NIHON KOHDEN BSM-2300A SERIES BEDSIDE MONITOR AND ACCESSORIES, MODEL BSM-2300A SERIES

K011918 · Nihon Kohden America, Inc. · MHX · Sep 12, 2001 · Cardiovascular

Device Facts

Record IDK011918
Device NameNIHON KOHDEN BSM-2300A SERIES BEDSIDE MONITOR AND ACCESSORIES, MODEL BSM-2300A SERIES
ApplicantNihon Kohden America, Inc.
Product CodeMHX · Cardiovascular
Decision DateSep 12, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1025
Device ClassClass 2
AttributesPediatric

Intended Use

The device is intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility. The device is intended to monitor the electrocardiogram to generate audible and/or visible alarms when an arrhythmia exists. The device is also intended to monitor heart rate, pulse rate, blood oxygen saturation (SpO2) noninvasive blood pressure (NIBP) invasive blood pressure (IBP), body temperature, carbon dioxide concentration (CO2 and EtCO2), and respiratory rate. The device may generate an audible and/or visual alarm when a measured rate falls outside preset limits. The device may also condition and transmit physiological signals via radio frequency. The device will be available for use by medical personnel on all patient populations.

Device Story

BSM-2300A Series Bedside Monitor is a multi-parameter patient monitor for clinical use by medical personnel. Device inputs include physiological signals (ECG, SpO2, NIBP, IBP, temperature, CO2, respiratory rate) via patient-connected sensors/cables. System processes these inputs to display waveforms and numerics on a color LCD touchscreen; provides arrhythmia detection and alarm generation for out-of-limit parameters. Features include optional thermal array printer, network communication, and RF signal transmission. Output allows clinicians to observe real-time patient status, facilitating timely clinical intervention. Benefits include continuous patient surveillance and automated alerting for critical physiological changes.

Clinical Evidence

Bench testing only. No clinical data provided. Device performance verified through environmental stress testing (temperature/humidity), electromagnetic interference/compatibility (EMC) testing, and safety standards testing (IEC 60601-1, 60601-1-1, 60601-1-2, 60601-2-30, 60601-2-34). Software validation confirmed operation within specifications.

Technological Characteristics

Multi-parameter bedside monitor with color LCD touchscreen. Includes thermal array printer, battery pack, and RF transmission capability. Complies with IEC 60601-1, 60601-1-1, 60601-1-2, 60601-2-30, 60601-2-34, and CISPR11. Patient-contacting accessories (e.g., SpO2 probes) utilize materials tested per ISO 10993. Software-driven system for signal processing and alarm management.

Indications for Use

Indicated for cardiac and vital signs monitoring of all patient populations in medical facilities. Monitors ECG (arrhythmia detection), heart rate, pulse rate, SpO2, NIBP, IBP, body temperature, CO2/EtCO2, and respiratory rate. Provides audible/visual alarms for arrhythmias or out-of-limit physiological measurements.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Koll918 P NIHON KOHDEN AMERICA, INC. June 19, 2001 # 510(k) NOTIFICATION BSM-2300A Series Bedside Monitor SEP 1 2 2001 # SECTION 2 - 510(K) SUMMARY ### Name and Address of Applicant Nihon Kohden America, Inc. Contact: Regulatory Affairs Manager 90 Icon Street Foothill Ranch, CA 92610 Phone: (949) 580-1555 Fax: (949) 580-1550 Device Name: BSM-2300A Series Bedside Monitor. Common names for the device include Bedside Device Name: Dow 2007 Coence Bouttor, Cardiac Monitor and Vital Signs Monitor. The classification Monitor, F Gallerinention Patient Monitor with Arrhytmia Detection and Alarms Legally Marketed Predicate: Nihon Kohden BMS-4100A Series Bedside Monitor per 510(k)# K001693. Description and Intended Use: The device is a multi-parameter monitor consisting of a color LCD touchscreen to display waveforms and numerics of monitored parameters, multi-parameter input unit (socket Screen to display waronomic and namovable battery pack. Options include a built-in thermal array pailer, visual alanm lindications card. The device is software driven. Both the device and the starside of recorder and hetwork communitor, direr it is display and record physiological data to provide cardiac predicale have the same intention association and the is intended to monitor the electrocaciliogram and vital visible and/or audible alarms when an arrhythmia exists. The device is also intended to monitor heart rate, pulse rate, blood oxygen saturation (SpO2) noninvasive blood pressure (NIBP) invasive monitor mean rate, pulse fate, plood oxygon oataration (<p>2) respiratory rate. The device may generate an blood pressure (fDT ); body tomporation of SS and SS (and St. This device may also condition addide and or visual alam who i modour calls Tato Talle Talle Taller will be available for use by medical personnel on all patient populations. ### Performance Testing - formanos Tooting The device complies with IEC 601-1 subclause 56.3(c) implemented by 21 CFR Part 898 1 The donion of the read on " Sead Wires and Patient Cables. To date, no other special controls r enomance olandards are known or established for this device. The device is designed to comply or performance candards and the standards: IEC 60601-1 (1988-12), Amendment 1 (1991-11), Amendment 2 (1995-03), IEC 60601-1-1 (1992-06), IEC 60601-1-2 (1993-05), CISPR11 Group 1, Allianonent = (1000-00), 12-27 (1994), JEC 60601-2-30 (1995-03), IEC 60601-2-34 (1994-12) - The device is not sterile. 트 - The device does not directly contact patients. Accessories that contact patients, such as probes and ■ The device are the same accessories as used with other legally marketed products or are comprised of the same component materials as the predicate accessories except as noted below. The disposable SpO2 probes include No.1776 3M Nonwoven Medical Tape which may contact the surface of intact skin. According to the manufacturer, 3M, the material was tested in accordance with ISO 10993 and Good Aooorang to the manafactively of adhesive materials within historically acceptable levels. - The device was subjected to environmental testing including temperature/humidity stress testing, . electromagnetic interference / electromagnetic compatibility testing and safety standards testing and performance testing procedures. Test criteria is established prior to testing based upon product specifications and applicable standards. The completed testing showed that the device met its product specifications and verified conformance to safety, reliability, and applicable standards. product operification and validation tested the operation of the software functions of the device. The results confirmed that the device performed within specifications. There are no significant changes in function, biocompatibility, performance or manufacturability compared to the predicate device that would affect the safety and effectiveness of the device as intended for use. Therefore, Nihon Kohden believes that the new BSM-2300 Series, is substantially equivalent to the predicate BSM-4100A Series Bedside Monitor. {1}------------------------------------------------ K011918 p.2/2 NIHON KOHDEN AMERICA, INC. June 19, 2001 ## 510(k) NOTIFICATION BSM-2300A Series Bedside Monitor ### SECTION 3 - PROPOSED LABELING #### Intended Use A. The device is intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility. The device is intended to monitor the vital signs monitoring withili a modical land/or audible alarms when an arrhythmia exists. The device is also intended to monitor heart rate, pulse rate, blood oxygen saturation (SpO₂) noninvasive blood pressure (NIBP) invasive blood pressure (IBP), body temperature, carbon nonimasive blood probodio (NBC) and respiratory rate. The device may generate an audible and/or visual alarm when a measured rate falls outside preset limits. #### B. Device/Package Labels The proposed product labels for the device are located in Attachment 2. #### C. Proposed Packaging Packaging for the device is depicted in Attachment 3. #### D. Instructions for Use The proposed instructions for use are provided with each packaged device and are presented in Attachment 9. #### Advertisement/Promotional Literature E. To date no advertisement or promotional literature for this device has been created for distribution in the United States. #### Contraindications, Precautions & Warnings F. Warnings and cautions are listed in the Operator's Manual as shown in Attachment 4. {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle. The text is in all caps and is evenly spaced around the circle. ### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 1 2 2001 Ms. Bonnie Bishop Nihon Kohden America, Inc. 90 Icon Street Foothill Ranch, CA 92610 Re: K011918 Trade Name: Nihon Kohden BSM-2300A Series Bedside Monitor and Accessories Regulation Number: 21 CFR 870.1025 Regulatory Class: III (three) Product Code: MHX Dated: June 19, 2001 Received: June 20, 2001 ### Dear Ms. Bishop: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this {3}------------------------------------------------ ### Page 2 - Ms. Bonnie Bishop response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, James E. Dillard III Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ G. Indications for Use Statement | 510(k) Number (if known): | K011918 | |---------------------------|---------| |---------------------------|---------| Device Name: BSM-2300A Series Bedside Monitors ### Indications for Use: The device is intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility. The device is intended to monitor the electrocardiogram to generate audible and/or visible alarms when an arrhythmia exists. The device is also intended to monitor heart rate, pulse rate, blood oxygen saturation (SpO2) noninvasive blood pressure (NIBP) invasive blood pressure (IBP), body temperature, carbon dioxide concentration (CO2 and EtCO2), and respiratory rate. The device may generate an audible and/or visual alarm when a measured rate falls outside preset limits. The device may also condition and transmit physiological signals via radio frequency. The device will be available for use by medical personnel on all patient populations. Dale Tell **Division of Cardiovascular & Respiratory Devices** 510(k) Number K011915 Prescription Use (Per 21 CFR 801.109)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...