K193596 · Reflow Medical · DQY · Feb 20, 2020 · Cardiovascular
Device Facts
Record ID
K193596
Device Name
Wingman 14, Wingman 14C, Wingman 18, Wingman 35
Applicant
Reflow Medical
Product Code
DQY · Cardiovascular
Decision Date
Feb 20, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Wingman Crossing Catheters (14/14C/18/35) are intended to be used in conjunction with steerable guidewires to access discrete regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices, including facilitation of the intraluminal placement of diagnostic/ interventional devices beyond peripheral stenotic lesions (including chronic total occlusions [CTOs]) and provide a conduit for delivery of saline solutions or diagnostic/therapeutic agents. The Wingman 14C Crossing Catheter is intended to be used in conjunction with steerable guidewires to access discrete regions of the coronary vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices and provide a conduit for delivery of saline solutions or diagnostic/therapeutic agents.
Device Story
Wingman Crossing Catheters are support catheters used in peripheral (all models) and coronary (14C model) interventions. Device consists of a support catheter, a concealed radiopaque beveled guide-tip, and an activating handle. Operated by physicians during percutaneous procedures; device is tracked over a steerable guidewire to the target lesion. Activating handle extends/retracts the beveled tip to facilitate crossing of stenotic lesions or chronic total occlusions (CTOs). Through-lumen allows guidewire exchange and delivery of saline or diagnostic/therapeutic agents. Provides mechanical support to guidewires; assists in navigating complex vascular anatomy. Benefits include improved success rates in crossing challenging lesions where standard guidewires fail.
Clinical Evidence
Prospective, multi-center, single-arm study (n=85 subjects, 86 CTO lesions). Primary effectiveness endpoint: 89.5% (77/86) success rate in crossing CTOs (lower 95% CI 82.5%). Primary safety endpoint: 4.8% (4/83) 30-day rate of MAE, perforation, embolization, or dissection (upper 95% CI 10.7%). Study met both primary endpoints.
Technological Characteristics
Polymer catheter shaft; concealed radiopaque beveled guide-tip; activating handle; through-lumen for guidewire/agent delivery. Compatible with 0.014" to 0.035" guidewires. Over-the-wire percutaneous delivery. No software or energy source.
Indications for Use
Indicated for patients requiring access to discrete regions of peripheral vasculature (all models) or coronary vasculature (14C model) to facilitate guidewire/device placement, exchange, and delivery of diagnostic/therapeutic agents or saline. Contraindicated for coronary and cerebral use (14/18/35 models) and cerebral use (14C model).
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 20, 2020
Krystal Santiago Director RA/QA 208 Avenida Fabricante #100 San Clemente, California 92672
Re: K193596
Trade/Device Name: Wingman 14, Wingman 14C, Wingman 18, Wingman 35 Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: PDU, DOY Dated: December 20, 2019 Received: December 23, 2019
Dear Krystal Santiago:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Interventional Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193596
Device Name Wingman Crossing Catheters
#### Indications for Use (Describe)
The Wingman Crossing Catheters (14/14C/18/3) are intended to be used in conjunction with steerable guidewires to access discrete regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices, including facilitation of the intraluminal placement of diagnostic/ interventional devices beyond peripheral stenotic lesions (including chronic total occlusions [CTOs]) and provide a conduit for delivery of saline solutions or diagnostic/therapeutic agents.
The Wingman 14C Crossing Catheter is intended to be used in conjunction with steerable guidewires to access discrete regions of the coronary vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices and provide a conduit for delivery of saline solutions or diagnostic/therapeutic agents.
The Wingman 14, 18, & 35 Crossing Catheter is contraindicated for use in the coronary and cerebral vasculature. The Wingman 14C Crossing Catheter is contraindicated for use in the cerebral vasculature.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> </span> |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
| Submitter | Reflow Medical, Inc.<br>208 Avenida Fabricante<br>#100<br>San Clemente, CA 92672<br><br>Contact person: Krystal Santiago<br>Phone: (310) 707.5882 |
|--------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared | February 8, 2020 |
| Device | Name of the device: Wingman Crossing Catheter<br>Common of usual name: Percutaneous Catheter<br>Classification name: Percutaneous Catheter<br>Regulatory Class: 2<br>Product Code: PDU, DQY |
| Legally marketed device<br>to which your firm is<br>claiming equivalence | Predicate Device: Medtronic Viance Crossing Catheter K120533<br>Reference Devices: Wingman Crossing Catheter<br>K132420 (Wingman 14)<br>K190393 (Wingman 14C)<br>K160848 (Wingman 18)<br>K173661 (Wingman 35) |
| Description of the device | The Wingman Crossing Catheter (14/18/35) is a device intended to<br>provide additional support to a steerable guidewire when accessing<br>discrete regions of the peripheral vasculature.<br>The Wingman Crossing Catheter (14C) is a device intended to provide<br>additional support to a steerable guidewire when accessing discrete<br>regions of the coronary vasculature and peripheral vasculature.<br><br>The device consists of a support catheter, with a concealed radiopaque<br>beveled guide-tip, and activating handle. The through-lumen of the<br>device can serve as a conduit for the delivery of diagnostic and<br>therapeutic agents. |
| Intended use of the<br>device | The Wingman Crossing Catheters (14/14C/18/35) are intended to be used<br>in conjunction with steerable guidewires to access discrete regions of the<br>peripheral vasculature. It may be used to facilitate placement and<br>exchange of guidewires and other interventional devices, including<br>facilitation of the intraluminal placement of diagnostic/ interventional<br>devices beyond peripheral stenotic lesions (including chronic total<br>occlusions [CTOs]) and provide a conduit for delivery of saline solutions<br>or diagnostic/therapeutic agents.<br><br>The Wingman 14C Crossing Catheter is intended to be used in<br>conjunction with steerable guidewires to access discrete regions of the<br>coronary vasculature. It may be used to facilitate placement and<br>exchange of guidewires and other interventional devices and provide a<br>conduit for delivery of saline solutions or diagnostic/therapeutic agents.<br>The Wingman 14/18/35 Crossing Catheters are contraindicated for use in<br>the coronary and cerebral vasculature. The Wingman 14C Crossing<br>Catheter is contraindicated for use in the cerebral vasculature. |
{4}------------------------------------------------
### Summary of the technological characteristics of your device compared to the predicate device
The technological characteristics of the subject Wingman Crossing Catheter are identical to the technological characteristics of the Wingman Crossing Catheter previously cleared K132420 (Wingman 14), K190393 (Wingman 14C), K160848 (Wingman 18), K173661 (Wingman 35) version of the device. The Wingman device is clinically substantially equivalent to the Medtronic Viance Crossing Catheter (K120533).
At a high level, the subject and predicate devices are based on the following same technological elements:
- all delivered to the target site using an over-the-wire percutaneous technique
- all have a through lumen to allow passage and exchange of guidewires
- all have a smooth inner lumen to provide reduced friction for guidewire movement ●
- all have a polymer catheter shaft with specific geometry to control the torque and push ● movements associated with lesion crossing
- . all use a specialized distal tip to facilitate crossing of the lesion
The following technological differences exist between the subject and predicate devices:
- . The subject Wingman device has a slightly differing distal tip configuration in comparison to the Medtronic Viance catheter, however both tip designs have been shown to achieve the same clinical outcomes.
- The subject Wingman Crossing Catheters are available in a wider product range encompassing ● 0.014-0.035" guidewire compatibility vs the standard 0.014" Medtronic Viance Catheter.
### A brief discussion of the nonclinical tests submitted
No additional nonclinical tests are submitted.
## A brief discussion of the clinical data submitted
This was a non-randomized, prospective, multi-center, single-arm study designed to evaluate the safety and effectiveness of the Wingman device to cross CTOs in infrainguinal arteries. The study was conducted both in the United States and outside the United States in Germany and Austria.
At baseline, the mean age of the study subjects was 71.4 ± 9.3 years, and 65.9% (56/85) was male. The mean body mass index (BMI) was 27.7 ± 5.3. A history of smoking was elicited in 76.5% (65/85) of subjects, and 30.6% (26/85) diabetic. A history of peripheral interventions was reported in 63.5% (54/85) of subjects and the most common Rutherford Category was 3, present in 70.6% (60/85) of subjects.
Among 85 subjects with 86 CTO lesions that could not be traversed with standard guidewires and techniques, 77 (89.5%) were successfully crossed with the Wingman study device. The lower 95% confidence limit for successful CTO crossing (the primary effectiveness endpoint) was 82.5%, meeting the primary effectiveness endpoint.
These results were achieved with a 4.8% (4/83) 30-day rate of MAE, perforation, embolization, or dissection; the primary safety endpoint. The upper 95% confidence limit for the primary safety endpoint was 10.7%, meeting the primary safety endpoint.
Both the primary endpoints were met, suggesting that the Wingman study device was safe and effective for crossing CTO lesions unable to be traversed with standard guidewires and techniques.
Conclusions
The conclusions drawn from the nonclinical and clinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed devices identified.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.