Microcatheter

K232647 · Surmodics,Inc. · DQY · Sep 29, 2023 · Cardiovascular

Device Facts

Record IDK232647
Device NameMicrocatheter
ApplicantSurmodics,Inc.
Product CodeDQY · Cardiovascular
Decision DateSep 29, 2023
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2

Intended Use

The Surmodics Microcatheter is intended to provide support to facilitate the placement of guidewires in the coronary and peripheral vasculatures and can be used to exchange one guidewire for another. This microcatheter is also intended to assist in the delivery of contrast media into the coronary and peripheral vasculatures. Do not use this microcatheter other than for use in the coronary and peripheral vasculatures.

Device Story

Sterile, single-use, percutaneous microcatheter; provides support for guidewire placement/exchange and contrast media delivery in coronary and peripheral vasculatures. Features single-lumen design; 1.9 Fr distal/2.6 Fr proximal outer diameter; compatible with 0.014" guidewires; available in 135 cm and 150 cm lengths. Utilizes Preside hydrophilic coating for lubricity. Operated by clinicians in interventional settings to access target vascular sites. Output is physical support for guidewires and conduit for contrast agents; aids clinical decision-making by facilitating navigation through complex vascular anatomy. Benefits patient by enabling access to discrete vascular regions.

Clinical Evidence

No clinical studies were required; substantial equivalence supported by bench testing and biocompatibility data.

Technological Characteristics

Single-lumen percutaneous catheter; 1.9 Fr distal/2.6 Fr proximal outer diameter; 0.014" guidewire compatibility; 135 cm/150 cm lengths. Features Preside hydrophilic coating. Sterile, single-use. No software or electronic components.

Indications for Use

Indicated for use in coronary and peripheral vasculatures to provide support for guidewire placement, guidewire exchange, and delivery of contrast media.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ July 29, 2024 Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. Surmodics. Inc. Jodie Christe Regulatory Affairs Associate II 9924 West 74th Street Eden Prairie, Minnesota 55344 Re: K232647 Trade/Device Name: Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: DQY Dear Jodie Christe: The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change for your device cleared on September 29, 2023. Specifically, FDA is updating this SE Letter because FDA inadvertently indicated that the SE determination also included review and clearance of a predetermined change control plan (PCCP). However, your 510(k) submission did not include a PCCP, so FDA is providing this administrative correction. Please see the attached revised clearance letter. Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Lydia Glaw, OHT2: Office of Cardiovascular Devices, 301-796-1456, Lydia.Glaw(@fda.hhs.gov. Sincerelv. # Samuel G. Raben -S for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {1}------------------------------------------------ July 29, 2024 Image /page/1/Picture/1 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is a stylized depiction of an eagle, while the FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are placed side-by-side, indicating a connection between the two organizations. Surmodics. Inc. Jodie Christe Regulatory Affairs Associate II 9924 West 74th Street Eden Prairie, Minnesota 55344 ## Re: K232647 Trade/Device Name: Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: DQY Dated: August 30, 2023 Received: August 31, 2023 Dear Jodie Christe: The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change for your device cleared on September 29, 2023. We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {2}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). > Sincerely, Samuel G. Raben -S for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K232647 Device Name Microcatheter Indications for Use (Describe) The Surmodics Microcatheter is intended to provide support to facilitate the placement of guidewires in the coronary and peripheral vasculatures and can be used to exchange one guidewire for another. This microcatheter is also intended to assist in the delivery of contrast media into the coronary and peripheral vasculatures. Do not use this microcatheter other than for use in the coronary and peripheral vasculatures. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------|---------------------------------------------------------------------------------| | <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary Date Prepared: August 30, 2023 ## Submitters Name/Contact Person ## 510(k) Submitter and Contact for Routine Correspondence Jodie Christe Regulatory Affairs Associate II Surmodics, Inc. 9924 West 74th Street Eden Prairie, MN 55344 Phone (952)-500-7065 Email: jchriste(@surmodics.com ## 510(k) Submitter Establishment Registration Number 3001374820 | General Information | | |----------------------------|-------------------------| | Trade Name: | Microcatheter | | Common / Usual Name: | Microcatheter | | Classification Name | Cardiovascular | | Regulation/Product Code | 21 CFR 870.5150 | | Device Panel | Cardiovascular | | Regulatory Classification: | Class II | | Product Code: | DQY | | Predicate Device: | Microcatheter (K173560) | {5}------------------------------------------------ #### Device Description 8.1 The Surmodics Microcatheter is a sterile, single use, percutaneous device intended to provide support to facilitate the placement of guidewires in the coronary and peripheral vasculature. Additionally, the microcatheter can be used for contrast media injection. The catheter is a single-lumen microcatheter with a 1.9 Fr distal outer diameter, a 2.6 Fr proximal outer diameter and is compatible with a 0.014" guidewire. The proximal end of the catheter shaft has a larger inner diameter, outer diameter and wall thickness and tapers to the distal end. The catheter is available in working lengths of 135 cm and 150 cm to facilitate access to various target sites. #### 8.2 Indication for Use The Surmodics Microcatheter is intended to provide support to facilitate the placement of guidewires in the coronary and peripheral vasculatures and can be used to exchange one guidewire for another. This microcatheter is also intended to assist in the delivery of contrast media into the coronary and peripheral vasculatures. Do not use this microcatheter other than for use in the coronary and peripheral vasculatures. #### 8.3 Comparison of Technological Characteristics The Surmodics Microcatheter device is substantially equivalent to the legally marketed predicate device in design, intended use, principles of operation, catheter materials, sizes and sterility. The Surmodics Microcatheter and the predicate device are intended to access discrete regions of the coronary and peripheral vasculature to facilitate the placement of or exchange of guidewires or to deliver contrast media. Both devices have similar dimensions and similar accessory compatibility. The only change from the predicate device is a modified lubricous hydrophilic coating. The modified hydrophilic top coat formulation, known as Preside, was developed to provide better customer experience. {6}------------------------------------------------ #### 8.4 Performance Bench Testing Results of design verification and biocompatibility testing demonstrates that the technological and material differences identified do not raise new questions of safety or effectiveness compared to the predicate device. The Surmodics Microcatheter has been evaluated through the following tests: - Coating Integrity - - -Particulate - -Track Force - Simulated Use (part of track force testing) - - Pinch testing lubricity - - -Biocompatibility ### 8:5 Clinical Studies and Testing No clinical studies were required for the Surmodics Microcatheter. #### 8.6 Conclusion Based on the device description, materials, technological characteristics, and accompanying performance and biocompatibility data it can be concluded that the device modifications made to the Surmodics Microcatheter are substantially equivalent to the predicate device and the device will continue to function per its intended use.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...