Zenith Support, 95 cm, Zenith Support, 105 cm, Zenith Support, 115 cm
K173709 · Inneuroco, Inc. · DQY · Jan 26, 2018 · Cardiovascular
Device Facts
Record ID
K173709
Device Name
Zenith Support, 95 cm, Zenith Support, 105 cm, Zenith Support, 115 cm
Applicant
Inneuroco, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Jan 26, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The Zenith Support Catheter is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.
Device Story
Zenith Support Catheter is a single-lumen, variable-stiffness catheter designed for peripheral and neurovascular access. It features a polymer shaft reinforced with a stainless steel cross-coil, a PTFE-lined lumen, and a hydrophilic coating. A radiopaque platinum/iridium marker band is located at the distal end. The device is inserted through a guide catheter or vascular sheath to provide a reinforcing conduit for other intravascular devices. It is supplied with a Tuohy-Borst hemostasis valve and a split-sheath introducer. Used in clinical settings by physicians, the device facilitates the delivery of interventional tools. By providing a stable, lubricious pathway, it aids in navigating complex vasculature, potentially improving procedural success and reducing vessel trauma.
Clinical Evidence
Bench testing only. No clinical data presented. Non-clinical evaluation included biocompatibility (ISO 10993), radiographic detectability, visual inspection, working length, hub compatibility, leak testing (air/liquid), particulate testing, simulated use, burst pressure, tensile strength, corrosion, packaging integrity, kink resistance, torque, and coating integrity. All tests met predetermined acceptance criteria.
Technological Characteristics
Materials: Polymer shaft, stainless steel cross-coil reinforcement, PTFE-lined lumen, platinum/iridium marker band. Dimensions: 0.074" ID, 0.086" nominal OD; lengths 95, 105, 115 cm. Features: Hydrophilic coating, luer-tapered hub. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Standards: ISO 10993 (biocompatibility), ISO 10555-1 (catheter performance), ISO 594-1/2 (luer), ASTM F1929/F88 (packaging).
Indications for Use
Indicated for the introduction of interventional devices into the peripheral and neuro vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
K230775 — SOFIA EX Intracranial Support Catheter · MicroVention, Inc. · Sep 14, 2023
K191237 — CERENOVUS Large Bore Catheter · Codman & Shurtleff, Inc. · Nov 8, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 26, 2018
InNeuroCo, Inc. Marianne Grunwaldt Director, Quality Assurance and Regulatory Affairs 4635 NW 103rd Avenue Sunrise, Florida 33351
Re: K173709
Trade/Device Name: Zenith Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DOY, DTL, DYB Dated: January 4, 2018 Received: January 5, 2018
Dear Marianne Grunwaldt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# William J. Heetderks -A 2018.01.26 16:50:00 -05'00'
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K173709
Device Name Zenith Support Catheter
Indications for Use (Describe)
The Zenith Support Catheter is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
### ADMINISTRATIVE INFORMATION
| Date of Summary Preparation: | December 1st, 2017 |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT INFORMATION | |
| Submitter/Manufacturer | InNeuroCo Inc<br>4635 NW 103rd Avenue<br>Sunrise, Florida 33351<br>Tel: 1-954-742-5988<br>Fax: 1-954-742-5989 |
| Primary Submission Contact | Marianne Grunwaldt<br>Cell:1-305-495-3883<br>Fax: 1-954-742-5988<br>Email: Marianne@InNeuroCo.com |
| Secondary Submission Contact | Ming Cheng Chew<br>Regulatory Consultant, Libra Medical,<br>8401 73rd Ave North, Suite 63<br>Brooklyn Park, MN 55428<br>Cell: 763-232-3701<br>Fax: 763-477-6357<br>Email: mcchew@libramed.com |
#### DEVICE INFORMATION
Device Trade or Proprietary Name Zenith Support Catheter
Device Common or Classification Name: Catheter, Percutaneous, 21 CFR 870.1250, Class II Hemostasis Valve, 21 CFR 870.4290, Class II Catheter Introducer, 21 CFR 870.1340, Class II
Product Code: DQY (Catheter) DTL (Hemostasis Valve) DYB (Catheter Introducer) Inc.
{4}------------------------------------------------
### IDENTIFICATION OF THE LEGALLY MARKETED DEVICE TO WHICH EQUIVALENCE IS BEING CLAIMED
| Name of Predicate Device | Name of Manufacturer | 510(k) Number |
|--------------------------|----------------------|---------------|
| Zenith 074 | InNeuroCo | K171672 |
## DEVICE DESCRIPTION
The InNeuroCo Zenith Support product consists of a catheter, a hemostasis valve, and a split sheath introducer. The Zenith Support catheter is a single lumen, variable stiffness catheter that has an outer polymer shaft and is reinforced with a Stainless Steel cross coil. The Zenith Support catheter has a radiopaque Platinum/Iridium marker band on the distal end. The Zenith Support is available with an internal diameter of 0.074 inches and a nominal outer diameter of 0.086 inches. There are three working lengths available: 95 cm, 105 cm, and 115 cm in either diameter. The Zenith Support catheter has a PTFE-lined lumen throughout the catheter shaft. The catheter is flexible and has a hydrophilic coating. The Zenith Support catheter is inserted through a guide catheter or vascular sheath, provides access to the target site, and once in place, provides a reinforcing conduit for other intravascular devices. Accessories included with the device are a Tuohy-Borst hemostasis valve with an extension luer and a split sheath introducer. The Zenith product line is supplied sterile, non-pyrogenic, and intended for single use only.
## INDICATIONS FOR USE / INTENDED USE
The Zenith Support Catheter is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.
| | Predicate Device<br>InNeuroCo, Inc.<br>Zenith 074 | Reference<br>Device<br>InNeuroCo SDA | InNeuroCo, Inc.<br>Zenith Support |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K171672 | K161262 | K173709 |
| Classification | 21CFR870.1250,<br>Class II | 21CFR870.1250,<br>Class II | 21CFR870.1250,<br>Class II |
| Product Code | DQY | DQY | DQY |
| Review Panel | Cardiovascular | Cardiovascular | Cardiovascular |
| | Predicate Device<br>InNeuroCo, Inc.<br>Zenith 074 | Reference<br>Device<br>InNeuroCo SDA | InNeuroCo, Inc.<br>Zenith Support |
| Indications For<br>Use | The Zenith is<br>indicated for the<br>introduction of<br>interventional devices<br>into the peripheral and<br>neuro vasculature. | The Super Distal<br>Access (SDA) is<br>indicated for the<br>introduction of<br>interventional<br>devices into the<br>peripheral and<br>neuro vasculature. | The Zenith<br>Support Catheter<br>is indicated for<br>the introduction<br>of interventional<br>devices into the<br>peripheral and<br>neuro vasculature. |
| Components<br>Supplied | Zenith Catheter, Peel<br>Away Introducers,<br>Hemostasis Valve,<br>Scout Introducer | Catheter, Peel<br>Away Introducer,<br>Hemostasis<br>Valve | Zenith Support<br>Catheter, Peel<br>Away Introducer,<br>Hemostasis Valve |
| Catheter Shaft<br>Material | Polymeric Catheter | Polymeric<br>Catheter | Polymeric<br>Catheter |
| Inner Liner | PTFE | PTFE | PTFE |
| Catheter Shaft<br>Reinforcement | Stainless Steel | Stainless<br>Steel/Nitinol | Stainless Steel |
| Reinforcement<br>pattern | Cross Coil | Braid | Cross Coil |
| Lubricious<br>Coating | Hydrophilic Coating | Hydrophilic<br>Coating | Hydrophilic<br>Coating |
| Radiopaque<br>Marker Band | Platinum/ Iridium | Platinum/ Iridium | Platinum/ Iridium |
| Packaging | Tyvek/Nylon Pouch,<br>polyethylene support<br>tube, packaging card,<br>SBS carton | Tyvek/Nylon<br>Pouch,<br>polyethylene<br>support tube,<br>packaging card,<br>SBS carton | Tyvek/Nylon<br>Pouch,<br>polyethylene<br>support tube,<br>packaging card,<br>SBS carton |
| Working<br>Lengths | 115, 125, 132 cm | 105, 115, 125 cm | 95, 105, 115 cm |
| | Predicate Device<br>InNeuroCo, Inc.<br>Zenith 074 | Reference<br>Device<br>InNeuroCo SDA | InNeuroCo, Inc.<br>Zenith Support |
| Proximal<br>Internal<br>Diameter (ID) | 0.074 inches | 0.062 inches | 0.074 inches |
| Distal ID | 0.074 inches | 0.060 inches<br>distal | 0.074 inches |
| Proximal Outer<br>Diameter | 0.086 inches | 0.079 inches | 0.086 inches |
| Distal Outer<br>Diameter | 0.086 inches | 0.072 | 0.086 inches |
| Peel Away<br>Introducer | Aid in catheter tip<br>introduction into<br>hemostasis valve | Aid in catheter tip<br>introduction into<br>hemostasis valve | Aid in catheter tip<br>introduction into<br>hemostasis valve |
| Hemostasis<br>Valve | Minimizes blood loss | Minimizes blood<br>loss | Minimizes blood<br>loss |
| Luer Tapered<br>Hub | Yes | Yes | Yes |
| Compatible<br>Guidewire | 0.038 inches | 0.038 inches | 0.038 inches |
| Scout<br>Introducer | Yes | No | No |
| Sterilization | EO SAL 10-6 | EO SAL 10-6 | EO SAL 10-6 |
| Pyrogenicity | Nonpyrogenic | Nonpyrogenic | Nonpyrogenic |
| Biocompatibility | Meets ISO 10993-<br>1:2009 | Meets ISO<br>10993-1:2009 | Meets ISO<br>10993-1:2009 |
# COMPARISON TO THE PREDICATE DEVICE
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
## SUMMARY OF NON-CLINICAL DATA
Each non-clinical test was analyzed to determine whether or not re-testing should be conducted for the Zenith Support Catheter. The results of the analysis and re-testing showed that the Zenith Support Catheter is substantially equivalent to the predicate device.
{8}------------------------------------------------
| Test | Test Method<br>Summary | Acceptance<br>Criteria | Re-Tested? | Analysis |
|--------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibilit<br>y-Material<br>Mediated<br>Pyrogen | Testing<br>completed per<br>ISO 10993-11 | The test article<br>extracts must not<br>cause a febrile<br>reaction greater<br>than 0.5°C in any<br>individual<br>subject. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | There are no<br>material changes<br>from the predicate<br>device.<br>Manufacturing<br>process is identical<br>to the predicate.<br>Only the segment<br>lengths were<br>changed and pitch<br>adjustments were<br>within the predicate<br>device's tolerance |
| Biocompatibilit<br>y- Cytotoxicity<br>MEM Elution | Testing<br>completed per<br>ISO 10993-5 | The cultures<br>treated with the<br>test article must<br>not have a<br>reactivity grade<br>greater than 2. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | There are no<br>material changes<br>from the predicate<br>device.<br>Manufacturing<br>process is identical<br>to the predicate.<br>Only the segment<br>lengths were<br>changed and pitch<br>adjustments were<br>within the predicate<br>device's tolerance |
| Biocompatibilit<br>y-Hemolysis<br>ASTM Method,<br>extract human<br>blood | Testing<br>completed per<br>ISO 10993-4 | The hemolytic<br>index above the<br>negative control<br>article must be<br>less than 5%. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | There are no<br>material changes<br>from the predicate<br>device.<br>Manufacturing<br>process is identical<br>to the predicate.<br>Only the segment<br>lengths were<br>changed and pitch<br>adjustments were<br>within the predicate<br>device's tolerance |
| Biocompatibilit<br>y-Hemolysis,<br>ASTM method, | Testing<br>completed per<br>ISO 10993-4 | The hemolytic<br>index above the<br>negative control | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | There are no<br>material changes<br>from the predicate<br>device. |
| Test | Test Method<br>Summary | Acceptance<br>Criteria | Re-Tested? | Analysis |
| direct contact<br>(human blood) | | article must be<br>less than 5%. | | Manufacturing<br>process is identical<br>to the predicate.<br>Only the segment<br>lengths were<br>changed and pitch<br>adjustments were<br>within the predicate<br>device's tolerance |
| Biocompatibilit<br>y-Unactivated<br>Partial<br>Thromboplastin<br>Time | Testing<br>completed per<br>ISO 10993-4 | There must be no<br>statistical<br>decrease between<br>the UPTT of<br>plasma exposed<br>to the test article<br>and to the<br>negative or<br>untreated control. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | There are no<br>material changes<br>from the predicate<br>device.<br>Manufacturing<br>process is identical<br>to the predicate.<br>Only the segment<br>lengths were<br>changed and pitch<br>adjustments were<br>within the predicate<br>device's tolerance |
| Biocompatibilit<br>y-Complement<br>Activation | Testing<br>completed per<br>ISO 10993-4 | There must be no<br>statistical increase<br>between either the<br>C3a or SC5b-9<br>concentrations in<br>plasma exposed<br>to the test article<br>as compared to<br>the negative and<br>untreated<br>controls. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | There are no<br>material changes<br>from the predicate<br>device.<br>Manufacturing<br>process is identical<br>to the predicate.<br>Only the segment<br>lengths were<br>changed and pitch<br>adjustments were<br>within the predicate<br>device's tolerance |
| Biocompatibilit<br>y-Dog<br>Thromboresista<br>nce | Testing<br>completed per<br>ISO 10993-4 | The test articles<br>must receive a<br>thrombus<br>formation score<br>less than or equal<br>to that of the<br>control. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | There are no<br>material changes<br>from the predicate<br>device.<br>Manufacturing<br>process is identical<br>to the predicate.<br>Only the segment<br>lengths were<br>changed and pitch<br>adjustments were<br>within the predicate<br>device's tolerance |
| Test | Test Method<br>Summary | Acceptance<br>Criteria | Re-Tested? | Analysis |
| Biocompatibilit<br>y-Maximization<br>Sensitization | Testing<br>completed per<br>ISO 10993-10 | The test article<br>must elicit a<br>positive response<br>in less than 10%<br>of the test<br>animals. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | changed and pitch<br>adjustments were<br>within the predicate<br>device's tolerance<br>There are no<br>material changes<br>from the predicate<br>device.<br>Manufacturing<br>process is identical<br>to the predicate.<br>Only the segment<br>lengths were<br>changed and pitch<br>adjustments were<br>within the predicate<br>device's tolerance |
| Biocompatibilit<br>y-<br>Intracutaneous<br>Toxicity/Reacti<br>vity | Testing<br>completed per<br>ISO 10993-10 | The test article<br>extracts must not<br>induce a<br>significantly<br>greater biological<br>reaction than the<br>control. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | There are no<br>material changes<br>from the predicate<br>device.<br>Manufacturing<br>process is identical<br>to the predicate.<br>Only the segment<br>lengths were<br>changed and pitch<br>adjustments were<br>within the predicate<br>device's tolerance |
| Biocompatibilit<br>y-Acute<br>Systemic<br>Toxicity Test | Testing<br>completed per<br>ISO 10993-11 | The test article<br>extracts must not<br>induce a<br>significantly<br>greater biological<br>reaction than the<br>control. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | There are no<br>material changes<br>from the predicate<br>device.<br>Manufacturing<br>process is identical<br>to the predicate.<br>Only the segment<br>lengths were<br>changed and pitch<br>adjustments were<br>within the predicate<br>device's tolerance |
| Test | Test Method<br>Summary | Acceptance<br>Criteria | Re-Tested? | Analysis |
| Animal<br>Testing-<br>Angiographic<br>Evaluation | Product<br>evaluated<br>within a<br>porcine model<br>for clinically<br>significant<br>injury. | No units tested<br>can cause vessel<br>injury. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | The tips, materials,<br>and transitions are<br>identical to the<br>predicate, and the<br>initial evaluation for<br>the predicate<br>included the use of<br>the device without<br>the Scout Introducer |
| Chemical<br>Compatibility | Catheter<br>exposed to<br>chemicals<br>readily<br>available in a<br>clinical setting. | Chemicals have<br>no negative effect<br>on the catheter or<br>accessories. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | There are no<br>material changes<br>from the predicate<br>device, and pitch<br>adjustments are<br>within the tolerance<br>of the predicate<br>device |
| Radiographic<br>Detectability | Testing<br>completed per<br>ISO 10555-1 | Product shall be<br>visible under<br>fluoro imaging. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | The marker bands<br>are identical to the<br>predicate |
| Visual<br>Inspection | Testing<br>completed per<br>ISO 10555-1 | Catheter shall<br>appear free from<br>damage,<br>including a<br>rounded tip and<br>smooth transition<br>points. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | There are no<br>material changes<br>from the predicate<br>device, and pitch<br>adjustments are<br>within the tolerance<br>of the predicate<br>device |
| Working<br>Length | Testing<br>completed per<br>ISO 10555-1 | Test samples<br>should be within<br>existing working<br>length<br>specification. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | The segment lengths<br>do not impact the cut<br>to length step in<br>addition the working<br>length test compares<br>the output to the<br>target length |
| Hub<br>compatibility | Testing<br>completed per<br>ISO 594-1 and<br>ISO 594-2 | Hub shall meet<br>existing Luer<br>specifications. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | The hub is identical<br>to the predicate<br>device |
| Outside<br>diameter | Testing<br>completed per<br>ISO 10555-1 | Test samples<br>should be within<br>existing outside | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | OD is unaffected by<br>the working length<br>change |
| Test | Test Method<br>Summary | Acceptance<br>Criteria | Re-Tested? | Analysis |
| Leak – Air | Testing<br>completed per<br>ISO 10555-1 | Test samples<br>should be within<br>existing Air -<br>Leak<br>specifications. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | There are no<br>material changes<br>from the predicate<br>device, and pitch<br>adjustments are<br>within the tolerance<br>of the predicate<br>device |
| Leak – Liquid | Testing<br>completed per<br>ISO 10555-1 | Test samples<br>should be within<br>existing Leak -<br>Liquid<br>specifications. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | There are no<br>material changes<br>from the predicate<br>device, and pitch<br>adjustments are<br>within the tolerance<br>of the predicate<br>device |
| Particulates | Testing<br>completed per<br>USP 788 | Test samples<br>should be within<br>existing<br>Particulate<br>specifications. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | There are no<br>material changes<br>from the predicate<br>device, and pitch<br>adjustments are<br>within the tolerance<br>of the predicate<br>device |
| Simulated Use -<br>Bench | Zenith<br>underwent<br>simulated use<br>testing by a<br>physician in a<br>benchtop<br>model | Test samples<br>must meet<br>predetermined<br>user needs | Yes | Zenith Support<br>Catheter test samples<br>met the acceptance<br>criteria for<br>Simulated Use-<br>Bench to<br>demonstrate that it is<br>substantially<br>equivalent to the<br>predicate device |
| Catheter Burst | Testing<br>completed per<br>ISO 10555-1 | Test sample burst<br>pressures must<br>meet or exceed<br>existing minimum<br>burst pressure<br>specification. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter. | There are no<br>material changes<br>from the predicate<br>device, and pitch<br>adjustments are<br>within the tolerance<br>of the predicate<br>device |
| Test | Test Method<br>Summary | Acceptance<br>Criteria | Re-Tested? | Analysis |
| Tensile | Testing<br>completed per<br>ISO 10555-1 | Test sample<br>ultimate tensile<br>strength must<br>meet or exceed<br>existing tensile<br>strength<br>specifications. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | The bonds and joints<br>are the same as the<br>predicate Zenith 074<br>device. Any pitch<br>adjustments are<br>within the tolerance<br>of the predicate<br>device. |
| Corrosion | Testing<br>completed per<br>ISO 10555-1 | Test samples shall<br>exhibit no<br>evidence of<br>corrosion. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | There are no<br>material changes<br>from the predicate<br>device, and pitch<br>adjustments are<br>within the tolerance<br>of the predicate<br>device |
| Packaging -<br>Dye Leak | Testing<br>completed per<br>ASTM F1929-<br>12 | Test sample shall<br>not exhibit any<br>visual leaks or<br>channels | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | The packaging is<br>identical to the<br>cleared reference<br>SDA device |
| Packaging -<br>Peel | Testing<br>completed per<br>ASTM F88-09 | Test sample<br>tensile strength<br>must meet or<br>exceed existing<br>tensile strength<br>specifications. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | The packaging is<br>identical to the<br>cleared reference<br>SDA device |
| Kink Resistance | Samples are<br>subjected to<br>different<br>diameters until<br>kink is<br>observed. | Test sample kink<br>resistance must<br>meet or exceed<br>existing Kink<br>Resistance<br>specifications. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | All transitions are<br>the same, only the<br>location of the<br>transition changes,<br>in addition, the pitch<br>adjustments are<br>within the Zenith<br>074 tolerance |
| Torque | Conditioned<br>samples are<br>torqued to<br>failure | Test sample<br>torque results<br>must meet or<br>exceed existing<br>torque<br>specifications. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | There are no<br>material changes<br>from the predicate<br>device, and pitch<br>adjustments are<br>within the tolerance<br>of the predicate<br>device |
| PTFE Liner<br>inspection | Zenith was<br>challenged to | Test sample liner<br>adhesion must | Testing was not<br>repeated for the | There are no<br>material changes |
| Test | Test Method<br>Summary | Acceptance<br>Criteria | Re-Tested? | Analysis |
| | demonstrate<br>liner<br>adherence. | meet or exceed<br>existing PTFE<br>Liner inspection<br>specifications. | Zenith Support<br>Catheter | from the predicate<br>device, and pitch<br>adjustments are<br>within the tolerance<br>of the predicate<br>device |
| Hydrophilic<br>Coating<br>Integrity | Conditioned<br>samples were<br>repeatedly<br>exposed to<br>friction to<br>demonstrate<br>that the<br>hydrophilic<br>coating is not<br>affected. | Test sample<br>results must meet<br>or exceed existing<br>Hydrophilic<br>Coating Integrity<br>specifications. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | There are no<br>material changes<br>from the predicate<br>device, and pitch<br>adjustments are<br>within the tolerance<br>of the predicate<br>device |
| Labeling<br>Legibility | Label is legible<br>after printing. | Test samples shall<br>demonstrate text<br>legibility. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter<br>substantially<br>equivalent to the<br>predicate device. | The label is almost<br>identical to the<br>predicate device,<br>except that the Scout<br>Introducer is<br>removed from the<br>labeling |
| Barcode | Barcode is<br>readable with a<br>standard<br>barcode reader. | Test samples shall<br>demonstrate<br>readily readable<br>barcodes | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | Barcodes are similar<br>to the predicate |
| Sterilization | ISO 11135 and<br>AAMI TIR 28 | Sterilization load<br>shall pose an<br>equal or lesser<br>challenge to<br>sterilize than<br>existing sterile<br>product loads. | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | The Zenith Support<br>Catheter was<br>adopted into the<br>Zenith Family<br>Sterilization Cycle |
| Shelf Life | ASTM F1980 | Aged test samples<br>must meet or<br>exceed existing<br>specifications | Testing was not<br>repeated for the<br>Zenith Support<br>Catheter | There are no<br>material changes<br>from the predicate<br>device, therefore<br>shelf life testing<br>conducted for the<br>Zenith 074 still<br>applies |
# Table 1: Non-Clinical Testing Analysis
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
{14}------------------------------------------------
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.