VITALMAX 4000 CONFIGURED/VITALMAX 4000 MODULAR/VITALMAX 4100/MINIPACK 3100/3000/VITALMAX 800+/MINIPACK 300

K970445 · Pace Tech Medical Monitors, Inc. · MHX · Jan 30, 1998 · Cardiovascular

Device Facts

Record IDK970445
Device NameVITALMAX 4000 CONFIGURED/VITALMAX 4000 MODULAR/VITALMAX 4100/MINIPACK 3100/3000/VITALMAX 800+/MINIPACK 300
ApplicantPace Tech Medical Monitors, Inc.
Product CodeMHX · Cardiovascular
Decision DateJan 30, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1025
Device ClassClass 2

Intended Use

The Vitalmax 4000 is intended for use in the hospital/clinical environment to measure and monitor the following parameters: ECG waveform NIBP systolic, diastolic, and mean arterial pressure values (MEAN or MAP) Blood oxygen saturation (SpO2 or Pulse oximetry) SpO2 waveform Pulse (SpO2 and NIBP) signal strength Pulse (SpO2 and NIBP) or heart (ECG) rate Temperature End-tidal Co2 concentration (Et Co2) Co2 waveform Minimum inspired Co2 concentration (inCo2) Respiration rate Additional options offered are: Built-in printer with two channels of annotated waveform or text form of vital signs Thoracic impedance respiration Invasive pressure waveforms: systolic, diastolic, and mean values (IP, IP2) Second temperature channel and Δ temp Analog output The Vitalmax 4000 is available in configured and modular form. The Vitalmax 4000 configured monitor is available in various models which offer the options of Co2, thoracic impedance respiration, printer, and invasive pressures. The vitalmax 4000 is not intended to be used as apnea monitor or during MRI (Magnetic Resonance Imaging).

Device Story

Multi-parameter patient monitors (Vitalmax/Minipack series) for hospital/clinical use; measure ECG, NIBP, SpO2, temperature, CO2, respiration, invasive pressure, and anesthetic gases. Inputs: physiological signals via sensors/electrodes. Processing: signal conditioning and parameter calculation; output: real-time waveforms, numeric vital signs, and trends on integrated display or printer. Operated by clinicians; provides continuous monitoring to assist in patient status assessment and clinical decision-making. Benefits: real-time physiological data for patient safety.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Multi-parameter patient monitoring system; includes ECG, NIBP, SpO2, CO2, invasive pressure, and gas analysis modules. Features integrated display, optional thermal printer, and analog output. Connectivity includes standard physiological sensor interfaces. Software-based parameter calculation and waveform processing.

Indications for Use

Indicated for hospital/clinical environment patient monitoring of ECG, NIBP, SpO2, pulse, temperature, CO2, respiration, invasive pressure, and anesthetic gas concentrations. Contraindicated for use as an apnea monitor or during MRI procedures.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Related Devices

Submission Summary (Full Text)

{0} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Iihan M. Bilgutay President Pace Tech Medical Monitors 510 Garden Avenue North Clearwater, FL 34615-4126 Re: K970445 Vitalmax 4000 Configure/Vitalmax 4000 Modular/Vitalmax 4100 Regulatory Class: II (Two) Product Code: 74 MHX Dated: November 5, 1997 Received: November 10, 1997 Dear Mr. Bilgutay: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {1} Page 2 - Mr. Iihan M. Bilgutay This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html." Sincerely yours, Thomas J. Callahan Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2} # INDICATION FOR USE STATEMENT In accordance with the FDA new requirement, January 1, 1996 the "Indication For Use", attached, please find the indication for use for all of the devices included under the Premarket notification #K970445. These devices are: Vitalmax 4000 Configured / Vitalmax 4000 Modular / Vitalmax 4100 / Minipack 3000 / Minipack 3100 / Minipack 300 / Vitalmax 800 PLUS. ## Vitalmax 4000 Configured / Modular The Vitalmax 4000 is intended for use in the hospital/clinical environment to measure and monitor the following parameters: - ECG waveform - NIBP systolic, diastolic, and mean arterial pressure values (MEAN or MAP) - Blood oxygen saturation (SpO2 or Pulse oximetry) - SpO2 waveform - Pulse (SpO2 and NIBP) signal strength - Pulse (SpO2 and NIBP) or heart (ECG) rate - Temperature - End-tidal Co2 concentration (Et Co2) - Co2 waveform - Minimum inspired Co2 concentration (inCo2) - Respiration rate ## Additional options offered are: - Built-in printer with two channels of annotated waveform or text form of vital signs - Thoracic impedance respiration - Invasive pressure waveforms: systolic, diastolic, and mean values (IP, IP2) - Second temperature channel and $\Delta$ temp - Analog output The Vitalmax 4000 is available in configured and modular form. The Vitalmax 4000 configured monitor is available in various models which offer the options of Co2, thoracic impedance respiration, printer, and invasive pressures. The vitalmax 4000 is not intended to be used as apnea monitor or during MRI (Magnetic Resonance Imaging). (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number K970445 {3} # Vitalmax 4100 The Vitalmax 4100 CN is intended for use in the hospital/clinical environment to measure and monitor the following parameters: - ECG waveform - NIBP systolic, diastolic, and mean arterial pressure values (MEAN or MAP) - Blood oxygen saturation (SpO2 or Pulse oximetry) - SpO2 waveform - Pulse (SpO2 and NIBP) signal strength - Pulse (SpO2 and NIBP) or heart (ECG) rate - Temperature - End-tidal Co2 concentration (Et Co2) - Co2 waveform - Minimum inspired Co2 concentration (inCo2) - Respiration rate - Nitrous oxide concentration (N2O) The Vitalmax 4100 G and Vitalmax 4100 A include these additional parameters: - End-tidal expired Halothane, Enflurane, or Isoflurane agent gas concentration (etHAL, etENF, etISO) - Inspired Halothane, Enflurane, or Isoflurane agent gas concentration (inHAL, inENF, inISO) - Fractional inspired oxygen in gas concentration (Fio2 or %o2) Additional options offered for both models are: - Thoracic impedance respiration - Invasive pressure waveform, systolic, diastolic and mean values (IP1, IP2) - Second temperature channel and temperature difference [Δ temp] - Built-in strip chart recorder - Sevoflurane, and Desflurane (etSev, inSev, etDes, inDes) - Analog Output The Vitalmax 4100 is not intended to be used as apnea monitor or during MRI (Magnetic Resonance Imaging). # Minipack 3000/3100 The Minipack 3000 and Minipack 3100 are intended for use in the hospital/clinical environment to measure and monitor the following parameters: - ECG waveform - NIBP systolic, diastolic, and mean arterial pressure values (MEAN or MAP) - Blood oxygen saturation (SpO2 or Pulse oximetry) - SpO2 waveform - Pulse (SpO2 and NIBP) signal strength {4} - Pulse (SpO2 and NIBP) or heart (ECG) rate - Temperature - End-tidal Co2 concentration (Et Co2) - Co2 waveform - Minimum inspired Co2 concentration (inCo2) - Respiration rate Additional options offered are: - Built-in printer with two channels of annotated waveform or text form of vital signs - Thoracic impedance respiration - Invasive pressure waveforms: systolic, diastolic, and mean values (IP1, IP2) - Second temperature channel and Δ temp - Analog output The vitalmax 3000 and Vitalmax 3100 are not intended to be used as apnea monitors or during MRI (Magnetic Resonance Imaging). ## Minipack 300/ Vitalmax 800 PLUS The Minipack 300 and Vitalmax 800 PLUS are intended for use in the hospital/clinical environment to measure and monitor the following parameters: - ECG waveform - NIBP systolic, diastolic, and mean arterial pressure values (MEAN or MAP) - Blood oxygen saturation (SpO2 or Pulse oximetry) - SpO2 waveform - Pulse (SpO2 or NIBP) signal strength - Pulse (SpO2 and NIBP) or heart (ECG) rate - Temperature - End-tidal Co2 concentration (Et Co2) - Co2 waveform - Minimum inspired Co2 concentration (inCo2) - Respiration rate Additional options offered are: - Thoracic impedance respiration - Add-on 27 ccolumn thermal printer (Minipack 300 Series) - Built-in printer with two channels of annotated waveform (Vitalmax 800 PLUS) - Analog output The Minipack 300 and Vitalmax 800 PLUS are not intended to be used as apnea monitors or during MRI (Magnetic Resonance Imaging).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...