K130858 · Reverse Medical Corporation · DQY · Oct 11, 2013 · Cardiovascular
Device Facts
Record ID
K130858
Device Name
REVERSE MEDICAL MICROCATHETER
Applicant
Reverse Medical Corporation
Product Code
DQY · Cardiovascular
Decision Date
Oct 11, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Intended Use
The Reverse Medical Microcatheter-027 is intended for use in neuro, peripheral and coronary vasculature for the infusion of diagnostic agents such as contrast media, and therapeutic agents such as occlusion coils.
Device Story
The Reverse Medical Microcatheter-027 is a single-lumen, flexible, wire-reinforced composite catheter designed for accessing distal tortuous vasculature. It features a hydrophilic coating to reduce friction and dual radiopaque markers at the distal tip for fluoroscopic visualization. The device includes a shaping mandrel for physician-performed steam-shaping of the tip and a peel-away introducer. It is operated by physicians in clinical settings (neuro, peripheral, coronary). The catheter connects to standard luer-compatible accessories. By facilitating the delivery of diagnostic and therapeutic agents, the device assists in interventional procedures, potentially improving patient outcomes in vascular treatments.
Clinical Evidence
Bench testing only. No clinical data. Verification testing included biocompatibility (ISO 10993), sterilization validation (ISO 11135), and mechanical performance (dimensional, tip buckling, lubricity, flow rate, pressure, kink resistance, tensile strength, and navigation/pushability). All tests met acceptance criteria.
Technological Characteristics
Single-lumen, wire-reinforced composite catheter; variable stiffness; hydrophilic coating; dual radiopaque markers; luer adapter. Sterilized via EtO (SAL 10^-6). Complies with ISO 10993 for biocompatibility and ISO 11135 for sterilization.
Indications for Use
Indicated for use in neuro, peripheral, and coronary vasculature for infusion of diagnostic agents (contrast media) and therapeutic agents (occlusion coils).
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
K110813 — HEADWAY 27 MICROCATHETER · MicroVention, Inc. · Aug 5, 2011
K050599 — TRACKER EXCEL -14 PRE-SHAPED MICROCATHETER · Boston Scientific Corp · Jun 3, 2005
K122684 — REVERSE MEDICAL MICROCATHETER · Reverse Medical Corporation · Dec 7, 2012
K251668 — SurfRider 13 Microcatheter · Kaneka Americas Holding, Inc. · Jan 5, 2026
K252569 — Carrier XL Delivery Catheter · Balt USA, LLC · Oct 22, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
K130858
Image /page/0/Picture/1 description: The image shows the logo for Reverse Medical. The text "REVERSE MEDICAL" is in bold, black font. There is a curved line underneath the text that ends in an arrow pointing upwards.
# 510(k) Summary according to 807.92
Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Reverse Medical Corporation is providing the summary of Substantial Equivalence for the Reverse Medical Microcatheter-027.
Sponsor /Applicant Name and Address 5.1
Reverse Medical Corporation 13700 Alton Parkway, Suite 167 Irvine, CA 92618
> 5.2 Sponsor Contact Information
Linda D'Abate, Vice President RA/CA/QA Phone: 949-215-0660 x202 Mobile 714-235-6608 Fax: 949-215-0661 Email: Idabate@reversemed.com
> 5.3 Date of Preparation of 510(k) Summary
October 4, 2013
5.4 Device Trade or Proprietary Name
Reverse Medical Microcatheter-027
ર .5 Device Common/Usual or Classification Name
Catheter, Percutaneous (Product Code: DQY) (21 CFR.870. 1250)and Catheter, Infusion (Product Code KRA) (21 CFR 870.1210) and Diagnostic Intravascular Catheter (DQO, 21CFR870.1200)
Identification of the Legally Marketed Devices to which Equivalence is Being 5.6 Claimed:
| Name of Predicate Devices | Name of Manufacturer<br>(City, State) | 510(k)<br>Number |
|-----------------------------------|---------------------------------------|------------------|
| Headway™ 27 Microcatheter | MicroVention, Inc.<br>Tustin, CA | K110813 |
| Excelsior® XT-27™ Microcatheter | Stryker Neurovascular<br>Fremont, CA | K113778 |
| Reverse Medical Microcatheter-021 | Reverse Medical Corp.<br>Irvine, CA | K122684 |
#### 5.7 Device Description
The Reverse Medical Microcatheter-027 is a single lumen, flexible stiffness composite catheter. The catheter shaft has a hydrophilic coating to reduce friction during use. The Reverse Medical Microcatheter-027 dimensions are included on the individual device labels. The Reverse Medical Microcatheter-027 inner lumen can accommodate guidewires up to 0.025 inches inner diameter to
{1}------------------------------------------------
access distal tortuous vasculature. Dual radiopaque markers at the distal portion of the catheter facilitate fluoroscopic visualization.
Each Reverse Medical Microcatheter-027 is provided with accessories, which include a shaping mandrel and peel away introducer within a Tyvek™ pouch.
The shaping mandrel allows the catheter tip to be steam shaped by the physician for proper adjustment to the anatomy prior to use. Test data is presented in this submission for both the 027 and 021 Reverse Medical Microcatheters to support the inclusion of the shaping mandrel and the peel away introducer.
The Reverse Medical Microcatheter-027 incorporates a standard luer adapter to facilitate the attachment of accessories. The catheter is provided sterile, non-pyrogenic, and is intended for single use only.
#### 5.8 Intended Use
The Reverse Medical Microcatheter-027 is intended for use in neuro, peripheral, and coronary vasculature for the infusion of diagnostic agents such as contrast media, and therapeutic agents such as inclusion coils.
{2}------------------------------------------------
| | MicroVention<br>Headway™ 27<br>Microcatheter | Excelsior® XT-27™<br>Microcatheter | Reverse Medical<br>Microcatheter-021 | Reverse Medical<br>Microcatheter-027 |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Number | K110813 | K113778 | K122684 | TBD |
| Classification | Class II, DQY | Class II, DQY | Class II, DQY and KRA | Class II, DQY and KRA |
| Indication | Intended for use in<br>the peripheral, coro-<br>nary and neurovascu-<br>lature for the infusion<br>of diagnostic agents,<br>such as contrast<br>media and therapeutic<br>agents such as<br>occlusion coils. | Intended to assist in<br>delivery of diagnostic<br>agents (such as con-<br>trast media), thera-<br>peutic agents, and<br>non-liquid intervene-<br>tional devices (such as<br>stents) intended for<br>use in the neurovascu-<br>lature and with a<br>catheter of 0.027" I.D. | Intended for use in<br>neuro, peripheral and<br>coronary vasculature<br>for the infusion of<br>diagnostic agents such<br>as contrast media, and<br>therapeutic agents<br>such as occlusion coils. | Intended for use in<br>neuro, peripheral and<br>coronary vasculature<br>for the infusion of<br>diagnostic agents such<br>as contrast media, and<br>therapeutic agents<br>such as occlusion coils. |
| Shaft<br>Materials | Coaxial lumen braided<br>shaft variable stiffness<br>catheter with<br>radiopaque marker on<br>distal end. | Semi-rigid proximal<br>shaft that transitions<br>into the flexible distal<br>shaft with single or<br>dual radiopaque mark-<br>ers at the distal end. | Single lumen, wire<br>reinforced shaft,<br>variable stiffness<br>catheter with dual<br>radiopaque markers<br>on distal end. | Single lumen, wire<br>reinforced shaft,<br>variable stiffness<br>composite catheter<br>with dual radiopaque<br>markers on distal end. |
| Proximal End<br>Configuration | Luer Hub | Luer Hub | Luer Hub | Luer Hub |
| Radiographic<br>markers/radi<br>opacity | Dual radiopaque<br>markers at distal tip. | Single or dual<br>radiopaque markers at<br>distal end of shaft. | Dual radiopaque<br>markers at distal tip. | Dual radiopaque<br>markers at distal tip. |
| Packaging | Polyethylene hoop<br>and PET/PE/Tyvek<br>pouch inside SBS<br>carton. | Polyethylene hoop<br>and PET/PE/Tyvek<br>pouch inside SBS<br>carton. | Polyethylene hoop<br>and PET/PE/Tyvek<br>pouch inside SBS<br>carton. | Polyethylene hoop<br>and PET/PE/Tyvek<br>pouch inside SBS<br>carton. |
| Sterilization | EtO | EtO | EtO | EtO |
| Peel Away<br>introducer | | Yes | Yes (data presented in<br>this submission) | Yes |
| Catheter Tip<br>Shaping<br>Mandrel | Yes | Yes | Yes (data presented in<br>this submission) | Yes |
#### 5.9 Comparison to Predicate Devices
{3}------------------------------------------------
#### Summary of Non-Clinical Data 5.10
## 5.10.1 Biocompatibility and Sterilization
The Reverse Medical Microcatheter-027 is classified as an Externally Communicating Device, Circulating Blood, Limited Contact (s24 hours). Results of the testing demonstrate that the blood-contacting materials are biocompatible.
Blood-contacting materials were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO 10993-1 guidelines "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing." The Reverse Medical Microcatheter-027 successfully passed all of the following biocompatibility tests:
| Test | Results | Conclusion |
|----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| Cytotoxicity<br>L929 MEM Elution Test | MEM Elution test scored a grade 0 (No cell<br>lysis) per ISO 10993-5. | Non-Cytotoxic |
| Sensitization<br>Kligman Maximization | Sensitization test scored a grade 0 (no visible<br>change) per ISO 10993-10 | Non-Sensitizing |
| Systemic Toxicity (Acute)<br>ISO Acute Systemic<br>Injection Test | Acute Systemic Injection Test Articles scored<br>0 with no toxicity or animal weight loss for<br>both the cottonseed oil and saline extracts<br>per ISO 10993-11 | Non-Toxic |
| Hemocompatibility:<br>Complement Activation | Under the conditions of the C3a assay, the<br>test article exhibited activation at 9421<br>ng/ml. This was 5.8%of the normalized C3a<br>concentration produced by CVF. Under the<br>conditions of the SC5b-9 assay, the test<br>article exhibited activation at 8836 ng/ml.<br>This was 0.1%of the normalized SC5b-9<br>concentration produced by CVF. All<br>biomaterials have the potential to affect the<br>make-up of the complement activation<br>components of the blood. | No greater biological<br>response than<br>corresponding control |
| | | |
| Hemolysis | The test article with negative control exhibited a hemolytic grade score of zero and is considered non-hemolytic. | Non- Hemolytic |
| Inactivated Partial<br>Thromboplastin Time | The negative and positive controls met the criteria for a valid assay. The variance between duplicate readings was less than 15%. The clotting times were as follow:<br>Positive control: 107 seconds<br>Negative control: 300 seconds<br>Reference material: 300 seconds<br>Test Article: 290 seconds<br><br>The test article is considered to be a minimal activator of the intrinsic coagulation pathway. The test article was considered to pass the test. | Non Activator |
| <i>In vivo</i> Thrombogenicity | Implantation of the test and control devices resulted in no adverse effects or clinical signs. | Non-Thrombogenic |
| Pyrogenicity<br>USP Material Mediated<br>Rabbit Pyrogen Test | Material-mediated pyrogenicity test was non-pyrogenic with no individual rabbit at any time having a temperature rise of greater than or equal to 0.5 deg. C per ISO 10993-11 | Non- Pyrogenic |
| EtO Residuals<br>Ethylene Oxide and<br>Ethylene Chlorohydrins<br>Residuals | ETO residuals met the criteria in accordance with ISO 10993-7 Part 7 | Passed, acceptable limits |
| Genotoxicity/Bacterial<br>Reverse Mutation | A statistically significant increase in the number of revertant colonies was not observed with the test article extracts as compared to the negative controls in both non-activated and S9 activated conditions. | Non-Mutagenic |
| Genotoxicity - Mouse<br>lymphoma | The test article extracts (saline and DMSO)<br>did not produce significantly more revertant<br>colonies than the negative controls. | Non-Mutagenic |
| Genotoxicity - In vivo<br>micronucleus | There was no statistically significant increase<br>in the number of micronucleus in the test<br>article extract (saline and sesame oil) vs.<br>negative control group. | Non-Mutagenic |
{4}------------------------------------------------
{5}------------------------------------------------
Sterilization conditions have been validated according to ANSI / AAMI / ISO 11135, Sterilization of Health Care Products-Part 1: Requirement, for development, validation and routine control of a sterilization process for medical devices to provide a Sterility Assurance Level (SAL) of 10 °.
## 5.10.2 Design Verification (Bench-Top Testing)
The physical, mechanical, and performance testing of the Reverse Medical Microcatheter-027 demonstrate that the product is substantially equivalent to the currently marketed predicate devices. Design verification testing was conducted to evaluate the physical and mechanical properties of the Reverse Medical Microcatheter-027 with and without steam-shaping capabilities. All studies were conducted in accordance with Reverse Medical Design Control procedures. All testing was performed on units that were sterilized and met all inspection criteria. Tests on the Reverse Medical Microcatheter-027 included:
| Test Descriptions | Acceptance Criteria | Requirements | Pass/Fail |
|-------------------------------------|------------------------------------------------------------------------------------------------------------|--------------------------------------|-----------|
| Dimensional/Visual<br>Inspection | Per test protocol rev B | Pass/Fail | Pass |
| Tip Buckling Test | Peak Force Perform the same or better than<br>the predicate devices | 95/90 | Pass |
| Coating Lubricity Test | Frictional force<br>Coating Length | 95/90 | Pass |
| Flexibility/Shaft Stiffness<br>Test | The distal tip and proximal shaft flexibility<br>should perform comparably or better than the<br>predicate | Pass/Fail | Pass |
| Priming Volume Test | For comparison to predicate | Equal to or better than<br>predicate | Pass |
| Flow Rate Test | For comparison to predicate | Equal to or better than<br>predicate | Pass |
| Guidewire Compatibility<br>Test | Free movement of appropriately sized<br>guidewires<br>Frictional force | 95/90 | Pass |
{6}------------------------------------------------
| Guide Catheter<br>Compatibility Test | Free movement inside an appropriately sized<br>Guide Catheter. Frictional force measurement | 95/90 | Pass |
|-----------------------------------------------|---------------------------------------------------------------------------------------------|-----------|------|
| Catheter Air Leakage Test<br>-per ISO 10551-1 | Per ISO 10551-1 | Pass/Fail | Pass |
| Catheter Liquid Leakage<br>Test | No leaks in accordance with protocol | Pass/Fail | Pass |
| Dynamic Pressure Test | No leaks due to dynamic pressure test.<br>Peak Pressure: | Pass/Fail | Pass |
| Static Pressure Test | Burst pressure in accordance with protocol | 95/90 | Pass |
| Kink Resistance Test | Distal kink resistance<br>Medial/proximal: for characterization<br>purposes. | 95/90 | Pass |
| Torque Strength Test | X revolutions without failure. | Pass/Fail | Pass |
| Corrosion Resistance | No corrosion on metallic components. | Pass/Fail | Pass |
| USP Particulate Testing | Report total of particles. Compare to<br>predicate devices. | USP 788 | Pass |
| Tensile Strength | Distal/medial ; medial/proximal<br>Proximal/hub<br>Proximal/hub | 95/90 | Pass |
{7}------------------------------------------------
The physical, mechanical, and performance testing of the Reverse Medical Microcatheter-027 demonstrate that the product is safe and effective for its labeled indications and is Substantially Equivalent to the currently marketed predicate devices
| Test Descriptions | Acceptance Criteria | Requirement | Pass/Fail |
|-------------------------------------------|---------------------------------|-------------|-----------|
| Navigation/ Accessibility/<br>Pushability | Comparable to predicate devices | Pass/Fail | Pass |
| Micro-devices<br>Compatibility | Comparable to predicate devices | Pass/Fail | Pass |
In vitro Test Results for Reverse Microcatheter-027 Without Steam-Shaping Capabilities
## Performance/Functional Verification Test Results for Reverse Microcatheter-027 and 021 With Steam-Shaping Capabilities
| Test Descriptions | Acceptance Criteria | Requirements | Pass/Fail |
|-----------------------------------------------------|--------------------------------------------------|--------------|-----------|
| Dimensional and Visual<br>Inspection | Per test protocol rev B. | Pass/Fail | Pass |
| Steam Shaping Capabilities | No Damage<br>Shape Inspection: at various angles | Pass/Fail | Pass |
| Coating Lubricity Test - Post<br>Steam Shaping Test | Frictional force | Pass/Fail | Pass |
| Static Pressure Test - Post<br>Steam Shaping Test | Burst pressure | 95/90 | Pass |
| Tensile Strength - Post Steam<br>Shaping Test | Distal/Medial<br>Medial/Proximal<br>Proximal/Hub | 95/90 | Pass |
## In vitro Bench Test Results for Reverse Microcatheter-027 and 021 Post Steam-Shaping
| Navigation/ Accessibility/<br>Pushability | Comparable to predicate | Pass/Fail | Pass |
|-------------------------------------------|--------------------------------------------------------------------|-----------|------|
| Micro-devices Compatibility | Can allow uses of other micro-devices, in<br>simulated flow model. | | |
{8}------------------------------------------------
#### 5.11 Substantial Equivalence
The performance of the Reverse Medical Microcatheter-027 with and without steam-shaping capabilities in this submission demonstrates that the product is substantially equivalent to the performance of the predicate devices. The equivalence was shown through comparison of component materials and specifications, performance, biocompatibility testing, and sterilization validation.
The Reverse Medical Microcatheter-027 is substantially equivalent in intended use, design, technology/principles of operation, materials, and performance to the predicate devices. Differences between the devices do not raise any significant issues of safety or effectiveness.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure with three flowing lines extending from the figure's head, representing health, services, and the human aspect. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the emblem.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G6(19 Silver Spring, MI) 20993-0002
October 11, 2013
Reverse Medical Corporation Mr. Jeffrey Valko President/Chief Executive Officer 13700 Alton Parkway, Suite 167 Irvine. CA 92618
Re: K130858
Trade/Device Name: Reverse Medical Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Catheter, Perculancous Regulatory Class: Class II Product Code: DQY. KRA and DQO Dated: September 6, 2013 Received: September 6, 2013
Dear Mr. Valko:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA `s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any with all the Act 's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical
{10}------------------------------------------------
Page 2 - Jeffrey Valko
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYowIndustry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Joyce M. Whang -S
for Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
## Indications for Use
510(k) Number: K130858
Device Name: Reverse Medical
Indications For Use:
The Reverse Medical Microcatheter-027 is intended for use in neuro, peripheral and coronary vasculature for the infusion of diagnostic agents such as contrast media, and therapeutic agents such as occlusion coils.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
# Joyce M. Whang -S
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.