MODIFICATION TO: POCKETVIEW ECG SOFTWARE

K013311 · Micromedical Industries, Ltd. · MHX · Jan 3, 2002 · Cardiovascular

Device Facts

Record IDK013311
Device NameMODIFICATION TO: POCKETVIEW ECG SOFTWARE
ApplicantMicromedical Industries, Ltd.
Product CodeMHX · Cardiovascular
Decision DateJan 3, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1025
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

PocketView ECG software is a version of Cardioview™ 3000 software, a Windows-based program intended to interpret electrocardiograms. PocketView ECG software receives, displays and stores a single or standard 12 Lead Simultaneous ECG recording using a proprietary digital data transmission protocol. The device contains proprietary software algorithms to receive, store, and interpret the ECG signal. PocketView ECG Software allows the ECG information to be displayed on a commercially available personal digital assistant (PDA) running the Windows CE operating software.

Device Story

PocketView ECG software is a modification of Cardioview™ 3000; enables 12-lead simultaneous ECG data display on Windows CE-based personal digital assistants (PDAs). Device receives ECG signals via proprietary digital transmission protocol; stores and interprets signals using proprietary algorithms. Used by clinicians to view ECG information on mobile PDAs; facilitates clinical decision-making through automated interpretation of ECG recordings.

Clinical Evidence

Bench testing only. Testing verified that the modified device displayed acceptable performance compared to the predicate device.

Technological Characteristics

Software-based ECG interpretation system. Operates on Windows CE-based PDAs. Utilizes proprietary digital data transmission protocol for ECG signal reception. Software algorithm performs signal interpretation.

Indications for Use

Indicated for the interpretation of single or standard 12-lead simultaneous ECG recordings. Intended for prescription use.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KO13311 ### 510(k) SUMMARY ### Submitted by: JAN 0 3 2002 Micromedical Industries Ltd Date Prepared: October 1, 2000 ### Proposed Device: PocketView version of Cardioview™ 3000 software #### Predicate Device: Cardioview™ 3000 software ### Device Description: The proposed device is modification to the Cardioview™ 3000 software that allows the 12 Lead Simultaneous Cable to be linked to commercially available personal digital assistants (PDA) running the Windows CE operating system. ### Statement of Intended Use: PocketView ECG software is a version of Cardioview™ 3000 software, a Windows-based program intended to interpret electrocardiograms. PocketView ECG software receives, displays and stores a single or standard 12 Lead Simultaneous ECG recording using a proprietary digital data transmission protocol. The device contains proprietary software algorithms to receive, store, and interpret the ECG signal. PocketView ECG Software allows the ECG information to be displayed on a commercially available personal digital assistant (PDA) running the Windows CE operating software. ## Summary of Technological Characteristics or New Device to Predicate Devices The technological features of PocketView ECG do not differ significantly from Cardioview™™ 3000 software. The predicate device and the modified device are identical with the exception that the modified device has the connectivity feature allowing the ECG information to be displayed on a commercially available personal digital assistant (PDA) running Windows CE operating system. #### CONFIDENTIAL {1}------------------------------------------------ # Discussion of Nonclinical Tests; Conclusions Drawn from Nonclinical Tests Nonclinical testing was performed to evaluate the modification to the predicate device. Testing verified that the modified device displayed acceptable performance. CONFIDENTIAL {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three abstract shapes that resemble a person with outstretched arms, symbolizing care and protection. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 0 3 2002 Mr. Stephen Cresswell MicroMedical Industries, Ltd. 11 Technology Drive Labrador, Queensland AUSTRALIA Re: K013311 Trade Name: PocketView ECG Software Regulation Number: 21 CFR 870.1025 Regulation Name: Patient Physiological Monitor Regulatory Class: Class III (three) Product Code: MHX Dated: November 12, 2001 Received: December 4, 2001 Dear Mr. Cresswell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Stephen Cresswell Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Odile Tel Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indication for Use 510(k) Number: K013311 PocketView ECG Software Device Name: Indication for Use: PocketView ECG software is a version of Cardioview™ 3000 software, a Windows-based Pocket VIEW DOO soltware is a version di Station and personal digital assistant (PDA). program micindod to merprovers displays and stores a single or standard 12 Lead Focker FCG sontware roost to, is transmitted either locally or transtelephonically using a proprietary digital data transmission protocol. The device contains proprietary software algorithms to receive, store, analyze, and interpret the ECG signal. ## (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) Over-The-Counter Use __ OR
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...