M3290A INTELLIVUE INFORMATION CENTER SOFTWARE RELEASE E.01 AND INTELLIVUE CLINICAL NETWORK ON VLAN

K040955 · Philips Medical Systems, Inc. · MHX · May 12, 2004 · Cardiovascular

Device Facts

Record IDK040955
Device NameM3290A INTELLIVUE INFORMATION CENTER SOFTWARE RELEASE E.01 AND INTELLIVUE CLINICAL NETWORK ON VLAN
ApplicantPhilips Medical Systems, Inc.
Product CodeMHX · Cardiovascular
Decision DateMay 12, 2004
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1025
Device ClassClass 2
AttributesSoftware as a Medical Device, Pediatric

Intended Use

Indicated for central monitoring of multiple adult, pediatric, and neonatal patients; and where the clinician decides to monitor cardiac arrhythmia of adult, pediatric, and neonatal patients and/or ST segment of adult patients to gain information for treatment, to monitor adequacy of treatment, or to exclude causes of symptoms.

Device Story

M3290A IntelliVue Information Center Software (Release E.01) and IntelliVue Clinical Network facilitate central monitoring of physiological data from multiple patients. System enables operation on customer-provided IEEE 802.1q compatible VLAN network infrastructure. Clinicians use system to monitor cardiac arrhythmias (all ages) and ST segments (adults) to inform treatment decisions, assess treatment adequacy, or investigate symptoms. Device processes network-transmitted physiological signals; displays data at central station for clinical review. Supports clinical decision-making by providing centralized alarm and waveform visualization.

Clinical Evidence

Bench testing only. No clinical data provided. Verification and validation included system-level tests, integration tests, environmental tests, and safety testing based on hazard analysis. Pass/Fail criteria were derived from predicate device specifications; results demonstrated substantial equivalence.

Technological Characteristics

Software-based central monitoring system. Operates on IEEE 802.1q compatible VLAN network infrastructure. Provides arrhythmia and ST segment alarm detection. Connectivity via networked physiological monitoring system.

Indications for Use

Indicated for central monitoring of adult, pediatric, and neonatal patients; cardiac arrhythmia monitoring for all ages; ST segment monitoring for adults. Prescription use only.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## MAY 1 2 2004 #### 8.0 510 (k) Summary summary of 510(k) safety and effectiveness information រ ន This submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92. 1. The submitter of this premarket notification is: Philips Medical Systems This summary was prepared on 12 April, 2004. - 2. The name of this device is the M3290A IntelliVue Information Center Software Release E.01 and IntelliVue Clinical Network on VLAN. Classification names are as follows: | Classification | ProCode | Description | |----------------|---------|--------------------------------------------------------------| | None | 74 MHX | Physiological Monitor, Patient<br>Monitor | | 870.1025, II | 74 DSI | Arrhythmia Detector and Alarm | | 870.1025, II | 74 MLD | Monitor, ST Alarm | | 870.2800, II | 74 DSH | Recorder, Magnetic Tape, Medical | | 870.2300, II | 74 MSX | System, Network and Communication,<br>Physiological Monitors | - 3. The new device is substantially equivalent to the previously cleared IntelliVue Information Center Software and IntelliVue Clinical Network cleared under K993171, K993907, K001057, K001664, K011093, K021422, K023698, K030038, K031403, & K040357. - 4. The modification is a change that permits operation of the IntelliVue Clinical Network on customer provided IEEE 802.1q compatible VLAN network infrastructure. - 5. The new device has the same Indications for Use as the legally marketed predicate device. For central monitoring of multiple adult, pediatric, and neonatal patients; and where the clinician decides to monitor cardiac arrhythmia of adult, pediatric, and neonatal patients and/or ST seqment of adult patients to gain information for treatment, to monitor adequacy of treatment, or to exclude causes of symptoms . - 6. The new device has the same technological characteristics as the legally marketed predicate device. - 7. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the new device with respect to the predicate. Testing involved system level tests, integration tests, environmental tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that IntelliVue Clinical Network on VLAN infrastructure meets all reliability requirements and performance claims. {1}------------------------------------------------ Image /page/1/Picture/11 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features an eagle with its wings spread, symbolizing protection and strength. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 1 2 2004 Philips Medical Systems c/o Mr. Dave Osborn Quality Program Manager 3000 Minuteman Road Andover, MA 01810 #### Rc: K040955 KU40933 Trade Name: M3290A Intellivue Information Center Software, Release E.01, and Intellivue Clinical Network on VLAN Regulation Number: 21 CFR 870. 1025 Regulation Name: Patient Physiological Monitor (With Arrhythmia Detection or Alarms) Regulatory Class: Class II (two) Product Code: MHX Dated: April 12, 2004 Received: April 13, 2004 Dear Mr. Osborn: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your boctor > I <<<<> pc device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosure) is regars actment date of the Medical Device Amendments, or to continered prior to May 20, 1978, the excordance with the provisions of the Federal Food, Drug, de vices mat nave been rechasined in avere approval of a premarket approval application (PMA). and Cosmetor rear (11ct) the device, subject to the general controls provisions of the Act. The I ou may, mereleve, maniel of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can may be subject to such additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2 - Mr. Dave Osborn Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that I DA s issuation of a backed complies with other requirements of the Act that IDA has made a determination administered by other Federal agencies. You must or any Federal statutes and regulations daminding, but not limited to: registration and listing (21 comply with an the Act 3 requirements, with 801); good manufacturing practice requirements as set CFK Fatt 607), labeling (21 OFR Part 820); and if applicable, the clectronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. product launtion control pro herems (2) ====================================================================================================================================== Fins icher while anow you to begin maining of substantial equivalence of your device to a legally premaired notification. THPT BAT maining of casification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific at nee 101 ) 594-4646. Also, please note the regulation entitled, Comact the Office of Comment of Comment of notification" (21CFR Part 807.97) you may obtain. I whisoralium by reference to premiesponsibilities under the Act may be obtained from the Oinci general information on your respectional and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Dna Heiseku for Bram D. Zuckerman, M Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ #### Indications for Use # 510(k) Number (if known): _ K040 455 M3290A IntelliVue Information Center Software, Device Name: M3290A Incellivae IntelliVue Clinical Network on VLAN ### Indications for Use: mandate is our Indicated for central monitoring on and clinician decides to monitor and neonatur pacients) and neonatal patients cardiae arriferma of adult patients to gain information for and/OL ST Segment of adaily of treatment, or to exclude causes of symptoms. Prescription Use yes (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use No (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Dna Xuelu Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...