Passport Series Patient Monitors (including Passport 17m, Passport 12m and T1)

K152902 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · MHX · Feb 18, 2016 · Cardiovascular

Device Facts

Record IDK152902
Device NamePassport Series Patient Monitors (including Passport 17m, Passport 12m and T1)
ApplicantShenzhen Mindray Bio-Medical Electronics Co., Ltd.
Product CodeMHX · Cardiovascular
Decision DateFeb 18, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1025
Device ClassClass 2
AttributesPediatric

Intended Use

The Passport 17m and Passport 12m patient monitors are intended for monitoring, displaying, alarming, and transferring of multiple physiological parameters including ECG (3-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, OT analysis, and heart rate (HR)), respiration rate (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure(IBP), pulmonary artery wedge pressure (PAWP), cardiac output (C.O.), continuous cardiac output (CCO), mixed/central venous oxygen saturation (Sv02/Scv02), carbon dioxide (C02), Oxygen (02), anesthetic gas (AG), impedance cardiograph (ICC), bispectral index (BIS), and respiration mechanics (RM). The equipment also provides an interpretation of resting 12-lead ECG. All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following: · The arrhythmia detection, ST Segment analysis of Mortara algorithm, BIS, RM, CCO, SvO2/ScvO2, and PAWP monitoring are intended for adult and pediatric patients only; · ST Segment analysis of Mindray algorithm is intended for adult patients only; · C.O. monitoring is restricted to adult patients only; • ICG monitoring is only for use on adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg. The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use. The T1 Patient Monitor is intended for monitoring, displaying, reviewing, and transferring of multiple physiological parameters including ECG (3-lead, or 12-lead selectable, arrhythmia detection, ST Segment analysis, QT analysis, and heart rate (HR)), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), pulmonary artery wedge pressure (PAWP), and carbon dioxide (CO2). The equipment also provides an interpretation of resting 12-lead ECG. All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following: • The arthythmia detection, ST Segment analysis of Mortara algorithm, and PAWP monitoring are intended for adult and pediatric patients; · ST Segment analysis of Mindray ECG algorithm is intended for adult patients only. The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.

Device Story

Passport Series (12m, 17m, T1) are multiparameter patient monitors; used in healthcare facilities by clinical professionals. Devices accept inputs from various physiological sensors (ECG, SpO2, NIBP/IBP, CO2, etc.); process signals to display waveforms, numerics, and alarms; provide automated interpretation of resting 12-lead ECG. T1 functions as a standalone monitor or as a module for 12m/17m. Clinicians use real-time data and alarms to monitor patient status; output supports clinical decision-making regarding patient care. Benefits include continuous physiological surveillance and early detection of adverse events.

Clinical Evidence

Bench testing only. Functional and system-level testing conducted to evaluate modifications and ensure compliance with consensus standards and FDA Class II Special Controls Guidance for Arrhythmia Detector and Alarm.

Technological Characteristics

Multiparameter monitors with integrated touchscreen displays (17", 12", 5"). Connectivity via wireless modules (ASUS, Silex, Cyberlink). Power via AC/DC or rechargeable Lithium-ion batteries. Supports modular measurement inputs. Software includes ECG algorithms (Mindray and Mortara) for arrhythmia, ST-segment, and QT analysis. Data storage via Compact Flash or SD cards.

Indications for Use

Indicated for monitoring, displaying, alarming, and transferring physiological parameters (ECG, Resp, Temp, SpO2, PR, NIBP, IBP, PAWP, C.O., CCO, SvO2/ScvO2, CO2, O2, AG, ICG, BIS, RM) in adult, pediatric, and neonatal patients in healthcare facilities. Specific exclusions apply for certain advanced parameters (e.g., C.O., ICG, BIS, RM) based on patient age or size.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing right, representing health and human services. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 18, 2016 Shenzhen Mindray Bio-medical Electronics Co., Ltd Yanhong Bai Manager Regulatory Affairs Mindray Building, Keji 12th Road South, High-tech Industrial Park, Nanshan Shenzhen. 518057 CN Re: K152902 Trade/Device Name: Passport Series Patient Monitors (Passport 12m, Passport 17m and T1) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX, DSI, MLD, DRT, DXN, DSK, FLL, DOA, DPZ, CCK, CBO, CBS, CBR, CCL, DSB, DXG, OLW Dated: February 4, 2016 Received: February 5, 2016 Dear Yanhong Bai: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Kenneth J. Cavanaugh -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K152902 #### Device Name Passport Series Patient Monitors (Passport 12m, Passport 17m and T1) Indications for Use (Describe) Passport 12m and 17m Patient Monitors: The Passport 17m and Passport 12m patient monitors are intended for monitoring, displaying, alarming, and transferring of multiple physiological parameters including ECG (3-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, OT analysis, and heart rate (HR)), respiration rate (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure(IBP), pulmonary artery wedge pressure (PAWP), cardiac output (C.O.), continuous cardiac output (CCO), mixed/central venous oxygen saturation (Sv02/Scv02), carbon dioxide (C02), Oxygen (02), anesthetic gas (AG), impedance cardiograph (ICC), bispectral index (BIS), and respiration mechanics (RM). The equipment also provides an interpretation of resting 12-lead ECG. All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following: · The arrhythmia detection, ST Segment analysis of Mortara algorithm, BIS, RM, CCO, SvO2/ScvO2, and PAWP monitoring are intended for adult and pediatric patients only; · ST Segment analysis of Mindray algorithm is intended for adult patients only; · C.O. monitoring is restricted to adult patients only; • ICG monitoring is only for use on adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg. The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use. T1 Patient Monitor: The T1 Patient Monitor is intended for monitoring, displaying, reviewing, and transferring of multiple physiological parameters including ECG (3-lead, or 12-lead selectable, arrhythmia detection, ST Segment analysis, QT analysis, and heart rate (HR)), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), pulmonary artery wedge pressure (PAWP), and carbon dioxide (CO2). The equipment also provides an interpretation of resting 12-lead ECG. All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following: • The arthythmia detection, ST Segment analysis of Mortara algorithm, and PAWP monitoring are intended for adult and pediatric patients; · ST Segment analysis of Mindray ECG algorithm is intended for adult patients only. The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. {3}------------------------------------------------ ## 510(K) SUMMARY In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Passport Series Patient Monitors is provided below. | Device Common Name: | Patient Monitor | |----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Proprietary Name: | Passport Series Patient Monitors (including Passport 17m,<br>Passport 12m and T1) | | Submitter: | SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>Shenzhen 518057, P.R. China<br>Tel: +86 755 81885635<br>Fax: +86 755 26582680 | | Contact: | Yanhong Bai<br>Manager Regulatory Affairs<br>SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>518057, P.R. China<br>Tel: +86 755 81885635<br>Fax: +86 755 26582680<br>E-mail: baiyanhong@mindray.com | | Date Prepared: | September 30, 2015 | | Classification Regulation: | 21 CFR 870.1025, Class II, Arrhythmia detector and alarm<br>(including ST-segment measurement and alarm) | | Panel: | Cardiovascular | #### Classification Regulation, Classification Name and Product Codes: | Product<br>Code | Regulation<br>Number | Panel | Regulation Description | Device Common Name | |-----------------|----------------------|----------------|--------------------------------------------------------------------------------------|-----------------------------------------------------------------------------| | Primary | | | | | | MHX | 21 CFR<br>870.1025 | Cardiovascular | Arrhythmia detector and<br>alarm (including ST-<br>segment measurement and<br>alarm) | monitor, physiological, patient<br>(with arrhythmia detection or<br>alarms) | | Secondary | | | | | | Product<br>Code | Regulation<br>Number | Panel | Regulation Description | Device Common Name | | DSI | 21 CFR<br>870.1025 | Cardiovascular | Arrhythmia detector and<br>alarm (including ST-<br>segment measurement and<br>alarm) | detector and alarm, arrhythmia | | MLD | 21 CFR<br>870.1025 | Cardiovascular | Arrhythmia detector and<br>alarm (including ST-<br>segment measurement and<br>alarm) | monitor, st segment with alarm | | DRT | 21 CFR<br>870.2300 | Cardiovascular | Cardiac Monitor<br>(including<br>cardiotachometer and rate<br>alarm) | monitor, cardiac (incl.<br>cardiotachometer & rate alarm) | | DXN | 21 CFR<br>870.1130 | Cardiovascular | Noninvasive blood<br>pressure measurement<br>system | system, measurement, blood-<br>pressure, non-invasive | | DSK | 21 CFR<br>870.1110 | Cardiovascular | Blood pressure computer | computer, blood-pressure | | FLL | 21 CFR<br>880.2910 | Cardiovascular | Clinical electronic<br>thermometer | thermometer, electronic, clinical | | DQA | 21 CFR<br>870.2700 | Cardiovascular | Oximeter | Oximeter | | DPZ | 21 CFR<br>870.2710 | Cardiovascular | Ear oximeter | oximeter, ear | | CCK | 21 CFR<br>868.1400 | Anesthesiology | Carbon dioxide gas<br>analyzer | analyzer, gas, carbon-dioxide,<br>gaseous-phase | | CBQ | 21 CFR<br>868.1500 | Anesthesiology | Enflurane gas analyzer | analyzer, gas, enflurane,<br>gaseous-phase (anesthetic<br>concentration) | | CBS | 21 CFR<br>868.1620 | Anesthesiology | Halothane gas analyzer | analyzer, gas, halothane,<br>gaseous-phase (anesthetic conc.) | | CBR | 21 CFR<br>868.1700 | Anesthesiology | Nitrous oxide gas<br>analyzer | analyzer, gas, nitrous-oxide,<br>gaseous phase (anesthetic conc.) | | CCL | 21 CFR<br>868.1720 | Anesthesiology | Oxygen gas analyzer | analyzer, gas, oxygen, gaseous-<br>phase | | DSB | 21 CFR<br>870.2770 | Cardiovascular | Impedance<br>plethysmograph | plethysmograph, impedance | | DXG | 21 CFR<br>870.1435 | Cardiovascular | Single-function,<br>preprogrammed<br>diagnostic computer | computer, diagnostic, pre-<br>programmed, single-function | | OLW | 21 CFR<br>882.1400 | Neurology | Electroencephalograph | index-generating<br>electroencephalograph software | {4}------------------------------------------------ {5}------------------------------------------------ | Primary Predicate Device: | K143195 - Passport Series Patient Monitors (Including Passport 17M, Passport 12m, and T1); Shenzhen Mindray Bio-Medical Electronics Co., Ltd | |---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------| | Reference Predicates: | K101521 - Philips ST/AR ST and Arrhythmia Software Model; Philips Medical Systems | | | K150352 - V Series Monitoring System; Shenzhen Mindray Bio-Medical Electronics Co., Ltd | | | K150632 - Hypervisor IX Monitoring System; Shenzhen Mindray Bio-Medical Electronics Co., Ltd | ### Indications for Use: ## Passport 12m and 17m Patient Monitors: The Passport 17m and Passport 12m patient monitors are intended for monitoring, displaying, reviewing, alarming, and transferring of multiple physiological parameters including ECG (3lead , 5-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, OT analysis, and heart rate (HR)), respiration rate (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure(IBP), pulmonary artery wedge pressure (PAWP), cardiac output (C.O.), continuous cardiac output (CCO), mixed/central venous oxygen saturation (SvO2/ScvO2), carbon dioxide (CO2), Oxygen (O2), anesthetic gas (AG), impedance cardiograph (ICG), bispectral index (BIS), and respiration mechanics (RM). The equipment also provides an interpretation of resting 12-lead ECG. All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following: - The arrhythmia detection. ST Segment analysis of Mortara algorithm, BIS, RM, o CCO, SvO2/ScvO2, and PAWP monitoring are intended for adult and pediatric patients only; - ST Segment analysis of Mindray algorithm is intended for adult patients only; ● - C.O. monitoring is restricted to adult patients only; - ICG monitoring is only for use on adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg. The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use. ## T1 Patient Monitor: The T1 Patient Monitor is intended for monitoring, displaying, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, or 12-lead selectable, arrhythmia detection, ST Segment analysis, QT analysis, and heart rate (HR)), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), pulmonary artery {6}------------------------------------------------ wedge pressure (PAWP), and carbon dioxide (CO2). The equipment also provides an interpretation of resting 12-lead ECG. All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following: - The arrhythmia detection. ST Segment analysis of Mortara algorithm, and PAWP ● monitoring are intended for adult and pediatric patients; - ST Segment analysis of Mindrav ECG algorithm is intended for adult patients ● only. The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use. ### Device Description: The subject Passport Series Patient Monitors includes three monitors: - Passport 12m Patient Monitor ● - o Passport 17m Patient Monitor - o T1 Patient Monitor All of the devices in the family are multiparameter monitors indicated for monitoring, displaying, reviewing, alarming, and transferring multiple physiological parameters. The Passport 12m and 17m Patient monitors provide patient monitoring capabilities based on the user selected modules that are plugged into the main monitor. The T1 patient monitor is one of the available modules that can be plugged into the Passport 17m or 12m monitor. The T1 can also be used as a standalone monitor and when used as a standalone monitor, it provides a subset of the functions that are provided by Passport 17m and 12m. #### Performance Data: - o To establish the substantial equivalence of the Passport Series Patient Monitors (including Passport 12m, Passport 17m and T1), Mindray conducted functional and system level testing on the subject devices. The testing provided an evaluation of the performance of the device relevant to each of the modifications to the subject devices since their previous clearance. The functional and system level testing showed that the devices continue to meet specifications and the performance of the device is equivalent to the predicate. - In addition. Mindray has conducted testing to ensure the subject devices meet relevant ● consensus standards. - Mindray has also followed the FDA Class II Special Controls Guidance Document: ● Arrhythmia Detector and Alarm issues on October 28, 2003. #### Substantial Equivalence: {7}------------------------------------------------ Comparison of Indications - Both the predicate device and the subject Passport 12m, Passport 17m and T1 devices are multiparameter patient monitors intended to be used in healthcare facilities under the direction of clinical professionals. The indications for use of the subject devices have been modified to include the new feature of QT analysis. Although this feature is not present in the predicate device, it is present in the cleared Philips ST/AR ST and Arrhythmia software (K101521), and thus does not constitute a new intended use for a multi-parameter monitor. In conclusion, the minor changes to the indications for use do not change the fundamental intended use of the Passport 12m. Passport 17m and T1 as multiparameter monitors. Comparison of Technological Characteristics - The table below compares the key technological feature of the subject devices to the primary predicate device (the Mindray Passport Series Patient Monitors K143195). The features in grey are the features that have been modified since their previous clearances and that are the subject of this 510(k). | | Predicate Device (K143195) | | | Subject Devices | | | |-----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|------|------| | Feature | 17m | 12m | T1 | 17m | 12m | T1 | | Integrated<br>display and<br>touchscreen | 17"<br>1280*1024<br>pixels | 12"<br>800*600<br>pixels | 5", 480*272 pixels | Same | Same | Same | | Secondary<br>display | Independent<br>control and<br>display | Display is<br>linked to<br>integrated<br>display | Independent<br>display and<br>control via a VGA<br>port in the T1<br>docking station | Same | Same | Same | | Additional<br>display<br>features | The minitrends diagram, OxyCRG diagram, other<br>monitor view, and calculation can be viewed<br>when using an external LCD screen | Same | Same | Same | Same | Same | | Wireless<br>module | The ASUS<br>module is<br>used for<br>connecting<br>to a network<br>wirelessly.<br>constructing<br>a monitoring<br>network<br>with a<br>central<br>monitoring<br>system<br>(CMS). | Support<br>wireless<br>function<br>using the<br>ASUS and<br>Silex<br>wireless<br>module. | Support wireless<br>function using the<br>Cyberlink wireless<br>module | Same | Same | Same | | Module rack | Independent of the patient<br>monitor, provides 8<br>standard module slots to<br>extend the measurement<br>capabilities of the system | Same | Not supported | Same | Same | Same | | | Predicate Device (K143195) | | | Subject Devices | | | | Feature | 17m | 12m | T1 | 17m | 12m | T1 | | Power supply | Two<br>rechargeable<br>Lithium-ion<br>battery or<br>AC power<br>supply | One<br>rechargeable<br>Lithium-ion<br>battery or<br>AC power<br>supply | One rechargeable<br>Lithium-ion<br>battery or DC<br>power supply or<br>AC power supply | Same | Same | Same | | Battery | Chargeable Lithium-Ion,<br>11.1 VDC, 4500 mAh, 350 g | Chargeable Lithium-Ion,<br>11.1 VDC, 4500 mAh, 350 g | Chargeable<br>Lithium-Ion, 7.4<br>VDC, 2500 mAh | Same | Same | Same | | External<br>memory card | Compact Flash | Compact Flash | Secure Digital | Same | Same | Same | | Data<br>Recorder | The thermal recorder<br>records patient information,<br>measurement numerics, up<br>to three waveforms, etc. | | Not supported | Same | Same | Same | | Speaker | Give alarm tones (45 to 85 dB), key tones, QRS<br>tones; support PITCH TONE and multi-level<br>tone modulation | | Give alarm tones (45 to 85 dB), key tones, QRS<br>tones; support PITCH TONE and multi-level<br>tone modulation | Same | Same | Same | | Supports T1<br>as a module | Supported | Supported | / | Same | Same | / | | ECG<br>(Mindray<br>Algorithm) | 3-lead , 5-lead or 12-lead selectable, arrhythmia<br>detection, ST segment analysis, an interpretation<br>of resting 12-lead ECG, J-point Auto detection,<br>Dual channel Pace detection and heart rate (HR) | | | Added QT analysis | | | | ECG<br>(Mortara<br>Algorithm) | 3-lead , 5-lead or 12-lead selectable, arrhythmia<br>detection, ST segment analysis, J-point Auto<br>detection, and heart rate (HR) | | | Same | | | | Arrhythmia<br>Analysis<br>(Mindray<br>Algorithm) | Asystole, VFib/VTac, Vtac, Vent. Brady,<br>Extreme Tachy, Extreme Brady, PVC, Couplet,<br>Bigeminy, Trigeminy, R on T, VT>2, PVCs,<br>Tachy, Brady, Missed Beats, Vent. Rhythm,<br>PNP, PNC, Multif. PVC,<br>Nonsus. Vtac, Pause, Irr. Rhythm | | | Added AFib | | | | Arrhythmia<br>Analysis<br>(Mortara<br>Algorithm) | Asystole, Vfib, Vtac, Vent. Rhythm, Couplet,<br>VT>2, Bigeminy, Trigeminy, R on T, Multif.<br>PVC, Irr. Rhythm, Tachy, Brady, Missed Beats,<br>PNP, PNC | | | Same | | | | Adjustable<br>QRS<br>Detection<br>Threshold<br>(Mindray<br>Algorithm) | Not supported | Not supported | Not supported | Added | | | | Respiration<br>rate (Resp) | Measurement range:<br>Adult: 0 to 120 rpm; Pediatric, neonate: 0 to 150 rpm.<br>Accuracy:<br>7 to 150 rpm: ±2 rpm or ±2%, whichever is<br>greater; | Measurement range:<br>Adult: 0 to 120 rpm; Pediatric, neonate: 0 to 150 rpm.<br>Accuracy:<br>7 to 150 rpm: ±2 rpm or ±2%, whichever is<br>greater; | Measurement range:<br>Adult: 0 to 120 rpm; Pediatric, neonate: 0 to 150 rpm.<br>Accuracy:<br>7 to 150 rpm: ±2 rpm or ±2%, whichever is<br>greater; | Same | Same | Same | | | Predicate Device (K143195) | | | Subject Devices | | | | Feature | 17m | 12m | T1 | 17m | 12m | T1 | | Temperature<br>(Temp) | Measurement range: 0 to 50°C (32 to 122°F)<br>Accuracy: ±0.1°C or ±0.2°F (without probe) | | | Same | Same | Same | | Pulse oxygen<br>saturation<br>(SpO2) | Mindray SpO2 Module<br>Measurement range: 0 to 100%<br>Accuracy:70 to 100%: ±2% (adult/pediatric<br>mode)<br>70 to 100%: ±3% (neonate mode)<br>0% to 69%: Not specified.<br><br>Masimo SpO2 Module<br>Measurement range: 1 to 100%<br>Accuracy:70 to 100%: ±2% (measured without<br>motion in adult/pediatric mode)<br>70 to 100%: ±3% (measured without motion in<br>neonate mode)<br>70 to 100%: ±3% (measured with motion)<br>1% to 69%: Not specified.<br><br>Nellcor SpO2 Module<br>Measurement range: 0 to 100%<br>Accuracy:70 to 100%: ±2% (adult/pediatric)<br>70 to 100%: ±3% (neonate)<br>0% to 69%: Not specified. | | | Same | Same | Same | | Pulse rate<br>(PR) | PR from Mindray SpO2 Module<br>Measurement range: 20 to 254 bpm<br>Accuracy:±3 bpm<br><br>PR from Masimo SpO2 Module<br>Measurement range: 25 to 240 bpm<br>Accuracy:±3 bpm (measured without motion)<br>±5 bpm (measured with motion)<br><br>PR from Nellcor SpO2 Module<br>Measurement range: 20 to 300 bpm<br>Accuracy:20 to 250 bpm: ±3 bpm<br>251 to 300 bpm, not specified<br><br>PR from IBP Module<br>Measurement range: 25 to 350 bpmbpm<br>Accuracy:±1 bpm or ±1%, whichever is greater" | | | Same | Same | Same | | Non-invasive<br>blood<br>pressure<br>(NIBP) | Measurement range:<br>Adult<br>Systolic: 40 to 270<br>Diastolic: 10 to 210<br>Mean: 20 to 230<br><br>Pediatric<br>Systolic: 40 to 200<br>Diastolic: 10 to 150<br>Mean: 20 to 165<br><br>Neonate<br>Systolic: 40 to 135<br>Diastolic: 10 to 100<br>Mean: 20 to 110<br><br>Accuracy:<br>Max mean error: ±5 mmHg<br>Max standard deviation: 8 mmHg | | | Same | Same | Same | | | Predicate Device (K143195) | | | Subject Devices | | | | Feature | 17m | 12m | T1 | 17m | 12m | T1 | | Invasive<br>blood<br>pressure<br>(IBP) | Measurement range: -50 to 300 mmHg<br>Accuracy:±2% or ±1 mmHg, whichever is<br>greater (without sensor) | | | Same | Same | Same | | Pulse<br>Pressure<br>Variation<br>(PPV) | Supported feature of IBP | | | Same | Same | Same | | Cardiac<br>output (C.O.) | The temperature change is<br>displayed as a curve in the<br>C.O. split screen, and the<br>monitor calculates the C.O.<br>value from this curve. The<br>monitor is capable of<br>storing 6 measurements. | | Not supported | Same | Same | Same | | Continuous<br>cardiac<br>output (CCO) | CCO/SvO2 interface<br>module is used to interface<br>with Edwards Vigilance II<br>monitor / Vigileo Monitor<br>which measures continuous<br>cardiac output (CCO) and<br>mixed venous oxygen<br>saturation (SvO2). | | Not supported | Same | Same | Same | | Mixed/central<br>venous<br>oxygen<br>saturation<br>(SvO2/ScvO2) | CCO/SvO2 interface<br>module is used to interface<br>with Edwards Vigilance II<br>monitor / Vigileo Monitor<br>which measures mixed<br>venous oxygen saturation<br>(SvO2) and central venous<br>oxygen saturation (ScvO2). | | Not supported | Same | Same | Same | | Carbon<br>dioxide<br>(CO2) | Sidestream CO2 Module:<br>Measurement range: 0 to 99 mmHg<br>Accuracy:<br>0 to 40 mmHg: ±2 mmHg<br>41 to 76 mmHg: ±5% of the reading<br>77 to 99 mmHg: ±10% of the reading<br>Microstream CO2 Module:<br>Measurement range: 0 to 99 mmHg<br>Accuracy:<br>0 to 38 mmHg: ±2 mmHg<br>39 to 99 mmHg: ±5% of the reading ±5% of the<br>reading+0.08% of (the reading-38)<br>Mainstream CO2 Module:<br>Measurement range: 0 to 150 mmHg<br>Accuracy:<br>0 to 40 mmHg: ±2 mmHg<br>41 to 70 mmHg: ±5% of the reading<br>71 to 100 mmHg: ±8% of the reading<br>101 to 150 mmHg: ±10% of the reading | | | Same | Same | Same | | Feature | Predicate Device (K143195) | | | Subject Devices | | | | | 17m | 12m | T1 | 17m | 12m | T1 | | Oxygen (O2) | Oxygen values are<br>calculated based on user<br>input and not directly<br>measured. | | Not supported | Same | Same | Same | | Anesthetic<br>gas (AG) | The AG module can<br>identify two anesthetic<br>gases in a mixture<br>automatically and<br>distinguish between them<br>according to their<br>contributions to the MAC<br>value for display as the<br>primary and secondary<br>anesthetic agents. | | Not supported | Same | Same | Same | | Impedance<br>cardiograph<br>(ICG) | Measurement range:<br>SV: 5 to 250 ml<br>HR: 44 to 2m<br>C.O. 1.4 to 15 L/min<br><br>Accuracy:<br>SV: Not specified.<br>HR: ±2 bpm<br>C.O. Not specified. | | Not supported | Same | Same | Same | | Bispectral<br>index (BIS) | Measured parameters:<br>EEG, BIS, BIS L, BIS R | | Not supported | Same | Same | Same | | Respiration<br>mechanics<br>(RM) | Measurement range: Adult:<br>0 to 120 rpm; Pediatric,<br>neonate: 0 to 150 rpm.<br>Accuracy: 7 to 150 rpm:±2<br>rpm or ±2%, whichever is<br>greater;<br>0 to 6 rpm: Not specified." | | Not supported | Same | Same | Same | | 2 New<br>neonatal pre-<br>wired<br>electrode | Not supported | | | Added | | | #### Device Comparison Table {8}------------------------------------------------ {9}------------------------------------------------ {10}------------------------------------------------ {11}------------------------------------------------ #### Substantial Equivalence Conclusion: Based on the detailed comparison of specifications for each of the modifications to the previously cleared Passport Series devices (K143195), the performance testing and conformance with applicable standards, the Passport Series Patient Monitors (including Passport 12m, Passport 17m and T1) can be found substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...