DATEX-OHMEDA S/5TM FM WITH L-FICU04 AND L-FICU04A SOFTWARE AND N-FCREC MODULE

K043276 · Datex-Ohmeda · MHX · Dec 23, 2004 · Cardiovascular

Device Facts

Record IDK043276
Device NameDATEX-OHMEDA S/5TM FM WITH L-FICU04 AND L-FICU04A SOFTWARE AND N-FCREC MODULE
ApplicantDatex-Ohmeda
Product CodeMHX · Cardiovascular
Decision DateDec 23, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1025
Device ClassClass 2

Intended Use

The S/5™ FM with L-FICU04 and L-FICU04A and N-FCREC Module are intended for multiparameter patient monitoring. The S/5 FM with L-FICU04 or L-FICU04A software is indicated for monitoring of hemodynamics (including arrhythmia and ST-segment analysis) and respiratory status of all hospital patients. Extension module N-FCREC (option N-FCREC or N-FC) is indicated for monitoring CO2 and respiration rate of all hospital patients. CO2 measurements are indicated for patients weighing over 5kg (11lbs). The S/5 FM Monitor and N-F(C)(REC) Extension Module are indicated for use by qualified medical personnel only.

Device Story

Modular multiparameter patient monitor for hospital use; operated by physicians/nurses via command board or remote controller (REMCO). Inputs: ECG, impedance respiration, invasive/non-invasive blood pressure, pulse oximetry, temperature (via E-PSM(P) module); CO2 (via N-F(C)(REC) side-stream gas analyzer using 4.2-4.3µm IR filters). Device processes signals to calculate heart rate, respiration rate, ST segments, and arrhythmia; displays numeric values, waveforms, and trends. Features audible/visual alarms. Supports stand-alone or networked (hardwired/wireless) configurations for central viewing/archiving. Thermal array printer (recorder) provides hard-copy traces/data. Benefits: continuous patient status monitoring, enabling timely clinical intervention.

Clinical Evidence

Bench testing only. Verification and validation of specifications performed against mandatory and voluntary standards including IEC 60601-1, IEC 60601-2-27, IEC 60601-2-30, IEC 60601-2-34, IEC 60601-2-49, ISO 9919, ISO 9918, and various AAMI standards (EC57, EC11, EC13). No clinical data presented.

Technological Characteristics

Modular multiparameter monitor; side-stream gas analyzer (miniC) using 4.2-4.3µm IR filters; thermal array printer. Connectivity: hardwired/wireless LAN. Standards: IEC 60601-1, IEC 60601-2-27/30/34/49, ISO 9919/9918, EN 1060, ANSI/AAMI EC11/13/57. Software: L-FICU04/A (includes arrhythmia analysis).

Indications for Use

Indicated for monitoring hemodynamics (arrhythmia, ST-segment) and respiratory status in all hospital patients. CO2 monitoring indicated for patients >5kg (11lbs). For use by qualified medical personnel only.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a sequence of handwritten alphanumeric characters. The sequence appears to be 'K043276'. The characters are written in a bold, slightly irregular style, giving them a hand-drawn appearance. The contrast between the characters and the background is high, making them easily readable. Page 1 of 4 DEC 2 3 2004 # Premarket Notification 510(k) Summary As required by section 807.92 Datex-Ohmeda S/5™ FM with L-FICU04 and L-FICU04A Software and N-FCREC Module ## GENERAL COMPANY INFORMATION as required by 807.92(a)(1) ## COMPANY NAME/ADDRESS/PHONE/FAX: Datex-Ohmeda 86 Pilgrim Road Needham, MA 02492 USA Tel: 781-449-8685 Fax: 781-433-1344 #### NAME OF CONTACT: Mr. Joel Kent DATE: C November 23, 2004 ## DEVICE NAME as required by 807.92(a)(2) #### TRADE NAME: Datex-Ohmeda S/5TM FM with L-FICU04 and L-FICU04A Software and N-FCREC Module ## COMMON NAME: Patient Monitor #### CLASSIFICATION NAME: The following Class II classifications appear applicable: | MHX | Monitor, Physiological,Patient (With Arrhythmia Detection or Alarm) | 870.1025 | |----------------------------------------------------------|---------------------------------------------------------------------|----------| | MLD | Monitor, ST segment with Alarm | 870.1025 | | CCK | Analyzer, gas, carbon-dioxide, gaseous-phase | 868.1400 | | The following Class I classifications appear applicable: | | | | DSF | Paper Chart Recorder | 870.2810 | {1}------------------------------------------------ ## NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3) The Datex-Ohmeda S/5TM FM with L- FICU04 and L- FICU04A software is substantially equivalent to the predicate Datex-Ohmeda S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A software (K022740). The N-F(C)(REC) Extension module used with the FM Monitor allows measuring CO2 (module versions N-FC and N-FCREC) and printing on a recorder chart (module versions N-FREC and N-FCREC). The CO2 measurement of the N-FC(REC) is based on the CO2 measurement of the predicate M-miniC module (K023454). The recorder option is based on the predicate M-REC module, originally cleared in the Datex AS/3 AM Monitor (K933285) and the Datex AS/3 Compact Monitor (K933156) submissions. #### DEVICE DESCRIPTION as required by 807.92(a)(4) The S/5™ FM Monitor with L-FICU04(A) software is a patient monitor, which measures and displays patient physiological parameters in hospital environments. The S/5™ FM uses two types of plug-in measurement modules: Extension module N-F(C)(REC) with CO2 or Recorder option, or both, and Patient side Module E-PSM(P), that measures ECG, impedance respiration , invasive and non-invasive blood pressures, pulse oximetry and temperature. The E-PSM(P) is the subject of a separate 510(k). The miniC unit of the N-FC(REC) Extension module, included in this 510(k) submission, is a side stream gas analyzer. It monitors the Carbon dioxide (CO2) inhaled and exhaled by the patient, by measuring the absorption of CO2 at 4.2-4.3 micrometer using narrow band IR filters. It can also monitor the respiration rate as the frequency of peak (end tidal) CO2 measurements per minute. All concentrations are measured and displayed breath by breath. The MiniC unit of the Extension module is equipped with the mini D-fend water separation system. There are two software options available for the S/5 FM: L-FICU04 and L-FICU04A. The L-FICU04A is equipped with extended arrhythmia analysis capability. Other than arrhythmia analysis capabilities, this software option is identical to the L-FICU04. Commonly, the software is collectively referred to as L-FICU04(A). The E-PSM(P) and N-F(C)(REC) Modules placed in the S/5 FM are automatically recognized by the monitor. Monitoring can begin when patient cables are connected to the module plug-in jacks. The modules perform parameter measurement and minor data processing. The S/5™ FM displays measured parameters in the form of numeric values, waveforms and trends on the screen. Audible and visual alarms are used to indicate patient status. The priority profile of an alarm depends on the parameter. The L-FICU04(A) software performs some module-related tasks like arrhythmia analysis, ST value calculation, heart rate calculation, impedance and respiration rate calculation. All the module communication is also handled in the main software. The S/5™ FM is operated by command board keys. Typically pressing a key results in a pop up menu appearing on the screen. Selections can then be made easily from the menu using a unique ergonomically designed pointing device on the keyboard called a ComWheel™. In addition of the standard command board there is an optional type of keyboard, a hand-held Remote controller (REMCO), which is directly connected to the S/5™ FM via a long cord but provides more flexibility in controlling the monitor while the doctor or nurse is handling other patient care needs. The Recorder is a thermal array printer with which you can print from one to three traces and numerical information with a selectable speed. The S/5™ FM Monitor can be in a stand-alone or networked configuration. If networked, measurement data (parameter-associated numbers, waveforms or trends and patient care documentation) is sent to the network for central viewing of the network, monitor-to-monitor viewing, or for archiving. Software I .- FICU04 and L-FICU04A include support for wireless LAN communication between the monitor and S/5 Central station. The networking can thus be hardwired or wireless. {2}------------------------------------------------ #### INTENDED USE as required by 807.92(a)(5) #### Intended use: The S/5™ FM with L-FICU04 and L-FICU04A and N-FCREC Module are intended for multiparameter patient monitoring. #### Indications for use: The S/5 FM with L-FICU04 or L-FICU04A software is indicated for monitoring of . hemodynamics (including arrhythmia and ST-segment analysis) and respiratory status of all hospital patients. Extension module N-FCREC (option N-FCREC or N-FC) is indicated for monitoring CO2 and respiration rate of all hospital patients. CO2 measurements are indicated for patients weighing over 5kg (11lbs). The S/5 FM Monitor and N-F(C)(REC) Extension Module are indicated for use by qualified medical personnel only. ## SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6) The Datex-Ohmeda S/5™ FM with L- FICU04 and L- FICU04A software is substantially equivalent to the predicate Datex-Ohmeda S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A software (K022740), as indicated by the similarities discussed below. The S/5™ FM Monitor is a modular multiparameter patient monitor providing connections to measurement modules, in a similar way as the predicate device. The general construction, indications for use and intended use of the S/5™ FM Monitor are almost the same as for the predicate device (K022740). The basic model of the monitor is the S/5™ FM with L-FICU04 software. When equipped with extended bedside arrhythmia analysis capability, the monitor is called S/5™ FM with L-FICU04A. The arrhythmia analysis functionality of the S/5TM FM with L-FICU04A is identical to the functionality of the predicate device S/5™ Compact Critical Care Monitor with L-CICU02A software (K022740). The N-F(C)(REC) Extension module used with the FM Monitor allows measuring CO2 (module versions N-FC and N-FCREC) and printing on a recorder chart (module versions N-FREC and N-FCREC). The CO2 measurement of the N-FC(REC) is based on the CO2 measurement of the predicate M-miniC module (K023454). The recorder option is based on the predicate M-REC module, originally cleared in the Datex AS/3 AM Monitor (K933285) and the Datex AS/3 Compact Monitor (K933156) submissions. The S/5™ FM Monitor and its predicate device both have the same basic user interface and alarms. They are manufactured using the same processes, and have the same safety and effectiveness, as indicated by the thorough and successful testing of the FM Monitor, documented in this submission. Based on the above and a detailed analysis in other documentation included in this 510(k) notification and attachments, it is evident that the main features and indications for use of the S/5™ FM Monitor with I - L-FICU04 and L- L-FICU04A software are substantially equivalent in safety and effectiveness compared to the predicate. {3}------------------------------------------------ #### SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3) Datex-Ohmeda S/5TM FM with L-FICU04 and L-FICU04A Software and N-FCREC Module complies with the safety standards below and is therefore safe and effective for the intended use. The device has been thoroughly tested through validation and verification of specifications. Verification of compliance with the following mandatory and voluntary standards has been made: - IEC 60601-1:1988+ Amdt .: 1:1991 + Amdt. 2:1995 - EN 60601-1: 1990 + A1:1993+A2:1995+A13:1996 - CAN/CSA-C22.2 No.601.1-M90 +S1:1994+Amdt. 2:1998 - UL 2601-1:1997 - . IEC 60601-2-27:1994/EN 60601-2-27:1994 - · . IEC 60601-2-30:1999/EN 60601-2-30:2000 - IEC 60601-2-34:2000/EN 60601-2-34:2000 - IEC 60601-2-49:2001/EN 60601-2-49:2001 - ISO 9919:1992/EN865:1997 - . EN 1060-1:1995/EN 1060-3:1997 - EN 12470-4:2000 - ISO 9918:1993 / EN 864:1996 - . IEC/EN 60601-1-2: 2001 - . . IEC/EN 60601-1-4: 1996+A1:1999 - . ANSI/AAMI ES-1:1993 - . ANSI/AAMI EC57:1998 - . ANSI/AAMI EC11:1991 - . ANSI/AAMI EC13:2002 - IEC 60068-2 (Mech stress tests, Temp and Humidity tests). . #### CONCLUSION: The summary above shows that there are no new questions of safety and effectiveness for the Datex-Ohmeda S/5™ FM with L-FICU04 and L-FICU04A Software and N-FCREC Module as compared to the predicate device. {4}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 2 3 2004 Datex Ohmeda c/o Mr. Joel C. Kent Quality and Regulatory Affairs Manager 86 Pilgrim Road Needham, MA 02492 Re: K043276 K043270 Trade Name: Datex-Ohmeda S/5™ FM with L-FICU04 and L-FICU04A Software and Extension module N-FCREC Regulation Number: 21 CFR 870.1025 Regulation Name: Patient Monitor Regulatory Class: Class II (two) Product Code: MHX Dated: November 24, 2004 Received: November 26, 2004 Dear Mr. Kent: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Section 310(t) premier is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreative) to regard cate of the Medical Device Amendments, or to commerce prior to May 28, 1976, the enactment date of the Medical Device All Producti commente prior to May 20, 1976, the enaordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). alle Cosment Act (71ct) that to nor requestly subject to the general controls provisions of the Act. The I ou may, merciolo, manel and include requirements for annual registration, listing of general controls provisions of the ree, labeling, and prohibitions against misbranding and t adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is crassified (see above) the existing major regulations affecting your device can may be subject to such additional controls. Existing major and contribution may oc subject to sueritudions, Title 21, Parts 800 to 898. In addition, FDA may be found in the Oous acements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 - Mr. Joel C. Kent Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0295. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Bermuda for Bram D. Zuckerman, M.D. Director Division fo Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use 510(k) Number (if known): _ K043276 Device Name: Datex-Ohmeda S/5™ FM with L-FICU04 and L-FICU04A Software and Extension module N-FCREC Indications for Use: The S/5 FM with L-FICU04 or L-FICU04A software is indicated for monitoring of hemodynamics (including arrhythmia and ST-segment analysis) and respiratory status of all hospital patients. Extension module N-FCREC (option N-FCREC or N-FC) is indicated for monitoring CO2 and respiration rate of all hospital patients. CO2 measurements are indicated for patients weighing over 5kg (11lbs). The S/5 FM Monitor and N-F(C)(REC) Extension Module are indicated for use by qualified medical personnel only. Prescription Use × Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) R.H. Simmons (Division Sign-OH) Division of Cardiovascular Devices 510(k) Number K043274 Page __ of ___________________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...