K101479 · Reflow Medical · DQY · Aug 11, 2010 · Cardiovascular
Device Facts
Record ID
K101479
Device Name
WINGMAN EXTENDABLE TIP SUPPORT CATHETER
Applicant
Reflow Medical
Product Code
DQY · Cardiovascular
Decision Date
Aug 11, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Intended Use
The Wingman Extendable Tip Support Catheter is intended to be used in conjunction with steerable guidewires to access discreet regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices and provide a conduit for the delivery of saline solutions or diagnostic/therapeutic agents.
Device Story
The Wingman Extendable Tip Support Catheter is a percutaneous catheter designed to provide support to steerable guidewires during peripheral vascular procedures. The device features a support catheter body, a concealed radiopaque beveled guide-tip, and an activating handle. The through-lumen allows for guidewire placement/exchange and the delivery of diagnostic or therapeutic agents and saline. Used by physicians in clinical settings, the device assists in navigating complex vascular anatomy. By providing enhanced support, it facilitates the successful placement of interventional tools, potentially improving procedural efficiency and patient outcomes in peripheral vascular interventions.
Clinical Evidence
No clinical data. Substantial equivalence was demonstrated through bench testing and comparative simulated use testing on representative finished devices.
Technological Characteristics
Percutaneous catheter featuring a concealed radiopaque beveled guide-tip and an activating handle. Provides a through-lumen for guidewire support and fluid delivery. Class II device (21 CFR 870.1250).
Indications for Use
Indicated for use in patients requiring access to discrete regions of the peripheral vasculature to facilitate guidewire/interventional device placement, exchange, and delivery of diagnostic/therapeutic agents or saline.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
K120178 — WINGMAN EXTENDABLE TIP SUPPORT CATHETER · Reflow Medical · Feb 8, 2012
K193596 — Wingman 14, Wingman 14C, Wingman 18, Wingman 35 · Reflow Medical · Feb 20, 2020
K170996 — Wingman 14C Crossing Catheter · Reflow Medical, Inc. · Jun 9, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
#### 6. 510(k) Summary of safety and effectiveness
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT: | ReFlow Medical | |
|---------------------------|-----------------------------------------------------------------------------------------------------------|--------------|
| TRADE NAME: | Wingman Extendable Tip Support Catheter | |
| COMMON NAME: | Guide Catheter | |
| CLASSIFICATION<br>NAME: | Percutaneous Catheter | AUG 1 1 2010 |
| DEVICE<br>CLASSIFICATION: | Class 2, per 21 CFR 870.1250 | |
| PRODUCT CODE | DQY | |
| PREDICATE DEVICES: | Outback Catheter (K001577)<br>QuickCross Support Catheter (K 072750)<br>Gopher Support Catheter (K091345) | |
#### Substantially Equivalent To:
The Wingman Extendable Tip Support Catheter is substantially equivalent in intended use, principal of operation and technological characteristics to the Outback Catheter, the QuickCross Support Catheter and the Gopher Support Catheter.
#### Description of the Device Subject to Premarket Notification:
The Wingman Extendable Tip Support Catheter is a device intended to provide additional support to a steerable guidewire when accessing discrete regions of the peripheral vasculature. The device consists of a support catheter, with a concealed radiopaque beveled guide-tip, and activating handle. The through-lumen of the device can serve as a conduit for the delivery of diagnostic and therapeutic agents.
#### Indication for Use:
The Wingman Extendable Tip Support Catheter is intended to be used in conjunction with steerable guidewires to access discreet regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices and provide a conduit for the delivery of saline solutions or diagnostic/therapeutic agents.
#### Technical Characteristics:
The Wingman Extendable Tip Support Catheter has similar physical and technical characteristics to the predicate devices.
| ReFlow Medical | Page 11 of 70 | |
|--------------------------------------------|---------------|------------------------|
| Wingman Extendable Tip Support<br>Catheter | | Premarket Notification |
{1}------------------------------------------------
#### Performance Data:
All necessary verification and validation testing has been performed for the Wingman Extendable Tip Support Catheter to assure substantial equivalence to the predicate devices and demonstrate the device performs as intended. All testing was performed on test units representative of finished devices. Comparative simulated use testing demonstrated that the Wingman Extendable Tip Support Catheter is substantially equivalent to the predicate devices.
#### Basis for Determination of Substantial Equivalence:
Upon reviewing the safety and efficacy information provided in this submission and comparing intended use, principle of operation and overall technological characteristics. the Wingman Extendable Tip Support Catheter is determined by ReFlow Medical, to be substantially equivalent to existing legally marketed devices.
ReFlow Medical Wingman Extendable Tip Support Catheter
Page 12 of 70 Premarket Notification
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993=0002
Reflow Medical c/o Ms. Rebecca K. Pine Head of Regulatory Affairs/Quality Assurance 5311 Foxhound Way San Diego, CA 92130
AUG 1 1 2010
Re: K101479
Trade Name: Wingman Extendable Support Catheter Regulation Number: 21 CFR§ 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: July 15, 2010 Received: July 16, 2010
Dear Ms. Pine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Rebecca K. Pine
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
5. Indications for Use Statement
# K 101479
## INDICATIONS FOR USE STATEMENT
AUG 1 1 2010
510(k) Number (if known):
Device Name: Wingman Extendable Tip Support Catheter
Indications for Use:
The Wingman Extendable Tip Support Catheter is intended to be used in conjunction with steerable guidewires to access discreet regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices and provide a conduit for the delivery of saline solutions or diagnostic/therapeutic agents.
AND/OR
Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page of
(Division Sign-Off)
Division of Cardiovascu
510(k) Number K101479
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.