K974684 · Scimed Life Systems, Inc. · DQY · Feb 19, 1998 · Cardiovascular
Device Facts
Record ID
K974684
Device Name
SCIMED 7 FRENCH WISEGUIDE GUIDE CATHETER
Applicant
Scimed Life Systems, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Feb 19, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Intended Use
SCIMED 7 French Wiseguide guide catheter is intended for use in general intravascular and coronary applications. It provides a pathway through which medical instruments, such as stents, balloon dilatation catheters, guide wires or other therapeutic devices may be introduced. This device is not intended for use in the cerebral vasculature.
Device Story
7 French guide catheter; provides pathway for introduction of medical instruments (stents, balloon dilatation catheters, guide wires). Used in coronary and general intravascular procedures. Operated by physicians in clinical settings. Features three-layer shaft construction: inner layer for low friction; middle layer for kink resistance and torque control; outer layer for stiffness, backup support, and radiopacity. Distal tip radiopaque for fluoroscopic visualization. Single-use, sterile device.
Clinical Evidence
Bench testing only. Functional testing included pressure burst, tip bond tensile, shaft tensile, hub tensile, material adhesion, tip coefficient of friction, force transmitted by catheter tip, torque response, and dye flow.
Technological Characteristics
Three-layer catheter shaft; biocompatible materials; variable durometer outer shaft for transitional flexibility; radiopaque distal tip. Dimensions: 7 French. Sterile, single-use. No software or electronic components.
Indications for Use
Indicated for patients requiring general intravascular and coronary procedures for the introduction of therapeutic devices. Contraindicated for use in the cerebral vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
SCIMED® 8 French Wiseguide Guide Catheter
SCIMED® 7 French Triguide®-Max Guide Catheter
Cordis® Corporation 7 French Vista Brite Tip™ Guide Catheter
Related Devices
K961280 — SCIMED 6-10 F TRIGUIDE GUIDE CATHETERS 6F,7F,8F,9F & 10F · Scimed Life Systems, Inc. · May 15, 1996
K981788 — SCIMED 6 FRENCH WISEGUIDE GUIDE CATHETER · Scimed Life Systems, Inc. · Aug 18, 1998
K970823 — SCIMED 8 FRENCH WISEGUIDE GUIDE CATHETER · Scimed Life Systems, Inc. · Jun 3, 1997
K961999 — SCHNEIDER GUIDER SOFTIP GUIDING CATHETERS · Boston Scientific Scimed, Inc. · Nov 19, 1996
K963663 — 11F GUIDING CATHETER · Boston Scientific /International Technologies · Jan 13, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
K974684
## Section 4
## Summary of Safety and Effectiveness
# (Pursuant To Section 12 of the SAFE MEDICAL DEVICES ACT of 1990)
FEB 1 9 1998
#### I. General Provisions
| Submitter's Name<br>and Address | SCIMED Life Systems, Inc.<br>One SCIMED Place<br>Maple Grove, Minnesota 55311 |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Melanie Raska<br>(612) 494-2962 |
| Classification Name | Similar to Diagnostic Intravascular<br>Catheters (21CFR Part 870.1200) |
| Common or Usual Name | Coronary Guide Catheter |
| Proprietary Name | SCIMED® 7 French Wiseguide™ Guide<br>Catheter |
| Name of Predicate Devices | SCIMED® 8 French Wiseguide™ and 7<br>French Triguide® -Max Guide Catheters,<br>and Cordis® Corporation 7 French Vista<br>Brite Tip™ Guide Catheters |
#### III. Device Description
II.
The shaft of the 7 F Wiseguide guide catheter utilizes common biocompatible materials and consists of the following three layers: 1) the inner layer that provides a low coefficient of friction and facilitates passage of medical devices such as stents, balloon dilatation catheters, guide wires or other therapeutic devices, 2) the middle layer that extends from the shaft to the tip to provide kink resistance and torque control; and 3) the outer layer which provides stiffness, backup support, curve retention and radiopacity.
In addition, the outer primary catheter shaft is constructed of various material durometers providing a distal curve area of the catheter with transitional flexibility.
The distal tip is radiopaque to allow visualization under fluoroscopy during a procedure.
The devices will be provided sterile and are intended for one procedure use only.
{1}------------------------------------------------
## Section 4
# Summary of Safety and Effectiveness (cont.)
### IV. Intended Use
The 7 F Wiseguide catheter is designed to provide a pathway through which medical instruments, such as stents, balloon dilatation catheters, guide wires or other therapeutic devices may be introduced. This device is not intended for use in the cerebral vasculature.
#### Summary of Technological Characteristics: V.
The 7 F Wiseguide catheter is similar to SCIMED's currently marketed 8 F Wiseguide catheter.
## VI. Non-clinical Test Summary
Functional testing consisted of pressure burst, tip bond tensile, shaft tensile, hub tensile, material adhesion, tip coefficient of friction, force transmitted by catheter tip, torque response and dye flow. Test results verified that the 7 F Wiseguide catheter is adequate for its intended use. The 7 F Wiseguide catheter is considered substantially equivalent to guide catheters currently marketed by SCIMED and Cordis based on a comparison of intended use, the design, and the results of in-vitro testing and evaluation.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The eagle is positioned to the right of a circular emblem containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856
FEB 1 9 1998
Ms. Melanie Raska Requlatory Affairs Specialist Scimed Life System, Inc. Boston Scientific Corporation One Scimed Place Maple Grove, MN 55311-1566
Re : K974684 Scimed® 7 French Wiseguide™ Guide Catheter Regulatory Class: II (Two) Product Code: 73 DQY December 15, 1997 Dated: Received: December 16, 1997
Dear Ms. Raska:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. ----------------------------------------------------------------------------------------------------------------------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality and him System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the
{3}------------------------------------------------
#### Page 2 - Ms. Melanie Raska
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Notification SCIMED® 7 F Wiseguide Guide Catheter
## Section 2
Indications for Use
510(k) Number (if known) _
Device Name: SCIMED® 7 French Wiseguide™ Guide Catheter
Indications for Use:
SCIMED 7 French Wiseguide guide catheter is intended for use in general intravascular and coronary applications. It provides a pathway through which medical instruments, such as stents, balloon dilatation catheters, guide wires or other therapeutic devices may be introduced. This device is not intended for use in the cerebral vasculature.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR Over The Counter Use **__**
(Optional Format 1-2-96)
(Division Sign-Off) Division of Cardinvasc Dar and Neurological Devices 510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.