PQI · Wound Retraction And Protection System

General, Plastic Surgery · 21 CFR 878.4371 · Class 2

Overview

Product CodePQI
Device NameWound Retraction And Protection System
Regulation21 CFR 878.4371
Device ClassClass 2
Review PanelGeneral, Plastic Surgery
3rd-Party ReviewableYes

Identification

An irrigating wound retractor device is a prescription device intended to be used by a surgeon to retract the surgical incision, to provide access to the surgical wound, to protect and irrigate the surgical wound, and to serve as a conduit for removal of fluid from the surgical wound.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

In combination with the general controls of the FD&C Act, the Irrigating Wound Retractor Device is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible and evaluated for particulate matter. (2) Performance data must demonstrate the sterility and pyrogenicity of the patient-contacting components of the device. (3) Performance data must support shelf life by demonstrating continued functionality and sterility of the device over the identified shelf life. (4) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Performance testing must: (i) Characterize the tear resistance, tensile strength, and elongation properties of the barrier material; (ii) Demonstrate that the liquid barrier material is resistant to penetration by blood, and is non-flammable; (iii) Characterize the forces required to deploy the device; (iv) Characterize the device's ranges of operation, including flow rates and maximum suction pressures; (v) Demonstrate the ability of the device irrigation apparatus to maintain a user defined or preset flow rate to the surgical wound; and (vi) Demonstrate the ability of the device to maintain user defined or preset removal rates of fluid from the surgical wound. (5) The labeling must include or state the following information: (i) Device size or incision length range; (ii) Method of sterilization; (iii) Flammability classification; (iv) Non-pyrogenic; (v) Shelf life; and (vi) Maximum flow rate and suction pressure.

Recent Cleared Devices (2 of 2)

RecordDevice NameApplicantDecision DateDecision
K172132CleanCision Wound Retraction and Protection SystemPrescient Surgical, Inc.Oct 25, 2017SESE
DEN150038CleanCisionTM Wound Retraction and Protection SystemPrescient SurgicalDec 16, 2016DENG

Top Applicants

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...