CROSSBOSS CATHETER MODEL M-2000, STINGRAY ORIENTING BALLOON CATHETER MODEL M-1000, STINGRAY GUIDEWIRE MODEL M-3004 AND M

K102725 · Bridge Medical, Inc. · DQY · May 10, 2011 · Cardiovascular

Device Facts

Record IDK102725
Device NameCROSSBOSS CATHETER MODEL M-2000, STINGRAY ORIENTING BALLOON CATHETER MODEL M-1000, STINGRAY GUIDEWIRE MODEL M-3004 AND M
ApplicantBridge Medical, Inc.
Product CodeDQY · Cardiovascular
Decision DateMay 10, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2

Intended Use

The BridgePoint Medical System (consisting of the CrossBoss™ Catheter, Stingray™ Orienting Balloon Catheter, and Stingray™ Guidewire) is intended to facilitate the intraluminal placement of conventional guidewires beyond stenotic coronary lesions (including chronic total occlusions) prior to PTCA or stent intervention.

Device Story

BridgePoint Medical System comprises CrossBoss Catheter, Stingray Orienting Balloon Catheter, and Stingray Guidewire. Used in coronary/peripheral vasculature to facilitate guidewire placement beyond chronic total occlusions (CTO). CrossBoss catheter features coiled stainless steel shaft, hydrophilic coating, and torque device for manual manipulation; provides pushability and torque. Stingray balloon catheter features dual-lumen design with side-ports for guidewire steering and non-compliant balloon for fluoroscopic orientation. Stingray guidewire is 0.014" stainless steel with platinum/tungsten coil for radiopacity. Operated by physicians in clinical settings (e.g., cath lab) under fluoroscopic guidance. System enhances ability to navigate stenotic lesions; facilitates subsequent PTCA or stent placement. Benefits include successful guidewire crossing of complex lesions with reduced risk.

Clinical Evidence

Clinical evaluation included a 147-patient study in the United States. Study demonstrated the system successfully facilitates placement of guidewires beyond chronic total occlusions with no significant increase in patient risk. Bench testing (tensile, torque, kink resistance, trackability, corrosion, coating durability, particulate, balloon performance, dimensional) and biocompatibility testing were leveraged from identical predicate devices.

Technological Characteristics

System components: CrossBoss catheter (stainless steel coiled shaft, polyester/polyurethane exterior, hydrophilic coating), Stingray balloon catheter (dual-lumen, non-compliant balloon), Stingray guidewire (0.014" stainless steel, platinum/tungsten coil, PTFE/hydrophilic coating). Manual operation; fluoroscopic visualization. Single-use, disposable. No software or electronic components.

Indications for Use

Indicated for patients requiring intraluminal guidewire placement beyond stenotic coronary lesions, including chronic total occlusions, prior to PTCA or stent intervention.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### MAY 1 0 2011 ## 510(K) SUMMARY ؟ ﺗﺮﻳﺒﺎ This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. 510(k) Number: ______________________________________________________________________________________________________________________________________________________________ ### Applicant Information: | Date Prepared: | January 18, 2011 | |----------------|------------------| |----------------|------------------| Name: BridgePoint Medical Address: 2800 Campus Drive, #50 Plymouth, MN 55441 Phone: 763-225-8500 Fax: 763-225-8718 | Contact Person: | Jill Munsinger | |-----------------|-------------------------------------------| | Phone Number: | office: 763-225-8510 / cell: 651-270-0572 | | E-mail: | jmunsinger@bridgepointmedical.com | ### Device Information: | Trade Name | | Common Name Classification Name | Class | |---------------------------------------|------------------------|---------------------------------|----------| | CrossBoss™ Catheter | Percutaneous Catheter | Percutaneous Catheter | Class II | | StingrayTM Orienting Balloon Catheter | Percutaneous Catheter | Percutaneous Catheter | Class II | | Stingray™ Guidewire | Percutaneous Guidewire | Percutaneous Guidewire | Class II | ### Predicate Devices: The BridgePoint Medical System (comprised of the CrossBoss™ Catheter, Stingray™ Orienting Balloon Catheter, and Stingray™ Guidewire) is substantially equivalent (and in some cases identical) in intended use and/or method of operation and technical aspects to the following predicate devices: | Device | Reference 510(k) Number | |---------------------------------------|-------------------------------------------------| | LuMend FrontrunnerTM CTO Catheter | K013284, K023223, K023114, K031005, and K033535 | | CrossBossTM Catheter | K081130 and K091841 | | StingrayTM Orienting Balloon Catheter | K080987 and K101591 | | StingrayTM Guidewire | K081187 and K083727 | {1}------------------------------------------------ #### Device Description: The CrossBoss Catheter is a single use over the wire disposable percutaneous catheter consisting of a full length coiled stainless steel shaft with polyester and polyurethane exterior. The coiled shaft provides torque and makes it possible to push the device, and also provides a guidewire lumen. The distal shaft transitions to an enlarged (1mm diameter) rounded distal tip. This stainless steel tip provides an atraumatic element that is intended to enhance the catheter's ability to move within the vasculature with reduced risk of arterial tissue engagement while providing radiopaque visibility. The distal portion of the CrossBoss Catheter is hydrophilic coated to enhance lubricity. The proximal portion includes an internal stainless steel hypotube stiffener that provides additional push. A torque device, coaxially positioned over the proximal portion of the CrossBoss Catheter, provides a comfortable user interface for device manipulation. The torque device (similar to a guidewire torque device) is positionable along the proximal portion of the catheter and includes a torsion release safety mechanism. This safety mechanism insures the torque input generated by the user remains within the torsional operating strength of the catheter shaft. The Stingray Orienting Balloon Catheter is a single use, over-the-wire, disposable, dual lumen percutaneous catheter that facilitates the placement, support and steering of a guidewire into discrete regions of the coronary and peripheral vasculature through the central guidewire lumen or through one of two side-ports (identified by radiopaque markers). The side-ports connect with the central guidewire lumen and facilitate guidewire steering (at an angle to the central lumen) by allowing the guidewire to exit the catheter. The catheter contains a small non-compliant balloon segment used for fluoroscopic orientation on the distal tip of the flexible shaft. The Stingray Guidewire is a conventionally constructed 0.014" diameter, single use, disposable guidewire that consists of a full-length stainless steel shaft with proximal PTFE coating where the distal portion of the shaft is taper ground to provide distal flexibility. The distal portion also includes a coaxially positioned coil constructed of platinum/tungsten material for visibility under fluoroscopy. The coil is fixed to the stainless steel core wire via silver alloy solder and is coated with hydrophilic coating. The distal tip of the guidewire is supplied with an angled geometry which transitions to a conventional rounded tip. The core wire (~0.0035" diameter) extends approximately 0.007" distal of the rounded tip. #### Intended Use: The BridgePoint Medical System (consisting of the CrossBoss™ Catheter, Stingray™ Orienting Balloon Catheter, and Stingray™ Guidewire) is intended to facilitate the intraluminal placement of conventional guidewires beyond stenotic coronary lesions (including chronic total occlusions) prior to PTCA or stent intervention. {2}------------------------------------------------ #### Comparison to Predicate Device(s): The designs of the devices comprising the BridgePoint Medical System are identical to their respective predicate devices. Bench testing was previously conducted using the CrossBoss Catheter, Stingray Orienting Balloon Catheter, and Stingray Guidewire predicate devices in tests designed to address a chronic total occlusion application. Therefore, all prior bench performance testing (tensile, torque, kink resistance, trackability, corrosion, coating durability, particulate, balloon performance, dimensional, etc.) are directly applicable to this application. As well, all previous biocompatibility tests conducted for the aforementioned predicate devices is directly applicable to the devices that comprise the BridgePoint Medical System as they are identical. As demonstrated in both animal studies and human clinical evaluations, the devices used as a system functions to enhance medical practice in facilitating the placement of guidewires or other interventional devices beyond chronic total occlusions. A 147 patient study was conducted in the United States that demonstrated the BridgePoint Medical System could successfully allow the placement of guidewires beyond chronic total occlusions with no significant increase in risk to the patient. #### Summary: Based upon the intended use and descriptive information provided in this pre-market notification. the BridgePoint Medical System consisting of the CrossBoss™ Catheter, Stingray™ Orienting Balloon Catheter, and Stingray™ Guidewire has been shown to be substantially equivalent to the currently marketed predicate devices. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three parallel lines forming its body and wings. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 MAY 1 0 2011 BridgePoint Medical, Inc. c/o Ms. Jill Munsinger Regulatory Affairs 2800 Campus Drive, #50 Plymouth, MN 55441 Re: K102725/S003 Trade/Device Name: Crossboss™ Catheter Model M-2000, Stingray™ Orienting Balloon Catheter Model M-1000 & Stingray™ Guidewire Model M-3004 Regulation Number: 21 CFR 870.1250 & 870-1330 Regulation Name: Percutaneous Catheter & Catheter guide wire Regulatory Class: Class II Product Code: DQY & DQX Dated: April 8, 2011 Received: April 11, 2011 Dear Ms. Munsinger: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical In microal in microsite devices that have been reclassified in accordance with the provisions of the Federal Food, Dr u and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affection your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Ms. Jill Munsinger Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # 7. INDICATIONS FOR USE STATEMENT 510(k) Number: (TBA) #### Device Name: BridgePoint Medical System consisting of the CrossBoss™ Catheter, StingrayTM Orienting Balloon Catheter, and Stingray™ Guidewire #### Indications For Use: The BridgePoint Medical System (consisting of the CrossBoss™ Catheter, Stingray™ Orienting Balloon Catheter, and Stingray™ Guidewire) is intended to facilitate the intraluminal placement of conventional guidewires beyond stenotic coronary lesions (including chronic total occlusions) prior to PTCA or stent intervention. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 (Division Sign-Off) Division of Cardiovascular Devices K102-728 510(k) Number__
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...