SPEX SUPPORT CATHETER

K141649 · Reflow Medical · DQY · Nov 5, 2014 · Cardiovascular

Device Facts

Record IDK141649
Device NameSPEX SUPPORT CATHETER
ApplicantReflow Medical
Product CodeDQY · Cardiovascular
Decision DateNov 5, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2

Intended Use

The speX Support Catheter is intended to be used in conjunction with steerable guidewires to access discreet regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices and provide a conduit for the delivery of saline solutions or diagnostic/therapeutic agents

Device Story

Over-the-wire, single-lumen support catheter; used to provide additional support to steerable guidewires during peripheral vascular access. Features proximal female luer hub, flexible shaft with varying stiffness, and radiopaque tapered distal tip. Distal end includes lubricious coating. Operated manually by physician during percutaneous interventional procedures. Device facilitates guidewire/device exchange and serves as conduit for saline or diagnostic/therapeutic agents. Benefits patient by enabling navigation through complex peripheral anatomy and assisting in crossing occlusions via controlled forward advancement of tapered tip.

Clinical Evidence

Bench testing only. No clinical data. Testing included simulated use, tensile strength, torque, pressure, coating verification, corrosion, luer verification, dimensional/visual inspection, sterilization validation, and comprehensive biocompatibility (cytotoxicity, irritation, sensitization, pyrogenicity, systemic toxicity, hemolysis, PTT, complement activation, thrombosis).

Technological Characteristics

Single-lumen over-the-wire catheter; medical-grade polymers; braid-reinforced shaft; PTFE lining; radiopaque tapered distal tip; hydrophilic coating (distal 40cm); 4.6F OD; 0.035" guidewire compatible; luer hub handle; sterile; single-use.

Indications for Use

Indicated for use in peripheral vasculature to access discrete regions, facilitate guidewire/interventional device placement and exchange, and deliver saline or diagnostic/therapeutic agents. Compatible with 0.035" guidewires.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, composed of three intertwined human profiles facing right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 5, 2014 ReFlow Medical Ms. Rebecca Pine VP. Regulatory/Clinical Affairs & Quality Assurance 1003 Calle Sombra San Clemente, CA 92673 Re: K141649 Trade/Device Name: speX Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: October 2, 2014 Received: October 2, 2014 Dear Ms. Pine, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Bram D. Zuckerman -S Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use ### 510(k) Number (if known) K141649 #### Device Name speX Support Catheter, 50cm, 300006; speX Support Catheter, 90cm, 300007; speX Support Catheter, 135cm, 300008; speX Support Catheter. 150cm. 300009 ### Indications for Use (Describe) The speX Support Catheter is intended to be used in conjunction with steerable guidewires to access discreet regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices and provide a conduit for the delivery of saline solutions or diagnostic/therapeutic agents Type of Use (Select one or both, as applicable) 스 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ## FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Image /page/2/Picture/15 description: The image shows the FDA logo. The logo is in a light gray color. The letters FDA are large and stylized. # Bram D. Zuckerman -S 2014.11.05 06:17:25 -05'00' This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | APPLICANT: | ReFlow Medical | |------------------------|---------------------------------------------------------------------------------------------------------------------| | DATE PREPARED: | October 1, 2014 | | CONTACT PERSON: | Rebecca K Pine<br>ReFlow Medical<br>1003 Calle Sombra<br>San Clemente, CA 92673<br>Phone: (760) 809.5178 | | TRADE NAME: | speX Support Catheter | | COMMON NAME: | Guide Catheter | | CLASSIFICATION NAME: | Percutaneous Catheter | | DEVICE CLASSIFICATION: | Class 2, per 21 CFR 870.1250 | | PRODUCT CODE | DQY | | PREDICATE DEVICES: | Volcano Valet Microcatheter (K112035)<br>Quickcross Support Catheter (K072750)<br>Minnie Support Catheter (K082337) | ## Substantially Equivalent To: The speX Support Catheter is substantially equivalent in intended use, principal of operation and technological characteristics to the Volcano Valet Microcatheter (K112035), QuickCross Support Catheter (K072750), and the Minnie Support Catheter (K082337). ## Description of the Device Subject to Premarket Notification: The speX Support Catheter is a device intended to provide additional support to a steerable guidewire when accessing discrete regions of the peripheral vasculature. The device is an over the wire single lumen catheter with a proximal female hub and a distal flexible part with a radiopaque tapered tip. The distal end of the catheter has a lubricious coating. The speX Support Catheter is 0.035" guidewire compatible and has working lengths of: 50cm, 90c, 135cm and 150cm. ## Indication for Use: The speX Support Catheter is intended to be used in conjunction with steerable guidewires to access discreet regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices and {4}------------------------------------------------ provide a conduit for the delivery of saline solutions or diagnostic/therapeutic agents. # Technological Characteristics: The speX Support Catheter has similar physical and technical characteristics to the predicate device. All devices are highly flexible with a stabilizing radiopaque distal tip. In the case of the speX Support Catheter and the Volcano Catheter, both tips are shapeable. All devices utilize a luer hub handle. All devices are fabricated from medical-grade polymers with varying stiffness shafts. All devices are provided sterile and are intended for single-use. A summary of the similarities and differences between the speX Support Catheter and the identified predicates are shown in the table below: | | speX Support<br>Catheter | QuickCross<br>Support Catheter<br>(K072750) | Valet Mirocatheter<br>(K112035) | Minnie Support<br>Catheter<br>(K082337) | |----------------------------|--------------------------------------------------------------------------|---------------------------------------------------------------------------------|--------------------------------------------------------------------------|--------------------------------------------------------------------------| | Function | Support catheter | Support catheter | Support catheter | Support catheter | | Principle of<br>Operation | Controlled penetration<br>to create a channel<br>across the occlusion | Controlled<br>penetration to create<br>a channel across the<br>occlusion | Controlled<br>penetration to create a<br>channel across the<br>occlusion | Controlled<br>penetration to create a<br>channel across the<br>occlusion | | Anatomical<br>Site | Peripheral | Peripheral<br>Coronary | Peripheral<br>Coronary | Peripheral<br>Coronary | | Delivery to<br>Site | Percutaneous access;<br>over the wire | Percutaneous access;<br>over the wire | Percutaneous access;<br>over the wire | Percutaneous access;<br>over the wire | | Visualization | Radiographic imaging | Radiographic<br>imaging | Radiographic imaging | Radiographic imaging | | Flexibility | Highly flexible | Highly flexible | Highly flexible | Highly flexible | | Working<br>length | 50cm<br>90cm<br>135cm<br>150cm | 65cm-150cm | 100cm-150cm | 90cm-150cm | | Crossing<br>Mechanism | Distal Tip;<br>Radiopaque;<br>tapered | Distal Tip;<br>Radiopaque marker<br>Tapered | Distal Tip;<br>Radiopaque<br>tapered | Distal Tip;<br>Radiopaque<br>tapered | | Crossing<br>method | Forward (advancement)<br>penetration of tapered<br>tip | Forward<br>(advancement)<br>penetration of<br>tapered tip | Forward<br>(advancement)<br>penetration of tapered<br>tip | Forward<br>(advancement)<br>penetration of tapered<br>tip | | Use interface | Luer as handle | Luer as handle | Luer as handle | Luer as handle | | Luer<br>connector | Yes | Yes | Yes | Yes | | Manually<br>operated | Yes | Yes | Yes | Yes | | Tip Material | Polymer | Polymer | Polymer | Polymer | | Distal Tip OD | .022" | 0.02" - 0.041" | 0.02" | .0205"-0.0415" | | Shapeable tip | Yes | No, available in<br>straight or angled | Yes | No, available straight | | Catheter<br>Material | Polymer;<br>varying stiffness shaft;<br>braid reinforced; PTFE<br>lining | Polymer:<br>varying stiffness<br>shaft: stainless steel<br>braid (some models); | Polymer; varying<br>stiffness shaft, metal<br>braid | Polymer; metal braid<br>incorporated into<br>shaft | | Hydrophilic<br>coating | Yes, distal 40cm | Yes | Yes, distal 30cm | Yes, distal 40cm | | | speX Support<br>Catheter | QuickCross<br>Support Catheter<br>(K072750) | Valet Mirocatheter<br>(K112035) | Minnie Support<br>Catheter<br>(K082337) | | Catheter OD | 4.6F | 3.0-4.8F | 1.8-3.5F | 2.9-4.8F | | Single use | Yes | Yes | Yes | Yes | | Guidewire<br>compatibility | 0.035" | 0.014" -0.035" | 0.014" -0.035" | 0.014" -0.035" | | Targeted<br>injection | Yes- saline solutions,<br>diagnostic/therapeutic<br>agents | Yes- saline solutions,<br>diagnostic agents | Yes- saline solutions,<br>contrast agents | Yes- saline solutions,<br>diagnostic/therapeutic<br>agents | {5}------------------------------------------------ # Performance Data: All necessary testing has been performed for the speX Support Catheter to assure substantial equivalence to the predicate device and demonstrate the device performs as intended. All testing was performed on test units representative of finished devices. The device design was qualified through the following tests: - Simulated Use (baseline, aged) ● - Tensile Strength (baseline, aged) ● - Torque (baseline, aged) ● - Pressure Test (baseline, aged) - Coating verification (baseline, aged) - Corrosion Test (baseline, aged) - Luer Verification (baseline, aged) - Dimensional verification and visual inspections (baseline, aged) ● - Sterilization Validation ● - Biocompatibility ● - Cytotoxicity . - l Irritation - Sensitization - Materials Mediated Pyrogenicity - I Acute Systemic Tox - Hemolysis (direct, indirect) - PTT - Complement Activation - I Thrombosis The speX Support Catheter met all specified criteria and did not raise new safety or performance questions. # Basis for Determination of Substantial Equivalence: The data and information presented in this submission and the similarities and differences between the subject and predicate devices support a determination of substantial equivalence.. The speX Support Catheter is determined by ReFlow Medical, to be substantially equivalent to the identified predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...