K042117 · Velocimed, Inc. · DQY · Jan 7, 2005 · Cardiovascular
Device Facts
Record ID
K042117
Device Name
PROXIS SYSTEM, MODEL EPS 101
Applicant
Velocimed, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Jan 7, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Proxis System controls the flow of fluids in the coronary and periphery vasculature. This is achieved by the temporary occlusion of vessels and holding the column of fluid in the vessel stagnant. The stagnant column can be used to aid in the visualization of the lesion or be used as a means of local and temporary delivery of therapeutic solution(s). The safety and efficacy of this device as an embolic protection system has not been established. The Proxis Flow Control device is not indicated for use for embolic protection.
Device Story
Proxis System comprises Evacuation Sheath Catheter, Inflation System, Infusion Catheter, and Evacuation syringe. Sheath features two low-pressure compliant sealing balloons; proximal balloon remains in guide catheter; distal balloon resides in arterial vessel. Radiopaque markers assist placement. Physician inflates balloons to occlude vessel, creating stagnant fluid column. Stagnant column enables contrast dye visualization of lesions or localized delivery of therapeutic agents (anticoagulants, cardioplegia, thrombolytics). Evacuation syringe and infusion catheter facilitate aspiration of fluid and emboli/thrombi. Used in clinical settings by physicians. Benefits include improved lesion visualization and targeted drug delivery.
Clinical Evidence
Bench testing only. Performance verification included dimensional analysis, balloon compliance and integrity, tensile strength, torque strength, flexibility, and trackability. Device met all product specifications and applicable standards (ISO 10555, ISO 10993, ASTM D-4169, ISO 11607, ISO 11135).
Technological Characteristics
Intravascular catheter system with dual low-pressure compliant sealing balloons. Features hydrophilic coating. Radiopaque markers for visualization. Complies with ISO 10555 (sterile, single-use, balloon dilation), ISO 10993 (biocompatibility), and ISO 11135 (ETO sterilization). Mechanical operation via inflation system.
Indications for Use
Indicated for fluid flow control in coronary and peripheral vasculature via temporary vessel occlusion to facilitate lesion visualization or local delivery of therapeutic solutions. Not indicated for embolic protection.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
K063638 — MODIFICATION TO PROXIS SYSTEM, MODEL EPS 101 · St Jude Medical · May 11, 2007
K073700 — 6F PROXIS SYSTEM · St. Jude Medical, Inc. · Oct 2, 2008
K023548 — PROXIS FLOW-CONTROL DEVICE, MODELS EPS 101, IC 101 · Velocimed · Aug 4, 2003
K060651 — PROXIS SYSTEM, MODEL EPS101 · St Jude Medical · Sep 13, 2006
K052523 — PROXIS SYSTEM · Velocimed, Inc. · Sep 7, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Velocimed. The logo consists of a stylized image of a circle with three curved lines inside it, and a small dot below the circle. Below the image is the word "Velocimed" in a sans-serif font, with the trademark symbol to the right of the word.
11400 73rd Ave North, Suite 134 Minneapolis, MN 55369
K0421.17
763-463-4742 · Phone 763-488-9780 • Fax
### 510(K) SUMMARY SECTION 3.
#### 3.1 ADMINISTRATIVE INFORMATION
#### Name and address 3.1.1
| Submitted by: | Velocimed Inc<br>11400 73rd Avenue North, Suite 134<br>Minneapolis, MN 55369 |
|---------------|------------------------------------------------------------------------------|
|---------------|------------------------------------------------------------------------------|
| Contact Person: | John Carline |
|-----------------|--------------|
| Telephone No.: | 763-463-4742 |
| Facsimile No.: | 763-488-9780 |
Jan 4, 2005 Date Prepared:
# 3.1.2 Device Name
| Trade Name | Proxis System |
|---------------------|------------------------|
| Common Name | Percutaneous Catheter |
| Classification Name | Catheter, Percutaneous |
| Classification | Class II, DQY |
#### 3.2 SUBSTANTIAL EQUIVALENCE
The Proxis System with hydrophilic coating, the subject of this submission, is substantially equivalent to the Proxis System without hydrophilic coating (predicate device) which had been found to be substantially equivalent through the 510(k) notification process (K023548).
The Proxis System with hydrophilic coating is the same as the predicate device except for the addition of the hydrophilic coating.
The addition of the hydrophilic coating does not alter the intended use the Proxis System.
#### 3.3 INDICATION FOR USE
The Proxis System controls the flow of fluids in the coronary and periphery vasculature. This is achieved by the temporary occlusion of vessels and holding the column of fluid in the vessel stagnant. The stagnant column can be used to aid in the visualization of the lesion or be used as a means of local and temporary delivery of therapeutic solution(s). The safety and efficacy of this device as an embolic protection system has not been established. The Proxis Flow Control device is not indicated for use for embolic protection.
{1}------------------------------------------------
#### 3.4 DEVICE DESCRIPTION
The Proxis System has four major components: The Evacuation Sheath Catheter, the Inflation System, Infusion Catheter, and an Evacuation syringe.
The Evacuation Sheath Catheter has two low-pressure compliant sealing balloons that are inflated simultaneously. The proximal balloon stays within the guide catheter while the distal balloon resides in the arterial vessel. Radiopaque markers at the two balloon sites facilitate visualization and intravascular placement of the catheter prior to inflation. The Evacuation Sheath has sufficiently large inner diameter to accommodate standard therapeutic devices within its size range. The balloons are inflated using the Inflation Svstem.
Devices can be deployed through the Evacuation sheath to the target site before, during or after the sealing balloons are inflated to occlude the vessel. Infusing ~0.5cc of contrast dye through the guide catheter will produce a continuous "roadmap" of the lesion as an aid for the physician in guiding the therapeutic device to the lesion site.
Alternatively, while the vessel is occluded, therapeutic solutions like anticoagulant, cardioplegia and thrombolytics may be infused through the guide catheter and stagnated in the target vessel/lesion during the delivery of the therapeutic device or after the deployment of the therapeutic device.
The Evacuation syringe is provided for the removal of the fluid and the emboli/thrombi during aspiration. The infusion catheter may be used to infuse saline to augment the retrograde flow of fluid and the removal the emboli/thrombi.
#### 3.5 Performance Data
The Proxis System has been shown to meet the performance requirements of the product specification. Additionally, the Proxis System complies with the applicable section of the following product standards:
- ISO 10555 (Sterile, single-use intravascular catheter Part 1: General requirements . and Part 4: balloon dilation catheter)
- ISO 10993 (Biological evaluation of medical devices) .
- ASTM D-4169 (Packaging Integrity Testing) .
- ISO 11607 (Packaging for terminally sterilized medical devices) .
- ISO 11135: 1994(E) (Validation and routine control of ETO sterilization) .
Performance testing included dimensional verification, balloon compliance and integrity, catheter tensile strength, torque strength, flexibility and trackability. Test results demonstrate that the device meets or exceeds the requirements of these standards and performs substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN - 7 2005
Velocimed Inc. c/o Mr. John Carline Senior RA Specialist 11400 73td Avenue North, Suite 134 Minneapolis, MN 55369
> K042117 Proxis System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DOY Dated: December 21, 2004 Received: December 28, 2004
### Dear Mr. Carline:
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 – Mr. John Carline
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualles of a succession complies with other requirements of the Act
that FDA has made a determination that your device complies with other mast that FDA has made a decionmation an administered by other Federal agencies. You must of any Federal statures and regulations daminitives but not limited to: registration and listing (21 comply with the Act 3 requirements, mercesses canafacturing practice requirements as set CFK Part 807), labeling (21 CFR Part 820); good if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Decizens over device as described in your Section 510(k) I his letter will anow you to oegin maneting your antial equivalence of your device to a legally premarket notification. The PDA midning of basification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acrive for your acon 2011-01-20120. Also, please note the regulation entitled, Contact the Office of Obilipinance an (21 transmit (21CFR Part 807.97). You may obtain Misolaliung by reference to premainterial international the Act from the Division of Small other general International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Donna R. Lochner
( / Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Attachment A. Indications for use
510(k) Number: ¥042117
Device Name: Proxis System
### Indication for Use:
The Proxis System controls the flow of fluids in the coronary and periphery vasculature. This is achieved by the temporary occlusion of vessels and holding the column of fluid in the vessel stagnant. The stagnant column can be used to aid in the visualization of the lesion or be used as a means of local and temporary delivery of therapeutic solution(s). The safety and efficacy of this device as an embolic protection system has not been established. The Proxis Flow Control device is not indicated for use for embolic protection.
Prescription Use: X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use: (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Duna R. Willing
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number_k642117
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.