ADVANCE CS CORONARY SINUS INFUSION CATHETER

K131204 · Cook, Inc. · DQY · Jan 29, 2014 · Cardiovascular

Device Facts

Record IDK131204
Device NameADVANCE CS CORONARY SINUS INFUSION CATHETER
ApplicantCook, Inc.
Product CodeDQY · Cardiovascular
Decision DateJan 29, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Advance® CS Coronary Sinus Infusion Catheter is intended for temporary occlusion of the coronary sinus for infusion of contrast media, drugs, or therapeutic agents; or possible introduction of devices into the coronary venous system.

Device Story

Advance® CS Coronary Sinus Infusion Catheter; over-the-wire balloon catheter; 5.0 Fr size; compatible with 0.035 inch wire guide. Used by clinicians for temporary occlusion of coronary sinus; enables infusion of contrast media, drugs, or therapeutic agents; facilitates introduction of other devices into coronary venous system. Sterile, single-use device. Performance verified via bench testing (burst strength, sheath compatibility, inflation/deflation times, fatigue, torque, kink resistance) and acute porcine in vivo evaluation. Output is mechanical occlusion of the vessel; allows physician to perform targeted venous delivery or access.

Clinical Evidence

Bench testing included balloon burst strength, sheath compatibility, profile measurement, compliance, flow rate characterization, leakage, fatigue, bond/material strength, inflation/deflation time, torque, kink/flexibility, and particulate testing. Biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility) performed per ISO 10993-1. Acute in vivo animal evaluation (porcine coronary sinus) characterized safety and biologic response 48 hours post-inflation.

Technological Characteristics

5.0 Fr over-the-wire balloon catheter; compatible with 0.035 inch wire guide; 5, 6, or 7 Fr sheath compatibility. Mechanical sensing/actuation via balloon inflation. Sterile, single-use. Biocompatible materials per ISO 10993-1.

Indications for Use

Indicated for temporary coronary sinus occlusion to facilitate infusion of contrast media, drugs, or therapeutic agents, or to assist in the introduction of devices into the coronary venous system.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, bold, white letters on a black background. Below "COOK" is the word "MEDICAL" in smaller, white letters. The logo is simple and modern. COOK INCORPORATED 750 DANIELS WAY, P.O, BOX 489 BLOOMINGTON, IN 47402-0489 U.S.A. 612.339,2235 10 : 800.457.4500 WWW.COOKMEDICAL.COM # JAN 2 9 2014 ### 510(k) SUMMARY #### Submitted By: Jennifer Richardson, MA Cook Incorporated 750 Daniels Way P.O. Box 489 Bloomington, IN 47402 Phone: (812) 335-3575 x102370 Fax: (812) 332-0281 Date Prepared: December 20, 2013 #### Device: Advance® CS Coronary Sinus Infusion Catheter Trade Name: Common Name: Percutaneous Catheter Class II, Product Code: DQY, (21 CFR §870.1250) Classification: # Indications for Use: The Advance® CS Coronary Sinus Infusion Catheter is intended for temporary occlusion of the coronary sinus for infusion of contrast media, drugs, or therapeutic agents; or possible introduction of devices into the coronary venous system. #### Device Description: The Advance® CS Coronary Sinus Infusion Catheter is an over-the-wire catheter. The 5.0 Fr balloon catheter is compatible with a 0.035 inch diameter wire guide. The device will be supplied sterile and intended for one-time use. ### Comparison to Predicate Device: The proposed device is substantially equivalent to the predicate in terms of intended use, duration of use, principles of operation, and technological characteristics. # Discussion of Tests and Test Results: The device was subjected to the following tests to assure reliable design and performance under the specified testing parameters. These tests were comprised of: - Balloon Minimum Burst Strength Testing shows the balloons will burst at or above the 1. minimum rated burst pressure, with all failure modes being linear tears. The predetermined acceptance criteria were met. - Sheath Compatibility Qualitative and quantitative evaluations show that the balloons 2. are compatible with a 5, 6, or 7 Fr sheath. The predetermined acceptance criteria were met. - 3. Balloon Profile Measurement of the diameter of catheter shaft, bonds, and folded balloon shows that the device profile is compatible with a 5, 6. or 7 Fr sheath (0.074. 0.087, or 0.100 inch profile). The predetermined acceptance criteria were met. K131204 Page 1 of 2 {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for Cook Medical. The logo is black and white and features the word "COOK" in large, bold letters on the top line. Below that, the word "MEDICAL" is written in smaller letters. The logo is simple and professional. COOK INCORPORATED 750 DANIELS WAY, P.O. BOX 489 BLOOMINGTON, IN 47402-0489 U.S.A. HONE: 812.339.2235 TOLL FREE: 800.457.4500 •WWW.COOKMEDICAL.COM - 4. Balloon Compliance Testing shows that the balloon will inflate to the stated balloon size. The predetermined acceptance criteria were met. - 5. Dynamic Burst of the Distal Lumen Testing characterized saline and contrast agent flow rate vs. pressure curves. Testing shows that all test articles withstood at least pressures in excess of 300 psig regardless of injection media type. The predetermined acceptance criteria were met. - 6. Static Burst of the Distal Lumen Testing shows that similar test articles can be expected to fail at or above 100 psig. The predetermined acceptance criteria were met. - 7. Freedom from Leakage Testing shows that the balloon catheter will withstand pressurization to 200 psi for a time period of 15 seconds without leaking or significantly increasing in outside diameter. The predetermined acceptance criteria were met. - Balloon Fatigue Testing shows that balloons are free from leakage and damage on 8. inflation, withstanding 10 cycles of inflation. The predetermined acceptance criteria were met. - 9. Bond and Materials Strength Testing shows the tensile force during normal use should not fracture or rupture the balloon catheter bonds or materials. The predetermined acceptance criteria were met. - 10. Inflation / Deflation Time Testing shows that the balloon will inflate within 60 seconds and fully deflate within 60 seconds. The predetermined acceptance criteria were met. - 11. Torque Strength Testing shows that the catheter will withstand two full rotations without failure. - 12. Kink and Flexibility The testing shows that the catheter will not kink at a radius equal to or greater than 13mm. - 13. Particulate The testing shows that the balloon will not create particulates that exceed that of the reference limits. - 14. Dimensional Verification The testing shows that the devices will meet dimensional specifications. - 15. Animal Evaluation In vivo testing characterized the safety and biologic response approximately 48 hours following the acute inflation of the balloon in the porcine coronary sinus. - 16. Biocompatibility Testing (i.e., cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility) shows the device is biocompatible. In conformance with the applicable sections of ISO 10993-1, the predetermined acceptance criteria were met. # Conclusions Drawn from the Tests: The results of these tests provide reasonable assurance that the device is as safe and effective as the predicate device and support a determination of substantial equivalence. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol often associated with healthcare, featuring a staff with a snake winding around it. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 29, 2014 Cook Incorporated c/o Ms. Jennifer Richardson, MA 750 Daniels Way . P.O. Box 489 Bloomington, IN 47402 Re: K131204 Trade/Device Name: Advance® CS Coronary Sinus Infusion Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Advance® CS Coronary Sinus Infusion Catheter Regulatory Class: Class II Product Code: DQY Dated: December 30, 2013 Received: December 31, 2013 Dear Ms. Richardson: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {3}------------------------------------------------ Page 2 - Ms. Jennifer Richardson . or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. .fda.gov/MedicalDevices/ResourcesforYou/Industry/ucm141.htm Sincerely yours, M.A. Hilleman for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Cook Incorporated Advance® CS Coronary Sinus Infusion Catheter Traditional 510(k) 26 April 2013 # Indications for Use Statement 510(k) Number (if known): _K131204 Device Name: Advance CS Coronary Sinus Infusion Catheter # Indications for Use: The Advance CS Coronary Sinus Infusion Catheter is intended for temporary occlusion of the coronary sinus for infusion of contrast media, drugs, or therapeutic agents; or possible introduction of devices into the coronary venous system. Over-The-Counter Use X Prescription Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) M.A. Hillerman 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...