The Amulet™ Steerable Delivery Sheath is indicated to facilitate the delivery of the Amplatzer™ Amulet™ Left Arrial Appendage Occluder.
Device Story
The Amulet™ Steerable Delivery Sheath is a 14F, 75 cm long percutaneous catheter used to provide a pathway for delivering the Amplatzer™ Amulet™ Left Atrial Appendage Occluder. Operated by physicians in a clinical setting, the device features a bi-directional distal tip controlled by a deflection knob on the handle, allowing for 120° clockwise deflection. The sheath body is radiopaque for fluoroscopic visualization. The device is used to navigate cardiac anatomy to position the occluder. It does not include an integrated hemostasis valve or sideport. The output is the successful delivery and placement of the occluder, which benefits the patient by facilitating the closure of the left atrial appendage.
Clinical Evidence
Bench testing only. Performance was validated through a simulated-use model involving experienced physicians, confirming the device successfully delivers the Amplatzer Amulet Left Atrial Appendage Occluder. Biocompatibility, shelf-life, and mechanical performance (tensile, torque, system leak, dimensional, luer compatibility, handoff force, air introduction) were verified. Human factors evaluation confirmed no new user-device interactions or use errors compared to the predicate.
Technological Characteristics
14F, 75 cm long steerable catheter; bi-directional distal tip; deflection knob; radiopaque body; dual-curve geometry. Materials evaluated per ISO 10993-1, ISO 14971, and ASTM F2475. Sterilized via ethylene oxide (EO) per ISO 11135. Packaging tested per ASTM F2825, D4332, F2096, F1886, and F88.
Indications for Use
Indicated for patients requiring delivery of the Amplatzer™ Amulet™ Left Atrial Appendage Occluder.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The seal features a stylized depiction of a human figure, while the text is in a clear, sans-serif font, with "FDA" in a blue square.
September 29, 2023
Abbott Medical Dan Gapp Regulatory Affairs Project Manager 177 County Road B East St. Paul, Minnesota 55117
Re: K232690
Trade/Device Name: AmuletTM Steerable Delivery Sheath Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: DOY Dated: September 1, 2023 Received: September 1, 2023
Dear Dan Gapp:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by Katherine Katherine N. Trivedi -S N. Trivedi -S 0ate: 2023.09.29
for Rachel Neubrander Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232690
Device Name Amulet™ Steerable Delivery Sheath
Indications for Use (Describe)
The Amulet™ Steerable Delivery Sheath is indicated to facilitate the delivery of the Amplatzer™ Amulet™ Left Arrial Appendage Occluder.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Abbott. The logo consists of a stylized blue "A" on the left and the word "Abbott" in black font on the right. The "A" is a rounded square shape with a horizontal line through the middle.
# 510(k) SUMMARY
The 510(k) summary is submitted in accordance with 21 CFR 807.92 and the requirements of the Safe Medical Device Act (SMDA) of 1990.
# I. SUBMITTER INFORMATION
| Submitter Name: | Abbott Medical |
|--------------------|--------------------------------------------------|
| Submitter Address: | 177 County Road B East<br>St. Paul, MN 55117 USA |
| Phone: | (651) 756-5833 |
| Contact Person: | Dan Gapp |
| Date Prepared: | 28 September 2023 |
# II. DEVICE
| Name of Device: | Amulet™ Steerable Delivery Sheath |
|----------------------|------------------------------------------|
| Common Name: | Catheter Delivery Sheath |
| Classification Name: | Catheter, Percutaneous (21 CFR 870.1250) |
| Device Class: | II |
| Product Code: | DQY |
# III. PREDICATE DEVICES
Primary Predicate: Amplatzer™ Steerable Delivery Sheath (K220340, cleared 09 March 2022)
# IV. DEVICE DESCRIPTION
The Amulet™ Steerable Delivery Sheath is designed to provide a pathway through which a device may be delivered. The sheath is provided in one size (14F), with a working length of 75 cm and a bi-directional distal tip to provide positioning in the cardiac anatomy. The handle is equipped with a deflection knob to deflect the tip clockwise 120° and counterclockwise 0°. The body of the sheath is radiopaque for visibility under fluoroscopy and has a marker band located in the dilator eases penetration of tissue. The sheath and dilator utilize a dual curve in two dimensions, resulting in a three-dimensional geometry. The 14F flush adapter facilitates the attachment of additional components.
# V. INDICATIONS FOR USE
The Amulet™ Steerable Delivery Sheath is indicated to facilitate the delivery of the Amplatzer™ Amulet™ Left Atrial Appendage Occluder.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The Amulet Steerable Delivery Sheath incorporates substantially equivalent design, function, packaging, sterilization process, materials, fundamental technology, indication for use, and operating principles as those shared by the predicate Amplatzer Steerable Delivery Sheath (K220340).
A comparison of the Amulet Steerable Delivery Sheath and Amplatzer Steerable Delivery Sheath show these devices have substantially equivalent design characteristics, including a deflectable sheath, dilator and adapter. The difference between the subject device and predicate is the subject device does not include an integrated hemostasis valve and sideport.
# VII. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING
The following design verification and validation testing was provided in support of a substantial equivalence determination. Components not impacted by the design changes did not undergo additional testing.
#### Biocompatibility
The biocompatibility evaluation of the Amulet Steerable Delivery Sheath was conducted in accordance with the FDA Guidance: Use of International Standard ISO 10993-1 Biological evaluation of medical devices - Part 1:
{5}------------------------------------------------
# 510(k) SUMMARY
Evaluation and testing within a risk management process" (2020), ISO 10993-1, ISO 14971 and ASTM F2475. The device is considered an external communicating device with limited (≤24 hour) contact with circulating blood, therefore evaluation was conducted in the following categories:
- Hemocompatibility ●
- Cytotoxicity
- Sensitization
- Irritation ●
- Materials-Mediated Pyrogenicity ●
- Acute Systemic Toxicity
#### Design Verification
Bench testing, including shelf life testing, was conducted to demonstrate that the Amulet Steerable Delivery Sheath met all performance specifications, including:
- Tensile Testina ●
- Torque Testing ●
- System Leak
- Dimensional ●
- Device Compatibility ●
- Luer Compatibility .
- Handoff Force ●
- System Prep
- Visual Requirements ●
- Air Introduction ●
#### Sterilization
The Amulet Steerable Delivery Sheath is intended for single use only and is provided sterile via ethylene oxide (EO) gas to achieve a Sterility Assurance Level (SAL) of 10° per ISO 11135. The Amulet Steerable Delivery Sheath was adopted into routine cycles used to sterilize the primary predicate device, the Amplatzer Steerable Delivery System, based on a composition, packaging, design, pallet density, and bioburden levels per ANSI/AAMI TIR 28. An EO/ECH residuals assessment found the residuals to be acceptable with respect to the predicate device per ANSI/AAMI/ISO 10993-7, following 2X sterilization.
#### Packaging
Packaging verification studies were performed in complicable requirements of ASTM F2825, ASTM D4332, ASTM F2096; F1886/F1886M and ASTM F88/F88M. All device packaging met acceptance criteria following 2X sterilization, environmental conditioning and transport simulation.
#### Design Validation
A design validation study was performed to evaluate the Amulet Steerable Delivery Sheath to deliver an Amplatzer Amulet Left Atrial Appendage Occluder. In a benchtop simulated use model, physicians who have experience with left atrial appendage devices evaluated the performance of the Amulet Steerable Delivery Sheath to deliver the Amplatzer Amulet Left Atrial Appendage Occluder. The protocol acceptance criteria were met.
#### Human Factors
A Human Factors evaluation of the Amulet Steerable Delivery Sheath was performed to confirm the design changes did not introduce any new user-device interactions per the requirements of ANSI/AAM/IEC 62366-1. This evaluation did not observe any patterns of use errors or identify any new user-device interactions compared to the predicate device.
# VIII. CONCLUSION
Based on the indication for use, technological characteristics and non-clinical performance testing provided, the Amulet Steerable Delivery Sheath is substantially equivalent to the predicate Amplatzer Steerable Delivery
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the Abbott logo. The logo consists of a blue stylized letter "a" on the left and the word "Abbott" in black on the right. The letter "a" is made up of three horizontal lines and a vertical line, forming a stylized "a" shape. The word "Abbott" is in a bold, sans-serif font.
# 510(k) SUMMARY
Sheath (K220340). The Amulet Steerable Delivery Sheath should perform as intended in the specified use conditions.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.