physicians; functions as a system with inner catheters. Features braid-reinforced PEBAX shaft, atraumatic soft tip, and Siloxane coating for lubricity. Distal end includes gold marker bands and tungsten stripes for fluoroscopic visualization. Hub contains sideport with extension tubing for aspiration or flushing. Device is slittable to allow removal while leaving implanted leads in place. Benefits include streamlined lead delivery and access to venous anatomy.
Clinical Evidence
Bench testing only. Verification activities included functional design verification, compatibility testing, and a sterilization adoption assessment to confirm that modifications to shaft length, curve shape, and packaging did not impact safety or performance.
Technological Characteristics
Braid-reinforced, varying durometer PEBAX shaft; Siloxane coating; gold marker bands and tungsten stripes for fluoroscopy; molded proximal hub with sideport/3-way stopcock. Sterilized via EtO (ISO 11135-1, SAL 10^-6).
Indications for Use
Indicated for intracardiac access of the venous system of the heart to serve as a conduit for delivery of contrast medium, implantable left heart leads, and delivery tools, and for fluid support during implantation procedures.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Food and Drug Administration (FDA) and the Department of Health & Human Services. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The Department of Health & Human Services logo is a symbol with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" surrounding it.
June 2nd, 2023
Abbott Medical Mary Adams Regulatory Affairs Consultant 15900 Valley View Court Sylmar, California 91342
Re: K231311
Trade/Device Name: CPS Direct™ Universal slittable outer guide catheter (DS2C029) Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: May 5, 2023 Received: May 5, 2023
Dear Mary Adams:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Image /page/1/Picture/5 description: The image shows a digital signature. The name "Lydia S. Glaw -S" is written in a large font on the left side of the image. On the right side of the image, the text "Digitally signed by Lydia S. Glaw -S Date: 2023.06.02 12:11:54 -04'00'" is written in a smaller font.
Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
Submission Number (if known)
K231311
Device Name
CPS Direct™ Universal slittable outer quide catheter (DS2C029)
Indications for Use (Describe)
The CPS Direct™ Universal slittable outer guide catheter is designed for intracardiac access of the venous system of the heart and to serve as a conduit during implantation for the delivery of contrast medium and other devices (including implantable left heart leads and delivery tools), and support of fluids where minimizing blood loss is essential. In addition, the CPS Direct Universal slittable outer quide catheters can work with inner catheters as a system.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K231311
# 510(k) Summary
Prepared on: 2023-05-05
| Contact Details |
|-----------------|
|-----------------|
21 CFR 807.92(a)(1)
| Applicant Name | Abbott Medical | |
|-----------------------------------------------------------|----------------------------------------------------------------------------|--------------|
| Applicant Address | 15900 Valley View Court Sylmar CA 91342 United States | |
| Applicant Contact Telephone | (909) 991-5235 | |
| Applicant Contact | Ms. Mary Adams | |
| Applicant Contact Email | mary.adams2@abbott.com | |
| Device Name | 21 CFR 807.92(a)(2) | |
| Device Trade Name | CPS Direct <sup>™</sup> Universal slittable outer guide catheter (DS2C029) | |
| Common Name | Percutaneous catheter | |
| Classification Name | Catheter, Percutaneous | |
| Regulation Number | 870.1250 | |
| Product Code | DQY | |
| Legally Marketed Predicate Devices<br>21 CFR 807.92(a)(3) | | |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K130257 | CPS Direct <sup>™</sup> Universal Slittable Outer Catheter | DQY |
| Device Description Summary | 21 CFR 807.92(a)(4) | |
The device description of the CPS Direct™ Universal slittable outer guide catheter, 3D, is as follows.
· The CPS Direct™ Universal slittable outer quide catheter lead delivery during cardiac resynchronization therapy (CRT) procedures. The CPS Direct Universal provides access to the coronary venous system and acts as a conduit for contrast medium. The key design features of the CPS Direct™ Universal sittable outer quide catheter, 3D, has only undergone minimal changes to the shaft length, the curve design, and the package tray configuration. The diator packaged with the catheter has undergone a change in shaft length to accommodate the shaft length of the 3D catheter. The accessories listed in the device instructions for use have not changed. The key design features of the catheter are listed below:
- · Braid reinforced, varying durometer PEBAX shaft with molded proximal hub.
- · Atraumatic distal soft tip.
- · The outside surface and inside surfaces of the catheter shaft are coated with Siloxane to provide lubricity during use.
- · The distal end of the shaft has gold marker bands and tungsten stripes for fluoroscopic visibility.
- · Hub contains a sideport with extension tubing for contrast delivery, aspiration, or saline flush using a 3-way stopcock.
- · Accessories such as VBT used to assist the insertion of Abbott Devices (leads, quidewires, inner catheters, etc)
## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
21 CFR 807.92(a)(5)
The CPS Direct™ Universal slittable outer guide catheter is designed for intracardiac access of the heart and to serve as a conduit during implantation for the delium and other devices (including implantable left heart leads and delivery tools), and support of fluids where minimizing blood loss is essential. In addition, the CPS Direct Universal slittable outer quide catheters can work with inner catheters as a system.
# Indications for Use Comparison
{4}------------------------------------------------
The subject device CPS Direct™ Universal slittable outer guide catheter model DS2C029 (3D curve) has the following similarities to the predicate device model (DS2C029 extra wide curve) which previously received 510(k) concurrence (K130257):
- · Has the same indicated use,
- · Has the same fundamental scientific technology,
- · Incorporates the same basic catheter design,
- · Incorporates the same catheter construction,
- · Incorporates the same device and packaging materials, including sterile barrier,
- · Is sterilized using the same processes (EtO with SAL of at least 10^-6 per ISO 11135-1).
In summary, the CPS Direct Universal slittable outer guide catheter model DS2C029 (3D curve) described in this submission is substantially equivalent to the predicate device.
# Technological Comparison
### 21 CFR 807.92(a)(6)
The subject CPS Direct™ Universal slittable outer quide catheter model DS2C029 (3D curve) has the same technological characteristics as the currently marketed CPS Direct™ Universal slittable outer guide catheter model DS2C029 (X-Wide curve), with only minimal changes to the catheter, dilator and packaging design, and meets current regulatory requirements and standards.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The modifications to the CPS Direct™ Universal 3D sittable outer quide catheter and packaging, i.e., catheter length, catheter curve shape, and packaging tray, necessitated limited functional and packaging design verification (DV) testing, compatibility testing, and a sterilization adoption assessment study. Completion of all verification activities demonstrated that the modified catheter with associated components meets its predetermined design and that the subject CPS Direct™ Universal sittable outer quide catheter model DS2C029 (3D curve) is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.