M3290A INTELLIVUE INFORMATION CENTER SOFTWARE, RELEASE 1.00
K081983 · Philips Medical Systems · MHX · Aug 22, 2008 · Cardiovascular
Device Facts
Record ID
K081983
Device Name
M3290A INTELLIVUE INFORMATION CENTER SOFTWARE, RELEASE 1.00
Applicant
Philips Medical Systems
Product Code
MHX · Cardiovascular
Decision Date
Aug 22, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Intended Use
Indicated for central monitoring of multiple adult, pediatric, and neonatal patients; and where the clinician decides to monitor cardiac arrhythmia of adult, pediatric, and neonatal patients and/or ST segment of adult patients to gain information for treatment, to monitor adequacy of treatment, or to exclude causes of symptoms.
Device Story
M3290A IntelliVue Information Center (IIC) Software Release L.0 is a central monitoring system for physiological data. It receives patient data from IntelliVue MPxx patient monitors via a clinical network (including wireless). The system integrates the ST/AR J.0 algorithm for arrhythmia and ST-segment analysis. Features include manual laboratory data entry, ST baseline synchronization, wireless TAAP support, and HL7 parameter integration for VistA compliance. The system uses a SCADA communication protocol. Used in clinical settings, it allows clinicians to view real-time patient data, monitor cardiac status, and assess treatment efficacy. Output is displayed at a central station, enabling clinical decision-making regarding patient care and symptom management.
Clinical Evidence
Bench testing only. Verification and validation activities included system-level tests, performance tests, and hazard analysis. Pass/Fail criteria were based on specifications cleared for the predicate device. Results demonstrated that the software meets all reliability requirements and performance claims.
Technological Characteristics
Software-based central monitoring system. Connectivity via clinical network (wired/wireless) using SCADA protocol. Integrates ST/AR J.0 algorithm for arrhythmia and ST-segment analysis. Supports HL7 data exchange. Operates as a central station for physiological monitoring.
Indications for Use
Indicated for central monitoring of adult, pediatric, and neonatal patients; cardiac arrhythmia monitoring for all ages; ST segment monitoring for adult patients.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
K062271 — M3290A INTELLIVUE INFORMATION CENTER SOFTWARE, RELEASE J.00 (FOR M3140, M3145, M3150, M3151, M3154, M3155, M3169, M3170) · Philips Medical Systems · Nov 30, 2006
K040955 — M3290A INTELLIVUE INFORMATION CENTER SOFTWARE RELEASE E.01 AND INTELLIVUE CLINICAL NETWORK ON VLAN · Philips Medical Systems, Inc. · May 12, 2004
K023698 — M3290A INTELLIVUE INFORMATION CENTER SOFTWARE RELEASE E.O · Philips Medical Systems · Nov 19, 2002
K050742 — M3290A INTELLIVUE INFORMATION CENTER SOFTWARE RELEASE G.00 · Philips Medical Systems · Apr 5, 2005
K102495 — M3290B INTELLIVUE INFORMATION CENTER SOFTWARE, RELEASE A.0 · Philips Medical Systems · Sep 30, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white line drawing of a person's face. The face is slightly tilted to the right. The person has a round face, with a small nose and mouth. The person's eyes are closed. The person's hair is short and curly. The person is wearing a shirt with a collar.
## Philips Medical Systems
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR §807.92.
The submitter of this premarket notification is:
Claire Arakaki Regulatory Affairs Specialist Patient Monitoring Philips Medical Systems 3000 Minuteman Road, MS0480 Andover, MA 01810-1099
Tel: 978 659 4348 Fax: 978 685 5624 Email: claire.arakaki@philips.com
This summary was prepared on 8 July 2008.
The name of this device is the M3290A IntelliVue Information Center Software, Release L.0 Classification names are as follows:
| Classification | ProCode | Description |
|----------------|---------|--------------------------------------------------------------|
| none | 74 MHX | Physiological Monitor, Patient Monitor |
| 870.1025, II | 74 MLD | Arrhythmia Detector and Alarm |
| 870.1025, II | 74 MLD | Monitor, ST Alarm |
| 870.2800, II | 74 DSH | Recorder, Magnetic Tape, Medical |
| 870.2300, II | 74 MSX | System, Network and Communication,<br>Physiological Monitors |
- 2. The new device is substantially equivalent to the previously cleared M3290A Intelli Vue Information Center Software, Release J.0 marketed pursuant to K062271, K050742, K041741, K040955, K040357, K031403, K023698, K021422, K011093, K001057, K000854, K993907, K993171, and K964832.
- 3. The modification made to the device include the addition of the following features:
- . Support for the Wireless TAAP feature
- Addition of the Manual Data Laboratory (Lab) Entry application .
- Support of the ST baseline synchronization with the IntelliVue MPxx Patient Monitors
- Support for the printing of the predefined reports generated from the wireless IntelliVue MPxx . Patient Monitors
- . New data communication protocol used between the IntelliVue Patient Monitors and the IIC (SCADA protocol)
Tel: (978) 687 1501 Fax: 1978) 794 764
Confidential
{1}------------------------------------------------
- Support for the Philips patient monitors operating in the Customer Supplied Clinical Network . (wireless network)
- Addition of HL7 parameters for VistA compliance .
- Integration of the ST/AR J.0 algorithm (K080461) .
- న. The new device has the same Indications for Use and Intended Use as the legally marketed predicate devices.
- The new device has the same technological characteristics as the legally marketed 6. predicate devices.
- 7. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the new device with respect to the predicate. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that M3290A IntelliVue Information Center Software, Release L.0 meets all defined reliability requirements and performance claims.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 2 2008
Philips Medical Systems c/o Ms. Claire Arakaki Regulatory Affairs Specialist 3000 Minuteman Road, MS0480 Andover, MA 01810-1099
Re: K081983
Trade/Device Name: M3290A IntelliVue Information Center Software, Release L.0 Regulation Number: 21 CFR 870.1025 Regulation Names: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II Product Code: MHX, DSI, MLD, DSH, MSX Dated: August 6, 2008 Received: August 8, 2008
Dear Ms. Arakaki:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becared the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for woo cavaior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmotic For (110) that the device, subject to the general controls provisions of the Act. The r ou may, alerte or visions of the Act include requirements for annual registration, listing of general bonative profice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device ic may of receive for the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Claire Arakaki
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
ellomell
Bram D. Zuckerman
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): K08198 3
Device Name: M3290A IntelliVue Information Center Software Release L.00 (for M3140, M3145, M3150, M3151, M3154, M3155, M3169, M3170, and M3177)
Indications for Use:
Indicated for central monitoring of multiple adult, pediatric, and neonatal patients; and where the clinician decides to monitor cardiac arrhythmia of adult, pediatric, and neonatal patients and/or ST segment of adult patients to gain information for treatment, to monitor adequacy of treatment, or to exclude causes of symptoms.
Prescription Use Yes (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use No (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
lee
n of Cardiovascular Devices
Kogleg 510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.