K101619 · Schiller AG · MHX · Nov 12, 2010 · Cardiovascular
Device Facts
Record ID
K101619
Device Name
WELCH ALLYN 1500 PATIENT MONITOR
Applicant
Schiller AG
Product Code
MHX · Cardiovascular
Decision Date
Nov 12, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Intended Use
The Welch Allyn® 1500 Patient Monitor patient monitoring unit is designed for the monitoring of vital parameters such as ECG, SpO2, etCO2, non invasive blood pressure (NIBP), invasive blood pressure (IBP), temperature and respiration of a patient. The device is intended to be used by qualified doctors or trained medical personnel. The device is not suitable for transport. There is no danger for patients with pacemaker. The device is intended for the monitoring of one patient at a time. The device is not designed for sterile use nor is it designed for outdoor use. Do not use this monitor in areas where there is any danger of explosion or in the presence of flammable gases. The device is classified CF. It is defibrillation protected when the original accessories are used. However, as a safety precaution when possible, remove the electrodes before defibrillation. This product is not designed for direct cardiac application. The arrhythmia module is not intended for use with neonatal patients. The ST-analysis module is not intended for use with neonatal patients.
Device Story
The Welch Allyn 1500 Patient Monitor is an ultra-flat, LCD-equipped bedside monitor for vital sign tracking. It processes inputs from ECG, SpO2, etCO2, NIBP, IBP, temperature, and respiration sensors. The device utilizes integrated, previously cleared algorithms for signal analysis, including Mortara-based arrhythmia and ST-segment analysis. It is operated by qualified medical personnel in clinical settings for single-patient monitoring. Output is displayed on the integrated LCD screen to assist clinicians in real-time patient assessment. The device provides defibrillation protection and is intended for stationary use. It benefits patients by providing continuous, reliable monitoring of physiological status.
Clinical Evidence
No clinical data. Performance was established through bench testing, including electrical safety, electromagnetic compatibility (EMC), and verification of software and parameter-specific algorithms against recognized standards (IEC 60601-1, IEC 60601-2-27, IEC 60601-2-30, IEC 60601-2-34, IEC 60601-2-49, ISO 9919, EN 12470-4).
Technological Characteristics
Ultra-flat patient monitor with LCD display. Sensing principles: ECG, SpO2, etCO2, NIBP, IBP, temperature, respiration. Connectivity: Mains power for stationary use, internal battery. Safety: Class I, Type CF, defibrillation protected. Software: Programmable electrical medical system (IEC 60601-1-4). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-27, IEC 60601-2-30, IEC 60601-2-34, IEC 60601-2-49, ISO 9919, EN 12470-4.
Indications for Use
Indicated for monitoring vital parameters (ECG, SpO2, etCO2, NIBP, IBP, temperature, respiration) in neonates, children, and adults. Arrhythmia and ST-analysis modules are contraindicated for neonatal patients. Device is for use by qualified medical personnel only.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
K122253 — PROCARE MONITOR B20 · Ge Medical Systems China Co., Ltd. · Mar 22, 2013
K011918 — NIHON KOHDEN BSM-2300A SERIES BEDSIDE MONITOR AND ACCESSORIES, MODEL BSM-2300A SERIES · Nihon Kohden America, Inc. · Sep 12, 2001
K192343 — Patient Monitor · Shenzhen Creative Industry Co., Ltd. · Dec 31, 2019
K171580 — Monitor B125, Monitor B105 · Ge Medical Systems (China) Co., Ltd. · Nov 1, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
SCHILLER AG
Altgasse 68, Postfach
CH-6341 Baar
Tel: +41 (41) 766 42 42
Fax: +41 (41) 761 08 80
MWST Nr. 212858 sales@schiller.ch buving@schiller.ch http://www.schiller.ch
SCHILLER
101619
NOV 1 2 2010
# 510 (k) Summary of Safety and Effectiveness
### according to 21 CFR 807.92
#### Address A1
SCHILLER AG Altgasse 68 CH-6341 Baar Switzerland
| Contact Name: | Serkan Sezer | |
|---------------|----------------------------|--|
| Tel: | + 41 41 766 4252 | |
| Date: | 28. <sup>th</sup> May 2010 | |
#### A2 Device Name
| 1. Trade Name: | Welch Allyn 1500 Patient Monitor |
|-----------------|----------------------------------|
| 2. Common Name: | Monitoring System |
#### Legally Marketed Device АЗ
Legally Marketed Device to which this submitted device is compared:
### Parameters – 12-lead Rest-ECG Analysis
The 12-lead Rest-ECG Analysis algorithm used in the 1500 Patient Monitor has been FDAcleared in the SCHILLER ARGUS LCM K053207).
#### Parameters - Temperature
The Patient Temperature parameter circuitry and algorithm used in theWelch Allyn 1500 Patient Monitor has been FDA-cleared in the SCHILLER ARGUS LCM K053207).
#### Parameters - Respiration
The Respiration parameter circuitry and algorithm used in the 1500 Patient Monitor has been FDA-cleared in the SCHILLER ARGUS LCM K053207).
### Parameters --NIBP and Invasive Blood Pressure
The Not Invasive Blood Pressure parameter circuitry and algorithm used in the 1500 Patient Monitor has been FDA-cleared in the SCHILLER BP-200plus (K063814).
### Parameters -Invasive Blood Pressure
The Invasive Blood Pressure parameter circuitry and algorithm used in the 1500 Patient Monitor has been FDA-cleared in the SCHILLER ARGUS LCM K053207).
Page 1 OF 5
{1}------------------------------------------------
41 (41) 761 08 80
MWST Nr 212858 http://www.schiller.ch
### Parameters - Temperature
The Patient Temperature parameter circuitry and algorithm used in the 1500 Patient Monitor has been FDA-cleared in the SCHILLER ARGUS LCM K053207).Parameters - Pulse Oximetry (SpO2)
The SpO2 parameter circuitry and algorithm used in the 1500 Patient Monitor has been FDAcleared in the Nellcor Oximax NPB-40 Pulse Oximeter (K051352).
### Parameters - ECG Arrhythmia / ST-segment Analysis
The Mortara ECG Arrhythmia and ST-segment analysis algorithm used in the 1500 Patient Monitor has been FDA-cleared in the Datascope Passport 2 Vital Signs Monitor (K020550).
#### Parameters - end-tidal CO2
The end-tidal CO2 parameter circuitry and algorithm used in the 1500 Patient Monitor has been FDA-cleared in the Oridion Capnostream10 Bedside Monitor (K060065).
#### A4 Intended Use
The Welch Allyn® 1500 Patient Monitor patient monitoring unit is designed for the monitoring of vital parameters such as ECG, SpO2, etCO2, non invasive blood pressure (NIBP), invasive blood pressure (IBP), temperature and respiration of a patient.
- > The device is intended to be used by qualified doctors or trained medical personnel.
- > The device is not suitable for transport.
- 公 There is no danger for patients with pacemaker.
- The device is intended for the monitoring of one patient at a time. A
- A The device is not designed for sterile use nor is it designed for outdoor use.
- > Do not use this monitor in areas where there is any danger of explosion or in the presence of flammable gases.
- > The device is classified CF. It is defibrillation protected when the original accessories are used. However, as a safety precaution when possible, remove the electrodes before defibrillation.
- > This product is not designed for direct cardiac application.
- The arrhythmia module is not intended for use with neonatal patients. A
- The ST-analysis module is not intended for use with neonatal patients. A
Image /page/1/Picture/21 description: The image contains three different shapes. The first shape is a heart-like shape with a pointed bottom, and it is inside of a square. The second shape is a blob-like shape with some jagged edges, and it is also inside of a square. The third shape is a smaller heart-like shape with some protrusions on top.
Image /page/1/Picture/22 description: The image shows the text "page 2 of 5" written in cursive. The words are written in black ink on a white background. The number 2 is written after the word "page", and the number 5 is written after the word "of".
{2}------------------------------------------------
SCHILLER AG Altgasse 68, Postfach CH-6341 Baar Tel: +41 (41) 766 42 42
Fax: +41 (41) 761 08 80
MWST Nr. 212858 sales@schiller.ch buying@schiller.ch http://www.schiller.ch
Image /page/2/Picture/2 description: The image shows the word "SCHILLER" in a bold, outlined font. The letters are all capitalized and evenly spaced. A horizontal line is present underneath the word, emphasizing it.
A5 Table of Comparison (1)
| | | ARGUS LCM | WA1500PM |
|---|---------------------------|-------------------------------|-------------------------------|
| 1 | Dimensions: | 290x275x180mm 1) | 396 x 284 x 81 mm |
| 2 | Weight: | 4.6 kg / 9.2lbs 2) | 4.6 kg / 9.2lbs 2) |
| 3 | Environmental Conditions: | Indoor use | Indoor use |
| 4 | Operating temperature | +10° - 40° C | +10° - 40° C |
| 5 | Storage temperature | -10° - +50° C | -10° - +50° C |
| 6 | Relative humidity | 25% - 95%<br>(non condensing) | 25% - 95%<br>(non condensing) |
Discussion of Difference:
None of the above difference (1 ) can be considered as safety relevant difference.
We consider the submitted device to be as safe and effective as the Predicate SCHILLER ARGUS LCM (K 053207) device.
| Argus<br>LCM | WA1500PM | Module | Standards | Different |
|--------------|----------|------------|----------------------------------|--------------|
| X | X | Safety | IEC 60601-1 | Same |
| | X | Software | IEC 60601-1-4 | Same |
| X | X | EMC | IEC 60601-1-2 | New Standard |
| X | X | ECG | IEC 60601-1<br>IEC 60601-2-27 | Same |
| X | X | Monitoring | IEC 60601-2-34<br>IEC 60601-2-49 | Same |
| X | X | SPO2 | EN 9919 | New Standard |
| X | X | NIBP | IEC 60601-2-30 | Same |
| X | X | Temp. | EN 12470-4 | Same |
| X | X | CO2 | ISO 21647-1 | New Standard |
| X | X | IBP | IEC 60601-2-34 | Same |
1. Table of Comparison (2)
Image /page/2/Picture/10 description: The image contains three different shapes. The first shape is a heart, the second shape is a letter 'a', and the third shape is an apple. The heart and the letter 'a' are enclosed in a box.
Page 3 OF 5
{3}------------------------------------------------
+41 (41) 761 08 80
MWST Nr 212858 http://www.schiller.ch
#### Non-Clinical Tests BT
- 2. Electrical Safety and Reliability:
- > IEC 60601-1: General requirements for basic safety and essential performance. Protection Class I Type CF
- > IEC 60601-1-4: General requirements for collateral standard: programmable electrical medical systems.
- IEC 60601-1-6: General requirements for safety collateral standard usability A
- > IEC 60601-2-27: Particular requirements for the safety of electrocardiographic monitoring equipment
- > IEC 60601-2-30: Particular requirements for the safety, including essential performance, of automatic cycling non-invasive blood pressure monitoring equipment
- > IEC 60601-2-34: Particular requirements for the safety, including essential performance, of invasive blood pressure monitoring equipment
- > IEC 60601-2-49: Particular requirements for the safety of multifunction patient monitoring equipment
- > ISO 9919. Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use
- 3. Protection Class
- > Protection against electric shocks, Class I according to IEC/EN 60601-1 (with internal power)
- 4. IP- Protection
- > This device is not designed for outdoor use (IPX1)
- 5. EMC
- > IEC/EN 60601-1-2 (class A)
- 6. Additional Requirements
- > EN 1060-1 and 3 (non-invasive blood pressure recorders part 1)
- 7. Conformity
- > CE according to directive 93/42/EEC class Ilb
### 4) Data related to software quality
SCHILLER has reviewed its software development process following the guideline "reviewer guidance for computer controlled medical devices undergoing 510 (k) review". Device software requirements, software structure chart, software development, software revision/ modification, software identification, software verification, validation and testing are described in the data attached.
- Clinical Tests B2
- n.a.
Page 4 OF 5
{4}------------------------------------------------
SCHILLER AG Altgasse 68 Postfach CH-6341 Baar +41 (41) 766 42 42 Fax: +41 (41) 761 08 80
MWST Nr. 212858 sales@schiller.ch buying@schiller.ch http://www.schiller.ch
Image /page/4/Picture/2 description: The image shows the word "SCHILLER" in a bold, outlined font. Below the word is a thin, horizontal line. Under the line, the words "The Art of Diagnostics" are printed in a smaller, sans-serif font.
#### Summary of Performance Testing: ВЗ
The SCHILLER Welch Allyn 1500 Patient Monitoring and associated accessories have been tested and found to comply with recognized national and international performance, safety and electromagnetic compatibility standards for medical devices and product specifications listed in the SCHILLER Welch Allyn 1500 Patient Monitoring labeling.
A risk analysis, identifying potential hazards and documenting mitigation of the hazards, has been developed and verified/validated as part of SCHILLER AG, product development procedures. SCHILLER AG Quality System conforms to 21 CFR 820 and is Certified to ISO 9001:2008 and ISO 13485:2003
## Conclusion
As stated above, SCHILLER AG conclusion is that the SCHILLER Welch Allyn 1500 Patient Monitor is safe, effective, comply with the appropriate medical device standards and equivalent to the SCHILLER ARGUS LCMK053207; SP02 Module Nelcor K051352; CO2 Module Oridien K072295; ECG Library Mortara K020550; NIBPNodule SCHILLER K063814 all parameters currently on the market.
Baar (Switzerland) 28" May 2010
SCHILLER AG
Aligasse 68
CH-6341 Baar/Switzerland
Serkan Sezer
SCHILLER AG Quality and Regulatory Affairs Manager
Image /page/4/Picture/11 description: The image contains three different shapes. The first shape is a heart, the second shape is a kidney, and the third shape is a hand. The heart and kidney shapes are inside of a box, while the hand shape is not.
Paze
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines that suggest wings and a stylized body.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Serkan Sezer Quality and Regulatory Affairs Manager SCHILLER AG Altgasse 68 Baar, Zg SWITZERLAND 6341
NOV 1 2 2010
Re: K101619
Device Name: Welch Allyn® 1500 Patient Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Patient physiological monitor (with arrhythmia detection or alarms) Regulatory Class: Class II (Two) Product Codes: MHX Dated: October 19, 2010 Received: October 20, 2010
Dear Mr. Sezer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice; labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 – Mr. Serkan Sezer
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Arlene B. Boan
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________ K101619
NOV 1 2 2010
Welch Allyn 1500 Patient Monitor Device Name:
Indications For Use:
The Welch Allyn 1500 Patient Monitor is an ultra-flat patient monitor designed for neonates, children, adult and paediatric use. It has a LCD screen for comprehensive vital data monitoring.
Mains power supply is used for stationary use. The Welch Allyn 1500 Patient Monitor also provides full vital data monitoring during transport with the built-in battery.
The Welch Allyn® 1500 Patient Monitor patient monitoring unit is designed for the monitoring of vital parameters such as ECG, SpO2, etCO2, non invasive blood pressure (NIBP), invasive blood pressure (IBP), temperature and respiration of a patient.
- > The device is intended to be used by qualified doctors or trained medical personnel.
- > The device is not suitable for transport.
- > There is no danger for patients with pacemaker.
- > The device is intended for the monitoring of one patient at a time.
- > The device is not designed for sterile use nor is it designed for outdoor use.
- > Do not use this monitor in areas where there is any danger of explosion or in the presence of flammable gases.
- > The device is classified CF. It is defibrillation protected when the original accessories are used. However, as a safety precaution when possible, remove the electrodes before defibrillation.
- > This product is not designed for direct cardiac application.
- > The arrhythmia module is not intended for use with neonatal patients.
- > The ST-analysis module is not intended for use with neonatal patients.
page 1 of 2
{8}------------------------------------------------
Baar (Switzerland) 21th July 2010
SCHILLER AG
Altgasse 6
CH 6341 Baar/Switzerland
SCHILLER AG
Serkan Sezer
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ahlley Bovans for BDZ
(Division Sidn-Off) Division of Cardiovascular Devices
510(k) Number k101619
Page 1 of _2
Page 2 of 2
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.