INTELLIVUE PATIENT MONITOR
K131829 · Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · MHX · Aug 16, 2013 · Cardiovascular
Device Facts
| Record ID | K131829 |
| Device Name | INTELLIVUE PATIENT MONITOR |
| Applicant | Philips Medizinsysteme Boeblingen GmbH, Cardiac AN |
| Product Code | MHX · Cardiovascular |
| Decision Date | Aug 16, 2013 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 870.1025 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Intended Use
The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment. The MP5 and MP5SC monitors are also intended for use during patient transport inside the hospital environment; only the MP5 monitor is for use during patient transport outside of the hospital environment. The MP5 and MP5SC when used with the TRx4841A/TRx4851A IntelliVue Telemetry System Transceiver or with the IntelliVue Cableless Measurement Devices, are intended for use in a hospital environment and during patient transport inside the hospital environment. The monitors are only for use on one patient at a time. They are not intended for home use. Not therapeutic devices. The monitors are for prescription use only. The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11). ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients. The Predictive Temperature unit is intended for use with adult and pediatric patients in a hospital environment. The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only. The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients. The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.
Device Story
Multi-parameter patient monitors (MP5/MP5SC) collect physiological signals (ECG, invasive/non-invasive pressure, temperature, etc.) via sensors; process data through internal CPU; generate real-time alarms and store derived measurements. Used by healthcare professionals in hospitals and during transport. Output displayed on integrated screen; data communicated to central stations via IntelliVue Clinical Network. Modification adds support for Genius 2 Tympanic Temperature Probe. Clinical decision support tools (ProtocolWatch, IPI, PPV) assist providers in monitoring patient status; output informs clinical decisions regarding patient care and intervention. Benefits include continuous, centralized monitoring of critical physiological parameters.
Clinical Evidence
Bench testing only. Verification and validation activities included system-level testing, hazard analysis, and environmental/mechanical testing (temperature, humidity, shock, vibration, free fall). Testing confirmed compliance with IEC 60601-1 and IEC 60601-1-2 standards. No clinical trials were conducted for this modification.
Technological Characteristics
Multi-parameter monitor with integrated display and CPU. Supports ECG, invasive/non-invasive pressure, temperature, and gas analysis. Connectivity via IntelliVue Clinical Network. Software revision J.21. Complies with IEC 60601-1 (safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for monitoring physiological parameters in adults, pediatrics, and neonates in hospital environments or during transport. Includes ECG, ST-segment (adults only), predictive temperature (adults/pediatrics), Sepsis Protocol (adults), IPI (adults/pediatrics 1-12 years), and PPV (sedated adult patients on controlled ventilation). Not for home use.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
- IntelliVue Patient Monitors MP5 and MP5SC (K122439, K120366, K113657, K113441, K110474, K110622, K102562, K101449, K100939, K093268, K091927, K091395, K083517, K083228, K082633, K082583, K081793, K072070, K071426, K063725, K063315, K062392)
Reference Devices
- Genius 2 Tympanic Temperature Probe (K060649)
- TRx4841A/TRx4851A IntelliVue Telemetry System Transceiver
Related Devices
- K062392 — PHILIPS MP5 INTELLIVUE PATIENT MONITOR · Philips Medizin Systeme Boeblingen GmbH · Sep 6, 2006
- K102562 — INTELLIVUE PATIENT MONITOR, MODELS MP2, X2, MP5, MP5T, MP20, MP30, MP40, MP50, MP60, MP70, MP80, MP90, & MX800 · Philips Medizin Systeme Boblingen GmbH · Oct 21, 2010
- K041235 — PHILIPS MP20, MP30, MP40, MP50. MP60, MP70 AND MP90 INTELLIVUE PATIENT MONITORS · Philips Medizin Systeme Boeblingen GmbH · Jun 4, 2004
- K061610 — PHILIPS MODEL MP20, MP30, MP40, MP50, MP60, MP70, MP80 AND MP90 INTELLIVUE PATIENT MONITORS · Philips Medizin Systeme Boeblingen GmbH · Aug 14, 2006
- K050762 — PHILIPS INTELLIVUE PATIENT MONITOR, MODELS MP20, MP30, MP40, MP50, MP60, MP70 AND MP90 · Philips Medizin Systeme Boeblingen GmbH · Apr 22, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
## K131829 Page 1 of 5
AUG 1 6 2013
#### 510 (k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.
1. The submitter of this premarket notification is:
Herbert van Dyk Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany Tel: ++49 7031 463-1734 Fax: ++49 7031 463-2442 e-mail: herbert.van dyk@philips.com
This summary was prepared on June 14, 2013.
2. Names and classification of the devices
| Trade name: | IntelliVue Patient Monitors MP5 and MP5SC<br>Revision J.21. |
|--------------|-------------------------------------------------------------|
| Common name: | Multiparameter Patient Monitor |
Classification names are as follows:
| Device Panel | Classification | ProCode | Device Description |
|-------------------------------------------------|----------------|---------|------------------------------------------------------------------------------|
| Cardiovascular<br>Devices | §870.1025, II | DSI | Detector and alarm, arrhythmia |
| | §870.1025, II | MLD | Monitor, ST Segment with Alarm |
| | §870.1025, II | MHX | Monitor, Physiological, Patient<br>(with arrhythmia detection or<br>alarms) |
| | §870.1100, II | DSJ | Alarm, Blood Pressure |
| | §870.1110, II | DSK | Computer, Blood Pressure |
| | §870.1130, II | DXN | System, Measurement, Blood-<br>Pressure, Non-Invasive |
| | §870.1435, II | DXG | Computer, Diagnostic, Pre-<br>Programmed, Single-Function |
| | §870.1915, II | KRB | Probe, Thermodilution |
| | §870.2060, II | DRQ | Amplifier and Signal<br>Conditioner, Transducer Signal |
| | §870.2300, II | DRT | Monitor, Cardiac (incl.<br>Cardiotachometer & Rate Alarm) |
| | §870.2340, II | DPS | Electrocardiograph |
| | §870.2340, II | MLC | Monitor, ST Segment |
| | §870.2350, II | DRW | Electrocardiograph, Lead<br>Switching Adapter |
| | §870.2370, II | KRC | Tester, Electrode, Surface,<br>Electrocardiograph |
| | §870.2600, I | DRJ | System, Signal Isolation |
| | §870.2700, II | DQA | Oximeter |
| | §870.2770, II | DSB | Plethysmograph, Impedance |
| | §870.2800, II | DSH | Recorder, Magnetic tape,<br>Medical |
| | §870.2810, I | DSF | Recorder, Paper Chart |
| Device Panel | Classification | ProCode | Device Description |
| | §870.2850, II | DRS | Extravascular Blood Pressure<br>Transducer |
| | §870.2900, I | DSA | Cable, Transducer and<br>Electrode, incl. Patient<br>Connector |
| | - | MSX | System, Network and<br>Communication, Physiological<br>Monitors |
| | §870.2910, II | DRG | Transmitters and Receivers,<br>Physiological Signal,<br>Radiofrequency |
| Anesthesiology<br>Devices | §868.1400, II | CCK | Analyzer, Gas, Carbon Dioxide,<br>Gaseous-Phase |
| | §868.1500, II | CBQ | Analyzer, Gas, Enflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
| | §868.1500, II | NHO | Analyzer, Gas, Desflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
| | §868.1500, II | NHP | Analyzer, Gas, Sevoflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
| | §868.1500, II | NHQ | Analyzer, Gas, Isoflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
| | §868.1620, II | CBS | Analyzer, Gas, Halothane,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
| | §868.1700, II | CBR | Analyzer, Gas, Nitrous Oxide,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
| | §868.1720, II | CCL | Analyzer, Gas, Oxygen, Gaseous-<br>Phase |
| | §868.1880, II | BZC | Data calculator Pulmonary-<br>function |
| | §868.2375, II | BZQ | Monitor, Breathing Frequency |
| | §868.2480, II | LKD | Monitor, Carbon Dioxide,<br>Cutaneous |
| | §868.2500, II | KLK | Monitor, Oxygen, Cutaneous, for<br>Infant not under Gas Anesthesia |
| | §868.2775, II | KOI | Electrical peripheral nerve<br>Stimulator |
| General Hospital<br>and Personal Use<br>Devices | §880.2910, II | FLL | Thermometer, Electronic,<br>Clinical |
| Neurological<br>Devices | §882.1400, II | GWR | Electroencephalograph |
| | §882.1420, I | GWS | Analyzer, Spectrum,<br>Electroencephalogram Signal |
{1}------------------------------------------------
## K131829
# Page 2 of 5
- 3. The subject devices Philips IntelliVue Patient Monitors MP5 and MP5SC, both with software revision J.21 are substantially equivalent to previously cleared Philips IntelliVue Patient Monitors MP5 and MP5SC marketed pursuant to K122439, K120366, K113657, K113441, K110474, K110622, K102562, K101449, K100939,
{2}------------------------------------------------
#### Page 3 of 5
K093268, K091927, K091395, K083517, K083228, K082633, K082583,
K081793, K072070, K071426, K063725, K063315, K062392
4. Description of the Devices
The Philips IntelliVue Patient Monitor family comprises the multi-parameter patient monitor models: IntelliVue Patient Monitors MP5 and MP5SC, that consist of display units including built-in and central processing units (CPU) and physiological measurements. All monitors share the same system architecture and exactly the same software is executed on each monitor.
The IntelliVue Patient Monitors measure multiple physiological parameters such as surface ECG, invasive and non-invasive etc., generate alarms, record physiological pressure, -------------------------------------------------------------------------------------------------------------------------------------------------------------------signals, store derived data, and communicate derived data and alarms to central stations via the IntelliVue Clinical Network.
subject modification introduces the optional The Philips Tympanic Temperature Module (866149) for the IntelliVue Patient Monitors MP5 and MP5SC. This Tympanic Temperature extends the capability of the IntelliVue Patient Module Monitors MP5 and MP5SC to interface the Genius™ 2 Tympanic Temperature Probe (cleared by Covidien, K060649).
Additionally the software revision J.21 is made available for the IntelliVue Patient Monitors MP5 and MP5SC.
5. Intended Use
The subject devices IntelliVue Patient Monitors MP5 and MP5SC have the same Intended Use as the legally marketed predicate IntelliVue Patient Monitors MP5 and MP5SC. The Indications for Use of the subject Philips IntelliVue Patient Monitors MP5 and MP5SC have not changed as a result of the device modification.
The subject and predicate devices IntelliVue Patient Monitors MP5 and MP5SC have the following detailed Indications for Use Statements in their Instructions for Use:
The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
The monitors are intended to be used for monitoring and recording of, and to generate alams for, multiple
physiological parameters of adults, pediatrics, and neonates. monitors are intended for use by trained healthcare The professionals in a hospital environment.
The MP5 and MP5SC monitors are also intended for use during patient transport inside the hospital environment; only the MP5 monitor is for use during patient transport outside of the hospital environment. The MP5 and MP5SC when used with the TRx4841A/TRx4851A IntelliVue Telemetry System Transceiver or with the IntelliVue Cableless Measurement Devices, are intended for use in a hospital environment and during patient transport inside the hospital environment.
{3}------------------------------------------------
#### K131829
#### Page 4 of 5
The monitors are only for use on one patient at a time. They are not intended for home use. Not therapeutic devices. The monitors are for prescription use only. The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11). ST seqment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients. The Predictive Temperature unit is intended for use with adult and pediatric patients in a hospital environment. The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only. The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients. The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.
- 6. Technological Characteristics
The subject devices IntelliVue Patient Monitors MP5 and MP5SC have the same technological characteristics such as design, energy sources, portability, user interface, performance, and robustness as the legally marketed predicate IntelliVue Patient Monitors MP5 and MP5SC devices. The physical specifications of all devices and performed V&V testing have demonstrated the equivalence of the devices with respect to the intended use. Biocompatibility aspects of the Philips IntelliVue Patient Monitors MP5 and MP5SC are not affected and have not changed, the Covidien Genius™ 2 Tympanic Temperature Probe was cleared by Covidien under K060649.
- 7. Summary of Verification & Validation Activities
Testing according to the recognized consensus standards:
- . IEC 60601-1-2:2001 +A1:2004 (EMC)
- . IEC 60601-1:1988 +A1:1991+A2:1995 (General Safety) All applicable requirements have been met.
Environmental testing (temperature, humidity), mechanical testing (shock, vibration, and free fall).
All specified test requirements have been met. The tests confirmed that the IntelliVue Patient Monitors MP5 and MP5SC worked safely and according to their specifications and indicated claims during tests simulating general hospital conditions, handling and transport in hospital environments, and storage.
Software Functionality and as required by the Hazard Analysis testing on the IntelliVue Patient Monitors MP5 and MP5SC.
{4}------------------------------------------------
### K131829
## Page 5 of 5
All specified criteria have been met. The tests demonstrated that the IntelliVue Patient Monitors MP5 and MP5SC worked safely, effectively, and correctly in accordance with specifications and labeling claims in the intended environment of use.
- 8. Conclusion
Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified devices with respect to the predicate. Testing involved system level and as well as testing from the hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate devices and test results showed substantial equivalence. The results demonstrate that the Philips IntelliVue Patient Monitors MP5 and MP5SC meet all reliability requirements and performance claims.
{5}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 16, 2013
Philips Medizinsysteme Boeblingen Gmbh c/o Mr. Herbert Van Dyk Quality and Regulation Function Manager Hewlett-packard Str. 2 Boeblingen, GM D 71034
Re: K131829
Trade/Device Name: Intellivue Patient Monitor (MP5 and MPSSC) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II Product Code: MHX, DRT, DSI, MLD, DSJ, DSK, DXN, DXG, KRB, DRQ, DPS, MLC, DRW, KRC, DRJ, DQA, DSP, DSH, DSF, DRS, DSA, MSX, DRG, CCK, CBQ, NHO, NHP, NHQ, CBS, CBR, CCL, BZC, BZQ, LKD, KLK, KOI, GWR, GWS, FLL Dated: July 15, 2013 Received: July 17, 2013
Dear Mr. Herbert Van Dyk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading,
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{6}------------------------------------------------
Page 2 - Mr. Herbert Van Dyk
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hillemann
for
for Bram D. Zuckerman, Ph.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### 3.1 -Indications Statement
#### Indications for Use MP5 and MP5SC
510 (k) Number (if known) : __
Device Name: Philips IntelliVue patient monitors MP5 and MP5SC, software revision J.21.
MP5 and MP5SC IntelliVue Patient Monitors:
The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment.
The MP5 and MP5SC monitors are also intended for use during patient transport inside the hospital environment; only the MP5 monitor is use during patient transport outside of the hospital for environment. The MP5 and MP5SC when used with the TRx4841A/TRx4851A IntelliVue Telemetry System Transceiver or with the IntelliVue Cableless Measurement Devices, are intended for use in a hospital environment and during patient transport inside the hospital environment.
The monitors are only for use on one patient at a time. They are not intended for home use. Not therapeutic devices. The monitors are for prescription use only.
The ECG measurement is intended to be used for diagnostic recording rhythm and detailed morphology of complex cardiac complexes of (according to AAMI EC 11) .
ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.
The Predictive Temperature measurement is intended for use with adult and pediatric patients in a hospital environment.
The SSC Sepsis Protocol, in the Protocol Watch clinical decision support tool, is intended for use with adult patients only.
The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.
The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.
Over-The-Counter Use No Prescription Use Ycs_ AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M. A. Lillemann
Page 15