K102239 · GE Healthcare Finland Oy · MHX · Oct 18, 2010 · Cardiovascular
Device Facts
Record ID
K102239
Device Name
CARESCAPE MONITOR B650
Applicant
GE Healthcare Finland Oy
Product Code
MHX · Cardiovascular
Decision Date
Oct 18, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Intended Use
The CARESCAPE Monitor B650 is a multi-parameter patient monitor intended for use in multiple areas and intrahospital transport within a professional healthcare facility. The CARESCAPE Monitor B650 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE Monitor B650 is indicated for monitoring and recording of, and to generate alarms for, hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, cardiac output, temperature and mixed venous oxygen saturation), impedance respiration, airway gases (CO2, O2, N2O and anesthetic agents), spirometry, gas exchange, and neurophysiological (including electroencephalography, Entropy, Bispectral Index (BIS) and neuromuscular transmission) status. The CARESCAPE Monitor B650 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network. The CARESCAPE Monitor B650 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility. The CARESCAPE Monitor B650 is not intended for use during MRI.
Device Story
Multi-parameter patient monitor for professional healthcare facilities; monitors adult, pediatric, and neonatal patients. Inputs: physiological signals via connected modules (ECG, invasive/non-invasive pressure, pulse oximetry, cardiac output, temperature, venous oxygen saturation, impedance respiration, airway gases, spirometry, neurophysiological signals). Operation: processes inputs via CARESCAPE Software Platform (ESP) and EK-Pro V12 arrhythmia detection algorithm; displays data on 15-inch screen (optional touch). Output: real-time physiological data, diagnostic analysis, and alarms. Used by clinicians for patient status assessment; supports stand-alone use, remote viewing, and network integration with other medical devices/data management systems. Benefits: continuous patient monitoring, early detection of physiological changes, and centralized data access.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including module verification, system integration testing, performance testing, safety testing, and design/risk analysis.
Technological Characteristics
Multi-parameter monitor with 15-inch display, integrated keypad, and modular frame. Connectivity: cabled/wireless network interface for remote viewing and data management. Software: CARESCAPE Software Platform (ESP) and EK-Pro V12 arrhythmia detection algorithm. Power: battery-operated or AC. Intended for professional healthcare environment.
Indications for Use
Indicated for monitoring, recording, and alarming of hemodynamic, respiratory, gas exchange, and neurophysiological status in adult, pediatric, and neonatal patients within professional healthcare facilities. Contraindicated for use during MRI.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
DATEX-OHMEDA S/5 ANESTHESIA MONITOR WITH L-ANE07 AND L-ANE07A SOFTWARE (K092680)
Related Devices
K131223 — CARESCAPE MONITOR B650 · GE Healthcare Finland Oy · Aug 28, 2013
K191249 — CARESCAPE B450 · GE Healthcare Finland Oy · Jan 24, 2020
K191323 — Carescape B850 · GE Healthcare Finland Oy · Jan 29, 2020
K151063 — Monitor B40 · Ge Medical Systems China Co., Ltd. · Dec 15, 2015
K102004 — ENDEAVOUR MONITORING SYSTEM · Mindray DS USA, Inc. · Sep 10, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
K102239
P1/3
OCT 1 8 2010
CE Healthcare Finland Ov
Ki-000031 GES
FI-00031 GES
Ti-00031 GES
Ti-00031 GES
Ti-00031 GES
Ti-1-358 10 30411
F: +358 9 14633110
510(k) Summary
:
In accordance with 21 CFR 807.92 the following summary of information is provided:
| <b>Date:</b> | August 5, 2010 |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Submitter:</b> | GE Healthcare Finland Oy. Kuortaneenkatu 2 Helsinki, Finland FIN-00510 |
| <b>Primary Contact Person:</b> | Päivi Roiha Regulatory Affairs Leader GE Healthcare Finland Oy Kuortaneenkatu 2 Helsinki, Finland FIN-00510 Phone: + 358 10 394 3731 Fax: +358-92726532 E-mail: paivi.roiha@ge.com |
| <b>Secondary Contact Person:</b> | Joel Kent Manager, Quality and regulatory Affairs GE Healthcare Phone: 781-449-8685 Fax: 781-433-1344 |
| <b>Device:</b> | CARESCAPE <sup>TM</sup> Monitor B650 |
| <b>Common/Usual Name:</b> | multi-parameter patient monitor |
| <b>Classification Names:</b> | 21 CFR 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm) |
| <b>Product Code:</b> | MHX |
Product Code:
:
. . . .
. ·
Image /page/0/Picture/7 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circle. The logo is black and white.
: .
{1}------------------------------------------------
K102239
p2/3
## K092027 CARESCAPE™ Monitor B850 Predicate Device(s):
K061185 DATEX-OHMEDA S/5 COMPACT ANESTHESIA MONITOR WITH L-CANE05(A) SOFTWARE
K031320 EK-PRO ARRHYTHMIA DETECTION ALGORITHM (EK-Pro VII)
K092680 DATEX-OHMEDA S/5 ANESTHESIA MONITOR WITH L-ANE07 AND L-ANE07A SOFTWARE, USING F-CU8 OR F-CU5(P) MONITOR FRAME OPTIONS AND E-EXT EXTENSION MODULE AND E-REC RECORDER MODULE
- The CARESCAPE Monitor B650 is a multi-parameter patient monitor Device Description: including both new and existing subsystems. The CARESCAPE Monitor B650 includes the monitor itself, the CARESCAPE Software Platform (also called ESP software) and the battery. The CARESCAPE Monitor B650 itself has 15 inch display (optional touch screen) with integrated keypad and a frame for parameter measurement modules. A variety of options are available to the customer including additional displays, various input devices (keyboard, mouse, bar code reader, and corded remote control), and physiological parameter measurement modules, which are existing subsystems. The CARESCAPE Monitor B650 communicates to a variety of existing OEM medical devices. The CARESCAPE Monitor B650 interfaces to a variety of other existing patient monitoring systems via a cabled or wireless network interface. The CARESCAPE Monitor B650 includes features and subsystems that are optional or configurable.
The CARESCAPE Monitor B650 is a multi-parameter patient monitor Intended Use: intended for use in multiple areas and intrahospital transport within a professional healthcare facility.
> The CARESCAPE Monitor B650 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time.
The CARESCAPE Monitor B650 is indicated for monitoring and recording of, and to generate alarms for, hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, cardiac output, temperature and mixed venous oxygen saturation), impedance respiration, airway gases (CO2, O2, N2O and anesthetic agents), spirometry, gas exchange, and neurophysiological (including electroencephalography, Entropy, Bispectral Index (BIS) and neuromuscular transmission) status.
The CARESCAPE Monitor B650 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network.
The CARESCAPE Monitor B650 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
The CARESCAPE Monitor B650 is not intended for use during MRI.
Technology:
The CARESCAPE Monitor B650 is a new monitor that essentially is a combination of the features and parameters of two existing predicate monitor
{2}------------------------------------------------
K102239
p3/3
platforms. The predicate monitors are CARESCAPE™ Monitor B850 (K092027) and the DATEX-OHMEDA S/5 COMPACT ANESTHESIA MONITOR WITH L-CANE05(A) SOFTWARE (K061185).
The CARESCAPE Monitor B650 has the EK-Pro Arrhythmia Detection Algorithm, EK- Pro V12 that employs the same functional technology as the predicate device(s) in the monitoring of ECG parameter data. The predicate algorithms are EK-PRO ARRHYTHMIA DETECTION ALGORITHM, EK-Pro V11 (K031320) and DATEX-OHMEDA S/5 ANESTHESIA MONITOR WITH L-ANE07 AND L-ANE07A SOFTWARE, USING F-CU8 OR F-CU5(P) MONITOR FRAME OPTIONS AND E-EXT EXTENSION MODULE AND E-REC RECORDER MODULE (K092680).
The fundamental technology of the CARESCAPE Monitor B650 is the same as the predicate devices.
The CARESCAPE Monitor B650 device is as safe and effective the predicate devices.
Determination of Substantial Equivalence:
Summary of Non-Clinical Tests:
The CARESCAPETM Monitor B650 and its applications comply with voluntary standards as detailed in this premarket submission. The following quality assurance measures were applied to the development of the system:
- Risk Analysis .
- . Requirements Reviews
- . Design Reviews
- Testing on unit level (Module verification) .
- Integration testing (System verification) .
- Final acceptance testing (Validation)
- Performance testing (Verification)
- Safety testing (Verification)
Summary of Clinical Tests:
The subject of this premarket submission. CARESCAPE™ Monitor B650 did not require clinical studies to support substantial equivalence.
GE Healthcare considers the CARESCAPE™ Monitor B650 to be as safe, as Conclusion: effective, and performance is substantially equivalent to the predicate device(s).
22
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
GE Healthcare Finland Oy c/o Ms. Paiva Roiha Regulatory Affairs Leader Kuortaneenkatu 2 Helsinki, Finland FIN-00510
OCT 1 8 2010
K102239 Re:
Trade/Device Name: CARESCAPE™ Monitor B650 Regulatory Number: 21 CFR 870.1025 Regulation Name: Patient Physiological Monitor (with arrhythmia detection or alarms) Regulatory Class: II (two) Product Code: 74 MHX Dated: August 5, 2010 Received: August 9, 2010
Dear Ms. Roiha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Paiva Roiha
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K102239
## Indications for Use
510(k) Number (if known):
OCT 1 8 2010
## Device Name: CARESCAPE™ Monitor B650
Indications for use:
The CARESCAPE Monitor B650 is a multi-parameter patient monitor intended for use in multiple areas and intrahospital transport within a professional healthcare facility.
The CARESCAPE Monitor B650 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time.
The CARESCAPE Monitor B650 is indicated for monitoring and recording of, and to generate alarms for, hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, cardiac output, temperature and mixed venous oxygen saturation), impedance respiration, airway gases (CO2, 02, N20 and anesthetic agents), spirometry, gas exchange, and neurophysiological (including electroencephalography, Entropy, Bispectral Index (BIS) and neuromuscular transmission) status.
The CARESCAPE Monitor B650 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network.
The CARESCAPE Monitor B650 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
The CARESCAPE Monitor B650 is not intended for use during MRI.
Over-The-Counter Use Prescription Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
W.M.S.
510(k) Number K102235
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.