DEROYAL INDUSTRIES, INC. ELECTROSURGICAL CAUTERY PENCIL

K961647 · Deroyal Industries, Inc. · GEI · Oct 8, 1996 · General, Plastic Surgery

Device Facts

Record IDK961647
Device NameDEROYAL INDUSTRIES, INC. ELECTROSURGICAL CAUTERY PENCIL
ApplicantDeroyal Industries, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateOct 8, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Intended Use

The DeRoyal Cautery Pencil is indicated for use whenever an electrosurgical procedure requires the removal of tissue or the control of bleeding by either cutting or coagulation.

Device Story

Electrosurgical cautery pencil; used in surgical procedures for tissue removal and hemostasis. Device delivers high-frequency electrical current to surgical site. Operated by surgeons/clinicians in clinical settings. Tip design facilitates cutting and coagulation. Titanium Nitride coating provides anti-stick properties. Device connects to standard electrosurgical generators.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Materials: Stainless steel tip, PVC cable insulation. Coating: Titanium Nitride (anti-stick). Dimensions: 6.0" pencil body, 10' cable. Energy source: High-frequency electrical current from external electrosurgical generator. Principle: Electrosurgical cutting and coagulation.

Indications for Use

Indicated for patients undergoing electrosurgical procedures requiring tissue removal or hemostasis via cutting or coagulation.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} DEROYAL INDUSTRIES, INC. ELECTROSURGICAL CAUTERY PENCIL K 961647 510(k) Summary SUMMARY OF THE SAFETY AND EFFECTIVENESS INFORMATION UPON WHICH AN EQUIVALENCE DETERMINATION COULD BE BASED OCT - 8 1996 ## SUBMITTER INFORMATION NAME: DeRoyal Industries, Inc. ADDRESS: 200 DeBusk Lane Powell, TN 37849 TELEPHONE: (423) 938-7828 CONTACT: Debra F. Manning DATE OF PREPARATION: April 22, 1996 ## DEVICE NAMES NAME: DeRoyal Industries, Inc. Electrosurgical Cautery Pencil COMMON/USUAL NAME: Cautery Pencil CLASSIFICATION NAME (if known): Electrosurgical Cutting & Coagulation Device and Accessories (79 GEI) ## PREDICATE OR LEGALLY MARKETED DEVICES DeBusk Technology Corporation ## DEVICE DESCRIPTION The DeRoyal Industries, Inc. Electrosurgical Cautery Pencil functions in the same manner as predicate devices in that it is intended to be used in electrosurgical procedures to remove tissue and control bleeding. The device can be used for both cutting and coagulating. Device Design/Materials Used/Physical Properties: The DeRoyal Industries, Inc. Electrosurgical Cautery Pencil is made of materials commonly used for their purpose. The concept of use is that the device is intended to remove tissue and control bleeding by use of high frequency electrical current. ## DEVICE INTENDED USE The DeRoyal Cautery Pencil is indicated for use whenever an electrosurgical procedure requires the removal of tissue or the control of bleeding by either cutting or coagulation. ## TECHNOLOGICAL COMPARISON WITH PREDICATE OR LEGALLY MARKETED DEVICE(S) | Characteristic | DeRoyal Device | DeBusk Technology Device | Rationale | | --- | --- | --- | --- | | Length of Pencil Body | 6.0" | 6.0" | | | Length of Cable | 10'0" | 10'0" | | | Material of Outer Cable Insulation | PVC | PVC | | | Method of Manufacturing of Tip | Stamped and formed from sheet of stainless steel | Stamped and formed from sheet of stainless steel | | | Tip Coating | Titanium Nitride | None | Provides a non-toxic, anti-stick, lubricated surface. |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...