SCIMED 8 FRENCH WISEGUIDE GUIDE CATHETER
K970823 · Scimed Life Systems, Inc. · DQY · Jun 3, 1997 · Cardiovascular
Device Facts
| Record ID | K970823 |
| Device Name | SCIMED 8 FRENCH WISEGUIDE GUIDE CATHETER |
| Applicant | Scimed Life Systems, Inc. |
| Product Code | DQY · Cardiovascular |
| Decision Date | Jun 3, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.1250 |
| Device Class | Class 2 |
Intended Use
SCIMED guide catheters are intended for use in general intravascular and coronary applications. They provide a pathway through which medical instruments, such as balloon dilatation catheters, guide wires or other therapeutic devices may be introduced. These devices are not intended for use in the cerebral vasculature
Device Story
The SCIMED 8 F Wiseguide Guide Catheter is a diagnostic intravascular catheter used to provide a pathway for introducing medical instruments, such as balloon dilatation catheters or guide wires, into the coronary or general intravascular systems. The device consists of a three-layer shaft: an inner PTFE layer for low friction, a middle braided stainless steel wire layer for kink resistance and torque control, and an outer Pebax layer with radiopaque filler for stiffness and visualization. The distal section features multiple Pebax segments of varying durometers to provide transitional flexibility. The device is used by physicians in clinical settings during interventional procedures. It is provided sterile for single-procedure use. By facilitating the delivery of therapeutic devices, it enables coronary and intravascular interventions.
Clinical Evidence
No clinical data. Evidence consists of non-clinical bench testing including pressure burst, tip bond tensile, shaft tensile, hub tensile, tip coefficient of friction, force transmission, and torque testing. Biocompatibility and shelf-life testing were also performed.
Technological Characteristics
Three-layer shaft construction: inner PTFE, middle braided stainless steel wire, outer Pebax with radiopaque filler. Distal section uses multi-durometer Pebax for transitional flexibility. 8 F size (0.086" ID). Lengths 40-125 cm. Sterile, single-use. No electronic or software components.
Indications for Use
Indicated for patients requiring general intravascular and coronary procedures for the introduction of therapeutic devices. Contraindicated for use in the cerebral vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
- SCIMED® 8 French Cyber™ Guide Catheters
- Cordis® Corporation, 8 French Vista Brite Tip™ Guide Catheters
Related Devices
- K961280 — SCIMED 6-10 F TRIGUIDE GUIDE CATHETERS 6F,7F,8F,9F & 10F · Scimed Life Systems, Inc. · May 15, 1996
- K981788 — SCIMED 6 FRENCH WISEGUIDE GUIDE CATHETER · Scimed Life Systems, Inc. · Aug 18, 1998
- K974684 — SCIMED 7 FRENCH WISEGUIDE GUIDE CATHETER · Scimed Life Systems, Inc. · Feb 19, 1998
- K151396 — MIVI Mi-AXUS Guide Catheter · Mivi Neuroscience, Inc. · Dec 9, 2015
- K153703 — Guiding Catheter · Texasmedical Technologies, Inc. · May 12, 2016
Submission Summary (Full Text)
{0}
510(k) Notification
SCIMED® 8 F Guide Catheter
JUN - 3 1997
K970823
Section 4
Summary of Safety and Effectiveness
(Pursuant To Section 12 of the SAFE MEDICAL DEVICES ACT of 1990)
I. General Provisions
Submitter’s Name and Address
SCIMED Life Systems, Inc.
One SCIMED Place
Maple Grove, Minnesota 55311
Contact Person
Angela Raun
(612) 494-2456
Classification Name
Similar to Diagnostic Intravascular Catheters (21CFR Part 870.1200)
Common or Usual Name
Coronary Guide Catheter
Proprietary Name
SCIMED® 8 French Wiseguide™ Guide Catheter
II. Name of Predicate Devices
SCIMED® 8 French Cyber™ Guide Catheters and Cordis® Corporation,
8 French Vista Brite Tip™ Guide Catheters
III. Device Description
The shaft of the 8 F Wiseguide guide catheter utilizes common biocompatible materials and consists of the following three layers: 1) the inner polytetrafluoroethylene (PTFE) layer that provides a low coefficient of friction and facilitates easy passage of medical devices such as balloon dilatation catheters, guide wires or other therapeutic devices, 2) the middle layer which is made of braided stainless steel wire that extends from the shaft to the tip to provide kink resistance, torque control and support and 3) the outer layer, manufactured from Pebax® and a radiopaque filler, which provides stiffness, memory and radiopacity.
In addition, the distal 5.8 to 17.5 inches (depending on the curve style) of the outer primary catheter shaft is constructed of multiple sections of Pebax. These sections are composed of durometers of Pebax providing a distal curve area of the catheter with transitional flexibility. The catheter is assembled as components over a core mandrel. The catheter shaft is then heat fused, forming a composite catheter shaft. This process results in a continuous inner and middle layer and a seamless outer shaft layer.
{1}
510(k) Notification
SCIMED® 8 F Guide Catheter
# Section 4 Summary of Safety and Effectiveness, cont.
The distal tip is made of Pebax and is radiopaque to allow visualization under fluoroscopy during a procedure.
The catheters utilize a one piece Pebax hub/strain relief that is molded to the proximal end of the guide catheter shaft. The 8 F Wiseguide Guide Catheters have a 0.086" ID. The catheters will be available in lengths ranging from 40 to 125 cm, with optional side holes. The devices will be provided sterile and are intended for one procedure use only.
## IV. Intended Use
The SCIMED guide catheters are designed to provide a pathway through which medical instruments, such as balloon dilatation catheters, guide wires or other therapeutic devices may be introduced. These devices are not intended for use in the cerebral vasculature.
## V. Summary of Technological Characteristics:
The 8 F Wiseguide Guide Catheters are similar to SCIMED's currently marketed 8 F Cyber Guide Catheters with the exception of the distal flexible sections providing varying degrees of flexibility in the distal curve area.
## VI. Non-clinical Test Summary
Functional testing consisted of pressure burst, tip bond tensile, shaft tensile, hub tensile, tip coefficient of friction, force transmitted by the catheter tip, and torque testing. Biocompatibility and shelf life testing has also been conducted. Test results verified that the 8 F Wiseguide catheter is adequate for its intended use. The 8 F Wiseguide guide catheters are considered substantially equivalent to guide catheters currently marketed by SCIMED and Cordis based on a comparison of intended use, the design and the results of in-vitro testing and evaluation.
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
JUN 3 1997
Ms. Angela Raun
Sr. Regulatory Affairs Specialist
Scimed Life Systems, Inc.
One Scimed Place
Maple Grove, Minnesota 55311-1566
Re: K970823
SCIMED® 8 French Wiseguide™ Guide Catheter
Regulatory Class: II (two)
Product Code: DQY
Dated: March 5, 1997
Received: March 6, 1997
Dear Ms. Raun:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in
{3}
regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597.
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{4}
510(k) Notification
SCIMED® 8 F Guide Catheter
# Section 2
## Indications for Use
510(k) Number (if known) K970823
Device Name: SCIMED® 8 French Wiseguide™ Guide Catheter
Indications for Use:
SCIMED guide catheters are intended for use in general intravascular and coronary applications. They provide a pathway through which medical instruments, such as balloon dilatation catheters, guide wires or other therapeutic devices may be introduced. These devices are not intended for use in the cerebral vasculature
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ☑ (Per 21 CFR 801.109) ☐ OR ☐ Over The Counter Use ☐
(Division Sign-Off) (Optional Format 1-2-96)
Division of Cardiovascular, Respiratory, Neurological Devices
(OK) Number K970823