CPS DIRECT SL II SLITTABLE OUTER CATHETER, MODELS DS2C001, DS2C002, DS2C003, DS2C004, DS2C005
K092075 · St. Jude Medical, Cardiac Rhythm Management Divisi · DQY · Aug 7, 2009 · Cardiovascular
Device Facts
Record ID
K092075
Device Name
CPS DIRECT SL II SLITTABLE OUTER CATHETER, MODELS DS2C001, DS2C002, DS2C003, DS2C004, DS2C005
Applicant
St. Jude Medical, Cardiac Rhythm Management Divisi
Product Code
DQY · Cardiovascular
Decision Date
Aug 7, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The St. Jude Medical CPS Direct® SL II Slittable Outer Catheters Indications for Use intended use is designed for intracardiac access of the venous system of the heart and to serve as a conduit during implantation for the delivery of contrast medium and St. Jude Medical devices, including implantable left heart leads and delivery tools, and support of fluids where minimizing blood loss is essential. In addition, the outer guide catheters can work with inner catheters as a system.
Device Story
CPS Direct SL II Slittable Outer Catheter functions as a conduit for intracardiac venous access; facilitates delivery of contrast media, implantable left heart leads, and delivery tools; supports fluid administration while minimizing blood loss. Used in clinical settings by physicians; operates as a standalone guide catheter or in conjunction with inner catheters as a system. Provides a pathway for device implantation and diagnostic contrast delivery.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Slittable outer guide catheter; designed for intracardiac venous access; compatible with inner catheters; materials and design support fluid delivery and lead implantation.
Indications for Use
Indicated for patients requiring intracardiac access to the venous system of the heart for delivery of contrast media, St. Jude Medical devices (including left heart leads and delivery tools), and fluid support.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is an abstract image of an eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-0609 Silver Spring, MD 20993-0002
St. Jude Medical Cardiac Rhythm Management Division % Ms. Colleen Canan Senior Regulatory Submission Specialist 15900 Valley View Court Sylmar, California 91342
## AUG 0 7 2009
### Re: K092075
Trade/Device Name: CPS Direct® SL II Slittable Outer Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: II Product Code: DQY Dated: June 30, 2009 Received: July 9, 2009
Dear Ms. Canan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting
{1}------------------------------------------------
Page 2-Ms. Colleen Canan
(reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely
Image /page/1/Picture/7 description: The image contains a single, large, cursive letter 'f'. The letter is drawn with thick lines, and it has a loop at the top and a long tail extending downwards. The letter is black against a white background.
Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use Statement
210(k) Number (if known)
K092075
CPS Direct® SL II Slittable Outer Catheter Device Name
The St. Jude Medical CPS Direct® SL II Slittable Outer Catheters Indications for Use intended use is designed for intracardiac access of the venous system of the heart and to serve as a conduit during implantation for the delivery of contrast medium and St. Jude Medical devices, including implantable left heart leads and delivery tools, and support of fluids where minimizing blood loss is essential. In addition, the outer guide catheters can work with inner catheters as a system.
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use in X (Per 21 CFR 801. 109)
Over-The-Counter Use
ecmde
OR
(Division Sign-Off) Division of Cardiovascular Devices
K092075 510(k) Number