NIHON KOHDEN BEDSIDE MONITOR, BSM 9100A SERIES

K082785 · Nihon Kohden America, Inc. · MHX · Nov 26, 2008 · Cardiovascular

Device Facts

Record IDK082785
Device NameNIHON KOHDEN BEDSIDE MONITOR, BSM 9100A SERIES
ApplicantNihon Kohden America, Inc.
Product CodeMHX · Cardiovascular
Decision DateNov 26, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1025
Device ClassClass 2
AttributesPediatric

Intended Use

The device is intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility. The device is intended to produce a visual record of the electrical signals produced by the heart and monitor the electrocardiogram to generate visible and/or audible alarms when an arrhythmia exists. The device is also intended to monitor heart rate, pulse rate, blood oxygen saturation (SpO2), noninvasive blood pressure (NIBP), invasive blood pressure (IBP), body temperature, Cardiac Output (CO), oxygen concentration (FiO2), Carbon dioxide Concentration (C02) and EtCO2 respiratory rate, BIS and inspired and expired anesthetic agents and gases including CO2, O2, N2O, Halothane, Isoflurane, Enflurane, Sevoofburane and Desflurane. Anesthetic agents and gases are detected using the cleared AG-920RA Anesthetic Agent Detection System K020046 cleared for distribution on July 23, 2002. The device also monitors Ventilators, CO2 Monitors, TOF Monitors, BIS Monitors and CCO/SvO2 Monitors .The device may generate an audible and/or visual alsommore a measured rate falls outside preset limits.

Device Story

BSM-9100A Series Bedside Monitor is a multi-parameter patient monitor for use in medical facilities by clinicians. It processes physiological inputs including ECG, SpO2, NIBP, IBP, temperature, CO, respiratory gases, and anesthetic agents. The device displays up to 16 traces of real-time data, trends, and alarms. It performs ECG analysis for arrhythmia detection and generates audible/visual alarms when parameters exceed preset limits. Clinicians use the visual display and alarm outputs to assess patient status and make clinical decisions. The device supports connectivity to external monitors (Ventilators, BIS, CCO/SvO2). Benefits include continuous patient surveillance and early detection of physiological abnormalities.

Clinical Evidence

Bench testing only. No clinical data provided. Performance verified through electromagnetic, environmental, safety, and software validation testing confirming operation within specifications.

Technological Characteristics

Class II physiological monitor. Features 16-trace display, 8-channel IBP, and selectable QRS detection. Complies with IEC 60601-1 and 21 CFR 868 standards. Power requirement: 210VA. Connectivity includes external monitor integration. Software-based arrhythmia detection and alarm processing.

Indications for Use

Indicated for cardiac and vital signs monitoring of patients within a medical facility. Monitors heart rate, pulse, SpO2, NIBP, IBP, temperature, CO, FiO2, CO2, EtCO2, BIS, and anesthetic agents. Provides arrhythmia detection and alarms.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K082785 P1/3 # NIHON KOHDEN AMERICA, INC. SPECIAL 510(k) NOTIFICATION BSM-9100A Series Bedside Monitor # Name and Address of Applicant Nihon Kohden America, Inc. 90 Icon Street Foothill Ranch, CA 92610 Contact: Jack Coggan NOV 2 6 2008 Director, Regulatory Affairs (949) 580-1555 ex. 3325 Fax: (949) 580-1550 #### Trade/Device Name: BSM-9100A Series Bedside Monitor ## Common or Usual Name: Bedside Monitor, Patient Monitor, Cardiac Monitor, Vital Signs Monitor and Anesthesia Monitor. ### Classification Name: The device has been classified as Class II by the Cardiovascular Device Classification Panel under 21 CFR Part 870.1025, Physiological patient Monitor with Arrabythmia Detection and Alarms as MHX and under 21 CFR 870.2340, ECG Analysis System as LOS. Functions of the device have also been classified as Class II by the Anesthesiology Device Classification Panel. ## Legally Marketed Predicate Device: Nihon Kohden BSM-6000 Series Bedside Monitor and Accessories per 510(k) K080342 commercial distribution certification dated February 26, 2008. ## Intended Use: The device is intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility. The device is intended to produce a visual record of the electrical signals produced by the heart and monitor the electrocardiogram to generate visible and/or audible alarms when an arrhythminar rists. The device is also intended to monitor heart rate, pulse rate, blood oxygen santuation (SpO2), noninvasive blood pressure (NIBP), invasive blood pressure (IBP), body temperature, Cardiac Output (CO), oxygen concentration (FiO2), Carbon dioxide Concentration (C02) and EtCO2 respiratory rate, BIS and inspired and expired anesthetic agents and gases including CO2, O2, N2O, Halothane, Isoflurane, Enflurane, Sevoofburane and Desflurane. Anesthetic agents and gases are detected using the cleared AG-920RA Anesthetic Agent Detection System K020046 cleared for distribution on July 23, 2002. The device also monitors Ventilators, CO2 Monitors, TOF Monitors, BIS Monitors and CCO/SvO2 Monitors .The device may generate an audible and/or visual alsommore a measured rate falls outside preset limits. {1}------------------------------------------------ ### A summary of the technological characteristics of the device compared to the predicate device: Similarities: - . The indications of use are identical. - The fundamental technology has not changed, in that the beside monitor . technology remains the same. - The alarm display, electrosurgery interference filter, heart rate counting method . and VPC counting range remains the same. - The arrhythmia graphic trend, arrhythmia alarm and arrhythmia recall remains the . same. - . The SpO2 display range and declared range remains the same. - The NIBP noninvasive blood pressure methods, display measuring range and . alarm limits remains the same. - . Invasive Pressure (IBP) measuring method, range accuracy, alarm limits, pulse sync tone, pulse rate count range and counting accuracy remains the same. - . Cardiac Output (CO) measuring method, range, accuracy and alarm limits remains the same. - . External communication, user interface, recorder, ECG acquisition and analysis and environmental conditions remain the same. - . Number of ECG electrodes remains the same. - . The heart rate counting, arrhythmia recall, respiration method and user interface all remains the same. Minor Differences: - . Number of traces are increased to 16. - To the arrhythmia alarm parameters section we added the following arrhythmia . events: Extreme Brady, Extreme Tachy, V Brady, SV Tachy, Pause, VPC, V Rhythm. Trigeminy, Irregular RR, No Pacer Pulse and Pacer Non-Capture. - . To the Pulse Alarm Limits section we changed the Upper Limits to 31 to 250 bpm. Lower Limits to 30 to 249 bpm. - To the O2 Alarms Limits section we changed the Upper Limits to 11 to 100% and . the Lower Limits to 10 to 99%. - . To the Invasive Blood Pressure section we changed the number of channels to 8 on the new device. - To the Power Requirements section we changed to 210VA (Main Unit). . - The change of QRS detection which is divided into three and becomes selectable ◆ to be optimized for each patient (adult, child/neonate). {2}------------------------------------------------ K082785 P3/3 SPECIAL 510(k) NOTIFICATION BSM-9100A Series Bedside Monitor #### 510(k) Summary: - . The device is not sterile. - The device does not directly contact patients. Accessories that contact patients, . such as probes and thermistors, are the same accessories as used with other legally marketed products or are comprised of the same component materials as the predicate accessories. Therefore, good laboratory practice studies were not required per 21 CFR Part 58. - The device complies with the IEC 60601-1 standard and sub-clause 56.3 (c) . implemented by 21 CFR Part 868 Performance Standard for Electrode Lead Wires and Patient Cables. - The BSM-9100A Series Bedside Monitor device was subjected to tests for . electromagnetic, environmental, safety and performance testing procedures. These tests verified the operation of the device. The software validation tested the operation of the software function of the device. The results confirmed that the device performed within specifications. - Therefore, Nihon Kohden believes that the device is substantially equivalent to . the predicate device. {3}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV 2 6 2008 Nihon Kohden America, Inc. c/o Mr. Jack Coggan Director, Regulatory Affairs 90 Icon Street Foothill Ranch, CA 92610 Re: K082785 Trade/Device Name: BSM-9100A Series Bedside Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Patient Physiological Monitor (with arrhythmia detection or alarms) Regulatory Class: Class II Product Codes: MHX Dated: October 29, 2008 Received: November 3, 2008 Dear Mr. Coggan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices markcted in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {4}------------------------------------------------ #### Page 2 - Mr. Jack Coggan Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050, This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html. Sincerely yours, 01 Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### NIHON KOHDEN AMERICA, INC. #### G. Indications for Use Statement: # 510(k) Number (if known): KO8278 ک Device Name: BSM-9100A Series Bedside Monitor #### Indications of Use: The device is intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility. The device is intended to produce a visual record of the electrical signals produced by the heart and monitor the electrocardiogram to generate visible and/or audible alarms when an arrhythmia exists. The device is also intended to monitor heart rate, blood oxygen saturation (SpO2), noninvasive blood pressure (NIBP), invasive blood pressure (IBP), body temperature, Cardiac Output (CO), oxygen concentration (FiO2), Carbon dioxide Concentration (C02) and EtCO2 respiratory rate, BIS and inspired and expired anesthetic agents and gases including CO2, O2, N2O, Halothane, Isoflurane, Enflurane, Sevoffurane and Desflurane. Anesthetic agents and gases are detected using the cleared AG-920RA Anesthetic Agent Detection System K020046 cleared for distribution on July 23, 2002. The device also monitors Ventilators, CO2 Monitors, TOF Monitors, BIS Monitors and CCO/SvO2 Monitors .The device may generate an audible and/or visual alarm when a measured rate falls outside preset limits. | Prescription Use | X | AND/OR | Over The Counter Use | | |-----------------------------|---|--------|------------------------|--| | (Part 21 CFR 801 Subpart D) | | | (21 CFR 807 Subpart C) | | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE) | | for B Zuckerman | |------------------------------------|-----------------| | (Division Sign-Off) | 11/26/08 | | Division of Cardiovascular Devices | | | Page 10 of 28 | 510(k) Number | K082785 | |---------------|---------------|---------| |---------------|---------------|---------|
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...