6F MACH1/34356-XXX; 5F MACH1/35845-XXX; 7F MACH1/34357-XXX; 8F MACH1/34358-XXX
K020028 · Boston Scientific · DQY · Jan 22, 2002 · Cardiovascular
Device Facts
| Record ID | K020028 |
| Device Name | 6F MACH1/34356-XXX; 5F MACH1/35845-XXX; 7F MACH1/34357-XXX; 8F MACH1/34358-XXX |
| Applicant | Boston Scientific |
| Product Code | DQY · Cardiovascular |
| Decision Date | Jan 22, 2002 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 870.1250 |
| Device Class | Class 2 |
Intended Use
Mach1 Guide Catheters are intended for use in general intravascular and coronary applications. They provide a pathway through which medical instruments, such as balloon dilatation catheters, guide wires or other therapeutic devices may be introduced. These devices are not intended for use in the cerebral vasculature.
Device Story
Mach1 Guide Catheters provide a pathway for introducing medical instruments, such as balloon dilatation catheters and guide wires, into the coronary or peripheral vasculature. The device is a multi-layered catheter (inner, middle, outer) available in 5F, 6F, 7F, and 8F sizes with lengths of 40-125 cm. It features varying durometers of Arnitel (polyether-ester) to provide specific flexibility zones. The tip is reinforced or non-reinforced, constructed from Hytrel (6F, 7F, 8F) or Arnitel (5F). Used by physicians in clinical settings for interventional procedures; the device is sterile and single-use. It facilitates the delivery of therapeutic devices to target sites, aiding in coronary or peripheral interventions. It is not for use in cerebral vasculature.
Clinical Evidence
Bench testing only. Testing included shaft and distal segment tensile and elongation, tip tensile, tip deflection, hub-to-shaft tensile, pressure shaft burst and leak, torque response, lumen integrity, radiopacity, biocompatibility, package pouch burst, and shelf life. Results verified the devices are adequate for their intended use.
Technological Characteristics
Multi-layered catheter shaft (inner, middle, outer). Materials: Arnitel (polyether-ester) for shaft segments; Hytrel or Arnitel for distal tips. Sizes: 5F, 6F, 7F, 8F. Lengths: 40-125 cm. Features: varying durometer flexibility zones, optional side holes. Sterilization: sterile, single-use.
Indications for Use
Indicated for use in general intravascular and coronary applications to provide a pathway for medical instruments (e.g., balloon dilatation catheters, guide wires). Not intended for use in the cerebral vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
- Boston Scientific Scimed 6F Mach1 Guide Catheter (K010874)
Related Devices
- K033441 — BOSTON SCIENTIFIC 6F RUNWAY GUIDE CATHETER · Boston Scientific Corp · Nov 21, 2003
- K230156 — 5F Launcher Guide Catheter, 6F Launcher Guide Catheter, 7F Launcher Guide Catheter, 8F Launcher Guide Catheter · Medtronic Vascular · Jun 30, 2023
- K132409 — CONCIERGE GUIDING CATHETER · Merit Medical Systems, Inc. · Aug 29, 2013
- K153703 — Guiding Catheter · Texasmedical Technologies, Inc. · May 12, 2016
- K982883 — MEDTRONIC 7F AND 8F ZUMA GUIDING CATHETER · Medtronics Interventional Vascular · Sep 4, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
JAN 2 2 2002
K0200228
## Section 6
## Summary of Safety and Effectiveness
# (Pursuant To Section 12 of the SAFE MEDICAL DEVICES ACT of 1990)
#### General Provisions 6.1
| Submitter's Name<br>and Address | Boston Scientific Scimed, Inc.<br>One Scimed Place<br>Maple Grove, Minnesota 55311 |
|---------------------------------|------------------------------------------------------------------------------------|
| Contact Person | Todd Kornmann<br>(763) 494-2467 |
| Classification Name | Similar to Diagnostic Intravascular<br>Catheters (21CFR Part 970.1200) |
| Common or Usual Name | Coronary or Peripheral Guide Catheter |
| Proprietary Name | Boston Scientific 5F, 7F, and 8F Mach1<br>Guide Catheters |
| Name of Predicate Device | Boston Scientific Scimed |
Boston Scientific Scimed 6F Mach1 Guide Catheter
#### Device Description 6.3
6.2
The Boston Scientific 5F, 7F, and 8F Mach1 catheters are similar to the currently marketed 6F Mach1 Guide Catheter, with the addition of three French sizes to the catheter line.
The Boston Scientific Mach1 catheters are designed to provide a pathway through which medical instruments, such as balloon dilatation catheters, guide wires or other diagnostic or therapeutic devices, may be introduced. These devices are not intended for use in the cerebral vasculature. The Mach1 catheters will be offered in 5F, 6F, 7F, and 8F sizes and are available in lengths ranging from 40-125 cm, with optional side holes. The devices are provided sterile and intended for one procedure use only.
The Mach1 catheters are manufactured using similar construction techniques as other currently marketed Boston Scientific guide catheters. The catheter shaft consists of three layers: the inner, middle and outer layers. The outer layer is composed of four segments,(5F has three segments) inclusive of the proximal shaft segment, but not including the distal tip. The outer layer segments are composed of the same material,
{1}------------------------------------------------
(Arnitel) but use varying durometers of the Arnitel (polyether-ester) shaft materials providing specific flexibility zones over the length of the catheter. The tip consists of a reinforced tip portion and a non-reinforced tip portion. The 6F, 7F and 8F tip is made of Hytrel, while the 5F has an Arnitel tip.
#### 6.4 Intended Use
The Mach Guide Catheters are intended for use in general intravascular and coronary applications designed to provide a pathway through which medical instruments, such as balloon dilatation catheters, guide wires or other therapeutic devices may be introduced. These devices are not intended for use in the cerebral vasculature.
#### Summary of Technological Characteristics ર્ણને સિંહ સિંહ સાંદર્ભ સાંત કર્યા હતા. તાલુકામાં આવેલું એક ગામનાં છે. તે છે. તે છે. આ ગામનાં છે. આ ગામનાં છે. તે છે. આ ગામનાં છે. આ ગામનાં છે. આ ગામનાં છે. આ ગામનાં તાલુકામ
Same as currently marketed 6F Mach1 guide catheters, cleared for market under K010874, June 21, 2001.
#### Non-clinical Test Summary ર્ભ રહ્યું હતું. હ
Functional testing consisted of shaft and distal segment tensile and elongation, tip tensile, tip deflection, hub to shaft tensile, pressure shaft burst and leak, torque response, lumen integrity and radiopacity. Biocompatibility, package pouch burst, and product shelf life testing have also been conducted. Test results verified that the 5F, 7F and 8F Mach1 catheters are adequate for the intended use. The 5F, 7F and 8F Mach1 guide catheters are considered substantially equivalent to the currently marketed 6F Mach 1 guide catheter based on a comparison of the intended use, the device design, and the results of in-vitro testing and evaluation.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 2 2002
Mr. Todd Kornmann Sr. Regulatory Affairs Specialist Boston Scientific Scimed, Inc. One Scimed Place Maple Grove MN 55311
Re: K020028
> Trade/Device Name: 5F, 7F and 8F Mach1 Guide Catheters Regulation Number: 21 CFR 870.1250 Regulation Name: Catheter, Percutaneous Regulatory Class: Class II Product Code: DQY Dated: January 3, 2002 Received: January 4, 2002
Dear Mr. Kornmann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Todd Kornmann
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Of R Fart 877) would by systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally presided predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the O inter gollet a mall Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
J. O'Neill
Bram Zuckerman, M.D.
kerman. M. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Special 510(k) Notification Boston Scientific Scimed, Inc. 5F, 7F and 8F Mach1 Guide Catheter
### Section 4
Indications for Use
K020028 510(k) Number (if known) _
Device Name: 5F, 7F, and 8F Mach1 Guide Catheters
Indications for Use:
Mach1 Guide Catheters are intended for use in general intravascular and coronary Machi Gurac Cathersvide a pathway through which medical instruments, such as applications. They provide a painting and other therapeutic devices may be introduced. These devices are not intended for use in the cerebral vasculature.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Prescription Use | <span style="text-decoration: overline;">✓</span> |
|--------------------------------------------------|-----------------------------------------------------------------------|
| (Per 21 CFR 801.109) | |
| OR | Over The Counter Use <span style="text-decoration: overline;"></span> |
| | (Optional Format 1-2-96) |
| Division of Cardiovascular & Respiratory Devices | |
| 510(k) Number | K00028 |
Concurrence of CDRH, Office of Device Evaluation (ODE)