Sterile Single Use Electrosurgical Pencil with Non-Coated Electrode, Sterile Single Use Electrosurgical Pencil with Non-Stick Electrode

K231405 · Jiangsu Rong FU Kang Medical Instruments Co., Ltd. · GEI · Jan 18, 2024 · General, Plastic Surgery

Device Facts

Record IDK231405
Device NameSterile Single Use Electrosurgical Pencil with Non-Coated Electrode, Sterile Single Use Electrosurgical Pencil with Non-Stick Electrode
ApplicantJiangsu Rong FU Kang Medical Instruments Co., Ltd.
Product CodeGEI · General, Plastic Surgery
Decision DateJan 18, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Intended Use

The devices are used to cut and coagulate soft tissue by means of high frequency electrical current during an electrosurgical procedure.

Device Story

Handheld monopolar electrosurgical pencil; used in surgical procedures for soft tissue cutting and coagulation. Device comprises plastic handle, connecting cable, control buttons (CUT/COAG), internal PCB, and stainless steel electrode tip (uncoated or non-stick). Operator (surgeon/clinician) connects device to an electrosurgical generator unit (ESU). ESU delivers high-frequency electrical energy to the electrode tip; energy passes through target tissue to a return electrode (ground pad). Non-stick electrodes utilize silicon oxide coating. Device is disposable and supplied sterile via EO sterilization. Output provides thermal effect on tissue for surgical incision or hemostasis.

Clinical Evidence

Bench testing only. Ex-vivo study on kidney, liver, and muscle tissue compared thermal damage zones (length, width, depth) against predicate; results showed virtually identical thermal effects. Non-stick performance evaluated; subject device showed improved coating integrity (no holes/craters). Electrical safety and EMC testing (conducted/radiated emissions, ESD, immunity) performed by third-party lab. Biocompatibility testing performed per ISO 10993. Package integrity and functional performance confirmed via aging/stability testing.

Technological Characteristics

Monopolar electrosurgical pencil; stainless steel electrodes; silicon oxide (non-stick) or uncoated tips; ABS insulation/collar; blue PVC cable. Energy source: external ESU (high-frequency). Standards: IEC 60601-1, IEC 60601-2-2, ISO 11135, ISO 11607, ISO 10993. Sterilization: EO. Dimensions: 190mm total length, 145mm handle, 45mm effective electrode length.

Indications for Use

Indicated for cutting and coagulating soft tissue during electrosurgical procedures. Prescription use only.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font. January 18, 2024 Jiangsu Rong Fu Kang Medical Instruments Co., Ltd. Mr. Steve Livneh CEO/President 3rd floor, Building D6, 6 Dongsheng Road West Jiangyin, Jiangsu 214400 China Re: K231405 Trade/Device Name: Sterile Single Use Electrosurgical Pencil with Non-Coated and Non-Stick Electrode Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: July 17, 2023 Received: December 21, 2023 Dear Mr. Livneh: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. | Mark<br>Trumbore -S | Digitally signed by<br>Mark Trumbore -S<br>Date: 2024.01.18<br>10:07:00 -05'00' | |---------------------|---------------------------------------------------------------------------------| |---------------------|---------------------------------------------------------------------------------| Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K231405 Device Name Sterile Single Use Electrosurgical Pencil with Non-coated and Non-stick Electrodes Indications for Use (Describe) The devices are used to cut and coagulate soft tissue by means of high frequency electrical current during an electrosurgical procedure. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows a logo with a blue globe on the left and the text "荣孚康 RFK" on the right. The globe is a network of interconnected nodes, suggesting a global network or connection. A horizontal line is placed under the text, emphasizing the brand name. # 510(K) SUMMARY- K231405 ## 1. 510(k) Owner: Name: Jiangsu Rong Fu Kang Medical Instruments Co., Ltd Address: 3rd floor, Building D6, 6 Dongsheng Road West, Jiangyin, Jiangsu, China 214400 Prepared on: January 17, 2024 ## 2. 510(k) Contact Information: Please Contact: Mr. Steve Livneh, CEO/President (1)-720-277-6060 Email: info@rfkinetics.com Address: 1589 Roseanna Drive, Northglenn Colorado 80234 USA ### 3. Device: Trade Name: Sterile Single Use Electrosurgical Pencil with Non-coated and Nonstick Electrodes Common Name: RFK Disposable Electrosurgical Pencil with Standard and Non-Stick Tip Electrodes Device Models: For the pencil: EP-100-3, EP-100-5 For the electrodes: - 1. Non coated electrode AE-EBD - 2. Non-stick electrode: AE-EBDN Classification Name: 878.4400: Electrosurgical Cutting and Coagulation Device and Accessories Class: II Product Code: GEI ## 4. Predicate and Reference Devices: Predicate Device: K192542 Single Use Electrosurgical pencil with non-coated and non-stick electrode (Nonsterile and sterile) Reference Device: K170369 Disposable Electrosurgical Pencils. CP1001 Series {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows a logo with a blue globe on the left and the text "荣孚康 RFK" on the right. The globe has a network of white lines and dots on it, suggesting a global network or connection. A blue line is underneath the text "荣孚康 RFK". made by Cathay Manufacturing Corp. # 5. Product Description: The Jiangsu Rong Fu Kang Medical Instruments Co. Ltd (here and after RFK) Electrosurgical Pencils and Electrodes are handheld instruments that are used for cutting and coagulation of soft tissue. They consist of a plastic handle, a connecting cable and plug, control buttons, internal PCB, and either a stainless steel uncoated or non-stick electrode tip. The Electrosurgical Pencils and electrodes are disposable and are supplied sterile. The connecting cable is plugged into an Electro-Surgical Generator Unit (ESU) which provides the high-frequency energy to be deployed in the surgical procedure. The plastic handle has two buttons: cutting and coagulation. The cutting button activates the CUT mode of the ESU and the coagulation button activates the COAG mode of the ESU. The electrode tip makes contact with the target tissue and delivers high frequency energy from the ESU to the target organs for the purposes of cutting and coagulation. This energy then passes through the body tissues and returns to the ESU via a return electrode (ground pad). ## 6. Indications for Use: The devices are used to cut and coagulate soft tissue by means of high frequency electrical current during an electrosurgical procedure. ## 7. Technological Characteristics: The proposed devices have substantially equivalent construction and performance as the predicate devices. ## 8. Substantial Equivalence: The technological characteristics and performance testing of the subject and predicate devices are substantially equivalent. The following table shows the comparisons in more detail information among the subject device and the predicate devices. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows a logo with a blue globe on the left and the text "荣孚康 RFK" on the right. The globe is a sphere with white dots connected by lines, creating a network-like appearance. The text is in blue and is accompanied by a horizontal blue line underneath. #### FIGURE 5-1: SUMMARY OF TECHNOLOGICAL CHARACTERISTICS AND PERFORMANCE TESTING | | Subject Device: K231405 | Predicate Device: K192542 | |-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | The devices are used to cut and<br>coagulate soft tissue by means<br>of high frequency electrical<br>current during an electrosurgical<br>procedure. | The devices are used to cut and<br>coagulate soft tissue by means<br>of high frequency electrical<br>current during an electrosurgical<br>procedure. | | Device Description | This device is intended to be<br>used as a hand-switching<br>monopolar accessory in<br>conjunction with a compatible<br>electrosurgical generator for<br>tissue cutting and coagulation.<br>The distal end of the pencil has<br>a 3/32" (2.36mm) collet that will<br>accept a variety of<br>electrosurgical electrodes with<br>3/32" (2.38mm) diameter shafts. | This device is intended to be<br>used as a hand-switching<br>monopolar accessory in<br>conjunction with a compatible<br>electrosurgical generator for<br>tissue cutting and coagulation.<br>The distal end of the pencil has<br>a 3/32" (2.36mm) collet that will<br>accept a variety of<br>electrosurgical electrodes with<br>3/32" (2.38mm) diameter shafts. | | Classification | 878.4400 | 878.4400 | | Product Code | GEI | GEI | | OTC or Prescription | Prescription | Prescription | | Energy Type | High-Frequency Monopolar | High-Frequency Monopolar | | Sterile? | Yes | Yes | | Sterilization Type | EO Sterilization | EO Sterilization | | Single Use? | Yes | Yes | | Electrode Type | Monopolar | Monopolar | | Electrode Material | Stainless Steel | Stainless Steel | | Electrode non-stick coating | Silicon oxide | Teflon | | Electrode Insulation Material | ABS | ABS | | Electrode Dimensions | Standard Blade Electrode 2.38<br>mm x 70 mm | Standard Blade Electrode 2.38<br>mm x 70 mm | | Insulating collar Materials | ABS | ABS | | Handle Length | 145 mm | 165 mm | | Effective Electrode Length | 45 mm | 45 mm | | Total Device Length | 190 mm | 210 mm | | Cable Material | Blue PVC Cable | Blue PVC Cable | | Standards Applied | IEC 60601-1, IEC 60601-2-2,<br>ISO 11135, ISO 11607 | IEC 60601-1, IEC 60601-2-2,<br>ISO 11135, ISO 11607 | | Biocompatibility | Conforms to ISO 10993 | Conforms to ISO 10993 | | Labeling | Conforms to 21 CFR Part 801 | Conforms to 21 CFR Part 801 | | Performance | Conforms to IEC 60601-1, and<br>IEC 60601-2-10, Part 2<br>particular requirements for the<br>basic safety and essential<br>performance of high frequency<br>surgical equipment and high<br>frequency surgical accessories.<br>Performance conforms to the<br>requirements specified in<br>"Premarket Notification<br>(510(k)) Submissions for<br>Electrosurgical Devices for<br>General Surgery | Conforms to IEC 60601-1, and<br>IEC 60601-2-10, Part 2<br>particular requirements for the<br>basic safety and essential<br>performance of high frequency<br>surgical equipment and high<br>frequency surgical accessories.<br>Performance conforms to the<br>requirements specified in<br>"Premarket Notification<br>(510(k)) Submissions for<br>Electrosurgical Devices for<br>General Surgery | | Stability and Shelf Life | 3 Years | 3 Years | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows a logo with a blue globe on the left and the text "荣孚康 RFK" on the right. The globe is a network of interconnected nodes, suggesting a global network or connection. The text is in Chinese and English, with a horizontal line underneath it. ## 9. Non-Clinical Bench Testing: #### Thermal Effects on Tissue In accordance with the nonbinding FDA guidance on "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery", part XI-E, the thermal effects on animal tissue of our device was compared with the predicate device K192542 Ex-vivo kidney, liver, and muscle tissues were used and the thermal damage zone sizes were compared for length, width, and depth. These tests showed that the thermal effects of our electrode are virtually identical to those shown in the predicate device. The non-stick performance of our non-stick electrode was found to be similar to the predicate device but without occurring holes and craters as in the predicate device's non-stick coating. #### Third party electrical testing: The device K231405 successfully passed the following electrical tests: - Conducted Emmisions (150kHz-30MHz) - Radiated Emissions (30MHz-1GHz) - Radiated Emissions (1GHz-18GHz) - Electrostatic Discharge (ESD) - Radio frequency electromagnetic field radiation immunity - Power frequency magnetic field All other EMC tests are not applicable to our device. The testing was conducted by: Organization Name: Suzhou Yipin Quality Technical Service Co., Ltd. Institutional Address: No. 558, Fenhu Avenue, Lili Town, Wujiang District, {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows a logo with a blue globe on the left and the text "荣孚康 RFK" on the right. The globe is surrounded by a network of lines and dots, suggesting a global connection. The text is in Chinese and English, with a blue line underneath. The logo appears to represent a company or organization with a global reach. Suzhou City, Jiangsu Province Test Location Name: Suzhou Yipin Quality Technical Service Co., Ltd. Industrial Park Branch Test Location Address: South side of Building B, No. 5 Minsheng Road, Suzhou Industrial Park 215028 www.epintek.com Tel: 0512-67997780 #### Third party biocompatibility testing was performed by: Sanitation & Environment Technology Institute, Soochow University No.199 Ren'ai Road, Suzhou Industrial Park, China Postcode 215123, www.sudatest.com. Tel: Direct: +86 512 65880038, Free: 4001078828, Email: med(@sudatest.com Subject device package integrity and functional performance tests were performed in-house in our Jiangsu facilities after aging and real-time stability tests to confirm the proposed shelf life. ## 10.Conclusion: The device has been shown to be substantially equivalent to the predicate device in terms of performance and physical and technological characteristics. The devices passed all tests both in-house and by third party qualified labs. It has been deemed safe and conforms to the standards discussed above.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...