Last synced on 20 December 2024 at 11:05 pm

General Laparoscopic Power Morcellation Containment System

Page Type
Product Code
Definition
Intended for use as a bag containment system by qualified surgeons for tissue extraction and/or power morcellation during general laparoscopic procedures.
Physical State
A retractor and boot assembly with a tissue pouch/specimen bag.
Technical Method
A tissue pouch is placed through a multi-instrument port to allow for containment of tissue. The pouch is insufflated to create a working space for morcellation. A laparoscopic power morcellator and a laparoscope are placed through the port and into the pouch to enable visualization of the morcellation.
Target Area
Abdomen
Regulation Medical Specialty
General and Plastic Surgery
Review Panel
General and Plastic Surgery
Submission Type
510(K)
Device Classification
Class 2
Regulation Number
878.4825
GMP Exempt?
No
Summary Malfunction Reporting
Ineligible
Implanted Device
No
Life-Sustain/Support Device
No
Third Party Review
Not Third Party Eligible

CFR § 878.4825 General laparoscopic power morcellation containment system

§ 878.4825 General laparoscopic power morcellation containment system.

(a) Identification. A general laparoscopic power morcellation containment system is a prescription device consisting of an instrument port and tissue containment method that creates a working space allowing for direct visualization during a power morcellation procedure following a laparoscopic procedure for the excision of benign tissue that is not suspected to contain malignancy.

(b) Classification. Class II (special controls). The special controls for this device are:

(1) The patient-contacting components of the device must be demonstrated to be biocompatible.

(2) Performance testing must demonstrate the sterility of patient-contacting components of the device.

(3) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life.

(4) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:

(i) Demonstration of the device impermeability to tissue, cells, and fluids;

(ii) Demonstration that the device allows for the insertion/withdrawal of laparoscopic instruments while maintaining pneumoperitoneum;

(iii) Demonstration that the containment system provides adequate space to perform morcellation and adequate visualization of the laparoscopic instruments and tissue specimen relative to the external viscera;

(iv) Demonstration that compatible laparoscopic instruments and morcellators do not compromise the integrity of the containment system; and

(v) Demonstration that users can adequately deploy the device, morcellate a specimen without compromising the integrity of the device, and remove the device without spillage of contents.

(5) Training must be developed and validated to ensure users can follow the instructions for use.

(6) Labeling must include:

(i) A contraindication for use in gynecological procedures;

(ii) A contraindication against use of tissue that is known or suspected to contain malignancy;

(iii) The following boxed warning: “Warning: Information regarding the potential risks of a procedure with this device should be shared with patients. The use of laparoscopic power morcellators may spread cancer. The use of this containment system has not been clinically demonstrated to reduce this risk;”

(iv) A statement limiting use of device to physicians who have completed the training program; and

(v) A shelf life.

[86 FR 66458, Nov. 23, 2021]

General Laparoscopic Power Morcellation Containment System

Page Type
Product Code
Definition
Intended for use as a bag containment system by qualified surgeons for tissue extraction and/or power morcellation during general laparoscopic procedures.
Physical State
A retractor and boot assembly with a tissue pouch/specimen bag.
Technical Method
A tissue pouch is placed through a multi-instrument port to allow for containment of tissue. The pouch is insufflated to create a working space for morcellation. A laparoscopic power morcellator and a laparoscope are placed through the port and into the pouch to enable visualization of the morcellation.
Target Area
Abdomen
Regulation Medical Specialty
General and Plastic Surgery
Review Panel
General and Plastic Surgery
Submission Type
510(K)
Device Classification
Class 2
Regulation Number
878.4825
GMP Exempt?
No
Summary Malfunction Reporting
Ineligible
Implanted Device
No
Life-Sustain/Support Device
No
Third Party Review
Not Third Party Eligible

CFR § 878.4825 General laparoscopic power morcellation containment system

§ 878.4825 General laparoscopic power morcellation containment system.

(a) Identification. A general laparoscopic power morcellation containment system is a prescription device consisting of an instrument port and tissue containment method that creates a working space allowing for direct visualization during a power morcellation procedure following a laparoscopic procedure for the excision of benign tissue that is not suspected to contain malignancy.

(b) Classification. Class II (special controls). The special controls for this device are:

(1) The patient-contacting components of the device must be demonstrated to be biocompatible.

(2) Performance testing must demonstrate the sterility of patient-contacting components of the device.

(3) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life.

(4) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:

(i) Demonstration of the device impermeability to tissue, cells, and fluids;

(ii) Demonstration that the device allows for the insertion/withdrawal of laparoscopic instruments while maintaining pneumoperitoneum;

(iii) Demonstration that the containment system provides adequate space to perform morcellation and adequate visualization of the laparoscopic instruments and tissue specimen relative to the external viscera;

(iv) Demonstration that compatible laparoscopic instruments and morcellators do not compromise the integrity of the containment system; and

(v) Demonstration that users can adequately deploy the device, morcellate a specimen without compromising the integrity of the device, and remove the device without spillage of contents.

(5) Training must be developed and validated to ensure users can follow the instructions for use.

(6) Labeling must include:

(i) A contraindication for use in gynecological procedures;

(ii) A contraindication against use of tissue that is known or suspected to contain malignancy;

(iii) The following boxed warning: “Warning: Information regarding the potential risks of a procedure with this device should be shared with patients. The use of laparoscopic power morcellators may spread cancer. The use of this containment system has not been clinically demonstrated to reduce this risk;”

(iv) A statement limiting use of device to physicians who have completed the training program; and

(v) A shelf life.

[86 FR 66458, Nov. 23, 2021]